Cardiac Electronic Devices: Future Directions and Challenges

ORCID Icon, ORCID Icon, , ORCID Icon &
Pages 325-338 | Published online: 25 Sep 2020

Abstract

Cardiovascular implantable electronic devices (CIEDs) are essential management options for patients with brady- and tachyarrhythmias or heart failure with concomitant optimal pharmacotherapy. Despite increasing technological advances, there are still gaps in the management of CIED patients, eg, the growing number of lead- and pocket-related long-term complications, including cardiac device–related infective endocarditis, requires the greatest care. Likewise, patients with CIEDs should be monitored remotely as a part of a comprehensive, holistic management approach. In addition, novel technologies used in smartwatches may be a convenient tool for long-term atrial fibrillation (AF) screening, especially in high-risk populations. Early detection of AF may reduce the risk of stroke and other AF-related complications. The objective of this review article was to provide an overview of novel technologies in cardiac rhythm–management devices and future challenges related to CIEDs.

Introduction

Modern cardiology develops and progresses through innovations in technology and a deeper understanding of the pathophysiology of heart diseases. Indeed, marked advances have been made since 1958, when the first pacemaker was implanted.Citation1 Over the decades, cardiovascular implantable electronic devices (CIEDs) have become the cornerstone of management for patients with brady- or tachyarrhythmias and heart failure (HF) with reduced ejection fraction (EF).Citation2Citation5 This is associated with the emergence of complex stimulation systems — pacemakers (PMs), implantable cardioverter–defibrillators (ICDs), cardiac resynchronization therapy (CRT) — and the growing number of patients with different indications treated with CIEDs. Indeed, rhythm-management devices may improve the life expectancy and quality of life of these patients.Citation1

Impressive progress in the field of cardiac pacing has led to technical improvements in existing devices and leads, and new ones are constantly emerging. Despite this development, the systems may have a downside associated with early and late (>3 months after implantation) complications of using such CIEDs. Many are related to the weakest links, ie, the transvenous lead and subcutaneous pocket.Citation6,Citation7 Wireless technology and optimization of pacing systems have emerged to minimize potential CIED side effects. The effective interrogation and monitoring of patients with CIEDs to detect arrhythmias or system malfunction assumes even great importance.Citation8 The objective of this review article was to provide an overview of novel technologies in cardiac rhythm–management devices and future directions and challenges related to CIEDs.

Search Strategy

We performed a comprehensive literature search using electronic databases (PubMed, ClinicalTrials.gov) to identify relevant studies and systematic reviews reporting on cardiac rhythm–management devices. The following search terms were included (individually and in combination): cardiac implantable electronic devices, pacemaker, implantable cardioverter–defibrillator, cardiac resynchronization therapy, leadless cardiac pacemaker, wearable cardioverter–defibrillator, subcutaneous cardioverter–defibrillator, substernal lead, remote monitoring, atrial high-rate episodes, smartwatch, and wearable technology. Selected articles, clinical trials, and guideline documents were reviewed for inclusion.

Who is Appropriate for CIEDs?

A challenge is whether a patient appropriately qualifies for CIED implantation, despite the current guidelines.Citation2Citation5 SCD-HeFTCitation9 was a randomized controlled trial (RCT) conducted among 2,521 patients with HF, left ventricular EF (LVEF) ≤35%, and New York Heart Association class II or III. It was found that ICD therapy was related to a 23% reduction of mortality compared with patients treated with amiodarone or placebo.Citation9 However, the recent results from the DANISH trialCitation10 did not show a survival benefits among patients with nonischemic HF with ICD implanted as primary prevention of sudden cardiac death (SCD). The exception was the subgroup of patients aged <68 years, in which the risk of all-cause death was significantly lower in the ICD group than the control group (HR 0.64, 95% CI 0.45–0.90). Importantly, 31% of deaths were not related to cardiovascular events, which may also indicate less benefit from ICDs in older and frail patients.Citation10 Lee et alCitation11 reported that the presence of ICDs significantly reduced the rate of all-cause death (HR 0.640, 95% CI 0.448–0.915) in patients with ischemic cardiomyopathy, but not among individuals with nonischemic cardiomyopathy (HR 0.984, 95% CI 0.641–1.509) during follow-up of 3.5±1.8 years.Citation11

Beyond recommendations about ICD therapy in patients with ischemic cardiomyopathy and poor LVEF, diagnostic algorithms for the identification of patients with relatively preserved LV contractility at increased risk of major arrhythmic events have been proposed. The PRESERVE EF studyCitation12 was performed among 575 patients of mean age 57 years and LVEF 50.8%. Participants were assessed in two steps: if there were abnormalities on ECG (eg, premature ventricular complexes, unsustained ventricular tachycardia, late potentials, prolonged QTc), patients were referred to programmed ventricular stimulation (PVS). For those with induced ventricular tachyarrhythmia (VT), ICDs were implanted. The primary end point was the occurrence of a major arrhythmic event: sustained ventricular tachycardia/fibrillation, appropriate ICD therapy, or SCD. The study found that 35.5% had abnormal ECG findings, and 27% of those were inducible with PVS. ICDs were implanted in 37 patients (90.2% of inducible subgroup). During the 32-month follow-up, there were no SCDs among ICD patients, whereas nine appropriate ICD shocks were observed. A previous studyCitation13 showed that patients with hypertrophic cardiomyopathy and noninducible arrhythmia with PVS had longer event-free survival. Inducibility with PVS was an independent predictor of SCD or appropriate ICD therapy among patients with hypertrophic cardiomyopathy.Citation13

While the effectiveness of ICD therapy in patients with nonischemic dilated cardiomyopathy and reduced LVEF (≤35%) is debated, the selection of patients with dilated cardiomyopathy and well-maintained LV contractility (LVEF >35%) at risk of malignant cardiac arrhythmic events who may gain a survival benefit from ICD therapy represents another challenging area. Gatzoulis et alCitation14 used a two-step algorithm in another study — ReCONSIDER; (NCT04246450) — which is an ongoing prospective observational trial among patients with nonischemic cardiomyopathy aiming to recognize those with a truly high risk of SCD.Citation14 CMR GUIDE (NCT01918215)Citation15 is an ongoing RCT to assess myocardial fibrosis and related risk of SCD among patients with LVEF 36%–50% and evidence of fibrosis on optimal HF therapy. Patients are randomized to receive ICD (as primary SCD prevention) or an implantable loop recorder (ILR). The composite primary end point is time to SCD or hemodynamically significant VT.Citation15

Another debated issue is the optimal selection criteria for CRT responders, especially among patients with HF, without typical left bundle–branch block.Citation16 New strategies, such as leadless pacing, optimization of LV-lead position, multipolar LV pacing, alternative right ventricular (RV) pacing, eg, His-bundle pacing or cardiac contractility modulation, may positively impact on further CIED therapy.Citation17 MORE-CRT MPP-PHASE II (NCT02006069)Citation18 is an RCT to assess the impact of multipoint pacing in nonresponders to 6 months of standard biventricular pacing. Preferential LV-only pacing is also considered an alternative to standard biventricular pacing.Citation19 A prospective randomized study of CRT with preferential adaptive LV-only pacing (AdaptResponse, NCT02205359)Citation20 is assessing if the new pacing algorithm reduces the incidence of the combined end point of all-cause mortality and HF decompensation compared with conventional CRT among patients eligible for CRT. The AdaptivCRT algorithm optimizes the pacing method and atrioventricular/interventricular delays, based on the current patient’s activity and intrinsic conduction.Citation20

His-bundle pacing is the most physiological form of ventricular pacing, and appears to be a safe and effective method during long-term follow-up.Citation21 This approach is considered superior to standard RV pacing and may also improve clinical outcomes in patients with CRT indications.Citation22 The His-SYNC (NCT02700425)Citation23 pilot trial was the first RCT comparing His-bundle pacing for CRT (His-CRT) vs biventricular pacing (BiV-CRT) among 41 patients with standard indications for CRT. At 6-month follow-up, His-CRT resulted in QRS narrowing with a nonsignificant trend toward a higher rate of echocardiographic response (91% vs 54%, p=0.078) compared with BiV-CRT; however, there were no significant differences in mortality or cardiovascular hospitalization between the groups.Citation23 As such, large multicenter RCTs are necessary to evaluate the clinical efficacy of His-bundle pacing and also comparing His-CRT and BiV-CRT. It is also open whether patients who qualify for CRT have a survival benefit from ICD. RESET-CRT (NCT03494933) is an ongoing RCT to compare clinical outcomes among patients with a CRT PM vs CRT defibrillator.

A recent European Heart Rhythm Association (EHRA) consensus document on management of arrhythmias and cardiac electronic devices in critically ill and postsurgery patients highlighted the risks and challenges among CIED patients with a terminal illness.Citation24 In an EHRA survey,Citation25 73% of patients declared that CIED implantation improved their quality of life, whereas 36% had concerns about the device, mostly related to ICD shocks, daily activities, or impairment of the device. Indeed, the final decision about CIED implantation should take into account the patient’s age, frailty, cardiac condition, and other comorbidities, concomitant with personal values and preferences.

Therefore, it is often necessary to create novel algorithms for the selection of CRT responders, individualized risk scores for SCD, or procedure-related complications, which may result in a highly individualized approach and targeted CIED implantation. The future may bring patient-specific digital models to calculate the risk–benefit profile and create a simulation — virtual implantation. This might check whether the procedure is feasible and which device is favorable for each patient, but also guide a lead during the real procedure. In addition, CIEDs may be considered a cotreatment of other morbidities, such as hypertension.NCT03757377 is an ongoing RCT evaluating a new PM algorithm that may be useful for patients with indications for antibradycardia pacing and persistent hypertension despite pharmacotherapy.

The Weakest Links

Transvenous leads represent a major source of CIED complications — not only dislocation or mechanical damage but also tricuspid regurgitation, venous occlusion, superior vena cava syndrome, cardiac perforation, cardiac device–related infective endocarditis (CDRIE) — and subcutaneous pockets: hematoma, decubitus, inflammation.Citation6,Citation7,Citation26 Palmisano et alCitation27 reported a higher risk of all-cause death among patients with CIEDs and early complications — pneumothorax (HR 8.731, 95% CI 1.42–53.63) and pocket hematoma (HR 2.515, 95% CI 1.07–5.94) — whereas CDRIE was most markedly related to increased risk of cardiovascular death (HR 4.025, 95% CI 1.5–10.78) during median follow-up of 56.9 months. An EHRA international consensus on how to prevent, diagnose, and treat CIED-infectionsCitation28 states that prevention and careful consideration before implantation are the best treatment for CDRIE. Indeed, leadless cardiac PMs (LCPs) and extravascular cardioverter-defibrillators have been designed to minimize complications. Also, an absorbable, antibiotic-eluting envelope has been created to use with CIEDs as a prophylactic strategy to prevent CDRIE.Citation29 These new technologies have already found a place in everyday clinical practice.

Leadless Cardiac Pacemakers

An LCP is a small (volume 0.8 cm3) single-chamber PM that is implanted directly into the RV by a special catheter and introducer sheath via transfemoral access.Citation30 Therefore, it does not require the subcutaneous pocket or transvenous lead.Citation31 The first LCP was Nanostim (St Jude Medical), implanted worldwide between 2013 and 2016. The device was recalled in 2016, due to battery failures, but the concept of LCPs has been widely accepted.Citation32

At present, the only type of LCP available on the market is the Micra transcatheter pacing system. The pacing mode is similar to transvenous PMs, so an LCP may be used as an alternative device.Citation30,Citation33 However, the system is limited to the RV component, meaning that an LCP may be indicated only for patients requiring single-chamber pacing, eg, permanent atrial fibrillation (AF) with bradycardia or for those with low expected stimulation percentage.Citation30 As such, patients with missing or difficult venous access, with a history of CDRIE, and indications for ventricular single-chamber pacing are considered good LCP candidates. Importantly, the potential benefits of LCPs must be confronted with the limited data on the long-term follow-up, and also the procedure of device replacement or retrieval is still debated.Citation34 According to a national expert consensus document of the Austrian Society of Cardiology, LCP retrieval should not be recommended as a routine procedure and should be limited only to specific issues, ie, endocarditis or system upgrades.Citation34 One worldwide experience of 40 successful device retrievals revealed that it may be feasible and safe if performed with a special sheath and a snare catheter and introduced via femoral access. The most common reasons for extraction included elevated pacing threshold, endovascular infection, and indications for a system upgrade to a transvenous device.Citation35 If it is necessary to replace the battery, a new LCP may be implanted next to old devices without extraction of previous ones; however, current clinical experience is very limited.Citation34

The results of the Micra Transcatheter Pacing StudyCitation36 showed the safety and efficacy(primary safety and efficacy end points were reached in 96% and 98.3%, respectively) of LCPs among 725 patients who had undergone device implantation. Likewise, the Micra Post-Approval RegistryCitation37 reported a high rate of successful LCP implantations (99.1%) with a low risk of major complications (2.7%) among 1,817 patients. During follow-up of 12-months postprocedure, complication rates in LCP patients were significantly lower (HR 0.37, 95% CI 0.27–0.52) than a historical transvenous PM group. The most common complications in the LCP group were pacing issues (0.72%), groin injury (0.61%), cardiac effusion/perfusion (0.44%), and infection (0.17%).Citation37 In another study, El-Chami et alCitation38 reported on the safety and feasibility of LCPs, also in patients after PM extraction and a recent CDRIE.

Piccini et alCitation39 compared clinical outcomes among 720 patients successfully implanted with LCPs, based on ventricular pacing indications: individuals with AF (68.3%) and those without AF (31.7%). Reasons for selecting LCPs in the non-AF group included an expectation of infrequent pacing (66.2%) and advanced age (27.2%). During 24 months of follow-up, there were no significant differences between the groups in occurrence of the composite primary outcome (cardiac failure, PM syndrome, or LCP-related syncope).Citation39 In another study, the safety and mortality of LCP implantation was assessed and stratified by whether patients were precluded from transvenous PMs.Citation40 It was found that 19.4% of patients were ineligible for traditional PMs because of venous access issues or prior CDRIE. Both acute and total mortality at 36 months (2.75% vs 1.32% [p=0.022] and 38.1% vs 20.6% [p<0.001], respectively) were significantly higher in the precluded patients than the non-precluded; however, the mortality rate among precluded patients was similar to historical transvenous PM group.Citation40

Despite concerns regarding LCPs in frail elderly patients, because of implant-sheath size and risk of perforation, Micra implantation also appears to be safe and feasible among those individuals.Citation41,Citation42 However, RCTs directly comparing the efficacy and safety of LCPs vs transvenous PMs are needed. In the EHRA prospective survey,Citation43 the overall use of LCPs in daily clinical practice remains low, constituting only 9% of all procedures and 36% of single-chamber PM implants. LCPrecipients were more often male (74% vs 54%) and had a history of valvular heart disease (45% vs 35%), AF (65% vs 23%), and other comorbidities (66% vs 52%) than those with transvenous single-chamber PMs, but no significant association was observed with patients’ age.Citation43 LCP implantation was successful in 98% of recipients, and the only procedure-related complication was groin hematoma.Citation44 Indeed, leadless devices are still in development, and there are also the prototypes of dual-chamber systems,Citation45 which may be used in a wider group of patients. As such, LCPs are a potential game changer for modern CIEDs.

MARVEL (NCT03157297)Citation46 was a recent study of a new LCP algorithm to synchronize ventricle pacing with atrial sensing (synchronous atrioventricular [AV] pacing). Consequently, MARVEL 2 (NCT03752151)Citation47 revealed that the new algorithm provided successful AV-synchrony pacing (mean 89.2%) among 75 LCP recipients with sinus rhythm and AV block. Notably, the atrial sensing algorithms were safe, and there were neither pauses nor episodes of PM-mediated tachycardia.Citation47 The technology is currently used in a new Micra AV device (approved by the US Food and Drug Administration in January 2020), broadening potential indications to LCP implantation.Citation48 Further innovations, such as compatibility with extravascular ICDs, leadless CRT, renewable batteries, or less invasive implantation procedures, are highly anticipated.

Wireless Cardiac Resynchronization Therapy

The SELECT-LV studyCitation49 investigated the clinical efficacy and safety of wireless stimulation endocardially for CRT (WiSE-CRT) pacing via an LV endocardial electrode and a pulse generator (implanted subcutaneously). The trial was conducted among 35 patients with HF and indications for biventricular pacing who were nonresponders to traditional CRT or implantation of a coronary sinus lead was not possible. The feasibility and efficacy of WiSE-CRT were reported. Rates of successful implantation and effective CRT pacing were high (97.1%), with a substantial improvement (84.8%) in the clinical composite score at 6 months; however, the rate of early serious complications was 31.5% at 1-month postprocedure follow-up (8.6% within 24 hours and 22.9% between 24 hours and 1 month).Citation49 According to recent data from a Multicenter International Registry of the WiSE-CRT pacing system,Citation50 implantation of the device was feasible in 94.4% of patients and 70% of those reported improvement in of HF symptoms. Complication rates differed among the centers: 4.4%, 18.8%, and 6.7% within <24 hours, 1–30 days, and 1–6 months, respectively.Citation50 Likewise, the SOLVE-CRT trialCitation51 is an ongoing RCT assessing the safety and effectiveness of the WiSE-CRT in nonresponders to traditional CRT.

Extravascular Cardioverter–Defibrillators

Wearable Cardioverter–Defibrillators

European Society of Cardiology (ESC) guidelines for the management of patients with ventricular arrhythmias and prevention of SCDCitation5 state that a wearable cardioverter-defibrillator (WCD) attached to a vest may be considered a bridge in patients with transient impaired LVEF, post–myocardial infarction (<40 days), postpartum cardiomyopathy, myocarditis (until recovery), or for those awaiting heart transplantation. VESTCitation52 was an RCT of 2,302 subjects with acute myocardial infarction and LVEF ≤35% designed to compare a WCD group vs a no-device group, with both groups receiving optimal pharmacotherapy. The study found that the WCD did not have a significant impact on the composite primary end point (SCD and death from sustained VT) during 90 days postinfarct.Citation52 Adherence to wearing the defibrillator was an important issue that conditioned the results of the VEST study, as highlighted by the on-treatment and per-protocol analyses that found a benefit of the wearable defibrillator in the group of patients selected for high compliance to apply the device.Citation53 Data from the WEARIT-II RegistryCitation54 showed a high rate of VT — 3% among patients with ischemic and congenital heart disease and 1% among nonischemic patients — during 3 months of WCD use. In total, there were 120 VT episodes recorded in 41 patients, and 54% of those received appropriate WCD therapy. In sum, 840 patients (42%) were implanted with an ICD, and LVEF improvement was the most common reason for ICD removal.Citation54 The EHRA surveyCitation55 reported that WCDs were used as a temporary solution, mostly for patients after CDRIE and awaiting ICD reimplantation or with transient LVEF impairment, after recent myocarditis/myocardial infarction, or before heart transplantation. Notably, the most common contraindications for WCD use were life expectancy <12 months and noncompliance.Citation55

Subcutaneous Cardioverter–Defibrillator

Subcutaneous cardioverter-defibrillators (sICDs) are implantable devices comprising a subcutaneous pulse generator and subcutaneous lead to deliver a shock as VT therapy. According to the current ESC guidelines, sICDs may be an alternative for patients who require an ICD but do not have an indication for ventricular pacing, CRT, or antitachycardia pacing.Citation5 The American Heart Association guidelinesCitation4 recommend sICDs for patients without proper venous access or at high risk of CDRIE. Boersma et alCitation56 reported low periprocedural complication rates, with 99.6% successfully implanted devices and high defibrillation efficacy (99.2% during defibrillation testing) among 1,116 patients who had undergone sICD implantation as primary prevention of SCD. In another study, Boersma et alCitation57 indicated a low risk of infection in sICD recipients, even in individuals with a history of CDRIE and explanted transvenous ICDs.

Results from an EHRA prospective survey on sICD use showed that sICDs were favorable among younger patients and those with lead-related complications or elevated risk/history of CDRIE, taking into consideration patient preferences and active lifestyle. Of note, need for ventricular antitachycardia pacing, CRT, or permanent pacing were benefiting, the transvenous ICD.Citation58 Implantation time and periprocedural complication rates were similar in both subgroups.Citation59 Since 2010, when sICD became available for patients, this technology has been evolving.Citation60 Various ongoing clinical trials have been designed to assess new extravascular systems, eg, EV ICD (NCT04060680) and ASE (NCT03802110), for testing new shock configurations.

Substernal Leads

ASD2Citation61 was the first human study to evaluate a novel approach to ICD therapy — substernal leads. Pacing threshold, sensing, and defibrillation efficacy were assessed among 79 patients who had undergone lead implantation. The lead was placed into the substernal space via subxiphoid access, and a defibrillation-patch electrode or active can emulator (subcutaneous) was set in the left mid-axillary line. It was found that R-wave amplitudes were compliant with ICD sensing (median R-wave 2.4 mV), successful ventricular pacing rates were 97.4%, and defibrillation efficacy was >80% with a single shock of 30 J. The lower shock energy (compared with the sICD’s 80 J) may be beneficial for battery longevity.Citation61 Indeed, substernal lead therapy may be feasible, and results are promising. Nevertheless, neither long-term follow-up data nor risk of infection and lead extraction are available. Indeed, substernal lead therapy may be feasible, and results are promising. Nevertheless, neither long-term follow-up data nor risk of infection and lead extraction are available ().

Table 1 Comparison of Implantable Cardioverter Defibrillators

Remote Monitoring of Patients with Cardiac Electronic Implantable Devices

Patients with CIEDs usually have routine in-person appointments to verify whether the device is functioning adequately and for assessment of clinical findings every 6–12 months.Citation62 In addition, unscheduled clinic visits may be necessary in cases of system malfunction or worsening of health status.Citation62 However, the conventional monitoring of CIEDs, resulting in limited contact with patients, is insufficient and outdated. Digital health-care models and remote control of devices are the future of modern medicine and cardiology. They involve patients taking an active role in their own clinical care, and such a personalized approach, including shared decision-making, is crucial ().Citation63

Table 2 Expected Advantages of Remote Monitoring of Cardiac Electronic Implantable Devices

Teletransmission systems transfer data recorded from the patient’s device to a database, where the data are available to the health-care team. New systems transmit the data via the patient’s smartphone or tablet. As such, telemonitoring allows assessment of the relevant technical parameters of the device, ie, battery status, electrode function, and system compatibility, on an ongoing basis. It also provides key clinical information, such as stimulation percentage, stored arrhythmic episodes, or current intracardiac electrograms.Citation8,Citation62 The daily telemetric care of a large population of such CIED patients may be considered triage of high-risk patients, with ongoing selection of individuals who require urgent medical intervention.Citation64

According to expert consensus, remote monitoring should be available for all patients with CIEDs as part of the standard follow-up strategy, but in particular for patients with HF and cardiac arrhythmias.Citation8,Citation62 There is urgent need for integrated medical care for patients with HF. Telemetric data may help to recognize current clinical status and device alarm. Of note, telemetric care is crucial to diagnose and monitor episodes of arrhythmia, especially life-threatening episodes of VT and ICD-delivered shocks (appropriate or inappropriate). As a result, the health-care team may efficiently modify pharmacotherapy or plan any further treatment strategy (). Importantly, the effectiveness of the therapy may be assessed on an ongoing basis.Citation64 A survey from the Health Economics Committee of the EHRACitation25 showed that early recognition of AF among PM patients, lead failure in ICD patients, and HF worsening in CRT patients were considered essential advantages of remote monitoring.

Table 3 Remote Monitoring of Patients with Cardiac Resynchronization Therapy

The opportunity comes from dual-chamber devices that can also monitor atrial rhythm, allowing the possibility of recognizing arrhythmias and AF. These recorded events are called atrial high-rate episodes (AHREs), more often of short duration, lasting a few seconds to minutes.Citation65,Citation66 The key issue is to diagnose real episodes of AF or atrial flutter, and thus it is necessary to evaluate stored current intracardiac electrograms and exclude false episodes ().Citation65Citation68 Among 2,718 individuals with HF and CIEDs, AHREs were found in 34.8%, whereas AF was confirmed in 91% by atrial electrograms.Citation69 In an observational study of 304 patients with HF and CRT with a defibrillator, AHREs were found in 57.9% of patients within 2.5 years postimplantation.Citation67 After inspection, 89.2% of AHREs were truly AF and 62% of these were AF de novo.Citation67

Table 4 Atrial High–rhythm Episodes: Common Causes of Atrial Fibrillation Misdetections,

A significant benefit of telemonitoring may be obtained by patients with previously undetected “silent” AF. An active search for real arrhythmia and related risk factors is a significant part of individualized approaches among patients with CIEDs. Miyazawa et alCitation70 revealed that age ≥65 years, diabetes mellitus, congestive HF, and left atrial volume index >34 mL/m2 were independent factors of AF de novo in patients with CIEDs. Indeed, detecting asymptomatic AF with an adequate therapeutic response may prevent potential complications, eg, stroke/thromboembolic events and AF-related morbidity. However, there remains a gap in the data regarding the management of patients with AHREs without a proper AF diagnosis.

How much AHRE is too much? This is the big question that derives from lack of knowledge on what the number of AHREs is and their minimum duration, which actually increases significantly the thromboembolism risk.Citation68,Citation71Citation73 One study among patients with dual-chamber CIEDs showed an increased risk of thromboembolism (HR 3.40, 95% CI 1.38–8.37) and all-cause death (HR 3.47, 95% CI 1.51–7.95) among patients with AHRE.Citation74 Of note, the West Birmingham Atrial Fibrillation Project reported that among CIED patients, a higher risk of thromboembolic events was related to comorbidities (ie, CHA2DS2VASc score), but not AHRE per se.Citation75 Indeed, Pastori et alCitation76 reported that AHREs ≥5 minutes (HR 1.788, 95% CI 1.247–2.562), diabetes mellitus (HR 1.909, 95% CI 1.358–2.683), HF (HR 2.203, 95% CI 1.527–3.178), and coronary artery disease (HR 1.862, 95% CI 1.293–2.681) were significantly related to the risk of major adverse cardiovascular events in patients with CIEDs. Furthermore, Boriani et alCitation77 found that among patients with AF duration ≥5 minutes and CIEDs, female sex (HR 3.43, 95% CI 1.05–11.18) and history of coronary artery bypass surgery(HR 4.34, 95% CI 1.44–13.13), but not higher burden of AF, were independent predictors of stroke.Citation77

New-onset AF is detected earlier in patients with remote monitoring than standard clinical care.Citation78Citation80 Also, treatment of arrhythmia is initiated significantly earlier (3 vs 54 days), but does not improve clinical outcomes of HF patients in term of stroke and bleeding prevention, based on introduction or termination of oral anticoagulation (OAC).Citation69 Perino et alCitation81 conducted a study of 2,101 patients with CIEDs on remote monitoring and AF duration >6 minutes (detected by device). Among these patients, OAC users had a significantly lower risk of stroke than a no-anticoagulation group (HR 0.68, 95% CI 0.47–0.97). Of note, a reduction in strokes with anticoagulation was reported in patients with AF duration >24 hours (HR 0.27, 95% CI 0.14–0.51).Citation81

There is a beneficial effect of OAC in stroke prevention in CIED patients with documented AF, while evidence of benefit is missing for AHRE (without formal AF diagnosis).Citation65,Citation68,Citation71,Citation72,Citation82,Citation83 An ongoing RCT, NOAH-AFNET 6 (NCT02618577),Citation84 is comparing edoxaban vs aspirin or no treatment among patients with AHRE (without AF) and two or more stroke risk factors. The primary outcome is time to first ischemic event or cardiovascular-related death.Citation84 The ARTESiA (NCT01938248)Citation85 trial is comparing apixaban vs aspirin in terms of stroke and systemic embolism in this group of patients.

The CASTLE-AFCitation86 trial was designed to compare the clinical outcomes of AF patients with HF and implanted ICDs. Subjects were randomized to catheter ablation or guideline-adherent medical therapy. The major exclusion criteria were heart-transplant candidacy or planned cardiovascular intervention. Patients who underwent catheter ablation had significantly lower risk of any-cause death (13.4% vs 25.0%, HR 0.53, 95% CI 0.32–0.86), cardiovascular-related death (11.2% vs 22.3%, HR 0.49, 95% CI 0.29–0.84), and hospitalization due to worsening of HF (20.7% vs 35.9%; HR 0.56, 95% CI 0.37–0.83), than the medical therapy group.Citation86

Noseworthy et alCitation87 assessed the generalizability of the CASTLE-AF trial among patients with AF and HF treated with ablation (n=7,465) or standard medical therapy (n=282,366). They found that only 7.8% of patients would have been eligible for CASTLE-AF. Catheter ablation was related to a lower risk of the primary outcome among all patients (HR 0.81, 95% CI 0.76–0.87) vs standard medical therapy, specifically in the CASTLE-AF–eligible subgroup (HR 0.82, 95% CI 0.70–0.96), but not in patients who met the exclusion criteria.Citation87 Likewise, the CABANACitation88 trial revealed that among AF patients, catheter ablation did not significantly lower the primary composite end point of death, disabling stroke, major bleeding, or cardiac arrest compared with a medical therapy group. In an observational real-world patient study,Citation89 assessing catheter ablation was related to a reduction in the composite end point vs medical therapy (HR 0.75, 95% CI 0.70–0.81). The risk reduction associated with ablation was higher among CABANA-eligible patients (HR 0.70, 95% CI 0.63–0.77).Citation89

The possibility of remote monitoring may facilitate comprehensive care management for patients with CIEDs.Citation90 However, RCT-based data are heterogeneous in terms of telemonitoring effectiveness, ie, improvement of outcomes for HF patients with CIEDs ().Citation91Citation95 The benefits of monitoring may be dependent on the health-care team’s reaction to the transmitted data. Therefore, remote monitoring should not be considered a treatment per se, but may allow for a more appropriate medical response to device alerts.Citation95,Citation96 Consequently, further developments should be focused on improving the technical issues and efficiency of telemonitoring, eg, artificial intelligence to for triage of high-risk patients or integration of the data with electronic medical records or extra features to monitor potential comorbidities. Another challenge is the accessibility, feasibility, and adherence to therapeutic protocols by both groups — physicians and patients. Further studies are needed to identify novel functions with a positive impact on clinical outcomes.Citation97

Table 5 Studies of Remote Monitoring of Cardiac Electronic Implantable Devices

Novel Approaches Incorporating Cardiac Rhythm–Monitoring Technology

Implantable Loop Recorder

The ILR is a type of long-term cardiac rhythm–monitoring device that is inserted underneath the chest skin. It is a safe and effective tool in the diagnosis of unexplained syncope, arrhythmias, or cryptogenic stroke, and is indicated for patients with recurring symptoms but too infrequent to be diagnosed with conventional ECG monitoring techniques.Citation98,Citation99 In patients with recurrent symptomatic palpitations, the ILR may be a useful diagnostic approach for evaluation.Citation100 Of note, recent studies of ILRs have shown their feasibility to detect and record asymptomatic AF. One RCT (CRYSTAL-AF)Citation101 was conducted to compare ECG monitoring with an ILR vs conventional follow-up among 441 patients after cryptogenic stroke. The primary end point was time to first AF detection within 6 months. ILRs were superior to standard monitoring in detecting AF: AF was detected in 8.9% of patients in the ILR group vs 1.4% of patients in the control group (HR 6.4, 95% CI 1.9–21.7) during 6 months of follow-up.Citation101 The LOOP (NCT02036450) study was designed to determine whether initiation of OAC if AF were detected with ILRs would reduce the risk of stroke among 597 patients aged ≥70 years with stroke risk factors. First insights from the LOOP study revealed that during 40 months of follow-up, AF was detected in 209 (35%) patients.Citation102 In addition, AF was silent in 90% of patients at debut, and 87% never noticed AF-related symptoms. The average heart rate during AF was only moderately elevated to 96 beats/minute, whereas average daytime sinus rate was 72 beats/minute.Citation103

Wearable Technology

Wearable technologies provide the possibility of continuous rhythm monitoring and display a real-time heart rate, as well as a review of heart-rate trends.Citation104 For example, Dörr et alCitation105 conducted a study using smartwatches among 508 patients of mean age 76.4 years, 46.6% of whom had previously diagnosed AF. High sensitivity, specificity, and accuracy (93.7%, 98.2%, and 96.1%, respectively) in detecting AF was observed. The authors speculated that AF detection may be feasible with high diagnostic precision using commercial smartwatches. Likewise, attaining proper signal quality during algorithm use was highlighted as the main limitation.Citation105

The possibility of cardiac rhythm monitoring via smartwatches was assessed in two large population-based cohort studies in the US and China. The Apple Heart StudyCitation106,Citation107 involved 419,297 participants aged ≥22 years without a history of AF. The idea of the study was to find out if a mobile application could recognize irregular heart rate and AF from data collected on the Apple Watch. During follow-up, an irregular pulse was detected in 0.52% of participants. Consequently, they were monitored by ECG patch for 7 days for arrhythmia, and AF was diagnosed in 34%. Notably, AF was observed more often among participants aged ≥65 years (3.2% vs 0.16% in participants aged <40 years) and men (0.7% vs 0.26% in women). Likewise, 76% of notified participants contacted their health-care provider, 33% were referred to a specialist, and 36% required further investigations.

The Huawei Heart StudyCitation108 conducted in China on 187,912 participants of mean age 35 years found that 424 (0.23%) individuals were notified of suspected AF, and AF was finally confirmed by clinical evaluation in 87% of those. Researchers highlighted that AF suspicion and identification significantly increased with age. Of note, the majority of patients entered a program of integrated AF management, and approximately 80% of high-risk patients were successfully anticoagulated.Citation108 Subsequent to the Huawei Heart Study, mAFA IICitation109 was a cluster RCT to assess the impact of integrated care using a mobile AF application (mAFA) compared with usual care on clinical outcomes among 3,324 AFpatients with two or more stroke risk factors.Citation109 The trial demonstrated that rates of composite outcome (ischemic stroke/systemic thromboembolism, death, and rehospitalization) and rehospitalization were lower among mAFA patients vs the usual-care control group (1.9% vs 6.0%, HR 0.39, 95% CI 0.22–0.67 and 1.2% vs 4.5%, HR 0.32, 95% CI 0.17–0.60, respectively).Citation109 In an ancillary analysis of mAFAII, proactive assessment of bleeding risks using the HAS-BLED score resulted in mitigation of modifiable bleeding risk factors, fewer bleeding events and an increase in OAC uptake compared to usual care.Citation110 Another approach was illustrated with the NOMED-AF (NCT03243474) study,Citation111 which assessed the prevalence of AF and concomitant comorbidities in a population aged ≥65 years. Long-term ECG monitoring was enabled by special vests equipped with electrodes and recorder. Data were transmitted to the monitoring platform every 24 hours. The study identified high-risk populations, requiring more intense AF screening.Citation111

The data available show that wearable technology using photoplethysmographic sensors, which detect blood volume changes, may recognize irregular heart rhythm, including previously unknown AF ().Citation106Citation108 As such, smartwatches may be convenient tools for long-term AF screening in large populations, especially in high-risk patients. Early AF detection may reduce the burden of stroke and other AF-related complications. Future studies may demonstrate the utility of wearable technology in AF integrated care and optimize a holistic approach within individuals with AF. Likewise, the advance of artificial intelligence may support treatment and assist in diagnosis, management, and prediction of occurrence of arrhythmia or other heart diseases.Citation112

Table 6 Wearable Technologies – an Insight from Observational Studies

Conclusion

The future of cardiac rhythm–management devices looks bright, but incorporation and adoption of new technologies remains a challenge. There is an immense need for an individualized approach with targeted CIED implantation and personalized risk stratifications. Further studies are needed to identify novel strategies and technologies with a positive impact on patient diagnosis and treatment. Novel solutions may see a move away from transvenous leads to leadless systems, with combinations of different individualized pacing and monitoring functions. Relevant developments and innovations of devices, implantation processes, and postprocedure follow-up may improve clinical outcomes among patients.

Disclosure

ZK reports personal fees from Boehringer Ingelheim, Bayer, and Pfizer outside the submitted work, DJW consultancy and speaker fees for Medtronic, Boston Scientific, and St Jude and research grants from Boston Scientific, GYHL consultancy for Bayer/Janssen, BMS/Pfizer, Medtronic, Boehringer Ingelheim, Novartis, Verseon, and Daiichi Sankyo and speaking for Bayer, BMS/Pfizer, Medtronic, Boehringer Ingelheim, and Daiichi Sankyo (no fees directly received personally), and GB small speaker’s fees from Medtronic, Boston, Biotronik, Boehringer, and Bayer outside the submitted work. The authors report no other potential conflicts of interest for this work.

References

  • Wilkoff BL, Auricchio A, Brugada J, et al. HRS/EHRA Expert Consensus on the Monitoring of Cardiovascular Implantable Electronic Devices (CIEDs): description of techniques, indications, personnel, frequency and ethical considerations: developed in partnership with the Heart Rhythm Society (HRS) and the European Heart Rhythm Association (EHRA); and in collaboration with the American College of Cardiology (ACC), the American Heart Association (AHA), the European Society of Cardiology (ESC), the Heart Failure Association of ESC (HFA), and the Heart Failure Society of America (HFSA). Endorsed by the Heart Rhythm Society, the European Heart Rhythm Association (a registered branch of the ESC), the American College of Cardiology, the American Heart Association . Europace. 2008;10(6):707–725.
  • Ponikowski P, Voors AA, Anker SD, et al. 2016 ESC guidelines for the diagnosis and treatment of acute and chronic heart failure. Eur Heart J. 2016;37(27):2129–2200m. doi:10.1093/eurheartj/ehw128
  • Brignole M, Auricchio A, Baron-Esquivias G, et al. 213 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy. Eur Heart J. 2013;34(29):2281–2329.
  • Al-Khatib SM, Stevenson WG, Ackerman MJ, et al. 2017 AHA/ACC/HRS guideline for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death. Circulation. 2018;138(13):e272–e391.
  • Priori SG, Blomström-Lundqvist C, Mazzanti A, et al. 2015 ESC guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: the task force for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death of the Europe. Eur Heart J. 2015;36(41):2793–2867.
  • Kirkfeldt RE, Johansen JB, Nohr EA, Dan Jørgensen O, Nielsen JC. Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark. Eur Heart J. 2014;35(18):1186–1194. doi:10.1093/eurheartj/eht511
  • Bongiorni MG, Burri H, Deharo JC, et al. 2018 EHRA expert consensus statement on lead extraction: recommendations on definitions, endpoints, research trial design, and data collection requirements for clinical scientific studies and registries: endorsed by APHRS/HRS/LAHRS. Europace. 2018;20:1217. doi:10.1093/europace/euy050
  • Slotwiner D, Varma N, Akar JG, et al. HRS expert consensus statement on remote interrogation and monitoring for cardiovascular implantable electronic devices. Hear Rhythm. 2015;12(7):e69–e100.
  • Bardy GH, Lee KL, Mark DB, et al. Amiodarone or an implantable cardioverter–defibrillator for congestive heart failure. N Engl J Med. 2005;352(3):225–237. doi:10.1056/NEJMoa043399
  • Køber L, Thune JJ, Nielsen JC, et al. Defibrillator implantation in patients with nonischemic systolic heart failure. N Engl J Med. 2016;375(13):1221–1230. doi:10.1056/NEJMoa1608029
  • Lee TC, Qian M, Mu L, et al. Association between mortality and implantable cardioverter‐defibrillators by aetiology of heart failure: a propensity‐matched analysis of the WARCEF trial. ESC Hear Fail. 2019;6(2):297–307. doi:10.1002/ehf2.12407
  • Gatzoulis KA, Tsiachris D, Arsenos P, et al. Arrhythmic risk stratification in post-myocardial infarction patients with preserved ejection fraction: the PRESERVE EF study. Eur Heart J. 2019;40(35):2940–2949. doi:10.1093/eurheartj/ehz260
  • Gatzoulis KA, Georgopoulos S, Antoniou CK, et al. Programmed ventricular stimulation predicts arrhythmic events and survival in hypertrophic cardiomyopathy. Int J Cardiol. 2018;254:175–181. doi:10.1016/j.ijcard.2017.10.033
  • Gatzoulis KA, Dilaveris P, Arsenos P, et al. Arrhythmic risk stratification in nonischemic dilated cardiomyopathy: the ReCONSIDER study design – A two-step, multifactorial, electrophysiology-inclusive approach. Hell J Cardiol. 2020. doi:10.1016/j.hjc.2020.03.008
  • Selvanayagam JB, Hartshorne T, Billot L, et al. Cardiovascular magnetic resonance-guided management of mild to moderate left ventricular systolic dysfunction (CMR GUIDE): study protocol for a randomized controlled trial. Ann Noninvasive Electrocardiol. 2017;22(4):12420. doi:10.1111/anec.12420
  • Salden OAE, Vernooy K, van Stipdonk AMW, Cramer MJ, Prinzen FW, Meine M. Strategies to improve selection of patients without typical left bundle branch block for cardiac resynchronization therapy. JACC Clin Electrophysiol. 2020;6(2):129–142. doi:10.1016/j.jacep.2019.11.018
  • Halbfass P, Sonne K, Nentwich K, Ene E, Deneke T. Current developments in cardiac rhythm management devices. Clin Res Cardiol. 2018;107(S2):100–104. doi:10.1007/s00392-018-1313-4
  • Leclercq C, Burri H, Curnis A, et al. Rationale and design of a randomized clinical trial to assess the safety and efficacy of multipoint pacing therapy: more response on cardiac resynchronization therapy with multipoint pacing (MORE-CRT MPP-PHASE II). Am Heart J. 2019;209:1–8. doi:10.1016/j.ahj.2018.12.004
  • Dilaveris P, Antoniou CK, Manolakou P, et al. Comparison of left ventricular and biventricular pacing: rationale and clinical implications. Anatol J Cardiol. 2019;22(3):132–139.
  • Filippatos G, Birnie D, Gold MR, et al. Rationale and design of the adapt response trial: a prospective randomized study of cardiac resynchronization therapy with preferential adaptive left ventricular-only pacing. Eur J Heart Fail. 2017;19(7):950–957. doi:10.1002/ejhf.895
  • Zanon F, Abdelrahman M, Marcantoni L, et al. Long term performance and safety of His bundle pacing: a multicenter experience. J Cardiovasc Electrophysiol. 2019;30(9):1594–1601. doi:10.1111/jce.14063
  • Sharma PS, Vijayaraman P, Ellenbogen KA. Permanent His bundle pacing: shaping the future of physiological ventricular pacing. Nat Rev Cardiol. 2020;17(1):22–36. doi:10.1038/s41569-019-0224-z
  • Upadhyay GA, Vijayaraman P, Nayak HM, et al. On-treatment comparison between corrective His bundle pacing and biventricular pacing for cardiac resynchronization: a secondary analysis of the His-SYNC Pilot Trial. Hear Rhythm. 2019;16(12):1797–1807. doi:10.1016/j.hrthm.2019.05.009
  • Boriani G, Fauchier L, Aguinaga L, et al. European Heart Rhythm Association (EHRA) consensus document on management of arrhythmias and cardiac electronic devices in the critically ill and post-surgery patient, endorsed by Heart Rhythm Society (HRS), Asia Pacific Heart Rhythm Society (APHRS), Card. EP Eur. 2018;21(1):7–8.
  • Mairesse GH, Braunschweig F, Klersy K, Cowie MR, Leyva F. Implementation and reimbursement of remote monitoring for cardiac implantable electronic devices in Europe: a survey from the health economics committee of the European Heart Rhythm Association. EP Eur. 2015;17(5):814–818.
  • Poole JE, Gleva MJ, Mela T, et al. Complication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures: results from the REPLACE registry. Circulation. 2010;122(16):1553–1561. doi:10.1161/CIRCULATIONAHA.110.976076
  • Palmisano P, Guerra F, Dell’Era G, et al. Impact on all-cause and cardiovascular mortality of cardiac implantable electronic device complications: results from the POINTED registry. JACC Clin Electrophysiol. 2020;6(4):382–392. doi:10.1016/j.jacep.2019.11.005
  • Blomström-Lundqvist C, Traykov V, Erba PA, et al. European Heart Rhythm Association (EHRA) international consensus document on how to prevent, diagnose, and treat cardiac implantable electronic device infections—endorsed by the Heart Rhythm Society (HRS), the Asia Pacific Heart Rhythm Society (APHRS), th. EP Eur. 2019;22(4):515–549.
  • Tarakji KG, Mittal S, Kennergren C, et al. Antibacterial envelope to prevent cardiac implantable device infection. N Engl J Med. 2019;380(20):1895–1905. doi:10.1056/NEJMoa1901111
  • Sideris S, Archontakis S, Dilaveris P, et al. Leadless cardiac pacemakers: current status of a modern approach in pacing. Hell J Cardiol. 2017;58(6):403–410. doi:10.1016/j.hjc.2017.05.004
  • Wiles BM, Roberts PR. Design and evaluation of the micra transcatheter pacing system for bradyarrhythmia management. Future Cardiol. 2019;15(1):9–15. doi:10.2217/fca-2018-0077
  • Sperzel J, Hamm C, Hain A. Nanostim—leadless pacemaker. Herzschrittmachertherapie und Elektrophysiologie. 2018;29(4):327–333. doi:10.1007/s00399-018-0598-3
  • Groner A, Grippe K. The leadless pacemaker. J Am Acad Physician Assist. 2019;32(6):48–50. doi:10.1097/01.JAA.0000554750.85170.d4
  • Steinwender C, Lercher P, Schukro C, et al. State of the art: leadless ventricular pacing: a national expert consensus of the Austrian Society of Cardiology. J Interv Card Electrophysiol. 2020;57(1):27–37. doi:10.1007/s10840-019-00680-2
  • Afzal MR, Daoud EG, Cunnane R, et al. Techniques for successful early retrieval of the Micra transcatheter pacing system: a worldwide experience. Hear Rhythm. 2018;15(6):841–846. doi:10.1016/j.hrthm.2018.02.008
  • Reynolds D, Duray GZ, Omar R, et al. A leadless intracardiac transcatheter pacing system. N Engl J Med. 2016;374(6):533–541. doi:10.1056/NEJMoa1511643
  • El-Chami MF, Al-Samadi F, Clementy N, et al. Updated performance of the Micra transcatheter pacemaker in the real-world setting: a comparison to the investigational study and a transvenous historical control. Hear Rhythm. 2018;15(12):1800–1807. doi:10.1016/j.hrthm.2018.08.005
  • El‐Chami MF, Johansen JB, Zaidi A, et al. Leadless pacemaker implant in patients with pre‐existing infections: results from the Micra postapproval registry. J Cardiovasc Electrophysiol. 2019;30(4):569–574. doi:10.1111/jce.13851
  • Piccini JP, Stromberg K, Jackson KP, et al. Patient selection, pacing indications, and subsequent outcomes with de novo leadless single-chamber VVI pacing. Europace. 2019;21(11):1686–1693. doi:10.1093/europace/euz230
  • Garg A, Koneru JN, Fagan DH, et al. Morbidity and mortality in patients precluded for transvenous pacemaker implantation: experience with a leadless pacemaker. Hear Rhythm. 2020. doi:10.1016/j.hrthm.2020.07.035
  • Tachibana M, Banba K, Matsumoto K, Ohara M. The feasibility of leadless pacemaker implantation for superelderly patients. Pacing Clin Electrophysiol. 2020;43(4):374–381. doi:10.1111/pace.13894
  • Pagan E, Gabriels J, Khodak A, et al. Safety of leadless pacemaker implantation in the very elderly. Hear Rhythm. 2020. doi:10.1016/j.hrthm.2020.05.022
  • Boveda S, Marijon E, Lenarczyk R, et al. Factors influencing the use of leadless or transvenous pacemakers: results of the European Heart Rhythm Association Prospective Survey. EP Eur. 2020;22(4):667–673.
  • Lenarczyk R, Boveda S, Mansourati J, et al. Peri-procedural management, implantation feasibility, and short-term outcomes in patients undergoing implantation of leadless pacemakers: European snapshot survey. EP Eur. 2020;22(5):833–838.
  • Bereuter L, Gysin M, Kueffer T, et al. Leadless dual-chamber pacing: a novel communication method for wireless pacemaker synchronization. JACC Basic Transl Sci. 2018;3(6):813–823. doi:10.1016/j.jacbts.2018.07.009
  • Chinitz L, Ritter P, Khelae SK, et al. Accelerometer-based atrioventricular synchronous pacing with a ventricular leadless pacemaker: results from the Micra atrioventricular feasibility studies. Hear Rhythm. 2018;15(9):1363–1371. doi:10.1016/j.hrthm.2018.05.004
  • Steinwender C, Khelae SK, Garweg C, et al. Atrioventricular synchronous pacing using a leadless ventricular pacemaker: results from the MARVEL 2 study. JACC Clin Electrophysiol. 2020;6(1):94–106.
  • Medtronic MicraTM AV MC1AVR1 device manual. Available from: https://www.medtronic.com/us-en/healthcare-professionals/products/cardiac-rhythm/pacemakers/micra-pacing-system.html.
  • Reddy VY, Miller MA, Neuzil P, et al. Cardiac resynchronization therapy with wireless left ventricular endocardial pacing: the SELECT-LV study. J Am Coll Cardiol. 2017;69(17):2119–2129. doi:10.1016/j.jacc.2017.02.059
  • Sieniewicz BJ, Betts TR, James S, et al. Real-world experience of leadless left ventricular endocardial cardiac resynchronization therapy: a multicenter international registry of the WiSE-CRT pacing system. Hear Rhythm. 2020;17(8):1291–1297. doi:10.1016/j.hrthm.2020.03.002
  • Singh JP, Abraham WT, Auricchio A, et al. Design and rationale for the stimulation of the left ventricular endocardium for cardiac resynchronization therapy in non-responders and previously untreatable patients (SOLVE-CRT) trial. Am Heart J. 2019;217:13–22. doi:10.1016/j.ahj.2019.04.002
  • Olgin JE, Pletcher MJ, Vittinghoff E, et al. Wearable cardioverter–defibrillator after myocardial infarction. N Engl J Med. 2018;379(13):1205–1215. doi:10.1056/NEJMoa1800781
  • Olgin JE, Lee BK, Vittinghoff E, et al. Impact of wearable cardioverter‐defibrillator compliance on outcomes in the VEST trial: as‐treated and per‐protocol analyses. J Cardiovasc Electrophysiol. 2020;31(5):1009–1018. doi:10.1111/jce.14404
  • Kutyifa V, Moss AJ, Klein H, et al. Use of the wearable cardioverter defibrillator in high-risk cardiac patients data from the prospective registry of patients using the wearable cardioverter defibrillator (WEARIT-II registry). Circulation. 2015;132(17):1613–1619. doi:10.1161/CIRCULATIONAHA.115.015677
  • Lenarczyk R, Potpara TS, Haugaa KH, et al. The use of wearable cardioverter-defibrillators in Europe: results of the European Heart Rhythm Association survey. EP Eur. 2016;18(1):146–150.
  • Boersma LV, El-Chami MF, Bongiorni MG, et al. Understanding outcomes with the EMBLEM S-ICD in primary prevention patients with low EF study (UNTOUCHED): clinical characteristics and perioperative results. Hear Rhythm. 2019;16(11):1636–1644. doi:10.1016/j.hrthm.2019.04.048
  • Boersma L, Burke MC, Neuzil P, et al. Infection and mortality after implantation of a subcutaneous ICD after transvenous ICD extraction. Hear Rhythm. 2016;13(1):157–164. doi:10.1016/j.hrthm.2015.08.039
  • Boveda S, Lenarczyk R, Fumagalli S, et al. Factors influencing the use of subcutaneous or transvenous implantable cardioverter-defibrillators: results of the European Heart Rhythm Association prospective survey. EP Eur. 2018;20(5):887–892.
  • Lenarczyk R, Boveda S, Haugaa KH, et al. Peri-procedural routines, implantation techniques, and procedure-related complications in patients undergoing implantation of subcutaneous or transvenous automatic cardioverter-defibrillators: results of the European Snapshot Survey on S-ICD Implantation. EP Eur. 2018;20(7):1218–1224.
  • Kaya E, Rassaf T, Wakili R. Subcutaneous ICD: current standards and future perspective. IJC Hear Vasc. 2019;24.
  • Boersma LVA, Merkely B, Neuzil P, et al. Therapy from a novel substernal lead: the ASD2 study. JACC Clin Electrophysiol. 2019;5(2):186–196. doi:10.1016/j.jacep.2018.11.003
  • Yee R, Verma A, Beardsall M, Fraser J, Philippon F, Exner DV. Canadian Cardiovascular Society/Canadian Heart Rhythm Society Joint Position Statement on the use of remote monitoring for cardiovascular implantable electronic device follow-up. Can J Cardiol. 2013;29(6):644–651. doi:10.1016/j.cjca.2012.11.036
  • Frederix I, Caiani EG, Dendale P, et al. ESC e-Cardiology Working Group Position Paper: overcoming challenges in digital health implementation in cardiovascular medicine. Eur J Prev Cardiol. 2019;26(11):1166–1177. doi:10.1177/2047487319832394
  • Liberska A, Kowalski O, Mazurek M, et al. Day by day telemetric care of patients treated with cardiac resynchronisation therapy: first Polish experience. Kardiol Pol. 2016;74:741–748. doi:10.5603/KP.a2016.0019
  • Hohnloser HS, Capucci A, Fain E, et al. ASymptomatic atrial fibrillation and Stroke Evaluation in pacemaker patients and the atrial fibrillation Reduction atrial pacing Trial (ASSERT). Am Heart J. 2006;152(3):442–447. doi:10.1016/j.ahj.2006.02.016
  • Ziegler PD, Glotzer TV, Daoud EG, et al. Detection of previously undiagnosed atrial fibrillation in patients with stroke risk factors and usefulness of continuous monitoring in primary stroke prevention. Am J Cardiol. 2012;110(9):1309–1314. doi:10.1016/j.amjcard.2012.06.034
  • Jȩdrzejczyk-Patej E, Lenarczyk R, Mazurek M, et al. Can we rely on machines? Device-detected atrial high rates correspond well with atrial arrhythmias in cardiac resynchronization recipients. Europace. 2016;18(3):436–444. doi:10.1093/europace/euv095
  • Gorenek B, Bax J, Boriani G, et al. Device-detected subclinical atrial tachyarrhythmias: definition, implications and management. EP Eur. 2017;19(9):1556–1578.
  • Martin DT, Bersohn MM, Waldo AL, et al. Randomized trial of atrial arrhythmia monitoring to guide anticoagulation in patients with implanted defibrillator and cardiac resynchronization devices. Eur Heart J. 2015;36:1660–1668. doi:10.1093/eurheartj/ehv115
  • Miyazawa K, Kondo Y, Nakano M, et al. Risk factors for the development of incident atrial fibrillation in patients with cardiac implantable electronic devices. Eur J Intern Med. 2018;52:54–59. doi:10.1016/j.ejim.2018.02.019
  • Kirchhof P, Benussi S, Kotecha D, et al. 2016 ESC guidelines for the management of atrial fibrillation developed in collaboration with EACTS. Eur Heart J. 2016;37(38):2893–2962.
  • Mairesse GH, Moran P, Van Gelder IC, et al. Screening for atrial fibrillation: a European Heart Rhythm Association (EHRA) consensus document endorsed by the Heart Rhythm Society (HRS), Asia Pacific Heart Rhythm Society (APHRS), and Sociedad Latinoamericana de Estimulación Cardíaca y Electrofisiolog. Europace. 2017;19(10):1589–1623. doi:10.1093/europace/eux177
  • Khan AA, Boriani G, Lip GYH. Are atrial high rate episodes (AHREs) a precursor to atrial fibrillation? Clin Res Cardiol. 2020;109(4):409–416. doi:10.1007/s00392-019-01545-4
  • Miyazawa K, Pastori D, Li YG, et al. Atrial high rate episodes in patients with cardiac implantable electronic devices: implications for clinical outcomes. Clin Res Cardiol. 2019;108(9):1034–1041. doi:10.1007/s00392-019-01432-y
  • Li YG, Miyazawa K, Pastori D, Szekely O, Shahid F, Lip GYH. Atrial high-rate episodes and thromboembolism in patients without atrial fibrillation: the West Birmingham atrial fibrillation project. Int J Cardiol. 2019;292:126–130. doi:10.1016/j.ijcard.2019.04.055
  • Pastori D, Miyazawa K, Li Y, et al. Atrial high-rate episodes and risk of major adverse cardiovascular events in patients with cardiac implantable electronic devices. Clin Res Cardiol. 2020;109(1):96–102. doi:10.1007/s00392-019-01493-z
  • Boriani G, Lip GYH, Pietro R, et al. The increased risk of stroke/transient ischemic attack in women with a cardiac implantable electronic device is not associated with a higher atrial fibrillation burden. Europace. 2017;19(11):1767–1775. doi:10.1093/europace/euw333
  • Lorenzoni G, Folino F, Soriani N, Iliceto S, Gregori D. Cost-effectiveness of early detection of atrial fibrillation via remote control of implanted devices. J Eval Clin Pract. 2014;20(5):570–577.
  • Varma N, Epstein AE, Irimpen A, Schweikert R, Love C. Efficacy and safety of automatic remote monitoring for implantable cardioverter-defibrillator follow-up. Circulation. 2010;122(4):325–332. doi:10.1161/CIRCULATIONAHA.110.937409
  • Amara W, Montagnier C, Cheggour S, et al. Early detection and treatment of atrial arrhythmias alleviates the arrhythmic burden in paced patients: the SETAM study. Pacing Clin Electrophysiol. 2017;40(5):527–536. doi:10.1111/pace.13062
  • Perino A, Fan J, Askari M, et al. How much atrial fibrillation is too much? Treatment benefit of anticoagulation based on threshold of device-detected Af. J Am Coll Cardiol. 2019;73(9):290. doi:10.1016/S0735-1097(19)30898-8
  • Verma A, Cairns JA, Mitchell LB, et al. 2014 focused update of the Canadian cardiovascular society guidelines for the management of atrial fibrillation. Can J Cardiol. 2014;30(10):1114–1130.
  • Lip GYH, Banerjee A, Boriani G, et al. Antithrombotic therapy for atrial fibrillation: CHEST guideline and expert panel report. Chest. 2018;154(5):1121–1201. doi:10.1016/j.chest.2018.07.040
  • Kirchhof P, Blank BF, Calvert M, et al. Probing oral anticoagulation in patients with atrial high rate episodes: rationale and design of the Non–vitamin K antagonist Oral anticoagulants in patients with Atrial High rate episodes (NOAH–AFNET 6) trial. Am Heart J. 2017;190:12–18. doi:10.1016/j.ahj.2017.04.015
  • Lopes RD, Alings M, Connolly SJ, et al. Rationale and design of the Apixaban for the reduction of thrombo-embolism in patients with device-detected sub-clinical atrial fibrillation (ARTESiA) trial. Am Heart J. 2017;189:137–145. doi:10.1016/j.ahj.2017.04.008
  • Marrouche NF, Brachmann J, Andresen D, et al. Catheter ablation for atrial fibrillation with heart failure. N Engl J Med. 2018;378(5):417–427. doi:10.1056/NEJMoa1707855
  • Noseworthy PA, Van Houten HK, Gersh BJ, et al. Generalizability of the CASTLE-AF trial: catheter ablation for patients with atrial fibrillation and heart failure in routine practice. Hear Rhythm. 2020;17(7):1057–1065. doi:10.1016/j.hrthm.2020.02.030
  • Packer DL, Mark DB, Robb RA, et al. Effect of catheter ablation vs antiarrhythmic drug therapy on mortality, stroke, bleeding, and cardiac arrest among patients with atrial fibrillation: the CABANA randomized clinical trial. JAMA. 2019;321:1261–1274. doi:10.1001/jama.2019.0693
  • Noseworthy PA, Gersh BJ, Kent DM, et al. Atrial fibrillation ablation in practice: assessing CABANA generalizability. Eur Heart J. 2019;40(16):1257–1264. doi:10.1093/eurheartj/ehz085
  • Ontario HQ. Remote monitoring of implantable cardioverter-defibrillators, cardiac resynchronization therapy and permanent pacemakers: a health technology assessment. Ont Health Technol Assess Ser. 2018;18(7):1–199.
  • Hindricks G, Taborsky M, Glikson M, et al. Implant-based multiparameter telemonitoring of patients with heart failure (IN-TIME): a randomised controlled trial. Lancet. 2020;384(9943):583–590. doi:10.1016/S0140-6736(14)61176-4
  • Sardu C, Santamaria M, Rizzo MR, et al. Telemonitoring in heart failure patients treated by cardiac resynchronisation therapy with defibrillator (CRT-D): the TELECART study. Int J Clin Pract. 2016;70(7):569–576. doi:10.1111/ijcp.12823
  • Boriani G, Da Costa A, Quesada A, et al. Effects of remote monitoring on clinical outcomes and use of healthcare resources in heart failure patients with biventricular defibrillators: results of the MORE-CARE multicentre randomized controlled trial. Eur J Heart Fail. 2017;19(3):416–425. doi:10.1002/ejhf.626
  • Morgan JM, Kitt S, Gill J, et al. Remote management of heart failure using implantable electronic devices. Eur Heart J. 2017;38(30):2352–2360. doi:10.1093/eurheartj/ehx227
  • Tajstra M, Sokal A, Gadula-Gacek E, et al. Remote supervision to decrease hospitalization rate (RESULT) study in patients with implanted cardioverter-defibrillator. EP Eur. 2020;22(5):769–776.
  • Braunschweig F, Anker SD, Proff J, Varma N. Remote monitoring of implantable cardioverter-defibrillators and resynchronization devices to improve patient outcomes: dead end or way ahead? EP Eur. 2019;21(6):846–855.
  • Goette A, Auricchio A, Boriani G, et al. EHRA White Paper: knowledge gaps in arrhythmia management - Status 2019. Europace. 2019;21(7):993–994.
  • Brignole M, Moya A, de Lange FJ, et al. 2018 ESC guidelines for the diagnosis and management of syncope. Eur Heart J. 2018;39(21):1883–1948.
  • Seidl K, Rameken M, Breunung S, et al. Diagnostic assessment of recurrent unexplained syncope with a new subcutaneously implantable loop recorder. Europace. 2000;2(3):256–262. doi:10.1053/eupc.2000.0108
  • Giada F, Gulizia M, Francese M, et al. Recurrent Unexplained Palpitations (RUP) study. Comparison of implantable loop recorder versus conventional diagnostic strategy. J Am Coll Cardiol. 2007;49(19):1951–1956. doi:10.1016/j.jacc.2007.02.036
  • Sanna T, Diener H-C, Passman RS, et al. Cryptogenic stroke and underlying atrial fibrillation. N Engl J Med. 2014;370(26):2478–2486. doi:10.1056/NEJMoa1313600
  • Diederichsen SZ, Haugan KJ, Brandes A, et al. Incidence and predictors of atrial fibrillation episodes as detected by implantable loop recorder in patients at risk: from the LOOP study. Am Heart J. 2020;219:117–127. doi:10.1016/j.ahj.2019.09.009
  • Diederichsen SZ, Haugan KJ, Brandes A, et al. Natural history of subclinical atrial fibrillation detected by implanted loop recorders. J Am Coll Cardiol. 2019;74(22):2771–2781. doi:10.1016/j.jacc.2019.09.050
  • Ip JE. Evaluation of cardiac rhythm abnormalities from wearable devices. JAMA. 2019;321(11):1098–1099. doi:10.1001/jama.2019.1681
  • Dörr M, Nohturfft V, Brasier N, et al. The WATCH AF trial: smartWATCHes for detection of atrial fibrillation. JACC Clin Electrophysiol. 2019;5(2):199–208. doi:10.1016/j.jacep.2018.10.006
  • Turakhia MP, Desai M, Hedlin H, et al. Rationale and design of a large-scale, app-based study to identify cardiac arrhythmias using a smartwatch: the Apple Heart Study. Am Heart J. 2019;207:66–75. doi:10.1016/j.ahj.2018.09.002
  • Perez MV, Mahaffey KW, Hedlin H, et al. Large-scale assessment of a smartwatch to identify atrial fibrillation. N Engl J Med. 2019;381(20):1909–1917. doi:10.1056/NEJMoa1901183
  • Guo Y, Wang H, Zhang H, et al. Mobile Photoplethysmographic technology to detect atrial fibrillation. J Am Coll Cardiol. 2019;74(19):2365–2375. doi:10.1016/j.jacc.2019.08.019
  • Guo Y, Lane DA, Wang L, et al. Mobile Health (mHealth) technology for improved screening, patient involvement and optimising integrated care in atrial fibrillation: the mAFA (mAF‐App) II randomised trial. Int J Clin Pract. 2019;73(7). doi:10.1111/ijcp.13352.
  • Guo Y, Lane DA, Chen Y, Lip GYH. Regular bleeding risk assessment associated with reduction in bleeding outcomes: the mAFA-II randomized trial. Am J Med. 2020. doi:10.1016/j.amjmed.2020.03.019
  • Kalarus Z, Balsam P, Bandosz P, et al. NOninvasive monitoring for early detection of Atrial fibrillation: rationale and design of the NOMED-AF study. Kardiol Pol. 2018;76(10):1482–1485. doi:10.5603/KP.a2018.0193
  • Pevnick JM, Birkeland K, Zimmer R, Elad Y, Kedan I. Wearable technology for cardiology: an update and framework for the future. Trends Cardiovasc Med. 2018;28(2):144–150. doi:10.1016/j.tcm.2017.08.003
  • Uslan DZ, Sohail MR, St Sauver JL, et al. Permanent pacemaker and implantable cardioverter defibrillator infection: a population-based study. Arch Intern Med. 2007;167(7):669–675. doi:10.1001/archinte.167.7.669