Review of Evidence Supporting the Arista™ Absorbable Powder Hemostat

Pages 173-188 | Received 22 Nov 2023, Accepted 21 Mar 2024, Published online: 23 May 2024

Abstract

Background

Uncontrolled and diffuse bleeding is a dreaded event during open and laparoscopic surgery that may lead to postoperative complications, obstruction of the surgical field that reduces visualization, and prolonged operating times. Powder hemostats can be used to control bleeding and are easy to use, have a safe profile, and can achieve broad coverage area at a low cost.

Methods

A strategic literature search of peer-reviewed, English language studies was conducted to capture evidence on the clinical efficacy and safety of a Microporous Polysaccharide Hemosphere (MPH) based Hemostat (Arista™ Absorbable Hemostat (Arista™ AH)).

Results

Six preclinical studies were found which supported the use of MPH in various animal models of laparoscopic and open surgery, all of which demonstrated its safety and efficacy. Five single-arm and 11 comparative clinical studies similarly supported the efficacy and safety of MPH in various surgery types, including cardiac, renal, and dermatologic surgery.

Conclusion

Published evidence supports the safe and effective use of MPH across a variety of surgical settings.

Introduction

Uncontrolled and diffuse bleeding is a dreaded event during open and laparoscopic surgery and may lead to substantial morbidity and mortality.Citation1,Citation2 Bleeding-related complications can be more common than thought, with one study reporting rates of up to 47.4% during cardiac surgery.Citation1 Further, uncontrolled bleeding can prolong hospital stay and result in significant additional medical interventions and costs.Citation3–5 Accordingly, effective hemostasis can improve surgical outcomes, reduce complications and the need for surgical re-intervention, and lower costs of medical care.Citation6,Citation7

The selection of optimal hemostatic agent depends on a variety of factors. These factors include surgery type, source and the extent of bleeding, as well as the patient’s coagulation status and medical history.Citation8,Citation9 Hemostats should meet basic clinical needs, which include efficacy, providing reliable and quick hemostasis, and having a preferable safety profile.Citation9–11 Additional considerations in hemostat selection include ease of storage, required preparation time/being ready on demand, and cost. For control of localized bleeding, application of direct pressure and thermal-based methods (eg, electrosurgery) are typically the methods of choice due to their simplicity.Citation11,Citation12 However, these methods of hemostasis may not be effective in controlling diffuse bleeding, particularly when the site of bleeding is not visible.Citation8,Citation13 Diffuse bleeding during surgery is associated with several complications, including reduction in core temperature, thrombocytopenia, and hypovolemic shock.Citation8 Further, diffuse bleeding can visually obstruct surgical field and thereby lead to longer operating times and increased risk of patient and staff injury.Citation8

Given the unique challenges associated with diffuse bleeding, powder hemostats that can be easily and quickly applied over a large surface area provide advantages over other hemostat types in controlling diffuse bleeding in both open and laparoscopic surgery.Citation2,Citation11,Citation14 However, despite their frequent use in many surgical settings, there is a paucity of reviews focused specifically on powder hemostats for control of surgical bleeding, and factors influencing the selection of the optimal hemostat.

The Microporous Polysaccharide Hemosphere (MPH) based hemostat (Arista™ Absorbable Hemostatic Particles (Arista™ AH), Becton, Dickinson and Company (BD), Franklin Lakes, NJ) was the first flowable powder hemostatic agent available on the market in the United States (US). This review will summarize the available clinical and preclinical evidence for MPH and evaluate its safety and efficacy versus other available powder hemostatic agents in controlling bleeding.Citation8

Methods

A targeted literature review was conducted on April 30th, 2020 by an information specialist for English peer-reviewed articles evaluating the use of MPH and other powder hemostats in both preclinical and clinical settings. MEDLINE®, Embase, EMB Reviews, and Cochrane Central Register of Controlled Trials were searched for relevant articles. The goal was to find evidence demonstrating the safety and efficacy of MPH or other powder hemostats across surgery types. Search included key terms such as hemostatic agent, hemostasis, hemostatic technique, microporous or degradable.

Results

Overview of Powder Hemostats

Powder hemostats belong to the category of mechanical hemostats, which act as a molecular sieve through contact with the bleeding site and promote platelet aggregation.Citation8–10 These agents are used during surgery to control diffuse raw surface bleeding, hard to reach bleeding sites, oozing venous bleeding, bone bleeding, and needle-hole bleeding. Mechanical hemostats include agents derived from animal-based sources (collagen and gelatin), and those derived from plant sources (oxidized regenerative cellulose [ORC; Surgicel®] and MPH). Furthermore, mechanical hemostats are available in several forms including sponges, pads, foam, and powder.Citation8 In the US, MPH was the first flowable powder hemostat on the market and was approved for use in 2006. ORC was initially available as a fabric and has recently become available in the powder form. PerClot® is a modified starch powder hemostat that is available internationally and is expected to be submitted for approval for use in the US. Other powder hemostats that are available internationally include starch-based hemostats like 4DryField®, HaemoCer™, and StarSil®, and collagen-based Hemoblast.

Powder hemostats can be easily dispersed over a large surface area and are therefore commonly used to control diffuse bleeding in open and laparoscopic surgeries.Citation13,Citation15 Benefits of powder hemostats over other hemostat types (eg, flowables and sealants) include the following:

  • Simple application that does not require a long-learning curve

  • Control of bleeding at technically difficult-to-reach bleeding sites

  • Control of bleeding over a large surface area and when the exact site of bleeding cannot be identified

  • Compatible with autologous cell salvage machines

  • Safe and well tolerated, without any known toxicity

  • Effective in patients taking anticoagulants or anti-thrombotic medication

  • Fast-acting and achieve rapid control of bleeding in high-risk patients

  • Cost-effective

Although all powder hemostats offer benefits, these agents differ in terms of their characteristics, safety, efficacy, and ease of use. These characteristics are summarized in : Characteristics of mechanical hemostats approved for use in the US. The differentiating features of plant-based powder hemostats are discussed in the following section.Citation16

Table 1 Characteristics of Mechanical Hemostats Approved for Use in the US

Plant-Based Powder Hemostats

Plant-derived powder hemostats, which include starch-based (MPH) and cellulose-based (ORC) products, are commonly used as adjunctive hemostats due to their ease of use and efficacy.Citation16,Citation19 Plant-derived hemostats may be beneficial in some surgical settings because they do not contain any animal- or human-derived components, which can cause adverse reactions in some patients.Citation20,Citation21 Among plant-derived products, ORC, a cellulose-based hemostat, is well known and accepted because of its ease of use, and bactericidal properties*.Citation16 Although ORC is bactericidal against a wide range of pathogenic microorganisms, it is not intended as a substitute for systemically administered therapeutic or prophylactic antimicrobial agents to control or to prevent postoperative infections.Citation22 However, despite its benefits, ORC may be associated with safety issues. For example, ORC is not fully absorbed for two to five weeks after surgery.Citation16 The material remaining at the surgical site may create granulation formation in the late post-operative period and result in foreign body reactions. Case studies have reported complications related to excess ORC at the surgical site, including adhesion formation and hematoma around ORC after cardiac surgeryCitation23,Citation24 and paraplegia following thoracic surgery.Citation25 Additionally, the low pH of ORC contributes to its anti-bactericidal properties.Citation26 However, the low pH also increases the inflammation of the surrounding tissue and has been reported to delay wound healing.Citation27 Further, studies have reported that excess ORC may interfere with postsurgical diagnostic investigations, given that commonly used imaging methods (eg, CT or nuclear magnetic resonance) cannot differentiate excess ORC from pathological growth.Citation28,Citation29 The increased risk of complications associated with non-absorbed ORC is a known risk of the product and is included in the product’s Instructions for Use.Citation30

Starch Crystallinity as a Consideration in the Selection of Starch-Based Hemostats

Although all starch-based hemostats share certain features, they can be differentiated based on the unique properties of the starch they are derived from.Citation3,Citation21,Citation31 Starch is a semi-crystalline polymer consisting of amorphous (ie, non-crystalline) regions, which are easily degraded by amylases, and crystalline regions, which are degraded more slowly.Citation32 Therefore, starches with a higher amount of crystallinity are not digested or resorbed as easily as those with less crystallinity.Citation33 Complications due to surgical gloves coated with starch powder are well documented and indicate an inflammatory response caused by the crystalline starch powder resulting in foreign body reactions, adhesion formation, and granulomatous reactions.Citation33–36 These concerns over inflammatory reactions have led several regulatory agencies to ban powdered surgical gloves.Citation37

Although crystallinity of starch powder hemostats has not been well documented, one study showed that starch crystallinity was observed with PerClot®, HaemoCer™, and Starsil®, but not Arista™ AH.Citation35 Note that at the time of publication, PerClot®, HaemoCer™, and Starsil® are not available in the US. Starch crystallinity was evaluated by X-ray powder diffraction in non-degraded state and after 24-hour degradation with α-amylase using polarized light microscopy. As indicated in Nilsson and Gold, the crystalline granules were observed after exposure to in vitro amylase degradation for 24 hours, indicating that they are resistant to resorption.Citation35 These results were recently confirmed and expanded upon in a comparison of Arista™ AH and PerClot®.Citation38 In both an in vitro degradation assay and an in vivo preclinical model, Arista™ AH showed no crystallinity at any time point, no material detected by Day 3 and normal wound healing. In contrast, PerClot® PHS showed slow resorption and the local persistence out to the Day 28 endpoint of crystalline degradation products that were associated with the onset of chronic inflammation and an early foreign body response.

Advantages of AristaTM AH Absorbable Hemostatic Particles

Arista™ AH consists of Microporous Polysaccharide Hemospheres (MPH), which are derived from purified potato starch and are currently used as an absorbable hemostatic agent. MPH is delivered as a flowable powder engineered to rapidly dehydrate blood, enhancing clotting on contact. MPH provides hemostasis by absorbing water and low-molecular weight compounds from the blood to concentrate blood solids, creating a scaffold for the formation of fibrin clot.Citation39,Citation40 The clot formed by the swelling of MPH beads, platelets, and clotting proteins is more resilient than a naturally formed clot. Preclinical studies show that the MPH-enhanced clot is enzymatically broken down within 12 hours of application, leaving no residue in the surgical field.Citation40 Clinical studies show that MPH is absorbed from the surgical field 24–48 hours postoperatively.Citation41,Citation42 Notably, the MPH mechanism of action does not depend upon the patient’s coagulation status, making it suitable for patients on anti-coagulation therapy across a variety of surgical settings.

The MPH particles in Arista™ AH are made through the reaction of purified starch and epichlorohydrin and were developed to provide a biocompatible, purified starch composition.Citation43

In the US, Arista™ AH is indicated in most surgical procedures as an adjunctive hemostatic device to assist when control of capillary, venous, and arteriolar bleeding by pressure, ligature, and other conventional procedures are ineffective or impractical, or when added hemostasis is necessary.

Preclinical studies have shown that the properties and mechanisms of action of Arista™ AH are associated with several benefits:

  • Due to fast absorption (within 24 hours), Arista™ AH does not interfere with imagingCitation43

  • It is associated with a minimal, if any inflammatory responseCitation44

  • The starch in the proprietary MPH technology does not exhibit starch crystallinityCitation45

  • Arista™ AH powder is ready on demand to address diffuse surgical bleeding, with no preparation needed as an adjunct to conventional hemostasisCitation42

  • Arista™ AH provides broad area coverage on rough surfaces and hard-to-reach areas when adjunctive hemostasis is requiredCitation2

  • Complete hemostasis is achieved within minutesCitation46,Citation47

  • It is not associated with adhesion formationCitation48–50

  • Safety and efficacy have been demonstrated across a variety of surgical areas, including general surgery, obstetrics/gynecology, urology, cardiac, and orthopedic surgeryCitation2,Citation39,Citation41,Citation51–65

However, it should be noted that some of this data was generated in preclinical models and may not correlate to performance in humans. Preclinical and clinical evidence supporting the use of MPH in a variety of surgical settings are detailed in the following sections.

Preclinical Evidence of MPH

Studies in animal models generally show that MPH effectively achieves hemostasis with minimal complications in both laparoscopic and open surgeries. These studies, which consist of single-arm and comparative studies, are summarized in and .

Table 2 Summary of Preclinical Studies of Arista™ AH

Table 3 Comparative Animal Studies of Arista™ AH versus Surgicel®

Two single-arm studies of MPH, both in porcine models of laparoscopic surgery, demonstrated that MPH provides effective and safe hemostasis, with complete hemostasis being achieved with a single application of MPH.Citation46,Citation47 Studies that evaluated MPH versus control (no hemostat applied) also show that MPH provides effective hemostasis in a variety/multitude of surgical settings. A study in a rat model of vascular anastomosis showed that MPH significantly reduced the duration of clamping and qualitatively reduced perivascular foreign body reactions compared with control.Citation67 The study also showed that MPH did not affect vessel patency after anastomosis and did not cause toxicity to the vessel walls. Another study in a porcine model of laparoscopic renal surgery showed that MPH significantly reduced the time to hemostasis compared to control treatment (ie, compression alone), regardless of lesion size.Citation40 In a rat model of mastectomy, MPH significantly reduced seroma volume, amount of fibrous tissue, as well as albumin, lactate dehydrogenase (LDH) levels, and white blood cell count, indicating that MPH may reduce seroma volume after mastectomy.Citation66

Animal studies that compared plant-based MPH (starch) to ORC (cellulose) and animal-based hemostats (ie, gelatin, thrombin, and collagen) are summarized in . These studies show that MPH is the only agent that is completely cleared from the surgical site within 24 hours of application.Citation71 Studies that compared MPH with ORC reported no significant differences in the amount of hemorrhage or edema volume between the two agents.Citation71,Citation72 A study that evaluated the two agents in a rat model of abdominal surgery showed that MPH was associated with significantly lower adhesion scores than controls, whereas this difference was not seen with ORC.Citation73 Another study using a rat model for neurosurgery compared MPH, ORC Avitene®, and FloSeal®. This study showed that MPH was the only agent that was not associated with granulomatous formation at any point after surgery.Citation71 Finally, a study using a porcine model for hepatic surgery showed that both MPH and ORC achieved effective hemostasis in all cases. However, in this study, time to hemostasis was better with ORC.Citation19

Clinical Evidence for MPH

Published clinical evidence for MPH supports the findings of the animal studies. Single-arm studies, which include three prospective, observational studies, and three case studies are summarized in . Case studies show that MPH provides adequate control of bleeding in cardiac and cranial surgeries.Citation51,Citation52,Citation55 Two prospective, observational studies of patients undergoing sinus surgery have also demonstrated MPH effectively controls bleeding and achieves hemostasis in as little as 30–45 seconds after application.Citation53,Citation54 In both studies, MPH was well tolerated and no patients developed allergic reactions or systemic complications. Additionally, MPH was rapidly cleared from the surgical site and was undetectable one week after surgery.Citation53 Although not indicated in the US for use in neurosurgery, one neurosurgery study showed that MPH provided effective hemostasis, which was achieved in approximately one minute of application.Citation41 Additionally, MPH was undetectable at the surgical site one day postoperatively, indicating that it is rapidly absorbed in neural tissue.

Table 4 Single-Arm Clinical Studies of Arista™ AH

Studies comparing MPH with untreated controls generally show that use of MPH significantly improves clinical outcomes in a variety of surgery types, including cardiac, nasal, general, and orthopedic surgeries. These studies are summarized in . In a retrospective study of 240 patients undergoing cardiothoracic surgery, MPH significantly reduced hemostasis time, postoperative chest tube output in the first 48 hours, and the need for postoperative blood transfusion.Citation2 In another retrospective study of patients undergoing cardiac surgery, MPH was associated with a significantly lower rate of complications including hematoma and infections compared with control.Citation56 A randomized, controlled study of 40 patients undergoing nasal surgery reported that MPH significantly reduced postoperative bleeding compared with control, with no differences in the rate of infections, edema, synechiae formation, pain, obstruction, or nasal discharge between the groups.Citation57,Citation58 In a retrospective study of patients undergoing robot-assisted prostatectomy, operating room time was shorter in the MPH group than in the control group.Citation59

Table 5 Comparative Clinical Studies of Arista™ AH

Two studies did not show differences in clinical outcomes between MPH and control treatment. In a study of patients undergoing mastectomy, there were no significant differences between the MPH and control groups in the duration or quantity of serosanguinous drainage or postoperative complications.Citation39 In a retrospective chart review of patients undergoing primary total knee arthroplasty, there were no differences in the rate of hematomas or infections between the MPH and control groups.Citation61

Comparative studies of MPH versus other hemostats are limited, with considerable variability among published studies in terms of study design, surgery type, and reported outcomes. Currently, ongoing studies are being performed to better compare Arista to other hemostatic agents. The safety and effectiveness of MPH compared with those of a gelatin sponge with or without thrombin were evaluated in a prospective, multi-center, multi-specialty, randomized, non-inferiority, controlled study.Citation42 The study included 288 patients undergoing general (n = 144), orthopedic (n = 72), or cardiac (n = 72) surgery. The proportion of patients who achieved complete hemostasis of the first treated lesion with 5 minutes (3 minutes for cardiac surgery; primary endpoint), was significantly higher in the MPH group (90.3%) than in the gelatin sponge group (80.4%; P < 0.0001). The time to achieve complete hemostasis was significantly different between the MPH and gelatin sponge groups (P = 0.003). The proportion of patients who achieved complete hemostasis within one minute was higher in the MPH group (50.3%) compared with the gelatin sponge group (32.9%).

A retrospective study of patients undergoing partial nephrectomy surgery also compared MPH to FloSeal®, a gelatin/thrombin-based hemostatic matrix.Citation62 Both agents effectively achieved hemostasis, with no differences between the groups in the cool ischemia time, intra-operative blood loss, duration of hospital stay, or postoperative complications. A prospective, double-blind study of patients undergoing sinus surgery compared MPH with Sinufoam™, a carboxylmethyl cellulose-based foam/gel and Nexfoam®, a starch-based sponge hemostat.Citation63

Comparative clinical evidence shows that there were no significant differences between MPH and FloSeal® in renal surgery or nasal dressings in sinus surgery, indicating that MPH is equally effective in these surgical settings.Citation62,Citation63 However, it should be noted that these studies enrolled a relatively small number of participants (N = 23–48) and therefore the results may not have reached statistical significance. Therefore, further studies are needed to further evaluate the comparative efficacy of MPH and other hemostats.

Economic Benefits of Using an Effective Topical Hemostatic Agent

Evidence shows that selection of optimal hemostatic agents can lead to cost savings, resulting from reduced surgery time, reduced need for surgical re-interventions, fewer blood transfusions, and fewer post-operative complications.Citation6,Citation7,Citation17,Citation68,Citation74,Citation75 Cost savings associated with these outcomes can be substantial, with one study showing that, for a US hospital that performed 245 cardiac surgeries annually, the improved outcomes associated with a topical hemostat corresponded to a net annualized saving of $1,532,896.Citation68 Another study that evaluated cost savings associated with the use of hemostats in spine surgery in the US estimated hospital savings of $2445 per surgery.Citation74 Additionally, a study that evaluated economic benefits of a topical hemostat from the perspective of the National Health Service in the United Kingdom (UK) showed overall net savings to the NHS of ₤178,283 ($228,078 USD) per 100 cardiac surgery patients who experience intraoperative bleeding requiring hemostatic therapy.Citation17

Taken collectively, results of these studies indicate that selection of the appropriate hemostatic agent can provide substantial cost savings for hospitals as a result of improved clinical outcomes.

Discussion

As detailed in this review, MPH is a safe and effective hemostatic agent in various types of surgical procedures. Although all powder hemostats are efficacious at achieving hemostasis, MPH has a well-established safety profile and therefore addresses some of the limitations associated with other hemostats. For example, ORC has been associated with foreign body reactions, adhesion formation, and hematoma.Citation23,Citation24 In contrast, animal studies show that MPH reduced adhesion formation compared to the controlCitation73 and is not associated with foreign body reactionsCitation67 or granuloma formation.Citation71

Fast absorption time is a key benefit of MPH over other powder hemostats. This feature is particularly important when postoperative imaging is required, as residual material associated with some hemostats may interfere with diagnostic imaging and cause incorrect diagnoses such as abscesses or hematomas.Citation28,Citation29 Additionally, MPH is the only starch-based powder hemostat that has been shown not to exhibit starch crystallinity, which decreases the risk of adhesion formation and granulomatous reactions. Accordingly, MPH has a well-documented safety profile and is not associated with foreign body reactions or allergic reactions.

Overall, preclinical and clinical evidence indicates that MPH is a safe and effective powder hemostat and provides solutions in a multitude of bleeding situations and surgical settings. A summary of published clinical evidence for MPH is provided in . Notably, in two studies, MPH was associated with a shorter OR time compared with control (although statistical significance was not reached in one of the studies). Although comparative clinical evidence evaluating MPH versus ORC is limited, published studies generally show that MPH is associated with equally low rates of complications, including hematoma and infection, as other topical hemostats. However, additional studies with larger numbers of participants are ongoing to determine the benefits of MPH versus other powder hemostats.

Although there are few studies that evaluated the economic value of MPH, its clinical benefits, including reduced surgery duration, reduced need for re-interventions, fewer blood transfusions, and reduced rate of complications, can provide significant economic benefits. The rapid absorption of MPH may provide further economic benefits over hemostats that do not get rapidly absorbed from the surgical site. For example, residual ORC can interfere with diagnostic imaging, with residual hemostat appearing as a pathological condition and therefore necessitate further investigation and medical interventions.Citation28,Citation29 Therefore, hemostats that are completely absorbed by the tissue may be advantageous in terms of reducing the costs of additional medical procedures.

While the benefits of MPH have been established in the literature and discussed in this review, it should be noted that there are some limitations listed in the product Instructions for Use.Citation39 Due to its primary component being potato starch, it is recommended that no more than 50g of MPH be used in patients with diabetes, though this is not unique to MPH and is similar to other starch-based hemostats. Also, the safety and effectiveness has not been fully established in neurosurgical and ophthalmic procedures.

Finally, potential economic or surgical benefits of MPH may result from a reduced need for specialized storage, which reduces preparation time and potential costs.Citation42,Citation69 A survey of 200 US registered nurses reported that hemostatic agents that require mixing or preparation prior to application (ie, Floseal®, Thrombin, Tissel®, Sugiflo®, Gelfoam® Plus, and Vitagel®) required two to six minutes of preparation time.Citation69 Therefore, agents that are ready on demand, such as MPH, may save costs as a result of reduced nursing time, as well as a reduction in wasted product (ie, product that is mixed but unused).

Conclusions

In summary, published preclinical and clinical evidence supports the use of MPH as a topical hemostatic agent in a variety of surgical settings. Use of MPH may be particularly beneficial in situations where fast absorption of the hemostat is desired, such as cases that require post-operative imaging. More trials and ongoing studies will continue to better define the benefits of MPH.

Disclosure

Dr LyBarger reports consulting fees and speaker honoraria from Becton Dickinson (BD); speaker for GSK and Seagen, outside of the current work.

Acknowledgments

This work was sponsored by Becton Dickinson (BD). Author was reimbursed for work related to this paper.

References

  • Stokes ME, Ye X, Shah M, et al. Impact of bleeding-related complications and/or blood product transfusions on hospital costs in inpatient surgical patients. BMC Health Serv Res. 2011;11(1):135. doi:10.1186/1472-6963-11-135
  • Bruckner BA, Blau LN, Rodriguez L, et al. Microporous polysaccharide hemosphere absorbable hemostat use in cardiothoracic surgical procedures. J Cardiothor Surg. 2014;9(1):134. doi:10.1186/s13019-014-0134-4
  • Schreiber MA, Neveleff DJ. Achieving hemostasis with topical hemostats: making clinically and economically appropriate decisions in the surgical and trauma settings. AORN J. 2011;94(5):S1–20. doi:10.1016/j.aorn.2011.09.018
  • Corwin HL, Gettinger A, Pearl RG, et al. The CRIT Study: anemia and blood transfusion in the critically ill--current clinical practice in the United States. Crit Care Med. 2004;32(1):39–52. doi:10.1097/01.CCM.0000104112.34142.79
  • Boucher BA, Hannon TJ. Blood management: a primer for clinicians. Pharmacotherapy. 2007;27(10):1394–1411. doi:10.1592/phco.27.10.1394
  • Tackett SM, Sugarman R, Kreuwel HT, et al. Hospital economic impact from hemostatic matrix usage in cardiac surgery. J Med Econ. 2014;17(9):670–676. doi:10.3111/13696998.2014.928638
  • Martyn D, Meckley LM, Miyasato G, et al. Variation in hospital resource use and cost among surgical procedures using topical absorbable hemostats. Clinicoecon Outcomes Res. 2015;7:567–574. doi:10.2147/CEOR.S88698
  • AORN. Management of surgical hemostasis. An independent study guide; 2013.
  • Docimo G, Filograna Pignatelli M, Ferrandes S, et al. Role of absorbable polysaccharide hemostatic powder in the prevention of bleeding and wound events after thyroid surgery. J Clin Med. 2023;12(17):5684. doi:10.3390/jcm12175684
  • Neveleff DJ. Optimizing hemostatic practices: matching the appropriate hemostat to the clinical situation. AORN J. 2012;96(5):S1–S17. doi:10.1016/j.aorn.2012.08.005
  • Patrone R, Gambardella C, Romano RM, et al. The impact of the ultrasonic, bipolar and integrated energy devices in the adrenal gland surgery: literature review and our experience. BMC Surg. 2019;18(Suppl 1):123. doi:10.1186/s12893-018-0457-5
  • Gambardella C, Clarizia G, Patrone R, et al. Advanced hemostasis in axillary lymph node dissection for locally advanced breast cancer: new technology devices compared in the prevention of seroma formation. BMC Surg. 2019;18(Suppl 1):125. doi:10.1186/s12893-018-0454-8
  • Vecchio R, Catalano R, Basile F, et al. Topical hemostasis in laparoscopic surgery. G Chir. 2016;37(6):266–270. doi:10.11138/gchir/2016.37.6.266
  • Bruckner BA, Ngo U, Ramchandani M, et al. Application techniques of a novel hemostat in cardiac operations: HEMOBLAST. J Card Surg. 2019;34(9):849–853. doi:10.1111/jocs.14171
  • Tompeck AJ, Gajdhar AUR, Dowling M, et al. A comprehensive review of topical hemostatic agents: the good, the bad, and the novel. J Trauma Acute Care Surg. 2020;88(1):E1–E21. doi:10.1097/TA.0000000000002508
  • Chiara O, Cimbanassi S, Bellanova G, et al. A systematic review on the use of topical hemostats in trauma and emergency surgery. BMC Surg. 2018;18(1):68. doi:10.1186/s12893-018-0398-z
  • Vyas KS, Saha SP. Comparison of hemostatic agents used in vascular surgery. Expert Opin Biol Ther. 2013;13(12):1663–1672. doi:10.1517/14712598.2013.848193
  • Offodile AC, Chen B, Aherrera AS, et al. Microporous polysaccharide hemospheres potentiate ischemia-induced skin flap necrosis in a murine model. Plast Reconstr Surg. 2017;139(1):59e–66e. doi:10.1097/PRS.0000000000002907
  • MacDonald MH, Wang AY, Clymer JW, et al. An in vivo comparison of the efficacy of hemostatic powders, using two porcine bleeding models. Med Devices. 2017;10:273–279. doi:10.2147/MDER.S140663
  • Korell M, Ziegler N, De Wilde RL. Use of modified polysaccharide 4DryField PH for adhesion prevention and hemostasis in gynecological surgery: a two-center observational study by second-look laparoscopy. Biomed Res. Int. 2016;2016:1. doi:10.1155/2016/3029264
  • Spotnitz WD, Burks S. State-of-The-art review: hemostats, sealants, and adhesives II: update as well as how and when to use the components of the surgical toolbox. Clin Appl Thromb Hemost. 2010;16(5):497–514. doi:10.1177/1076029610363589
  • Ethicon. SURGICEL® Powder. Absorbable hemostatic powder (oxidized regenerated cellulose). 389894R06 - Surgicel® Powder IFU (Code 3013SP)(Non-CE marked); 2018. Available from: https://hostedvl106.quosavl.com/qb/. Accessed March 26, 2024.
  • Ibrahim MF, Aps C, Young CP. A foreign body reaction to Surgicel mimicking an abscess following cardiac surgery. Eur J Cardiothorac Surg. 2002;22(3):489–90; author reply 90. doi:10.1016/S1010-7940(02)00328-7
  • Eto K, Matsumoto M, Kubo Y, Kemmochi R. Superior vena cava syndrome caused by a swollen absorbable haemostat after repair of ischaemic mitral regurgitation. J Cardiothorac Surg. 2014;9(1):1. doi:10.1186/1749-8090-9-1
  • Dogan S, Kocaeli H, Doygun M. Oxidised regenerated cellulose as a cause of paraplegia after thoracotomy: case report and review of the literature. Spinal Cord. 2005;43(7):445–447. doi:10.1038/sj.sc.3101632
  • Schonauer C, Tessitore E, Barbagallo G, et al. The use of local agents: bone wax, gelatin, collagen, oxidized cellulose. Eur Spine J. 2004;13(Suppl 1):S89–96. doi:10.1007/s00586-004-0727-z
  • Rustagi T, Patel K, Kadrekar S, Jain A. Oxidized cellulose (Surgicel) causing postoperative cauda equine syndrome. Cureus. 2017;9:7.
  • Badenes D, Pijuan L, Curull V, Sanchez-Font A. A foreign body reaction to Surgicel® in a lymph node diagnosed by endobronchial ultrasound-guided transbronchial needle aspiration. Ann Thorac Med. 2017;12(1):55–56. doi:10.4103/1817-1737.197780
  • Patane F, Zingarelli E, Verzini A, Di Summa M. Complication due to excessive use of Surgicel. Eur J Cardiothorac Surg. 2001;20(5):1034. doi:10.1016/S1010-7940(01)00967-8
  • Ethicon. SURGICEL® ORIGINAL, SURGICEL® NU-KNIT™ and SURGICEL® FIBRILLAR™ Non CE Marked IFU; 2018. Available from: https://hostedvl106.quosavl.com/qb/. Accessed March 26, 2024.
  • Berkow AE. Surgery. Cornstarch peritonitis following the trail of the surgeon’s gloves. Postgrad Med. 1976;59(1):231–236. doi:10.1080/00325481.1976.11716541
  • Woods JA, Morgan RF, Watkins FH, Edlich RF. Surgical glove lubricants: from toxicity to opportunity. J Emerg Med. 1997;15(2):209–220. doi:10.1016/S0736-4679(96)00349-6
  • Edlich RF, Long WB, Gubler DK, et al. Dangers of cornstarch powder on medical gloves: seeking a solution. Ann Plast Surg. 2009;63(1):111–115. doi:10.1097/SAP.0b013e3181ab43ae
  • Edlich RF, Long WB, Gubler KD, et al. Citizen’s Petition to Food and Drug Administration to ban cornstarch powder on medical gloves: maltese cross birefringence. Am J Emerg Med. 2009;27(2):227–235. doi:10.1016/j.ajem.2008.10.009
  • Nilsson G, Gold J Characterization of crystallinity in starch hemostatic products. Materials for Tomorrow – Chalmers Annual Materials Science Initiative Seminar. Gothenburg, Sweden; 2013.
  • Suarez-Kelly LP, Pasley WH, Clayton EJ, et al. Effect of topical microporous polysaccharide hemospheres on the duration and amount of fluid drainage following mastectomy: a prospective randomized clinical trial. BMC Cancer. 2019;19(1). doi:10.1186/s12885-019-5293-1
  • Humphreys MR, Lingeman JE, Terry C, et al. Renal injury and the application of polysaccharide hemospheres: a laparoscopic experimental model. J Endourol. 2008;22(6):1375–1381. doi:10.1089/end.2008.0008
  • Capella-Monsonís H, Shridhar A, Chirravuri B. A comparative study of the resorption and immune response for two starch-based hemostat powders. J Surg Res. 2023;282:210–224. doi:10.1016/j.jss.2022.09.022
  • Bard Davol Inc. AristaTM AH Instructions for Use. PK3798864; 2015.
  • Ji X, Xing C, Shi X United States Patent. Patent No. US8,575,132 B2. Nov. 5, 2013; 2013.
  • Singh RK, Baumgartner B, Mantei JR, et al. Hemostatic comparison of a polysaccharide powder and a gelatin powder. J Invest Surg. 2019;32(5):393–401. doi:10.1080/08941939.2017.1423421
  • Murat FJ, Le CQ, Ereth MH, et al. Evaluation of microporous polysaccharide hemospheres for parenchymal hemostasis during laparoscopic partial nephrectomy in the porcine model. JSLS. 2006;10(3):302–306.
  • Humphreys MR, Castle EP, Andrews PE, et al. Microporous polysaccharide hemospheres for management of laparoscopic trocar injury to the spleen. Am J Surg. 2008;195(1):99–103. doi:10.1016/j.amjsurg.2007.03.006
  • Singh P, Vasques D, Deleon F. Microporous polysaccharide hemospheres for adhesion prevention: a randomized controlled trial. J Gynecol Surg. 2013;29(4):196–202. doi:10.1089/gyn.2013.0007
  • Tschan CA, Niess M, Schwandt E, Oertel J. Safety and efficacy of microporous polysaccharide hemospheres in neurosurgery. Neurosurgery. 2011;69(SUPPL. 1):ons49–ons63. doi:10.1227/NEU.0b013e3182155a52
  • Hermans MHE, Brown L, Darmoc M. Adhesion prevention in an intraperitoneal wound model: performance of two resorbable hemostats in a controlled study in rabbits. J Biomed Mater Res. 2012;100 B(6):1621–1626. doi:10.1002/jbm.b.32730
  • Erdogan H, Kelten B, Tuncdemir M, et al. Hemostasis vs. epidural fibrosis?: a comparative study on an experimental rat model of laminectomy. Neurologia I Neurochirurgia Polska. 2016;50(5):323–330. doi:10.1016/j.pjnns.2016.05.002
  • Bruckner BA, Loebe M. Microporous polysaccharide hemosphere absorbable hemostat (AristaAH) use in re-operative cardiac surgical procedures. US Cardiol. 2012;9(2):96–98. doi:10.15420/usc.2012.9.2.96
  • Bruckner BA, Loebe M. The use of an absorbable microporous polysaccharide hemosphere haemostat (AristaAH) in ventricular assist device implant and cardiac transplantation procedures. Eur Cardiol. 2012;8(2):125–127. doi:10.15420/ecr.2012.8.2.125
  • Sindwani R. Use of novel hemostatic powder MPH for endoscopic sinus surgery: initial impressions. Otolaryngology. 2009;140(2):262–263. doi:10.1016/j.otohns.2008.11.009
  • Phillips PP. Safety and efficacy of Arista powder in 514 nasal septal and sinus surgery patients. Otolaryngology. 2013;149(2):P135. doi:10.1177/0194599813495815a301
  • Galarza M, Porcar OP, Gazzeri R, Martnez-Lage JF. Microporous Polysaccharide Hemospheres (MPH) for cerebral hemostasis: a preliminary report. World Neurosurg. 2011;75(3–4):491–494. doi:10.1016/j.wneu.2010.10.008
  • Reynbakh O, Akhrass P, Souvaliotis N, et al. Use of MPH hemostatic powder for electrophysiology device implantation reduces postoperative rates of pocket hematoma and infection. Curr Med Res Opin. 2018;34(10):1861–1867. doi:10.1080/03007995.2018.1476847
  • Antisdel JL, West-Denning JL, Sindwani R. Effect of microporous polysaccharide hemospheres (MPH) on bleeding after endoscopic sinus surgery: randomized controlled study. Otolaryngology. 2009;141(3):353–357. doi:10.1016/j.otohns.2009.06.078
  • Antisdel JL, Matijasec JL, Ting JY, Sindwani R. Microporous polysaccharide hemospheres do not increase synechiae after sinus surgery: randomized controlled study. Am J Rhinol Aller. 2011;25(4):268–271. doi:10.2500/ajra.2011.25.3619
  • Nunez-Nateras R, Hurd KJ, Ferrigni EN, et al. Athermal nerve sparing robot-assisted radical prostatectomy: initial experience with microporous polysaccharide hemospheres as a topical hemostatic agent. World j Urol. 2013;31(3):523–527. doi:10.1007/s00345-011-0815-8
  • Gilbert DR, Angell J, Abaza R. Evaluation of absorbable hemostatic powder for prevention of lymphoceles following robotic prostatectomy with lymphadenectomy. Urology. 2016;98:75–80. doi:10.1016/j.urology.2016.06.071
  • Gleason S, Mehl D, Payne W, et al. Microporous polysaccharide hemosphere efficacy and safety in primary total knee arthroplasty. J Orthopa. 2019;16(1):19–24. doi:10.1016/j.jor.2018.11.006
  • Palacios DA, McDonald M, Miyake M, Rosser CJ. Pilot study comparing the two hemostatic agents in patients undergoing partial nephrectomy. BMC Res Notes. 2013;6(1):399. doi:10.1186/1756-0500-6-399
  • Antisdel JL, Meyer A, Comer B, et al. Product comparison model in otolaryngology: equivalency analysis of absorbable hemostatic agents after endoscopic sinus surgery. Laryngoscope. 2016;126(Supplement 2):S5–S13. doi:10.1002/lary.25678
  • Kunduz E, Aysan E, Idiz UO, et al. Evaluation of local hemostatic effect of microporous polysaccharide hemospheres products in thyroid surgery: a prospective randomized controlled study. Turk J Surg. 2019;35(1):49–53. doi:10.5578/turkjsurg.4162
  • Tan SR, Tope WD. Effectiveness of microporous polysaccharide hemospheres for achieving hemostasis in Mohs micrographic surgery. Dermatologic Surg. 2004;30(6):908–914. doi:10.1111/j.1524-4725.2004.30261.x
  • Benlier E, Top H, Aygit AC, et al. Microvascular anastomosis with minimal suture and arista: an experimental study. J Reconstruct Microsurg. 2007;23(6):311–315. doi:10.1055/s-2007-985213
  • Egeli T, Sevinc AI, Bora S, et al. Microporous polysaccharide hemospheres and seroma formation after mastectomy and axillary dissection in rats. Balkan Med J. 2012;29(2):179–183. doi:10.5152/balkanmedj.2012.005
  • Ereth MH, Schaff M, Ericson EF, et al. Comparative safety and efficacy of topical hemostatic agents in a rat neurosurgical model. Neurosurgery. 2008;63(4 SUPPL):ONS369–ONS72.
  • Hoffmann NE, Siddiqui SA, Agarwal S, et al. Choice of hemostatic agent influences adhesion formation in a rat cecal adhesion model. J Surg Res. 2009;155(1):77–81. doi:10.1016/j.jss.2008.08.008
  • Emmez H, Tonge M, Tokgoz N, et al. Radiological and histopathological comparison of microporous polysaccharide hemospheres and oxidized regenerated cellulose in the rabbit brain: a study of efficacy and safety. Turk Neurosurg. 2010;20(4):485–491. Turkish. doi:10.5137/1019-5149.JTN.3300-10.2
  • Ereth MH, Dong Y, Schrader LM, et al. Microporous polysaccharide hemospheres do not enhance abdominal infection in a rat model compared with gelatin matrix. Surg Infect. 2009;10(3):273–276. doi:10.1089/sur.2007.033
  • Antisdel JL, Janney CG, Long JP, Sindwani R. Hemostatic agent Microporous Polysaccharide Hemospheres (MPH) does not affect healing or intact sinus mucosa. Laryngoscope. 2008;118(7):1265–1269. doi:10.1097/MLG.0b013e31816c7bc9
  • Ereth M, Sibonga J, Oliver W, et al. Microporous polysaccharide hemospheres do not inhibit bone healing compared to bone wax or microfibrillar collage. Orthopedics. 2008;31(3):1–4.
  • Barkun AN, Adam V, Lu Y, et al. Using hemospray improves the cost-effectiveness ratio in the management of upper gastrointestinal nonvariceal bleeding. J Clin Gastroenterol. 2018;52(1):36–44. doi:10.1097/MCG.0000000000000709
  • Joshi MR, Latham J, Okorogheye G. Use of a flowable haemostat versus an oxidised regenerated cellulose agent in primary elective cardiac surgery: economic impact from a UK healthcare perspective. J Cardiothorac Surg. 2017;12(1):107. doi:10.1186/s13019-017-0660-y
  • Makhija D, Rock M, Xiong Y, et al. Cost-consequence analysis of different active flowable hemostatic matrices in cardiac surgical procedures. J Med Econ. 2017;20(6):565–573. doi:10.1080/13696998.2017.1284079
  • Wood E. Advance preparation tends to waste many absorbable hemostatic agents. OR Manag. 2016;3:32.
  • ACOG. Topical hemostatic agents at time of obstetric and gynecologic surgery. ACOG Committee Opinion. Obsetr Gynecol. 2020;136(4):e81–e9. doi:10.1097/AOG.0000000000004104
  • Ramirez MG, Niu X, Epstein J, Yang D. Cost-consequence analysis of a hemostatic matrix alone or in combination for spine surgery patients. J Med Econ. 2018;21(10):1041–1046. doi:10.1080/13696998.2018.1513261