Patient acceptability of the Tecnis® multifocal intraocular lens

&
Pages 403-410 | Published online: 24 Mar 2011

Abstract

Cataract surgery has evolved. The goal of the surgeon includes both restoration of vision and refinement of vision. Patients’ desire for spectacle independence has driven the market for presbyopia-correcting cataract surgery and development of novel intraocular lens (IOL) designs. The Tecnis® Multifocal Intraocular Lens incorporates an aspheric, modified anterior prolate IOL with a diffractive multifocal lens design. The design aims to minimize spherical aberration and improve range of focus. The purpose of this review is to assess patient acceptability of the Tecnis® multifocal intraocular lens.

Introduction

Presbyopia-correcting intraocular lenses are designed to provide freedom from glasses after cataract surgery. Monofocal intraocular lenses (IOL) are designed to provide excellent visual acuity at a single, fixed focal length; thus, if patients are corrected for distance, they require spectacle correction for reading vision.Citation1 The importance of independence from glasses should not be undervalued. In one study, 10% of patients with presbyopia reported a willingness to trade 5% of their life expectancy to be rid of presbyopia.Citation2 Thus restoration of both distance and near vision is a goal of the ophthalmologist. While medical treatments have failed to restore the accommodative abilities of the natural lens, modern cataract surgery with intraocular lens implantation is a specific time point where a surgical intervention to improve near vision is possible. However, replicating the optics of the young, accommodative lens remains a challenge.

Multifocal IOLs, developed in the last two decades, are a response to this challenge. Lenses are categorized as diffractive IOLs and refractive IOLs based on lens design. Comparison of multifocal and monofocal IOLs shows that multifocal IOLs improve near vision and lead to spectacle independence more often than monofocals.Citation3Citation8 However, symptoms of decreased visual acuity and photic phenomena are a source of patient dissatisfaction with multifocal lenses and are reported with higher frequency.Citation9

Many types of presbyopic-correcting lenses are available for implantation today. Currently in the United States, multifocal lens options include the ReZoom® lens (Abbott Medical Optics Inc, Santa Ana, CA), the ReSTOR® lens (SA60D3, Alcon Laboratories Inc, Fort Worth, TX), and the Tecnis® lens (ZM900, ZMA00, ZMB00, Abbott Medical Optics Inc, Santa Ana, CA). The alternative presbyopia-correcting lens is the Crystalens® (Eyeonics and 1CU, AlisoViego, CA; Human Optics, Erlangen, Germany). Alternative lens designs, including fully accommodating intraocular lenses, are currently available outside of the United States or are in clinical trial. In a recent review, Lane et al discussed the ReZoom® and ReSTOR® lenses in detail.Citation8 Briefly, the ReZoom® lens is a three-piece multifocal lens with multiple refractive optical zones. The annular rings (zones) provide focus at different distances and are proportioned to provide quality near, intermediate, and distance vision. The ReSTOR® lens is an apodized, diffractive one-piece lens. The diffractive grating is on the anterior surface of the lens. The Crystalens uses a hinged plate-optic design to allow anterior movement of the optic to provide a small range of accommodation. Multiple clinical trials and reviews of these lenses are available in the literature.

Evolution of Tecnis® lens design

The first generation of Tecnis® was a monofocal, aspheric, anterior-modified prolate lens (ZA9000 series). The Tecnis® lens was designed to minimize spherical aberrations and improve contrast sensitivity after cataract surgery with lens implantation.Citation10 In younger patients, the negative spherical aberration of the crystalline lens and the positive spherical aberration of the cornea together create a nominal overall aberration.Citation11 Spherical aberration increases with age as the lens gains positive aberration; this contributes to decreased visual function.Citation11 Visual quality with intraocular lenses has been shown to be worse than the natural lens in age-matched controls.Citation12 The Tecnis® lens was designed to add negative spherical aberration into the optical system, thus reducing the total spherical aberration. By reducing total aberration, the lens should reduce positive photic phenomena of glare and halos.Citation13 Researchers reported that the Tecnis® lens provided improved subjective quality of vision, corrected visual acuity (CVA), and contrast sensitivity at various distances and in varied lighting conditions.Citation10,Citation14,Citation15 The anterior-modified prolate design platform was used in the multifocal version of the Tecnis lens.

The multifocal Tecnis® lens diffraction pattern creates two major focal points that correspond to approximately 3.0 D on the spectacle plane. This design splits the light to be distributed among near and distant focus regardless of pupil size.Citation16,Citation17 The Tecnis® multifocal lens is a full diffractive optic versus an apodized diffractive optic of the ReSTOR® lens. Researchers have hypothesized that pupil size can affect the quality of vision; therefore, the full diffractive optic should make vision independent of pupil size. On the other hand, the apodized diffractive design (ReSTOR® lens) distributes light preferentially in the distance for large pupil sizes.Citation18 These contrasting features are based on different theories of how lens design relates to visual quality. In addition, the diffraction pattern of this lens is on the posterior surface of the lens (as compared to the ReSTOR® lens on the anterior surface of the lens) and the multifocal lens has the aspheric prolate technology of the monofocal lens platform. In this article, we will discuss the design of the Tecnis® lens and outcome measures that determine patient acceptability and lens efficacy of the Tecnis® multifocal lens.

Initially, the Tecnis lens was available only as a three-piece silicone lens (ZM900); within the past two years, a three-piece acrylic diffractive lens (ZMA00) and a one-piece acrylic diffractive lens (ZMB00) have been approved by the FDA. The majority of the research in the literature was performed with the ZM900 lens model. Hutz et al performed a randomized trial to compare the visual outcomes in patients implanted with the diffractive silicone multifocal in one eye and the diffractive Tecnis® ZMA00 acrylic multifocal IOL in their fellow eye.Citation19 Forty-two eyes of 21 cataract patients were randomized to either group and outcome measures were tested at six months and included uncorrected visual acuity (UDVA), best corrected distance visual acuity (CDVA) and corrected near visual acuity (CNVA), spherical equivalence, reading speed and acuity and photic phenomena. There were no significant differences between the two lens materials; both lenses provided excellent visual results.

Visual acuity

Outcomes based on visual acuity with multifocal lenses are divided into measurements of distance, intermediate, and near visual acuities. They are further subdivided into corrected and uncorrected (typically with spectacles) acuities. Near visual acuity is further subdivided into uncorrected, distance-corrected, or near-corrected visual acuities (UNVA, DCNVA, CNVA).Citation20 While this may seem excessive, distance corrected visual acuities at distance and near can be used to determine if refractive surgery (for the distance prescription) would resolve postoperative complaints of blurred vision.

Comparison studies, retrospective reviews, and case series of the Tecnis® multifocal lens (ZM900) have shown promising results (). Goes performed a prospective case series of 59 eyes of 30 patients with the Tecnis® ZM900 (one eye was excluded secondary to a history of amblyopia).Citation18 Ninety percent of patients achieved ≥20/30 UCVA (0.087 ± 0.085). All patients could read ≥J2 and 90% read J1 (0.133 ± 0.095). Near visual acuity also improved from one month to six months postoperatively. Fifteen eyes (25%) required refractive surgery for residual refractive error or postoperative astigmatism. Bautista et al performed a prospective case series of 250 eyes of 137 patients with cataracts.Citation21 Mean UCVA and BCVA were 0.144 ± 0.101 and 0.09 ± 0.03 logMAR (minimal angle of resolution) respectively, while 77.6% and 98.4% of eyes achieved ≥20/30. Uncorrected near visual acuity was ≥J2 in 96.8% and was ≥J1 in 83.2% of eyes. DCVA was ≥J1 in 95.6% of eyes. Similar to the other study, there was statistically significant improvement in visual acuity over time (attributed to neural adaptation).

Table 1 Summary of Tecnis® multifocal studies

In a recent retrospective review of 2500 eyes of 1558 patients implanted with the Tecnis® ZM900 (minimum of three months follow-up), 85% of patients achieved 20/30 BCVA at distance and 94% achieved J1 vision. Five percent of eyes required enhancement after initial surgery, but no eyes required lens exchange.Citation22

The Tecnis® multifocal lens showed equivalency or slightly worse visual acuity compared to monofocal lenses for distance visual acuity. The Tecnis® lens is superior to monofocal lenses for near vision (uncorrected or distance corrected). In the FDA nonrandomized clinical trial, bilateral ZM900 lens implantation was compared to the monofocalCeeOn® 911A (CEMN, AMO Inc) with 489 eyes divided between groups.Citation13 For distance vision, there was less than one Snellen line visual acuity difference for UCVA (7.54 ± 0.94 [20/22] versus 7.98 ± 1.06 [20/20]) and distance BCVA (8.35 ± 0.82 [20/18] versus 8.68 ± 0.78 [20/17]). Uncorrected UCVA and DCVA were significantly better in the Tecnis® group (P < 0.001) by four to five lines of near visual acuity. Palmer et al showed that the ZM900 resulted in slightly worse BCDVA compared to its monofocal counterpart (Z9000) (0.08 ± 0.05 logMAR versus 0.05 ± 0.05, respectively).Citation23

Multiple trials compare the Tecnis® multifocal lens (ZM900) to other multifocal lenses including the ReSTOR® (SA60D3), the Array® SA40N (Advanced Medical Optics, Santa Ana, CA), the ReZoom® (AMO), and the TwinSet® (Acri. Tec, Zeiss, Germany).Citation17,Citation23Citation27 In all studies, there were no statistically significant difference between multifocal intraocular lenses for distance visual acuities (corrected or uncorrected). Mester et al compared 50 eyes with bilateral diffractive ZM900 to 50 eyes with bilateral refractive multifocal Array SA40 (AMO).Citation17 Binocular uncorrected and corrected distance VA was not significantly different between the eyes; however, uncorrected and best-distance corrected near VA was significantly better in the Tecnis® group (P < 0.001). Hutz et al also compared the Tecnis® multifocal (ZM001, an older version of the ZM900) to the Array® SN40 and to the ReSTOR® SA60D3 to evaluate reading performance.Citation28 At one year, they evaluated reading speed (distance corrected and uncorrected) in 30 pts (n = 10 per group) at 40 cm, 60 cm and 80 cm under low (6 cd/m2) and bright (100 cd/m2) light conditions. Under low light conditions, the Tecnis® lens had significantly better near visual acuity outcomes. Under bright light conditions, the difference between groups was less. Mean reading speed with UCVA and DCVA was significantly faster in the Tecnis® group.

Cillino et al compared visual acuity outcomes in multifocal IOLs (Tecnis® ZM900, Array® SA40N, ReZoom®) with a monofocal IOL (AR40) in a randomized prospective clinical trial looking at 62 consecutive patients.Citation25 At twelve months postoperatively, the UCVA was ≥20/30 in all groups, and the BCVA was ≥20/20 in all groups without statistically significant differences. The Tecnis® group tended toward improved monocular UCVA and DCVA. The Tecnis® lens tended toward worse intermediate vision than either monofocal or refractive multifocal IOLs. Palmer et al also compared visual function in multifocal IOLs (Tecnis® ZM900, Twin-Set®, ReZoom®) with a monofocal IOL (Tecnis® Z9000) in a prospective, randomized trial.Citation22 At one and three months, binocular distance UCVA was similar in all groups; however, distance BCVA was better with the ReZoom® and monofocal Z9000 lens compared to both diffractive multifocal lenses. The lenses performed similarly for binocular NCVA. Not surprisingly, multifocal lenses performed better for distance corrected near VA than the monofocal lens. Other non-randomized trials support the conclusion that the Tecnis® ZM900, as compared to other multifocal diffractive or refractive lenses, provides equivalent or better visual acuity at distance and near, with or without correction.Citation26,Citation27

In summary, evidence from numerous studies supports the conclusion that the Tecnis® lens (ZM900) provides equivalent, if not improved, distance uncorrected visual acuity when compared to other multifocal lenses; it is controversial if the lens provides equivalent UCVA compared to monofocal lenses. Consistent with other diffractive lenses, the Tecnis® lens performed similarly or better at near when compared to refractive lenses. The significant advancement provided by this lens is the relative ‘pupil independence’ of the lens. Notably, diffractive lenses have previously been reported to provide worse intermediate visual acuity when compared to their refractive counterparts.Citation29,Citation30 Tecnis® has been shown in some studies to under perform its refractive counterparts and in others it has provided equivalent or better intermediate visual acuity.Citation24 Studies have also shown that this lens performs better in low light conditions compared with other multifocal lenses.Citation28

Patient satisfaction and spectacle independence

Patient acceptance of a presbyopic-correcting lens ultimately is determined by patient satisfaction and spectacle independence. Outcomes and optical features unique to each lens help guide discussions between the physician and patient regarding proper lens selection and realistic visual expectations. Proper preoperative counseling regarding expectations and limitations strongly influence patient selection and, ultimately, acceptance of the lens.

In the initial FDA nonrandomized clinical trial, Packer et al reported that 94.6% of patients would choose the Tecnis® ZM900 again versus 89.6% patients with the Cee-On® lens.Citation13,Citation20 Bautista et al reported on the patients’ subjective assessment of their vision. Thirty-nine percent of patients rated their vision as excellent, 51.2% rated their vision as good while 2.4% rated their vision as poor.Citation21 Satisfaction was similar regarding near or distance vision and improved over the duration of the study (150–210 days). Similarly, Mester et al assessed patient acceptance at four to nine months after lens implantation with the Array® SA40 and the Tecnis® ZM900.Citation17 Overall satisfaction was good with 88% of Tecnis® ZM900 patients and 78% of Array SA40 patients being very satisfied or satisfied. Forty-four percent of Tecnis® patients were very satisfied (versus 28% of Array® patients). No Tecnis® patients reported being unsatisfied. Goes reported that 96.4% of patients reported being “very satisfied” with the procedure and would choose the same lens again.Citation18 In summary, over 85% of patients report being satisfied with the Tecnis® lens.

In a randomized prospective trial comparing the Tecnis® ZM900, ReZoom®, Array SA40, and the monofocal AR40, patients showed no difference in overall rating of satisfaction between groups (P = 0.07).Citation25 However, patients completed a visual function questionnaire (VF-7) that assesses difficulties in vision-dependent activities of everyday life and has been shown to correlate with patient satisfaction after cataract surgery.Citation31 Multifocal lenses had significantly higher scores compared to the monofocal lens. Not surprisingly, analysis of the data revealed that scores were driven higher on questions about near visual tasks (small print, fine hand work).Citation25

Spectacle independence is a factor in patient satisfaction because it is the main reason patients choose multifocal lens implantation. Packer et al showed that patients with Tecnis® MIOL achieved spectacle independence 88.4% of the time versus 5.2% in CeeOn® patients (P < 0.0001).Citation13 On a rating scale of one to ten, the mean rating of vision without spectacles was 8.9 ± 1.4 for Tecnis® and 7.9 ± 2.0 for CeeOn (P < 0.0001). Mester et al reported spectacle independence in 82.6% of Tecnis® patients (versus 33.3% of Array® SA40).Citation17 Four Tecnis® patients required glasses (one for near correction) compared to 16 Array® patients (twelve for near correction). Palmer et al reported spectacle independence in 77% of ZM900 patients (versus 87.5% of TwinSet® patients, 44% of ReZoom® patients, and 4% of the monofocal control (Z9000) patients.Citation22 Cillino et al reported spectacle independence in 87.5% of Tecnis® patients.Citation25 And Goes et al reported spectacle independence in 92.8% of Tecnis® patients.Citation18 Across studies, spectacle independence ranged from 77% to 92.8% in patients with the Tecnis® MIOL thus significantly out performing monofocal lenses. Additionally, Tecnis® seemed to provide equivalent, if not improved, independence compared to other multifocal lenses.

Photic phenomena

Visual phenomena interfering with vision strongly influence patient satisfaction. When these phenomena interfere extensively with vision, a patient’s tolerance of the intraocular lens diminishes and an intraocular lens exchange may be required. Intraocular lenses can cause positive dysphotopsias (glare, halos, starbursts, etc) and negative dysphotopsias (shadows, dark crescents, etc). Multifocal lens design, with concentric rings of optical zones, creates positive dysphotopsias, also called photic phenomena, at higher rates than monofocal lenses.Citation6 Reported complaints of photic phenomena vary across studies ranging from 22.4% to 81%.Citation13,Citation17,Citation18,Citation21Citation23 Variability in reporting and outcomes could be secondary to method for reporting symptoms (solicited questions versus patient-generated complaints), completeness of chart notes, or study type variations.

In a prospective, noncomparison study of the ZM900 lenses, Bautista found that halos and glare were reported in 22.4% of patients (20.4% with mild symptoms, 2% with moderate symptoms, 0% with severe symptoms).Citation21 The symptoms improved (trend, not statistically significant) from initial postoperative visits to 150 to 210 days after surgery. Goes surveyed patients regarding occurrence of photic phenomena.Citation18 Although individual question results were not reported, the most common complaint was photic phenomena (halos and glare). At six months, no patients reported severe glare, 7.2% reported moderate glare, and 92.8% were asymptomatic. Between 3% and 4% of patients reported severe difficulty with night driving.

Studies that specifically asked patients about photic phenomena found significantly more frequent complaints. In the Akaishi et al study, severe glare and halos were reported in 6% and 2% of patients, respectively, and moderate glare and halos were reported in 26% and 16% of patients, respectively.Citation22 Palmer et al reported that dysphotopsias occurred in 81% of ZM900 eyes, 47% of TwinSet® eyes, 53% of ReZoom® eyes, and 48% of monofocal patients.Citation22 However, spontaneous, unprovoked patient complaints of dysphotopsias occurred in 16% to 19% of eyes with MIOLs versus 0% of eyes with monofocal lenses (P = 0.01).Citation23 At one year, Packer et al surveyed patients about difficulty with their vision. They found the most frequently reported symptoms to be night glare in 15.5% (10.3% moderate and 2.6% severe) and halos in 22.4% (mild 12.1%, moderate 5.2% and severe 5.2%).Citation13 The monofocal (Cee-on®) group reported glare and halos in 4.3% and 8.6% respectively. Photic phenomena are experienced by patients, but proper assessment of outcomes requires solicitation of symptoms by the physician.

Contrast sensitivity

Palmer et al compared contrast sensitivity of the Tecnis® ZM900, ReZoom®, TwinSet®, and Tecnis® monofocal Z9000 using the Functional Acuity Contrast Test (FACT) chart.Citation23 (The FACT chart uses Gaussian sine-wave gratings to measure contrast sensitivity of varied spatial frequencies and contrast levels.) With BSCVA under mesopic (10 lux) and scotopic (1 lux) luminance levels, the monofocal Z9000 group had the best results that were statistically significant. Other comparisons of the Tecnis® multifocal to monofocal lenses show a trend toward worse contrast sensitivity with the multifocal lens.Citation13 Similar to other multifocal lenses, the Tecnis® lens is not a perfect replica of the natural lens. Bautista et al found that the visual performance (under photopic conditions with BCVA) of the ZM900 lens underperformed compared to expected results from normal eyes on the Pelli–Robson test of contrast sensitivity.Citation21,Citation32 Inuoe et al designed model eyes with refractive and diffractive multifocal lenses to assess image quality for purposes of vitreoretinal surgery.Citation33 Not surprisingly, the lenses behaved consistently with their design – images through the near zones of the ReZoom® lens were blurred, a slight defocus occurred centrally in both diffractive lenses, and significant blur occurred in the peripheral portion of the full-diffractive lens (Tecnis® ZM900).

In the Palmer study, there were no statistically significant differences between the multifocal lenses in scotopic or mesopic conditions.Citation23 This result has been supported in other studies. Mester et al compared contrast sensitivity of the Tecnis® ZM900 to the Array lens using the FACT chart.Citation17 Similar to the other study, there were no statistically significant differences under photopic conditions; however under mesopic conditions, the Tecnis® group had improved contrast sensitivity at the highest frequencies (P < 0.05) Gierek-Ciuciura et al compared contrast sensitivity of ReZoom®, ReSTOR®, and Tecnis® ZM900 and found no difference in contrast sensitivity amongst these multifocal lenses.Citation26 Contrast sensitivity is a known potential drawback of multifocal lenses, and the Tecnis® lens is not an exception.

Neural adaptation

Tolerance to visual phenomena generated by multifocal lenses has been shown to improve over time. Researchers believe that the brain adapts to the altered visual input over time by a process called neural adaptation. Goes found significant near visual acuity improvement over six months even when refractive error remained stable (0.175 ± 0.122 logMAR to 0.127 ± 0.094 logMAR at one and six months, respectively (P = 0.005).Citation18 During this time they reported that complaints of dysphotopsias decreased, although no statistical values were given. Bautista et al reported statistically significant improvement of distance and near visual acuity with the ZM900 lens between initial, 1- to 3-month, and 5- to 7-month postoperative visits.Citation21 They reported that only a portion of patients improved over the course of the study.

Several studies hypothesize that neural adaptation plays a role in the visual outcomes of MIOLs.Citation17,Citation34,Citation35 Although results of neural adaptation training are controversial, Kaymak et al examined how training affected visual performance with ReSTOR® and Tecnis® multifocal IOLs.Citation36 Patients (eight bilateral ReSTOR and eight bilateral Tecnis® implantation) received computer-based visual training of one eye at six weeks after surgery. Although study numbers were small, there was a statistically significant improvement in distance and near visual acuity in the trained versus untrained eyes at six months (P < 0.01). No differences were noted between the two lenses. Thus, similar to other multifocal lenses, the Tecnis® lens is better tolerated over time.

Conclusion

In summary, evidence from numerous studies supports the conclusion that the Tecnis® lens (ZM900) provides equivalent, if not improved, distance uncorrected visual acuity when compared to other multifocal lenses. The studies reviewed conclude that patients with Tecnis® multifocal lenses are more likely to be spectacle independent (compared to other multifocal lenses). Patients with Tecnis® lenses experience photic phenomena and have similar contrast sensitivity to patients with other multifocal lenses. The change to an acrylic platform (ZMA00 and ZMB00) did not affect visual outcomes when compared to the silicone model (ZM900) and provided the option of a one piece lens design (ZMB00). The Tecnis® multifocal lens gives surgeons a novel alternative, in both lens design and platform, to use in multifocal lens implantation.

Financial support

Supported in part by Research to Prevent Blindness, Inc, New York, NY.

Disclosure

The authors have no financial interests in any of the products or topics mentioned in this report.

References

  • JavittJCWangFTrentacostDJOutcomes of cataract extraction with multifocal intraocular lens implantation: functional status and quality of lifeOphthalmology199710445895999111250
  • LuoBPBrownGCLuoSCThe quality of life associated with presbyopiaAm J Ophthalmol2008145461862218243152
  • LeylandMPringleEMultifocal versus monofocal intraocular lenses after cataract extractionCochrane Database Syst Rev20064CD00316917054162
  • JavittJCSteinertRFCataract extraction with multifocal intraocular lens implantation: a multinational clinical trial evaluating clinical, functional, and quality-of-life outcomesOphthalmology2000107112040204811054329
  • AvitabileTMaranoFCaninoEGLong-term visual results of bifocal intraocular lens implantationJ Cataract Refract Surg19992591263126910476512
  • SteinertRFVisual outcomes with multifocal intraocular lensesCurr Opin Ophthalmol2000111122110724823
  • PeposeJSMaximizing satisfaction with presbyopia-correcting intraocular lenses: the missing linksAm J Ophthalmol2008146564164818789794
  • LaneSSMorrisMNordanLMultifocal intraocular lensesOphthalmol Clin North Am20061918910516500531
  • WoodwardMARandlemanJBStultingRDDissatisfaction after multifocal intraocular lens implantationJ Cataract Refract Surg200935699299719465282
  • DenoyerALe LezMLMajzoubSQuality of vision after cataract surgery after Tecnis® Z9000 intraocular lens implantation: effect of contrast sensitivity and wavefront aberration improvements on the quality of daily visionJ Cataract Refract Surg200733221021617276260
  • GlasserACampbellMCPresbyopia and the optical changes in the human crystalline lens with ageVision Res19983822092299536350
  • ArtalPFerroMMirandaIEffects of aging in retinal image qualityJ Opt Soc Am A1993107165616628350155
  • PackerMChuYRWaltzKLEvaluation of the aspheric Tecnis® multifocal intraocular lens: one-year results from the first cohort of the food and drug administration clinical trialAm J Ophthalmol2010149457758420138609
  • KershnerRMRetinal image contrast and functional visual performance with aspheric, silicone, and acrylic intraocular lenses. Prospective evaluationJ Cataract Refract Surg20032991684169414522286
  • BellucciRScialdoneABurattoLVisual acuity and contrast sensitivity comparison between Tecnis® and AcrySof® SA60AT intraocular lenses: a multicenter randomized studyJ Cataract Refract Surg200531471271715899447
  • ChoiJSchwiegerlingJOptical performance measurement and night driving simulation of ReSTOR®, ReZoom®, and Tecnis® multifocal intraocular lenses in a model eyeJ Refract Surg200824321822218416255
  • MesterUHunoldWWesendahlTFunctional outcomes after implantation of Tecnis® ZM900 and Array® SA40 multifocal intraocular lensesJ Cataract Refract Surg20073361033104017531699
  • GoesFJVisual results following implantation of a refractive multifocal IOL in one eye and a diffractive multifocal IOL in the contralateral eyeJ Refract Surg200824330030518416266
  • HutzWWJackelRHoffmanPCComparison of visual performance of silicone and acrylic multifocal IOLs utilizing the same diffractive designActa Ophthalmol August 17, 2010. [Epub ahead of print]
  • KohnenTNew Abbreviations for Visual Acuity ValuesJ Cataract Refractive Surg2009351145
  • Palomino BautistaCCarmona GonzalezDCastillo GomezAEvolution of visual performance in 250 eyes implanted with the Tecnis® ZM900 multifocal IOLEur J Ophthalmol200919576276819787595
  • AkaishiLVazRVilellaGVisual performance of Tecnis® ZM900 diffractive multifocal IOL after 2500 implants: a 3-year followupJ Ophthalmol20102010 Pii717591. October 5, 2010. [Epub ahead of print]
  • Martinez PalmerAGomezFPEspanaAAVisual function with bilateral implantation of monofocal and multifocal intraocular lenses: a prospective, randomized, controlled clinical trialJ Refract Surg200824325726418416260
  • HutzWWEckhardtHBRohrigBIntermediate vision and reading speed with array, Tecnis®, and ReSTOR® intraocular lensesJ Refract Surg200824325125618416259
  • CillinoSCasuccioADi PaceFOne-year outcomes with new-generation multifocal intraocular lensesOphthalmology200811591508151618538402
  • Gierek-CiaciuraSCwalinaLBednarskiLMrukwa-KominekEA comparative clinical study of the visual results between three types of multifocal lensesGraefes Arch Clin Exp Ophthalmol2010248113314019763600
  • NgoCSinghMSngCVisual acuity outcomes with SA60D3, SN60D3, and ZM900 multifocal IOL implantation after phacoemulsificationJ Refract Surg201026317718220229949
  • HutzWWEckhardtHBRohrigBReading ability with 3 multifocal intraocular lens modelsJ Cataract Refract Surg200632122015202117137977
  • WalkowTLiekfeldAAndersNA prospective evaluation of a diffractive versus a refractive designed multifocal intraocular lensOphthalmology19971049138013869307630
  • RavalicoGParentinFPastoriGSpatial resolution threshold in pseudophakic patients with monofocal and multifocal intraocular lensesJ Cataract Refract Surg19982422442489530600
  • UusitaloRJBransTPessiTTarkkanenAEvaluating cataract surgery gains by assessing patients’ quality of life using the VF-7J Cataract Refract Surg199925798999410404378
  • MantyjarviMLaitinenTNormal values for the Pelli-Robson contrast sensitivity testJ Cataract Refract Surg200127226126611226793
  • InoueMNodaTMihashiTQuality of image of grating target placed in model of human eye with corneal aberrations as observed through multifocal intraocular lensesAm J Ophthalmology January 21, 2011. [Epub ahead of print]
  • Montes-MicoREspanaEBuenoIVisual performance with multifocal intraocular lenses: mesopic contrast sensitivity under distance and near conditionsOphthalmology20041111859614711718
  • PetermeierKSzurmanPSubjective and objective outcome following implantation of the apodized diffractive AcrySof ReSTOR®Ophthalmologe2007104539940417429652
  • KaymakHFahleMOttGIntraindividual comparison of the effect of training on visual performance with ReSTOR® and Tecnis® diffractive multifocal IOLsJ Refract Surg200824328729318416264