Purpose

To compare refractive outcomes, visual acuities, and satisfaction of patients between those treated with laser-assisted in situ keratomileusis (LASIK) using a Hansatome microkeratome (HM) and femto-assisted laser (FAL).

Methods

This was a retrospective analysis of 1,366 eyes in 687 patients who underwent LASIK with an HM (n=1,137) and an FAL (n=229) at the two centers of Hashmanis Hospital, Karachi, Pakistan. Refractive outcomes, including sphere, cylinder, and spherical equivalent in diopters (D), and visual acuities were assessed both preoperatively and at 1 month follow-up. Patient satisfaction was gauged by contacting the patient at the time of chart review.

Results

The postoperative median sphere, cylinder, and spherical equivalent values for those treated with FAL were 0.3±0.7 (−5.5–1.8), −0.5±0.6 (−5.0–1.0), and 0.0±0.7 (−6.0–1.6), respectively. For the HM arm, they were 0.0±1.28 (−10.8–6.8), −0.5±0.5 (−4.5–1.5), and −0.3±1.3 (−11.6–6.8), respectively. All preoperative values were statistically insignificant between the groups, while postoperative values were significant with P-values <0.001. Predictability and efficacy index was higher for the FAL (92.1%, 1.00) than the HM group (82.2%, 0.84). Similarly, patient satisfaction was slightly higher for those treated with FAL (93.3%) than HM (91.4%).

Conclusion

Our large retrospective analysis of eyes that have undergone LASIK using HM and FAL shows superior refractive outcomes in the latter, with special regard to procedural efficacy and predictability.

Introduction

Recently, the global prevalence of myopia has been shown to be on the increase, worldwide.Citation1 According to current trends, the international myopic burden is set to boom, with 4,578 million people or approximately half of the world population expected to be victims of this disorder.Citation2 In Pakistan, the prevalence of myopia among the adult population was found to be 36.5%.Citation3

The dynamic and multifaceted development of myopia in the region, coupled with technological advances over time, has led to the growing utilization of contemporary avenues of treatment for this disorder. Refractive surgery, particularly laser-assisted in situ keratomileusis (LASIK), has shown promising results in this regard.Citation4,Citation5

With the advent of Femto-LASIK, superior refractive outcomes in patients have been reported,Citation6 but not all researchers agree. Some show superior outcomes with the former,Citation7Citation10 others favor the use of a microkeratome,Citation11 and a few show no significant differences in either of the treatment modalities.Citation12,Citation13

This large-scale retrospective study compares the refractive outcomes following two distinct LASIK modalities in a large cohort of patients. To date, most studies comparing outcomes of LASIK with a Hansatome microkeratome (HM) and femto-assisted laser (FAL) have been conducted in western, developed countriesCitation7,Citation14 and have often compared outcomes in small sample sizes with limited refractive ranges.Citation12,Citation13,Citation15 To the best of our knowledge, this is the only study to assess the large-scale refractive outcomes of two distinct LASIK techniques in this region and one of the only in the world to include a wide range of refractive errors.

Methods

Patients

Thousand three-hundred sixty-six individual eyes of 687 patients were enrolled in this retrospective study, including 272 males and 415 females. Of this cohort, 1,137 eyes were treated with an HM and 229 with an FAL. The study period was from January 2013 to August 2016. All surgeries were conducted at the two centers of Hashmanis Hospital in Karachi, Pakistan. Three different surgeons took part in the HM group, while there was a single surgeon in the FAL cohort. Ethical approval for this retrospective study was granted by the Ethics Committee of the Hashmanis Hospital. Additionally, written informed consent for the surgery and inclusion in this study was obtained from all patients.

Inclusion/exclusion criteria

Prior to treatment, all individuals were assessed according to the study’s inclusion/exclusion criteria. Individuals older than 18 years with a stable refraction, a central corneal thickness (CCT) greater than 480 μm, a presumed residual stromal bed of >250 μm, and discontinued soft contact use for at least 1 week were deemed eligible. Informed consent was sought from all such individuals prior to surgery. Those with any ocular pathology, either active or residual, retinal pathologies including dystrophies or diabetic retinopathy, dry eyes with a Schirmer’s test II value of less than 2 mm, and those who were immunocompromised, nursing, or pregnant were excluded.

Prior to surgery, all patients underwent routine examinations including uncorrected visual acuity (UCVA), best-corrected visual acuity (BCVA), cycloplegic and subjective refractive error, slit-lamp examination, dilated retinal exam, ultrasonic pachymetry, keratometry, and corneal topography.

LASIK procedure and postoperative care

We performed surgeries in both eyes using the same procedure. Both surgeries were performed using the wavelight EX 500 machine (Alcon, Ft Worth, TX, USA), which was wavefront optimized. An HM (Bausch & Lomb, Rochester, NY, USA) was used to create flaps in one group, while a wavelight FS 200 Laser machine (Alcon) was employed in the other group undergoing LASIK with an FAL. We measured the CCT on the apex of the cornea, preoperatively. Then, once the flap was created, a second measurement was taken to calculate the thickness of the underlying stroma intraoperatively. In both instances, the Pocket II ultrasonic pachymeter (Quantel Medical, Inc., Bozeman, MT, USA) was used. A 140 μm and 120 μm flap was used in the HM and FAL groups, respectively. In those under FAL, a tissue separator was used to raise the flap, and in both arms, a balanced salt solution was used to irrigate the eye once the flap was placed back after the procedure.

Following surgery, these postoperative measures were employed: artificial tears 4 times a day for 3 weeks; moxifloxacin eye drops 4 times a day for 10 days; and combination drops with tobramycin and dexamethasone 4 times a day for 10 days.

Patients were followed up at 1 month postoperatively to review refractive outcomes including UCVA, spherical equivalent (SE), and sphere and cylinder values. The target refractive outcome was emmetropia in all patients that was either greater than −12.00 diopters (D) or less than 4.00 D. For those exceeding this value, the target was the refractive error subtracted from these values. The target outcome in both instances was dependent on the keratometry reading. Accordingly, procedural efficacy index was calculated as the proportion of the mean postoperative UCVA to the mean preoperative BCVA. Predictability was defined as percentage of eyes achieving mean SE within ±1.00 D.

Patient satisfaction

Patients were surveyed at the time of chart review and asked to rate their LASIK experience in the following classes: extremely satisfied, very satisfied, satisfied, and not satisfied.

Statistical analysis

AppSheet was used to enter data into Google Sheet from which the data were subsequently imported into SPSS 16.0 (SPSS Inc., Chicago, IL, USA). All subsequent analysis was done on this software. Descriptive statistics was used to calculate the median and standard deviation of all values; median values were used, as these data were not normally distributed. The Mann–Whitney U test was used to compare the sphere, cylinder, and SE between the two modalities. Graphs were constructed using SPSS and Microsoft Excel (Microsoft Corp., Redmond, WA, USA).

Results

Of the 1,366 eyes (687 patients) treated, 1,137 (570 patients) were in the HM group and 229 (117 patients) in the FAL. The mean age of patients in both groups was 25±5.8 and 27.0±7.3 years (), respectively.

Table 1 Preoperative data

Preoperative parameters included: sphere values of −4.3±2.8 D and −4.3±2.3 D; cylinder values of −1.0±1.0 D and −1.0±1.3 D; and SE values of −4.5±2.8 D and −4.5±2.5 D, among the HM and FAL groups, respectively (). No statistically significant difference existed in any of these values. displays the preoperative and postoperative astigmatism of the two treatment modalities. , on the other hand, shows the predicted versus the achieved SE.

Figure 1 Comparison of astigmatism.

Notes: (A) Hansatome microkeratome; (B) femto-assisted laser.
Figure 1 Comparison of astigmatism.

Figure 2 Comparison of attempted and achieved SE.

Notes: (A) Hansatome microkeratome; (B) femto-assisted laser.
Abbreviations: postop, postoperative; SE, spherical equivalent.
Figure 2 Comparison of attempted and achieved SE.

Postoperatively, the refractive outcomes for HM and FAL included: sphere values of 0.0±1.28 and 0.3±0.7; cylinder values of −0.5±0.5 and −0.5±0.6; and SE values of −0.3±1.3 and 0.0±0.7, respectively (). All differences in postoperative values were found to be statistically signifi-cant with P-values <0.001 ().

Table 2 Postoperative data

It was found that the FAL group yielded the most uniform outcomes in visual acuity, as almost all achieved a postoperative UCVA of 20/20 (), with a corresponding efficacy index of 1.00. This is in contrast with those in HM, which yielded a somewhat lower efficacy index of 0.84.

Figure 3 Comparison of postoperative UCVA and preoperative BCVA.

Notes: (A) Hansatome microkeratome; (B) femto-assisted laser.
Abbreviations: UCVA, uncorrected visual acuity; BCVA, best-corrected visual acuity.
Figure 3 Comparison of postoperative UCVA and preoperative BCVA.

The postsurgical refractive outcomes demonstrated good predictability (), with 92.1% of eyes that had undergone FAL achieving a mean SE within 1.00 D of the intended value. The same was achieved in 82.2% of eyes treated with an HM.

Figure 4 Comparison of refractive outcomes.

Notes: (A) Hansatome microkeratome; (B) femto-assisted laser.
Figure 4 Comparison of refractive outcomes.

Slightly more patients (93.3%) reported an overall level satisfaction in the FAL cohort, compared to HM (91.4%) (). The combined patient satisfaction rate was 91.8% for both procedures. 61.4% of patients were available for judging patient satisfaction in the HM and 64.1% in the FAL.

Table 3 Patient satisfaction

Discussion

Our study measured refractive outcomes across 1,366 eyes, of which a majority underwent LASIK with an HM (n=1,137) and the remainder with an FAL (n=229). The median age of patients in each treatment group was 25.0±5.8 and 27.0±7.3, which may be reflective of the increasing myopic burden among younger individuals.Citation16Citation18 Moreover, the preoperative data was similar for both the groups as indicated by the insignificant P-values.

The FAL cohort had significantly superior outcomes in two refractive domains: cylinder and SE (P<0.01). Sphere, on the other hand, showed better values in the HM group. This corresponded to a greater efficacy index among eyes treated with the FAL (n=1.00), compared with the HM (n=0.84). These results are consistent with Montes Mico et al’sCitation7 findings, where efficacy index for eyes treated with an FAL and a microkeratome were 1.07 and 1.00, respectively. Superior postoperative UCVA outcomes in eyes that have undergone FAL over microkeratome have been documented in other studies as well.Citation9,Citation10 Pajic et al,Citation15 in a prospective, randomized, and paired-eye study, also reported a lower efficacy index among eyes treated with microkeratome. However, significant differences in visual acuity between the two groups were only seen till the end of the first postoperative week as the delay in visual recovery in eyes treated with a microkeratome was attributed to a minimal interface fluid accumulation.Citation15 Our study, with a longer postoperative follow-up, shows a greater efficacy index in FAL at 1 month.

Interestingly, the aforementioned study correlated the lower efficacy index of eyes treated with microkeratome to the deviation in achieved versus intended flap thickness.Citation15 Several studies have shown that flaps produced using Femto-LASIK are more predictable, thinner, and have a favorable planar shape.Citation19Citation22 However, it is unclear whether the desirable flap properties translate into superior visual outcomes. Xia et alCitation8 found no significant differences in UCVA or BCVA in eyes treated with FAL or a mechanical microkeratome at any time postoperatively, despite achieving superior flap dimensions in the former group. Likewise, no significant differences in visual outcomes between the two flap cutting modalities were reported by Hashimoto et alCitation12 or Patel et alCitation13 in two separate randomized, controlled, paired-eye studies. However, the latter study reported greater flap thickness in eyes treated with Femto-LASIK, instead of blade LASIK, although the difference was not significant. One study was found to report a greater procedural efficacy in eyes treated with microkeratome LASIK over Femto LASIK.Citation11

Given the uncertainty surrounding the subject matter, meta-analyses have been conducted by Zhang et alCitation14 and Chen et alCitation23 to evaluate refractive outcomes of the two modalities. Neither of these reported any significant differences in efficacy, although a better predictability was documented by Chen et al.Citation23 In addition, both had conflicting reports on the prevalence of higher-order aberrations (HOAs) with Zhang et alCitation14 reporting significantly fewer HOAs in eyes that have undergone FAL, and Chen et alCitation23 reporting no significant differences in either group.

In our setting, eyes treated with FAL showed a greater procedural predictability of 92.1%, as opposed to 82.2% predictability of eyes treated with LASIK using an HM. These findings are consistent with those of other studies,Citation7,Citation23Citation25 and may be reflective of faster visual recoveries associated with Femto-LASIK.Citation24

Our study also assessed the reported satisfaction levels of patients postoperatively. It was found that both treatment modalities were received with high levels of satisfaction (): 91.4% reported satisfaction in the HM group and 93.3% in the FAL cohort. Similar high levels have also been reported in a previous study.Citation26

Limitations

This was a retrospective, observational study, and so all limitations associated with such methods apply. Additionally, our study was confined to the assessment of refractive outcomes following surgery, and we could not evaluate other pertinent outcomes like HOAs or contrast sensitivity. Also, our study did not report the flap thicknesses achieved with either treatment modality. Finally, we only had a 1 month follow-up and therefore cannot comment on the long-term effects of the procedure.

We recommend performing prospective studies where one surgeon performs surgeries on both modalities to get a more direct comparison. Such a model will account for any changes based on the skill of the surgeon.

Conclusion

We reported superior refractive outcomes in LASIK performed with an FAL as compared to one with an HM. Femto-LASIK also showed greater procedural efficacy and predictability in this large cohort of Pakistani patients. Additionally, slightly higher levels of patient satisfaction were reported in the FAL group.

Acknowledgments

We would like to thank Dr Azfar Nafees and Dr Khaliq ur Rehman for providing their surgical data for use in this study.

Disclosure

The authors report no conflicts of interest in this work.

References

  • PanCWRamamurthyDSawSMWorldwide prevalence and risk factors for myopiaOphthalmic Physiol Opt201232131622150586
  • HoldenBAFrickeTRWilsonDAGlobal prevalence of myopia and high myopia and temporal trends from 2000 through 2050Ophthalmology201612351036104226875007
  • ShahSPJadoonMZDineenBRefractive errors in the adult pakistani population: the national blindness and visual impairment surveyOphthalmic Epidemiol200815318319018569814
  • PillarAKruegerRAdvances in refractive surgery: June 2014 to July 2015Asia Pac J Ophthalmol201653212222
  • SolomonKDFernandez de CastroLESandovalHPLASIK world literature review: quality of life and patient satisfactionOphthalmology2009116469170119344821
  • HuhtalaAPietilaJMakinenPUusitaloHFemtosecond lasers for laser in situ keratomileusis: a systematic review and meta-analysisClin Ophthalmol20161039340427022236
  • Montes-MicoRRodriguez-GalieteroAAlioJLFemtosecond laser versus mechanical keratome LASIK for myopiaOphthalmology20071141626817070593
  • XiaLKYuJChaiGRWangDLiYComparison of the femtosecond laser and mechanical microkeratome for flap cutting in LASIKInt J Ophthalmol20158478479026309880
  • DurrieDSKezirianGMFemtosecond laser versus mechanical keratome flaps in wavefront-guided laser in situ keratomileusis: prospective contralateral eye studyJ Cataract Refract Surg200531112012615721704
  • TannaMSchallhornSCHettingerKAFemtosecond laser versus mechanical microkeratome: a retrospective comparison of visual outcomes at 3 monthsJ Refract Surg200925Suppl 7S668S67119705541
  • AlArfajKHanteraMMComparison of LASEK, mechanical microkeratome LASIK and Femtosecond LASIK in low and moderate myopiaSaudi J Ophthalmol201428321421925278800
  • HasimotoARGomesMFde SiqueiraMAMoreiraHFemtosecond laser versus mechanical microkeratome for LASIK flap creationArquivos Bras Oftalmol2013766335338
  • PatelSVMaguireLJMcLarenJWHodgeDOBourneWMFemtosecond laser versus mechanical microkeratome for LASIK: a randomized controlled studyOphthalmology200711481482149017350688
  • ZhangZHJinHYSuoYFemtosecond laser versus mechanical microkeratome laser in situ keratomileusis for myopia: metaanalysis of randomized controlled trialsJ Cataract Refract Surg201137122151215922108110
  • PajicBVastardisIPajic-EggspuehlerBGatzioufasZHafeziFFemtosecond laser versus mechanical microkeratome-assisted flap creation for LASIK: a prospective, randomized, paired-eye studyClin Ophthalmol201481883188925284975
  • Bar DayanYLevinAMoradYThe changing prevalence of myopia in young adults: a 13-year series of population-based prevalence surveysInvest Ophthalmol Vis Sci20054682760276516043848
  • LeeYYLoCTSheuSJLinJLWhat factors are associated with myopia in young adults? A survey study in Taiwan Military ConscriptsInvest Ophthalmol Vis Sci20135421026103323322575
  • AbdullahASJadoonMZAkramMPrevalence of uncorrected refractive errors in adults aged 30 years and above in a rural population in PakistanJ Ayub Med Coll Abbottabad201527181226182727
  • AhnHKimJKKimCKComparison of laser in situ keratomileusis flaps created by 3 femtosecond lasers and a microkeratomeJ Cataract Refract Surg201137234935721241920
  • PrakashGAgarwalAYadavAA prospective randomized comparison of four femtosecond LASIK flap thicknessesJ Refract Surg201026639240220677726
  • SuttonGHodgeCAccuracy and precision of LASIK flap thickness using the IntraLase femtosecond laser in 1,000 consecutive casesJ Refract Surg200824880280618856234
  • ZhouYZhangJTianLZhaiCComparison of the Ziemer FEMTO LDV femtosecond laser and Moria M2 mechanical microkeratomeJ Refract Surg201228318919422373033
  • ChenSFengYStojanovicAJankovMR2ndWangQIntraLase femtosecond laser vs mechanical microkeratomes in LASIK for myopia: a systematic review and meta-analysisJ Refract Surg2012281152422233436
  • StonecipherKIgnacioTSStonecipherMAdvances in refractive surgery: microkeratome and femtosecond laser flap creation in relation to safety, efficacy, predictability, and biomechanical stabilityCurr Opin Ophthalmol200617436837216900030
  • KezirianGMStonecipherKGComparison of the IntraLase femtosecond laser and mechanical keratomes for laser in situ keratomileusisJ Cataract Refract Surg200430480481115093642
  • EydelmanMHilmantelGTarverMSymptoms and Satisfaction of Patients in the patient-reported outcomes with laser in situ keratomileusis (PROWL) studiesJAMA Ophthalmol20171351132227893066