Use of a novel extended blink test to evaluate the performance of two polyvinylpyrrolidone-containing, silicone hydrogel contact lenses

, , , &
Pages 819-825 | Published online: 03 May 2018

Abstract

Background

Sustained digital display viewing reduces eye blink frequency and tear film stability. To retain water and preserve a smooth optical surface, contact lens manufacturers have integrated the humectant polyvinylpyrrolidone (PVP) into silicone hydrogel contact lenses. In this study, extended blink time (EBT) was used to assess visual stability over a prolonged blink interval of two PVP-containing silicone hydrogel lenses, samfilcon A (SAM) and senofilcon A (SEN).

Materials and methods

This randomized, bilateral, masked, crossover study assessed lens performance in ten subjects after 16 hours of wear. EBT, ie, the time elapsed between cessation of blinking and blur-out of a threshold letter on the acuity chart, was measured. At the end of the wear period, subjects reported duration of computer use and rated visual quality (VQ) and comfort while wearing the assigned lens, and the investigator evaluated lens surface wetting characteristics. Each lens was removed and immediately weighed to determine total water content.

Results

EBTs were 10.42 seconds for SAM and 8.04 seconds for SEN (p = 0.015). Subjective ratings of VQ after 16 hours of wear were 84.6 for SAM and 74.4 for SEN (p = 0.049). Comfort ratings were 85.9 for SAM and 80.2 for SEN (p > 0.05). Median times of computer use were 6–8 hours for both lens types. Post blink, 70.0% of SAM and 30.0% of SEN lenses were completely wet (p = 0.021). Total water content after wear was 43.7% for SAM and 35.5% for SEN (p < 0.001).

Conclusion

EBT measurement indicated more stable vision with the PVP-containing SAM polymer compared with the PVP-containing SEN polymer. The SAM polymer also demonstrated better surface wetting and maintained higher water content after a prolonged period of wear. EBT can be valuable in assessing vision stability of patients after hours of computer use.

Introduction

Contact lens wearers often report increased dryness and discomfort, as well as tired eyes toward the end of the day.Citation1 By contrast, symptoms of blurry or fluctuating vision often start in the morning and persist throughout the day.Citation2 The latter complication can be exacerbated by the use of the digital display devices pervasive in today’s society. From smart phones to tablets to televisions to personal computers, Americans spend over 10 hours viewing electronic device displays daily.Citation3,Citation4

The widespread use of digital displays represents a challenge for contact lens makers to design and manufacture contact lenses that perform well under conditions of prolonged digital display viewing.Citation5 The consequences of such prolonged digital display viewing upon both visual acuity and comfort were recently described.Citation6 Briefly, the digital display environment results in decreased blink frequency, increased tear evaporation, incomplete blinking, decreased tear stability, and tear film disruption in both naked eyesCitation7Citation10 and the eyes of the contact lens wearerCitation11,Citation12 alike. In the latter case, this manifests as blurry or fluctuating vision that may indicate a lens surface not retaining water.Citation13 The need for the development of new materials that better retain water becomes greater as the number of different digital devices used and daily screen viewing time increase.

Similarly, contact lens wear is also identified as a risk factor in computer vision syndrome (CVS), a combination of eye and vision problems associated with the use of computersCitation14,Citation15 and characterized by symptoms including neck/shoulder/back pain, eye strain, headache, blurred vision, and dry eyes.Citation16 Mechanisms contributing to CVS include decreased blink frequency and completeness of blink, increased tear evaporation rate, and decreased tear film stability.Citation9,Citation13 Commonalities in symptoms and mechanisms suggest that digital eye strain and CVS are related if not the same, with the distinction that digital devices are typically viewed at closer distance than are computer monitors.

In both digital eye strain and CVS, blink rate and blink characteristics change during prolonged periods of concentration.Citation17 For example, blink rate in normal, non-lens-wearing subjects decreased from nominal 17 blinks per minute at rest down to 4.5 blinks per minute when reading,Citation18 and from 18 ± 5.7 blinks per minute down to 3.6 ± 1.8 blinks per minute during video display viewing.Citation7 It should be noted that blink frequency varies widely, both within a study due to patient factors and between studies due to differences in specific study conditions. In a more recent study conducted after the advent of smart phones, blink rate in adapted habitual contact lens-wearing subjects similarly decreased from 19 blinks per minute while listening to music down to 6.2 blinks per minute while playing a video game when not wearing lenses.Citation12 The same subjects’ blink rate decreased from 27 blinks per minute down to 23 blinks per minute when wearing lenses, although this was not statistically significant. Under both conditions, subjects blinked less while playing the video game (blink amplitude [BA], defined as the maximum percentage closure of the palpebral aperture during one blink cycle, of 77 while listening to music versus 67 during game play when not wearing lenses; BA of 84 while listening to music versus 70 during game play when wearing lenses). Incomplete blinking during computer use was similarly reported and found to have greater effect upon CVS symptoms,Citation19 as tear breakup and evaporation during incomplete blinking contribute to lens water loss.Citation20

Soft contact lenses tend to lose water to some extent during wear.Citation21 Dehydration of high water lenses and concomitant reduction in visual acuity subsequent to reduced blinking or a drafty environment was reported three decades ago and termed as dehydration blur,Citation22 yet more recently the phenomenon has been reported to occur with low-water, silicone hydrogel lenses as well.Citation23 While lenses of lower water content have been reported to dehydrate less than those of higher water content, and silicone hydrogels (typically of lower water content) less than traditional hydrogelsCitation21,Citation24 (typically of higher water content),Citation25 recent advances in material and surface chemistry resulted in newer high water lenses such as nesofilcon A that does not follow this trend.Citation6 Further studies to advance understanding of the chemical and surface properties of lens polymers that maintain water in both low and high water traditional hydrogel and silicone hydrogel lenses will drive development of future lens generations.

The ramifications of contact lens dehydration upon both visual acuity and comfort were recently described.Citation6 Briefly, lens dehydration both disrupts the lens tear filmCitation26 and effects subtle changes in lens dimension;Citation27 these phenomena lead to light scatter and visual aberration,Citation13,Citation28,Citation29 as well as contact lens-induced dry eye.Citation30

One strategy that lens manufacturers use to improve contact lens water retention and preserve a smooth optical surface is to integrate wetting agents into lens polymers.Citation31 One such agent is Pluronic® F127 (BASF, Florham Park, NJ, USA), a polyethylene oxide (PEO)-polypropylene oxide-PEO block copolymer surfactant, which is incorporated into nesofilcon ACitation32 (Bausch and Lomb Incorporated, Rochester, NY, USA), a US Food and Drug Administration (FDA) Group II (high water content, nonionic) hydrogel material.Citation33 This lens was reported to maintain greater than 98% of its initial water over the course of a 16-hour wearing day;Citation34 as such, the lens maintains a better predicted retinal image quality over time compared with etafilcon A based upon an in vitro method to predict logMAR visual acuity.Citation35 Another such agent is polyvinylpyrrolidone (PVP) that has proven to be highly effective, as lenses incorporating it into the polymer matrix have found both clinical and commercial success.Citation36 N-vinyl pyrrolidone (NVP), the monomer from which PVP is formed, was used as a co-monomer in several commercial contact lens materials over the past two decades due to the water-attracting properties of the resultant polymers. Fully formed PVP was first incorporated into commercial silicone hydrogel lens materials as an additive in the early 2000s (galyfilcon ACitation37 and senofilcon A [SEN],Citation38 both from Vistakon Division of Johnson and Johnson Vision Care, Jacksonville, FL, USA). Recently, samfilcon A (SAM)Citation39 (Bausch and Lomb Incorporated) integrated the PVP wetting agent into the polymer matrix through a novel manufacturing process that results in PVP tightly compacted at the surface.Citation40 A PVP-rich surface can create an environment that attracts and retains water, resulting in a highly wettable and smooth surface.

While both polymers include PVP as an internal wetting agent, the lenses differ in their fabrication methods and relative amounts of PVP. In the manufacture of the SEN lens, fully formed high molecular weight PVP is added to the monomer mixture prior to synthesis. In the manufacture of the SAM lens, the silicone monomers and NVP polymerize sequentially, first forming the silicone hydrogel matrix and then forming PVP, the latter present in fourfold greater quantity compared with SEN.Citation41 While similar in chemistry, SEN and SAM lenses have been characterized as silicone surrounding PVP and PVP surrounding silicone, respectively.Citation40

Even with reports in published literature of subjective comfort while wearing both SEN and SAM lenses, assessment of visual stability is lacking. Because water loss increases as blinks become less frequent while viewing a video display,Citation29 it is of interest to know how PVP internal wetting agents affect not only lens water retention and comfort under these conditions, but also visual quality (VQ). Therefore, an extended blink test was developed, and SEN and SAM silicone hydrogel lenses that incorporate PVP were compared with respect to dehydration resistance, open eye visual stability, VQ, comfort, and lens wetting over a prolonged blink interval.

Materials and methods

This study was reviewed and approved by Southwest Independent Institutional Review Board (Fort Worth, TX, USA). Written approval of the protocol, informed consent form, and subject instructions were provided prior to initiation of the study. Eligible subjects gave written informed consent and complied with the study procedures.

Lenses

SEN (38% water content), an FDA Group V-CrCitation33 (low water content, nonionic, non-surface treated, semi-interpenetrating network) silicone hydrogel lens, and SAM (46% water content), an FDA Group V-CCitation33 (low water content, nonionic, hydrophilic monomer) silicone hydrogel lens, were tested and compared.

Lens water content determination

Lens water content was determined by a modification of the method outlined in the International Organization for Standardization standard.Citation42 Water content was measured and calculated for individual lenses rather than pooled lenses specified in the standard. This allowed for the calculation of water loss pursuant to lens wear for individual lenses.

Clinical evaluation

A total of 10 healthy current (9) and former (1) contact lens-wearing subjects participated in this randomized, “double-masked” evaluation. One subject was a former lens wearer who had reverted to eyeglasses, while another subject wore no lens in one eye and a nesofilcon A lens in the contralateral eye. All other subjects wore the same lens in each eye. Subjects’ habitual lenses are summarized in ; no subject wore either test lens habitually. Each subject wore both test lens types in randomized order, ie, subjects wore either SAM or SEN lenses in both eyes over a 16-hour wear period on the first test day and then the other lens type on a second test day. All subjects were asked how much time they spent working with a computer on the day of each test. Although all were coincidentally habitual computer users, this was not a study inclusion or exclusion criterion. All clinical and nonclinical assessments were made at the end of the wear period. While investigators did not intentionally view the manufacturer’s identifying markings on either lens, unconscious practitioner bias remains possible.

Table 1 Habitual lenses of test subjects

Both objective and complementary subjective evaluations of lens performance were used in this study. A pilot extended blink test was developed to assess visual stability over a prolonged blink interval. In this test, the time elapsed between cessation of blinking and blur-out of a threshold letter on the acuity chart was measured in seconds. The investigator also evaluated lens fit and surface wetting characteristics with normal blinking using a slit lamp. Wetting extent was rated in one of five categories: 1) absence of nonwetting areas, 2) <8% of surface area nonwetting, 3) <17% of surface area non-wetting, 4) <25% of surface area nonwetting, and 5) >25% of surface area nonwetting. At the end of the evaluation, lenses were removed and immediately weighed to determine total water content, and lens types were compared.

Subjects were asked to rate both their VQ and comfort at the end of the 16-hour wear period on a 0 to 100 scale, with a score of 100 considered the best. Each subject scored each lens in each eye individually. Subjects also reported the amount of time they spent at a computer while wearing the study lenses.

Statistical analysis

Extended blink time (EBT), subjective VQ, and subjective comfort ratings were averaged over each subject’s two eyes prior to analysis. The differences between lenses in these outcomes were tested for significance using linear mixed-effects models including the fixed effects of lens type and crossover phase and the random effect of subject. The difference between lenses in water content was tested for significance using a two-sided, one-sample t-test. The difference between lenses in the fraction of lenses completely wettable was tested for significance using McNemar’s test. Differences were considered significant if p ≤ 0.05.

Results

Subjects reported comparable total duration of computer use, typically 6–8 hours for both lens types ().

Table 2 Self-reported duration of computer use (n = 10 subjects for each lens type)

Performance characteristics of the tested lenses are reported in . The SAM lenses demonstrated better visual stability than did the SEN lenses, with longer elapsed time until blur during the extended blink test (10.42 ± 4.86 seconds for SAM versus 8.04 ± 2.69 for SEN, p = 0.015). The SAM lenses also demonstrated better subjective VQ at the end of the wear period than did the SEN lenses (VQ rating of 84.6 ± 16.8 for SAM versus 74.4 ± 19.8 for SEN, p = 0.049). Although not significantly different, comfort ratings also favored the SAM lenses (comfort rating of 85.9 ± 24.4 for SAM versus 80.2 ± 21.6 for SEN, p = 0.40).

Table 3 Lens performance characteristics after 16 hours of wear (n = 10 subjects for each lens type)

On examination with a slit lamp at the end of the 16-hour wear period and just before lens removal, 70.0% of the SAM lenses demonstrated a perfectly wettable surface (absence of any nonwetting areas) post blink, compared with only 30.0% of SEN lenses (p = 0.021). Just after lens removal, total water content was 43.7% ± 0.7% for SAM lenses and 35.5% ± 1.1% for SEN lenses (p < 0.001).

Discussion

Pervasive use of more digital devices for longer durations has greatly increased the incidence of symptoms of dryness, tired eyes, eye health, and vision problems,Citation43 due to primarily decreased blink frequency and incomplete blinking leading to tear film disruption during video display use.Citation7Citation12 This becomes especially problematic during contact lens wear and establishes the need for development of contact lenses that better maintain lens water and a more stable tear film.

Contact lens-wearing patient perceptions also illustrate the need for such lenses. A recent survey of 568 subjects (136 daily disposable and 432 planned replacement contact lens wearers) found high incidences of wearers concerned about blurry/fluctuating vision (56%), bothered by blurry/fluctuating vision (68%), and that found blurry/fluctuating vision to have a high negative impact on lens wearing experience (70%).Citation2 Blurry/fluctuating vision concerned subjects more than did any other negative symptom. To cope with this symptom, subjects typically blink more frequently. However, this reflex is suppressed during periods of intense concentration such as when viewing digital video displays.Citation17

When 287 of the original test subjects were specifically asked to track symptoms of dry eyes, tired eyes, and blurry/fluctuating vision throughout the day, the reported incidence of these symptoms increased. In the morning (when many are engaged in computer work), the reported incidence of blurry/fluctuating vision (35%) was greater than the reported incidences of both dry eyes (16%) and tired eyes (6%) ().Citation2 By evening, the reported incidence of tired eyes (64%) was greater than the incidences of both blurry/fluctuating vision (51%) and dry eyes (49%).

Figure 1 Incidence of reported adverse symptoms of blurry/fluctuating vision, dry eyes, and tired eyes throughout the wearing day.

Notes: Copyright ©2016. Bausch and Lomb Incorporated. Reproduced from Epstein AB, Wilson B, Reindel WT. How visual performance influences patients’ perceptions of contact lens wear. CL Spectrum. 2016;31(13):20–25.Citation2
Figure 1 Incidence of reported adverse symptoms of blurry/fluctuating vision, dry eyes, and tired eyes throughout the wearing day.

In a previous study, a novel in vitro method predicted logMAR score based on optical image quality as the lens dehydrates with time. The study demonstrated that the SAM lens achieved better predicted logMAR scores at time zero and over the full 30s study time course, with 1.3 line mean improvement at the 30s time point compared with the SEN lens (p < 0.01).Citation44 While the measurements provided an objective estimation of image quality, they were not on-eye and did not provide a true measure of vision loss.

In the present clinical study, the assessment provided an on-eye measure of vision loss. EBTs while wearing SEN or SAM lenses were measured and compared to determine the relative effectiveness of the PVP wetting agents to maintain water and optical clarity in between the extended blinks that occur during digital display viewing. The measured EBT of the test subjects was 30% greater when wearing SAM lenses than when wearing SEN lenses (10.42 ± 4.86 versus 8.04 ± 2.69, respectively; p = 0.015, ).

The EBT variable differs from the noninvasive tear film breakup time, which is the time between the end of a complete blink and the first appearance of breaks in a reflected grid pattern without respect to visual acuity changes prior to tear film breakup.Citation45 Torkildsen used a similar test (Inter-blink Interval Visual Acuity Decay) to evaluate the ability of two lubricant eye drops to delay loss of visual function in dry eye patients.Citation46 EBT offers greater insight into vision as it is the time between the end of a complete blink and first loss of visual acuity as the lens surface dehydrates on the open eye. LogMAR charts used for objective evaluation of contact lens visual performanceCitation47 prove useful in this regard as adapted in the present study to capture the kinetics of vision change during restricted blinking. The greater resilience in stable vision may be due to the higher PVP content and specific character of PVP at the surface of SAM.

At the end of the wear period, the VQ of the SAM lens was rated significantly higher than that of the SEN lens (84.6 ± 16.8 versus 74.4 ± 19.8; p = 0.049). The time course and mechanism of deterioration of optical quality and VQ upon tear breakup while wearing silicone hydrogel lenses and suppressing blinking indicate that the breakup of the tear film introduces refractive aberrations and reflective distortions from the interface between the air and tear film. With the optical changes, the contrast sensitivity decreases with the decrease in image quality.Citation48

Overall, SAM lenses maintained more water than did SEN lenses at the time of removal (43.7% ± 0.7% versus 35.5% ± 1.1%; p < 0.001). More water associated with higher PVP content and specific character of PVP at the surface of SAM may also help to explain differences in the assessment of surface wettability. After 16 hours of wear, significantly more SAM lenses had 100% wettable surfaces (70.0% versus 30.0%; p = 0.021). Loss of wetting at the lens surface can alter the local refractive index in areas of nonwetting, and thus introduce refractive aberrations that impact VQ.

Tear film instability has also been linked to hyperosmolarity, and it is suggested that the instability may result in transient spikes in osmolarity that lead to triggering of sensory neurons.Citation49 Although not significantly different, comfort ratings also favored the SAM over the SEN lens (85.9 ± 24.4 versus 80.2 ± 21.6; p = 0.40). A larger sample size would be required to determine if apparent greater comfort while wearing SAM lenses becomes significant. Further research into the role of material characteristics, tear film instability, hyperosmolarity, and discomfort is warranted.

Wetting agents are used in both traditional and silicone hydrogel lenses to promote water retention and inhibit lens dehydration,Citation31 but their effects upon visual stability are sparsely reported in published literature. Materials reported effective in promoting water retention include polyethylene glycol,Citation50 hyaluronic acid,Citation51 phosphorylcholine,Citation52 and PVP.Citation53 Clinically, PVP wetting agent incorporated into one contact lens material resulted in improved optical quality and reduced dryness relative to the same material without PVP after 1 hour of lens wear in symptomatic patients.Citation53 Although the present study was not designed to separate the effects of individual lens characteristics such as the wetting agent, the lens containing more PVP (SAM) outperformed the lens containing less PVP (SEN) with respect to maintenance of visual acuity during suppressed blink after 16 hours of wear. While both PVP-containing lenses were similar in comfort rating, digital device display-viewing wearers of SAM lenses might experience better vision than wearers of SEN lenses. This is significant because inhibited blinking as occurs during digital display viewing can result in a poor lens wearing experience and subsequent discontinuation of lens wear.Citation26

Conclusion

The extended blink test developed for this study demonstrated more stable vision with the PVP-containing SAM polymer compared to the PVP-containing SEN polymer (EBT = 10.42 ± 4.86 seconds versus 8.04 ± 2.69 seconds; p = 0.015). Further, subjects rated VQ better when wearing SAM lenses than when wearing SEN lenses (VQ rating of 84.6 ± 16.8 versus 74.4 ± 19.8; p = 0.049). SAM lenses maintained greater water content than did SEN lenses (43.7% ± 0.7% versus 35.5% ± 1.1%; p < 0.001). Better visual stability and higher water content reflected as a higher percentage of completely wetted SAM lenses compared with SEN lenses (70.0% versus 30.0%; p = 0.021). A trend toward greater overall comfort with SAM lenses was observed, but the difference was not significant (comfort rating of 85.9 ± 24.4 versus 80.2 ± 21.6; p = 0.40). While the extended blink test has not yet been rigorously validated, it nonetheless can be valuable in the development of lens polymers to meet today’s visual demands, as well as in the assessment of visual stability of patients after hours of computer use and other modes of digital device viewing.

Acknowledgments

This work was funded by Bausch and Lomb Incorporated. The abstract of this paper was presented at the American Academy of Optometry meeting held at Anaheim, CA, USA, on November 9–12, 2016, as a poster presentation (program number 165121) with interim findings. The poster’s abstract was published online and is available for download at https://www.aaopt.org/detail/knowledge-base-article/use-novel-extended-blink-test-evaluate-polyvinylpyrrolidone-pvp-contact-lens-containing-polymer.

Disclosure

Joseph Chinn is a paid consultant of Bausch and Lomb Incorporated. All other authors are employees of Bausch and Lomb Incorporated. The authors report no other conflicts of interest in this work.