Continuous 24-hour ocular dimensional profile recording in medically treated normal-tension glaucoma

, , , &
Pages 197-202 | Published online: 29 Jan 2015

Abstract

Purpose

To analyze the 24-hour ocular dimensional profile in normal-tension glaucoma (NTG) patients on medical treatment.

Methods

Consecutive, medically treated NTG subjects were recruited from a university eye center. Subjects were on a mean of 1.7±0.7 types of antiglaucoma medications and 56.6% were on a prostaglandin analog. A contact lens-based sensor device was worn in one eye of NTG patients to record the intraocular pressure (IOP)-related profile for 24 hours, recording the following: variability from mean over 24 hours, nocturnally and diurnally, as well as the number of peaks and troughs diurnally and nocturnally.

Results

In 18 NTG subjects, the nocturnal variability around the mean contact lens-based sensor device signal was 48.9% less than the diurnal variability around the mean. The number of peaks was 54.7% less during the nocturnal period than during the diurnal period. The rate of increase in the ocular dimensional profile when going to sleep was significantly greater than the rate of decrease upon waking (P<0.001).

Conclusion

In medically treated NTG subjects, there was more variability in the IOP-related pattern during the daytime and there were fewer peaks during sleep.

Background

Intraocular pressure (IOP) is one the most important modifiable risk factors for glaucoma progression.Citation1Citation3 IOP measurements in the ophthalmology clinic are often limited to office hours, which do not fully represent the nychthemeral IOP peaks and fluctuations.Citation4,Citation5 Previously, 24-hour IOP monitoring studies have reported fluctuation as high as 8.2±1.4 mmHg, but many of these studies measured IOP in the sitting position, overnight in a hospital environment, and over a few sampling periods, which does not truly reflect the physiological and normal setting of one’s daily living, especially when the subjects are awakened at night for IOP measurements.Citation6Citation10

Various studies have reported the significance of IOP fluctuation on glaucoma progression.Citation11,Citation12 While there are studies that report otherwise,Citation13,Citation14 much of the existing literature has only documented inter-visit or daytime IOP variability over a limited number of hours. This is at least partly due to limitations of how frequently repeated tonometric IOP measurements can be taken.

A contact lens-based sensor (CLS) is nowadays available for the recording of ocular dimensional profiles for up to 24 hours.Citation15 This device has been shown to be safe and tolerable in healthy subjects and glaucoma patients as well as to provide reproducible recording of 24-hour profiles.Citation16Citation18

The aim of this study was to analyze the 24-hour ocular dimensional profile in normal-tension glaucoma (NTG) patients on medical treatment.

Methods

This study adhered to the tenets of the Declaration of Helsinki. Informed patient consent and approval by the Institutional Review Board of the Hospital Authority of Hong Kong were obtained prior to study commencement.

This was a prospective cohort study from July 2012 to June 2013, conducted at a university hospital in Hong Kong. The study recruited consenting adults (age >18 years old) with unilateral or bilateral NTG who were currently on topical antiglaucoma medications. NTG was defined as open angle on gonioscopy; progressive thinning of the retinal nerve fiber layer (RNFL) on optical coherence tomography, with corresponding glaucomatous visual field changes on the Humphrey Visual Field Analyzer; and an IOP ≤21 mmHg on all clinical visits based on previous medical records. Cases with previous glaucoma surgery or laser treatment, active or previous corneal disease, and subjects with only one functional eye were excluded.

The SENSIMED Triggerfish® (Sensimed AG, Lausanne, Switzerland) is a soft silicone CLS that enables recording of the ocular dimensional profile over a 24-hour period with minimal disturbance to one’s daily routines and sleep cycles. Such 24-hour profiles are related to the 24-hour IOP profiles.Citation18 Dimensional changes are recorded in the corneoscleral area for 30 seconds every 5 minutes over 24 hours and each recording “burst” represents 300 data points, the medians of which are plotted as a single graph which makes up the 24-hour ocular dimensional profile measured in sensory output units of millivolt equivalents (mVeq). The device was CE-marked and thus approved for clinical use in 2009. In healthy volunteers and glaucoma patients, the CLS was found to be tolerable in normal activities of daily living and during sleep. A contact lens with base curve 8.7 mm was also found to be well adapted in most eyes, although the device also exists in steeper (8.4 mm) and flatter (9.0 mm) base curves.Citation16Citation18

The CLS was placed on the subject’s eye by an ophthalmologist in the outpatient clinic after a slit-lamp examination of the anterior segment and Goldmann applanation tonometry (GAT) by a single investigator. For those with unilateral disease, the CLS was placed on the eye with NTG. For those with bilateral disease, a random eye assignment by card shuffling was used to determine the eye for the CLS placement. The subject then returned home with lubricating eyedrops and carried on his or her daily activities (both indoor and outdoor), apart from showering or swimming (as the device cannot be in contact with water). Subjects continued their same regimen of antiglaucoma eyedrops and slept in their habitual position at night. Each subject carried a logbook to record sleeping times during the 24-hour period. After 24 hours, the subject returned to the clinic to have the CLS removed followed by a slit-lamp examination and GAT. The data recorded by the SENSIMED Triggerfish® device were uploaded into a computer database. CLS profile parameters were extracted following smoothing of the profile using locally weighted polynomial regression.

Definitions

  • CLS variability from mean: this variable reflects variability around the mean value of all raw (not smoothed) CLS measurements in the respective period (24-hour, diurnal, or nocturnal).

  • Diurnal/nocturnal variability: same as above but only during the awake and asleep periods, respectively, where the sleep cycle is determined from recorded sleep times in each individual’s logbook and verified by the eye-blinking frequency (present during waking) and the presence of rhythmic ocular pulsation waves (present during sleep).

  • Sleep-to-wake and wake-to-sleep slopes: calculated with a generalized linear model on raw measurements in units of mVeq/hour from 1 hour before sleep or wake time to 1 hour after sleep or wake time, respectively.

  • Number of peaks: a peak is defined as a local maximum point in the smoothed curve. The calculation of the number of peaks occurs as follows: each trough is noted as the start of a peak.

Statistical analysis

The pre- and post-CLS GAT IOPs were compared using the Wilcoxon signed-rank test. The following parameters, extracted from CLS profiles, were compared using the Mann–Whitney U-test:

  1. nocturnal versus diurnal mean variability from mean;

  2. wake-to-sleep versus sleep-to-wake slopes;

  3. number of nocturnal versus diurnal peaks; and

  4. 24-hour, diurnal, and nocturnal variability among prostaglandin users and nonusers.

Fisher’s exact test was used to compare the differences in the number of nocturnal peaks between those using a prostaglandin antiglaucoma medication versus those using other, non-prostaglandin antiglaucoma medications. Statistical significance was taken as P<0.05.

Results

In the 18 subjects enrolled in the study, there were seven males and eleven females. The mean age was 65.1±13.7 years. GAT IOPs before and after the CLS wear were 15.3±2.2 and 13.8±1.7 mmHg, respectively, while on the same antiglaucoma medications (P=0.05). The mean RNFL thickness was 72.9±9.8 μm and the mean deviation and pattern standard deviation on Humphrey visual field analysis were −6.3±4.5 decibels (dB) and 6.1±4.1 dB, respectively.

The mean number of antiglaucoma eyedrops was 1.7±0.7. In the 18 subjects, 38.9% (7/18) used a single prostaglandin eyedrop administered in the evening, 44.4% (8/18) used a twice-daily non-prostaglandin eyedrop, and 17.6% (3/17) used both a nocturnal prostaglandin eyedrop plus a twice-daily antiglaucoma eyedrop.

The mean 24-hour variability from the mean CLS signal was 75.9±21.5 (range: 35.7–120.9; 95% confidence interval [CI]: 62.8–86.7) mVeq. The mean nocturnal variability from the mean was 52.1±20.3 (range: 23.9–105.1; 95% CI: 41.6–62.5) mVeq, which was 48.9% less than the mean diurnal variability from the mean (77.6±20.6) (range: 43.9–118.4; 95% CI: 66.9–88.2) mVeq (P=0.002). There were no statistically significant differences in the mean diurnal, nocturnal, or 24-hour variability between those with and without prostaglandin analog treatment (all P>0.1) ().

Table 1 Mean 24-hour, diurnal, and nocturnal variability among subjects with and without prostaglandin analog treatment

The mean number of peaks during sleep and daytime was 6.4±2.3 and 9.3±2.4, respectively. The number of peaks was 54.7% less during the nocturnal period than the diurnal period (P=0.001). There was no significant difference between the frequencies of nocturnal peaks among those using a prostaglandin antiglaucoma medication versus those using other, non-prostaglandin antiglaucoma medications ().

Table 2 Frequency distribution of number of peaks during the nocturnal period in subjects with and without prostaglandin treatment

The mean sleep time of subjects was from 8.53 pm ±12.6 minutes to 7.06 am ±3.6 minutes. The study population had a mean positive linear slope on their CLS profile from wake-to-sleep with a mean of 53.2±42.9 mVeq/hour, signifying an increase in the ocular dimensional profile when going to sleep. Similarly, the population had a mean negative linear slope from sleep-to-wake with a mean of −42.5±47.6 mVeq/hour, signifying a decrease in the ocular dimensional profile upon waking. On comparing the two slopes, the rate of increase when going to sleep was significantly greater than the rate of decrease upon waking (P<0.001) ().

Table 3 SENSIMED Triggerfish® contact lens-based sensor 24-hour parameters in diurnal versus nocturnal periods

Discussion

The ocular dimensional profiles of all 18 NTG subjects were unique, with individual peaks occurring at different time intervals throughout the day and night; no two tracings were identical. The mean ocular dimensional profile of the study population increased during sleep (positive wake-to-sleep slope) and decreased upon waking (negative sleep-to-wake slope), with the mean rate of increase during sleep being significantly greater than the rate of decrease upon waking (P<0.001). We are inclined to agree with previous postulations, wherein the increase in nocturnal IOP was thought to be related to the gravitational pull of fluids to the eye and an increase in episcleral venous pressure during the supine sleep posture.Citation19 Through the use of the 24-hour CLS, we affirm that the phenomenon of increased IOP with supine posturing was consistent among NTG subjects under medical treatment. Other authors have also observed the persistence of posture-induced IOP changes despite treatment of NTG patients with IOP-lowering medication.Citation20 Similarly, a previous Japanese study reported, in a population of NTG subjects, that the IOP spikes recorded from the habitual positions (sitting up during the day and supine at night) were positively correlated to the spikes induced from a postural-change test and after a water-drinking test.Citation21 In contrast, Renard et al reported that, in 27 subjects with suspected NTG, 24-hour IOP monitoring revealed that 54.5% exhibited a diurnal acrophase and 36.4% exhibited a nocturnal acrophase, while 9.1% had no nychthemeral rhythm in the absence of medical treatment.Citation22 The existence of concomitant obstructive sleep apnea syndrome in 80% of assessed patients could have influenced the observed IOP rhythm in their cohort.Citation22

Before the availability of continuous 24-hour ocular dimensional profile recording, it has been reported that, in many glaucoma patients, the IOP peaks occurred outside of office hoursCitation23,Citation24 and that two-thirds of IOP peaks in untreated glaucoma patients occurred during the nocturnal period.Citation25 In our NTG population, we noted that there were 55% less peaks occurring nocturnally than diurnally (P=0.001). The reduction in the number of peaks at night in our NTG population could be related to the use of prostaglandin antiglaucoma eyedrops, since 57% of our subjects were at least on prostaglandin antiglaucoma eyedrops nocturnally. However, it may also be simply due to the absence of physical activity, eye blinking, and saccadic movements at night, as these parameters are known to cause IOP spikes and fluctuations.Citation26 Other authors have suggested that the nocturnal IOP and IOP spikes may be related to visual field progression in NTG subjects.Citation21

Study on the effect of medical treatment on IOP fluctuations has shown that prostaglandin antiglaucoma eyedrops offer the best sustained 24-hour IOP reduction, while brimonidine offers the least.Citation27 In our study, the mean nocturnal variability was significantly lower than the mean diurnal variability by 49% (P=0.002). Our findings are in agreement with a previous study by Pajic et alCitation28 reporting higher coefficients of variation in CLS data during the daytime than at nighttime in NTG patients. Pajic et al reported reduced variability in nocturnal CLS data independent of whether or not patients were on antiglaucoma medication.Citation28 We have not recorded CLS profiles in our patients without medical treatment and therefore do not know whether the same is valid for this patient cohort; however, it seems plausible that the lower level of activity during sleep provides a natural reduction in variability, which IOP-lowering medical treatment may further influence. Holló et al recorded CLS profiles in primary open-angle glaucoma and ocular hypertension patients and found no difference in 24-hour, diurnal, or nocturnal standard deviation of CLS values in the presence of prostaglandin analog treatment as compared to without treatment.Citation29 The variability of diurnal versus nocturnal CLS values was not evaluated. Thus, in this patient cohort, it seems that, while the overall ocular dimensional profile was higher at night, the number of peaks and variability from the mean were significantly reduced at night for NTG subjects on medical treatment. The use of prostaglandins did not significantly affect the CLS variability (P>0.1).

It was interesting to note that the GAT IOP after CLS wear (13.8±1.7 mmHg) was lower (P=0.05) than the IOP before the CLS wear (15.3±2.2 mmHg), although the readings were measured exactly 24 hours apart. This seemingly lower IOP after the CLS use was short of statistical significance and can be explained by day-to-day variations in IOP.

There were limitations in this study. There was no sample size calculation, and the relatively small sample size of the study did not allow for evaluation of differences in IOP-related pattern nor stratification by the type of treatment. Furthermore, only one 24-hour measurement with the CLS was performed, therefore we could not ascertain the reproducibility of the 24-hour IOP profile. At present, the ocular dimensional profile is something that cannot be translated into IOP and therefore has no validation for use in clinical practice. Additional clinical research needs to be done before we can state (based on evidence) that the CLS can be useful in optimizing the timing of medical therapy for individual patients. Nevertheless, this is one of the few publications in the literature analyzing the 24-hour ocular dimensional profile on a continuous basis in NTG subjects receiving medical treatment.

Conclusion

Continuous 24-hour ocular dimensional profiles recorded in NTG patients on medical treatment revealed that the profile increased during sleep and decreased upon waking. There was a 50% reduction in both the variability from the mean and number of peaks during sleep as compared to daytime.

Author contributions

All authors 1) have made substantial contributions to conception and design, acquisition of data, or analysis and interpretation of data; 2) have been involved in drafting the manuscript or revising it critically for important intellectual content; 3) have given final approval of the version to be published; and 4) agree to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved.

Acknowledgments

The authors would like to thank Ms Lee Man Yee and her dedicated team of nursing and clerical staff for their support in research.

Disclosure

This study was supported by the provision of the SENSIMED Triggerfish® CLS, other device items, and publication-related expenses by Sensimed. The authors report no other conflicts of interest in this work. They have no proprietary interests and had full discretion regarding the analysis of data and write-up of the manuscript.

References

  • [No authors listed]The effectiveness of intraocular pressure reduction in the treatment of normal-tension glaucoma. Collaborative Normal-Tension Glaucoma Study GroupAm J Ophthalmol199812644985059780094
  • HeijlALeskeMCBengtssonBHymanLBengtssonBHusseinMEarly Manifest Glaucoma Trial GroupReduction of intraocular pressure and glaucoma progression: results from the Early Manifest Glaucoma TrialArch Ophthalmol2002120101268127912365904
  • KassMAHeuerDKHigginbothamEJThe Ocular Hypertension Treatment Study: a randomized trial determines that topical ocular hypotensive medication delays or prevents the onset of primary open-angle glaucomaArch Ophthalmol20021206701713 discussion 829–83012049574
  • BarkanaYAnisSLiebmannJTelloCRitchRClinical utility of intraocular pressure monitoring outside of normal office hours in patients with glaucomaArch Ophthalmol2006124679379716769832
  • HughesESpryPDiamondJ24-hour monitoring of intraocular pressure in glaucoma management: a retrospective reviewJ Glaucoma200312323223612782841
  • FribergTRSanbornGWeinrebRNIntraocular and episcleral venous pressure increase during inverted postureAm J Ophthalmol198710345235263565513
  • BuguetAPyPRomanetJP24-hour (nyctohemeral) and sleep-related variations of intraocular pressure in healthy white individualsAm J Ophthalmol199411733423478129008
  • De ViveroCO’BrienCLaniganLHitchingsRDiurnal intraocular pressure variation in low-tension glaucomaEye (Lond)19948Pt 55215237835445
  • Rota-BartelinkAMPittAStoryIInfluence of diurnal variation on the intraocular pressure measurement of treated primary open-angle glaucoma during office hoursJ Glaucoma1996564104158946298
  • TajunisahIReddySCFathilahJDiurnal variation of intraocular pressure in suspected glaucoma patients and their outcomeGraefes Arch Clin Exp Ophthalmol2007245121851185717901971
  • AsraniSZeimerRWilenskyJGieserDVitaleSLindenmuthKLarge diurnal fluctuations in intraocular pressure are an independent risk factor in patients with glaucomaJ Glaucoma20009213414210782622
  • CollaerNZeyenTCaprioliJSequential office pressure measurements in the management of glaucomaJ Glaucoma200514319620015870600
  • MedeirosFAWeinrebRNZangwillLMLong-term intraocular pressure fluctuations and risk of conversion from ocular hypertension to glaucomaOphthalmology2008115693494017936908
  • BengtssonBHeijlADiurnal IOP fluctuation: not an independent risk factor for glaucomatous visual field loss in high-risk ocular hypertensionGraefes Arch Clin Exp Ophthalmol2005243651351815756577
  • LeonardiMPitchonEMBertschARenaudPMermoudAWireless contact lens sensor for intraocular pressure monitoring: assessment on enucleated pig eyesActa Ophthalmol200987443343719016660
  • LorenzKKorbCHerzogNTolerability of 24-hour intraocular pressure monitoring of a pressure-sensitive contact lensJ Glaucoma201322431131623524857
  • MansouriKMedeirosFATafreshiAWeinrebRNContinuous 24-hour monitoring of intraocular pressure patterns with a contact lens sensor: safety, tolerability, and reproducibility in patients with glaucomaArch Ophthalmol20121301215341539
  • MottetBAptelFRomanetJPHubanovaRPépinJLChiquetC24-hour intraocular pressure rhythm in young healthy subjects evaluated with continuous monitoring using a contact lens sensorJAMA Ophthalmol2013131121507151624158696
  • PrataTSDe MoraesCGKanadaniFNRitchRParanhosAJrPosture-induced intraocular pressure changes: considerations regarding body position in glaucoma patientsSurv Ophthalmol201055544545320637484
  • KiuchiTMotoyamaYOshikaTInfluence of ocular hypotensive eyedrops on intraocular pressure fluctuation with postural change in eyes with normal-tension glaucomaAm J Ophthalmol2007143469369517386282
  • SakataRAiharaMMurataHIntraocular pressure change over a habitual 24-hour period after changing posture or drinking water and related factors in normal tension glaucomaInvest Ophthalmol Vis Sci20135485313532023821194
  • RenardEPalombiKGronfierCTwenty-four hour (nyctohemeral) rhythm of intraocular pressure and ocular perfusion pressure in normal-tension glaucomaInvest Ophthalmol Vis Sci201051288288919684006
  • WeinrebRNLiuJHNocturnal rhythms of intraocular pressureArch Ophthalmol2006124226927016476898
  • BaggaHLiuJHWeinrebRNIntraocular pressure measurements throughout the 24 hCurr Opin Ophthalmol2009202798319240539
  • LiuJHWeinrebRNMonitoring intraocular pressure for 24 hBr J Ophthalmol201195559960021330554
  • DownsJCBurgoyneCFSeigfreidWPReynaudJFStrouthidisNGSalleeV24-hour IOP telemetry in the nonhuman primate: implant system performance and initial characterization of IOP at multiple timescalesInvest Ophthalmol Vis Sci201152107365737521791586
  • StewartWCKonstasAGNelsonLAKruftBMeta-analysis of 24-hour intraocular pressure studies evaluating the efficacy of glaucoma medicinesOphthalmology2008115711171122. e118082886
  • PajicBPajic-EggspuchlerBHaefligerIContinuous IOP fluctuation recording in normal tension glaucoma patientsCurr Eye Res201136121129113821978205
  • HollóGKóthyPVarghaPEvaluation of continuous 24-hour intraocular pressure monitoring for assessment of prostaglandin-induced pressure reduction in glaucomaJ Glaucoma2014231e6e1224370812