The MGuard coronary stent: safety, efficacy, and clinical utility

, , , &
Pages 533-539 | Published online: 18 Sep 2015

Abstract

Atheromatous and thrombotic embolization during percutaneous coronary revascularization is a feared complication that may cause impaired myocardial reperfusion even with a patent epicardial vessel. The MGuard stent is a cobalt chromium bare metal stent with a porous net attached to its outer surface that has been designed to prevent thrombus fragmentation and distal embolization during stent implantation. This review summarizes the available evidence supporting the use of the MGuard stent in different scenarios such as lesions with high thrombus burden, saphenous vein graft interventions, coronary perforations, or carotid lesions.

Introduction

Atheromatous and thrombotic embolization during percutaneous coronary intervention (PCI) is a feared complication that may cause impaired myocardial reperfusion even with a patent epicardial vessel.Citation1 Patients with acute coronary syndromes and those undergoing PCI of saphenous vein graphs (SVGs) are at highest risk of distal embolization, which frequently occurs after stent implantation due to thrombus fragmentation and protrusion by the stent struts. Inadequate myocardial reperfusion may occur in up to 50% of cases despite the use of optimal evidence-based coronary revascularization and is associated with a worse in-hospital and long-term prognosis.Citation2

Prevention of embolization rather than treatment should be the way forward as treatment is unlikely to succeed if no reflow is established. Thus, several pharmacological and technical strategies have been tested to minimize the distal embolization risk.Citation3 Distal protection devices failed to show reductions in infarct size or mortality in patients undergoing primary PCI,Citation4 although they seem to improve angiographic and clinical outcomes in PCI of degenerated SVG.Citation5 Similarly, catheter-based thrombus aspiration has been associated with controversial results; although early single-center studies suggested a potential benefit, recent trials have shown no reductions in infarct size or mortality.Citation6,Citation7

The concept of a stent covered with a membrane attached to its outer surface (to create a mechanical barrier to prevent thrombus fragmentation and filtration through the struts) has also been tested previously.Citation8 However, covered stents, compared with bare metal stents, were associated with an increased risk of myocardial infarction, thus prohibiting their use in this scenario. Importantly, the barrier created by the membrane was related to delayed endothelization of the stent and also to side branch occlusion, which could explain the increased rate of myocardial infarction.Citation9

The MGuard stent is a novel bare metal stent with a porous net attached to its outer surface that has been designed to trap thrombus debris reducing distal embolization, while pores allow normal endothelization of the stent through the membrane. The purpose of this review is to summarize all the relevant literature available about the use of this device.

The MGuard stent

The MGuard™ stent consists of a balloon-expandable close-cell design bare metal stent with a polyethylene terephthalate microfiber sleeve attached to its outer surface (). The first-generation stent used a 316L stainless steel frame with strut thickness of 100 µm, while the new MGuard™ Prime platform is manufactured of L605 cobalt chromium alloy (strut thickness 80 µm). The Micronet™ is identical on both stents, with a fiber width of 20 µm and an expanded aperture size of 150×180 µm.Citation10 Thus, the net creates custom-designed pores that are 5- to 40-fold smaller than the stent cells and appears to effectively act as a mechanical barrier decreasing or even preventing the thrombus protrusion and embolization to the distal vessel.Citation11 MGuard™ and MGuard™ Prime diameters range from 2.5 mm to 4.0 mm, while lengths range from 11 to 39 and 13 to 38 mm, respectively.

Figure 1 MGuard stent crimped onto a balloon.

Notes: Green arrow indicates the mesh sleeve. Red arrow indicates the metallic platform.
Figure 1 MGuard stent crimped onto a balloon.

Technical considerations

Guiding catheter, deliverability, and stent dislodgement

The MGuard stent is compatible with 0.014 in guide wires and 6 Fg guiding catheters. Crossing profile is slightly higher than newer-generation bare metal stents (1.1–1.3 mm for the stainless steel platform and 1.0–1.2 mm for the cobalt chromium platform), which together with the close-cell design may impair the deliverability. In a recent trial, the MGuard stent was unable to reach or cross the culprit lesion in nine of 217 cases (4.1%).Citation12 Of note, all nine cases occurred with the stainless steel platform. Similarly, a recent single-center experience reported 1.8% of cases in which the stent could not reach the lesion.Citation13

From January 2011 to April 2014, the manufacturer received 54 complaints of stent dislodgement from the delivery system, most occurring with the MGuard Prime stent. By contrast, no cases of net detachment from the stent were seen. Stent dislodgement was reported in two of 217 cases (0.9%) of the prospective, randomized, multicenter evaluation of a polyethylene terephthalate micronet mesh–covered stent (MGuard) in ST-segment elevation myocardial infarction (MASTER) trial.Citation12 Similarly, stent dislodgement occurred in six of 155 cases with the MGuard Prime stent (3.87%) at the IDE-MGuard for acute ST-elevation reperfusion II (MASTER II) trial,Citation14 which led the manufacturer in 2014 to a voluntary field safety corrective action. This included the temporary suspension of the MASTER II trial and the addition of a new step during the manufacturing process so-called thermoforming in which the balloon-shaping process forms “shoulders,” which are intended to provide additional securement of the stent to the delivery system.

Side branch occlusion

Due to its double-layer design (metallic platform + mesh sleeve fibers), the MGuard stent may increase the risk of side branch (SB) occlusion. SB compromise may be related to two mechanisms: first, the summation of both layers (the metallic struts and the sleeve fibers) may act as a mechanical barrier blocking the coronary flow to the SB; second, the thrombus nearly located to the bifurcation may shift from the main vessel to the SB. It is possible that SBs compromised by the second mechanism could slightly improve with pharmacological treatment. Nevertheless, if the occluded SB is clinically relevant, a bifurcation technique may be considered (). In our experience, wiring the SB should not be extremely complicated. By contrast, to advance a balloon across the net may require additional support measures such as the use of extension catheters.

Figure 2 Coronary angiogram (30° cranial oblique projection) showing a totally occluded right coronary artery before intervention (A). Despite exhaustive thrombus aspiration. Red arrow points to the thrombotic total occlusion. (B), a high thrombus burden (TIMI thrombus score =4) persisted at the bifurcation (red arrows). After MGuard deployment (green line), TIMI 3 flow was restored to the posterolateral branch, although the posterior descending artery persisted totally occluded (C). Following “reverse T stenting,” final TIMI flow was 3 in both the posterior descending artery and the posterolateral branch branches (D, final result).

Abbreviation: TIMI, thrombolysis in myocardial infarction.
Figure 2 Coronary angiogram (30° cranial oblique projection) showing a totally occluded right coronary artery before intervention (A). Despite exhaustive thrombus aspiration. Red arrow points to the thrombotic total occlusion. (B), a high thrombus burden (TIMI thrombus score =4) persisted at the bifurcation (red arrows). After MGuard deployment (green line), TIMI 3 flow was restored to the posterolateral branch, although the posterior descending artery persisted totally occluded (C). Following “reverse T stenting,” final TIMI flow was 3 in both the posterior descending artery and the posterolateral branch branches (D, final result).

MGuard stent in patients with ST-segment elevation myocardial infarction

After promising preliminary results in two small studies, efficacy and safety of the MGuard stent in patients undergoing primary PCI were first assessed by Piscione et alCitation15 (). This was a multicenter Italian prospective registry that included 100 patients with ST-segment elevation myocardial infarction (STEMI). Thrombectomy was performed in 10% of the patients and predilation in 42% of cases. Complete myocardial reperfusion including myocardial blush grade =3 and ST-segment resolution at 60 minutes was achieved in 90% of patients. In hospital mortality occurred in seven patients (five presenting with cardiogenic shock preprocedure). Two subacute stent thromboses were also reported. Major adverse cardiovascular events (MACEs) up to 2 years were observed in 7.9% of the patients. Similarly, the MAGICAL study was a prospective multicenter international trial that enrolled 60 consecutive patients with STEMI.Citation16 In this study, a high rate of complete epicardial and myocardial perfusion was also achieved (thrombolysis in myocardial infarction [TIMI] 3 flow in 93% of patients, myocardial blush grade 3 in 73%, ST-segment resolution in 68%). Cardiac mortality up to 3 years was very low (7%).Citation17

Table 1 Summary of studies with the MGuard stent in patients with STEMI

As opposed to previous studies in whom manual aspiration was performed in <20% of cases, our group first reported results of the MGuard stent in selected cases of failed manual aspiration.Citation13 The study included 56 consecutive patients in whom a high thrombus burden persisted after aggressive manual aspiration. After MGuard stent deployment, >89% of cases had thrombus score =0, and >70% ST resolution was observed in 58.7% of patients. Only one stent thrombosis (acute) was reported, and 9-month MACE rate was very low (3.6%).

Importantly, a recent retrospective study compared results of consecutive procedures performed with either bare metal stents or MGuard stents.Citation18 Due to differences in baseline characteristics between groups, a propensity matching was performed. After adjustment, patients treated with the MGuard stent presented with restenosis more frequently than those treated with bare metal stents. These results, however, have not been confirmed in other recent studies.Citation19

Two large-scale randomized trials have compared the MGuard stent to conventional PCI. The MASTER I trial was a prospective, multicentre, international, randomized study in which 433 subjects presenting with STEMI were randomized in a 1:1 fashion to the MGuard stent versus bare metal or drug-eluting stents.Citation12 Device success was greater in the control arm (95.9% for MGuard vs 99.1% for control; P=0.003), but better reperfusion indices were obtained in the study group, as shown by the primary endpoint of the study, ST segment resolution (57.8% vs 44.7%; P=0.008), and incidence of TIMI 3 flow (91.7% vs 82.9%, P=0.006). Consequently, mortality at 30 days (0% vs 1.9%, P=0.06) and 1 year (1% vs 3.3%, P=0.09) tended to be lower in the MGuard group. Subgroup analyses showed the highest differences favoring the MGuard stent in cases with large thrombus burdenCitation20 or long delay time.Citation21 However, incidence of repeat revascularization was higher than the control group (8.6% vs 0.9%, P=0.0003).Citation22 The MASTER II trial was an international, multicenter, randomized trial designed to enrol 1,114 patients to show superiority of the MGuard stent vs conventional PCI in clinical events such as death or myocardial infarction. However, due to a higher than expected frequency of stent dislodgement, the study was voluntarily suspended in April 2014 after enrolment of 310 patients (155 patients per group). Analysis of the enrolled cohort showed no benefit of the MGuard stent in any of the analysed endpoints (ST-segment resolution was 56.9% vs 59.3% in the control group, P=0.68; mortality was 0.6% vs 1.9% in the control group, P=0.62).Citation14 Similarly, a pooled analysis of the MASTER I and II trials (n=743) failed to show better myocardial reperfusion (ST-segment resolution 57.5% for the MGuard vs 50.7% for the control group, P=0.07), although mortality at 30 days was lower (0.3% vs 1.9%, P=0.03).Citation14

Thus, data from randomized trials suggest that the use of MGuard stent may reduce distal embolization and may improve survival at 30 days compared with other commercially available coronary stents, although further randomized studies are warranted to confirm this hypothesis. Also, further research is needed to understand if the restenosis rate is only related to the lack of drug elution or, by contrast, if there is also an antigenic stimulus caused by the polyethylene terephthalate mesh.Citation23

MGuard stent in saphenous vein grafts

Compared with atherosclerotic plaques in native coronary arteries, SVG plaques are softer, longer, more friable, and contain necrotic debris and cholesterol crystals. Thus, percutaneous revascularization of SVG is associated with high rates of distal embolization, no-reflow phenomena, and periprocedural myocardial infarction. To avoid this complication, several drugs and devices have been tested. Graft stents have shown discouraging results and distal protection devices, although effective, are associated with an increase of procedural length and costs.Citation24,Citation25

The performance of the MGuard stent has been tested in two small multicenter prospective registries. The first-in-man study included 29 patients (17 with SVG intervention, mean SVG age =12.6 years).Citation26 Device and procedural success was 100%. Final TIMI 3 flow was observed in all cases, and periprocedural myocardial infarction was observed in a single case. However, an extension of the study with 23 SVG interventions showed a high rate of MACE (23%), although no stent thrombosis or cardiac deaths were observed up to 20 months of follow-up.Citation27 Similarly, the INSPIRE trialCitation28 reported a high rate of procedural success of the MGuard stent in 16 SVGs, achieving final TIMI 3 flow in all cases. Nevertheless, in agreement with the previous study, a high frequency of repeated revascularization of the target vessel was observed at 12 months of follow-up (19%).

MGuard stent in other scenarios

Coronary perforations

Graft stents seem to be an effective tool to seal severe perforations that have persisted despite prolonged balloon inflation. However, the high profile and low flexibility of previous covered stents compromise the deliverability of these devices, particularly in emergency situations.Citation9 Moreover, the mechanical barrier created by the membrane may preclude the endothelization of the stent, which eventually may be associated with stent thrombosis or with restenosis at the stent edges. Our experience, although limited, suggests that the MGuard stents can also be successfully used to seal coronary perforations.Citation29 The net, although pored, appears to effectively act as a mechanical barrier, decreasing or even preventing the leakage to the pericardial cavity. Nevertheless, additional mechanism might be involved, such as the stretching of the vessel or the compression of the disrupted tissue. Moreover, the cobalt chromium platform would be more deliverable than the graft stent allowing also successful deployment in tortuous vessels, whereas the porous net would allow endothelization of the stent. Importantly, when MGuard stents are used in this bailout situation, high rates of restenosis may be expected. The mechanism, although unknown, might also be related to the inflammatory process involving the perforation.

Coronary aneurysms

Coronary artery aneurysms are a rare condition with a reported incidence of 0.14%–4.9% of patients undergoing coronary angiography and 0.3%–5.3% in patients after PCI. The majority of coronary aneurysms are atherosclerotic in origin and, in most cases, they are asymptomatic. However, potential complications associated with these abnormalities include thrombus formation, distal embolization, and, rarely, rupture.

Current available evidence for MGuard stent is limited to two case reports,Citation30,Citation31 in which the MGuard stent was successfully implanted, achieving complete exclusion of the aneurysm. However, the percutaneous exclusion of coronary aneurysm is not generally recommended, and therefore the use of the MGuard stent in this scenario should not be advised. Moreover, in our experience, although limited to two cases, MGuard stent implantation did not achieve total exclusion of the aneurysm in the early phase and was related to severe restenosis at late follow-up.

Carotid arteries

To avoid microembolization after stent implantation in carotid artery stenosis, a mesh-covered stent may be implanted. Notably, for carotid arteries, the manufacturer has mounted the same porous net over a nitinol self-expandable platform (so-called CGuard stent). Although the clinical experience is very limited, recent results of the CARENET trial have shown a very low rate of new ischemic ipsilateral lesions (as assessed by diffusion-weighted magnetic resonance imaging) as well as no major adverse cardiac and cerebrovascular events at 30 days.Citation32 However, although these results are promising, more data are needed before making recommendations about the safety/efficacy of this device.

Perspectives

Considering the clinical benefits associated with drug-eluting stents in patients with STEMI and also due to the relatively high rate of restenosis of the MGuard stent, the manufacturer decided in 2014 to definitely cancel the MASTER II trial and to start development of the drug-eluting version of the MGuard stent. This new device will elute sirolimus from the stent platform (rather than the net) and will be available in late 2015.

Conclusion

The MGuard stent is a useful tool to avoid distal embolization during PCI of thrombotic lesions. The micronet attached to its outer surface prevents thrombus fragmentation and filtration during stent deployment. The enhanced myocardial reperfusion achieved with the use of the MGuard stent may be associated with a reduced mortality compared with other commercially available bare metal or drug-eluting stents. However, its use is associated with some limitations that must be highlighted: 1) the stent deliverability (even with the cobalt chromium version) may be slightly worse than other coronary stents that may be related to a higher crossing profile, and it must be taken into account when treating severely calcified or tortuous vessels; 2) due to its double-layer design (metallic platform + mesh sleeve fibers), the MGuard stent may compromise side branches, although bifurcation techniques can be performed to recover the compromised side branch; and 3) restenosis rate may be higher than with other commercially available coronary stents.

The MGuard stent can also be used in SVGs when conventional PCI with distal protection devices cannot be performed, as well as to seal coronary perforations that persist despite prolonged balloon inflation. In these cases, however, a high rate of repeated revascularization may be expected. The new sirolimus-eluting version of the MGuard stent, which is currently under development, is invented to minimize the restenosis rate of the MGuard Prime stent and will probably be available in late 2015 in Europe.

Disclosure

The authors report no conflicts of interest in this work.

References

  • ClaussenPAAbdelnoorMKvakkestadKMEritslandJHalvorsenSPrevalence of risk factors at presentation and early mortality in patients aged 80 years or older with ST-segment elevation myocardial infarctionVasc Health Risk Manag20141068368925525366
  • van ‘t HofAWLiemASuryapranataHHoorntjeJCde BoerMJZijlstraFAngiographic assessment of myocardial reperfusion in patients treated with primary angioplasty for acute myocardial infarction: myocardial blush grade. Zwolle Myocardial Infarction Study GroupCirculation19989723230223069639373
  • NiccoliGKharbandaRKCreaFBanningAPNo-reflow: again prevention is better than treatmentEur Heart J201031202449245520837571
  • StoneGWWebbJCoxDADistal microcirculatory protection during percutaneous coronary intervention in acute ST-segment elevation myocardial infarction: a randomized controlled trialJAMA200529391063107215741528
  • CohenDJMurphySABaimDSCost-effectiveness of distal embolic protection for patients undergoing percutaneous intervention of saphenous vein bypass grafts: results from the SAFER trialJ Am Coll Cardiol20044491801180815519010
  • Fernandez-RodriguezDRegueiroABrugalettaSOptimization in stent implantation by manual thrombus aspiration in ST-segment-elevation myocardial infarction: findings from the EXAMINATION trialCirc Cardiovasc Interv20147329430024867891
  • JollySSCairnsJAYusufSRandomized trial of primary PCI with or without routine manual thrombectomyN Engl J Med2015372151389139825853743
  • StankovicGColomboAPresbiteroPRandomized evaluation of polytetrafluoroethylene-covered stent in saphenous vein grafts: the Randomized Evaluation of polytetrafluoroethylene COVERed stent in Saphenous vein grafts (RECOVERS) TrialCirculation20031081374212821546
  • RomagueraRWaksmanRCovered stents for coronary perforations: is there enough evidence?Catheter Cardiovasc Interv201178224625321766425
  • CostaJRJrAbizaidADudekDSilberSLeonMBStoneGWRationale and design of the MGuard for acute ST elevation reperfusion MASTER trialCatheter Cardiovasc Interv201382218419023008177
  • DziewierzADudekDAdvantages of MGuard coronary stent systemMinerva Cardioangiol2012601334022322572
  • StoneGWAbizaidASilberSProspective, randomized, multicenter evaluation of a polyethylene terephthalate micronet mesh-covered stent (MGuard) in ST-segment elevation myocardial infarction: the MASTER trialJ Am Coll Cardiol Epub2012928
  • RomagueraRGomez-HospitalJASanchez-ElviraGMGuard mesh-covered stent for treatment of ST-segment elevation myocardial infarction with high thrombus burden despite manual aspirationJ Interv Cardiol20132611723419104
  • StoneGWThe MASTER II trial. Comparison of the MGuard embolic protection stent with standard stent in acute myocardial infarctionPresented at: The International Conference for Innovations MeetingDecember; 2014Tel Aviv, Israel
  • PiscioneFDanziGBCasseseSMulticentre experience with MGuard net protective stent in ST-elevation myocardial infarction: safety, feasibility, and impact on myocardial reperfusionCatheter Cardiovasc Interv201075571572119937780
  • DudekDDziewierzARzeszutkoLMesh covered stent in ST-segment elevation myocardial infarctionEuroIntervention20106558258921044911
  • DudekDDziewierzAKleczynskiPLong-term follow-up of mesh-covered stent implantation in patients with ST-segment elevation myocardial infarctionKardiol Pol201472214014524142749
  • Fernandez-CisnalACid-AlvarezBAlvarez-AlvarezBReal world comparison of the MGuard Stent versus the bare metal stent for ST elevation myocardial infarction (the REWARD-MI study)Catheter Cardiovasc Interv2015851E1E924905554
  • AmorosoGVosNSVan der HeydenJAA prospective, post-market study with the Mguard Prime Embolic Protection Stent in ST-segment elevation myocardial infarction: The International MGuard Prime Observational Study (IMOS Prime)Catheter Cardiovasc Interv Epub201537
  • CostaRAAbizaidALotanCImpact of thrombus burden on outcomes after standard versus mesh-covered stents in acute myocardial infarction (from the MGuard for acute ST elevation reperfusion trial)Am J Cardiol2015115216116625465924
  • DudekDBrenerSJRakowskiTEfficacy of an embolic protection stent as a function of delay to reperfusion in ST-segment elevation myocardial infarction (from the MASTER Trial)Am J Cardiol2014114101485148925277335
  • DudekDDziewierzABrenerSJMesh-covered embolic protection stent implantation in ST-segment-elevation myocardial infarction: final 1-year clinical and angiographic results from the MGUARD for acute ST elevation reperfusion trialCirc Cardiovasc Interv201582e00148425603802
  • LungHLHuangLHLinHCShyurSDAllergic contact dermatitis to polyethylene terephthalate meshJ Investig Allergol Clin Immunol2009192161162
  • StoneGWRogersCHermillerJRandomized comparison of distal protection with a filter-based catheter and a balloon occlusion and aspiration system during percutaneous intervention of diseased saphenous vein aorto-coronary bypass graftsCirculation2003108554855312874191
  • LeeMSParkSJKandzariDESaphenous vein graft interventionJACC Cardiovasc Interv20114883184321851895
  • KaluskiEHauptmannKEMullerRTsaiSKlapholzMGrubeECoronary stenting with MGuard: first-in-man trialJ Invasive Cardiol2008201051151518829994
  • GrubeEHauptmannKEMullerRUrielNKaluskiECoronary stenting with MGuard: extended follow-up of first human trialCardiovasc Revasc Med201112313814621640931
  • MaiaFCostaJRJrAbizaidAPreliminary results of the INSPIRE trial with the novel MGuard stent system containing a protection net to prevent distal embolizationCatheter Cardiovasc Interv2010761869220578092
  • RomagueraRGomez-HospitalJACequierANovel use of the Mguard mesh-covered stent to treat coronary arterial perforationsCatheter Cardiovasc Interv2012801757822511569
  • PatilPSethiAKaulUStent thrombosis with an aneurysm 7 years after a drug eluting stent implantationIndian Heart J201466221621924814120
  • DanziGBPomidossiGACasoloFCentolaMFerraresiRLotanCA new device to seal large coronary aneurysms: a case reportJ Med Case Rep2010423820682025
  • SchoferJEvaluation of a PET mesh covered stent in patients with carotid artery disease: results of the first in man CARENET trialPresented at: the Transcatheter Cardiovascular Therapeutics ConferenceSeptember; 2014Washington DC, USA
  • LindefjeldDSGuardaEMéndezMMicrovascular coronary flow comparison in acute myocardial infarction angioplasty treated with a mesh covered stent (MGUARD stent) versus bare metal stent: MICAMI-MGUARDCardiovasc Revasc Med20131414823337378