Optic disc measurements using the Heidelberg Retina Tomograph in amblyopia

, , , , , & show all
Pages 1025-1028 | Published online: 14 Sep 2010

Abstract

Purpose:

To investigate the characteristics of optic disc parameters in amblyopic eyes in which retinal involvement is uncertain.

Methods:

A total of 44 patients with a history of unilateral amblyopia (27 patients with persistent amblyopia and 17 patients with resolved amblyopia) were examined using the Heidelberg Retina Tomograph (HRT) II. Parameters examined included disc area, cup area, cup volume, rim area, rim volume, cup-to-disc area ratio, and mean retinal nerve fiber layer thickness.

Results:

In patients with persistent amblyopia, the amblyopic eyes were significantly more hyperopic than the fellow eyes. In the HRT parameters, there were no significant differences between the amblyopic and fellow eyes. In addition, after adjusting for refraction, the presence of strabismus, and the disc area, there was no significant difference in any HRT parameter between the amblyopic eyes of patients with persistent amblyopia and the previously amblyopic eyes of patients with resolved amblyopia.

Conclusions:

We did not find any strong evidence for the deformity of the optic disc of amblyopic eyes.

Introduction

Amblyopia is a visual disorder characterized by a subnormal visual acuity and contrast sensitivity in one or both eyes, caused by either visual deprivation or abnormal binocular interaction.Citation1 Amblyopia has generally been considered to be due to the abnormal development of the visual cortex in children.Citation2 Histologic and functional studies of the lateral geniculation nucleus (LGN) have shown that amblyopia also affects the LGN,Citation3Citation7 although some previous studies investigating the function of the LGN of amblyopic monkeys have indicated the contrary.Citation8 While previous studies have shown little evidence of retinal involvement in amblyopia,Citation2,Citation9 recent studies of the optic discs in amblyopia have suggested that anomalous optic discs might be found.Citation10Citation12 Therefore, it is still unclear whether or not the retina is entirely normal in amblyopia.

The Heidelberg Retina Tomograph (HRT), a cofocal scanning laser ophthalmoscopic device, has been used to evaluate quantitatively the three-dimensional surface topography of the optic nerve head and the surrounding nerve fiber layer.Citation13Citation15 HRT can provide objective measurements of the optic nerve head, such as cup-to-disc ratio, cup volume, rim area, and thickness of the nerve fiber layer. Therefore, HRT may be an ideal device by which to analyze the optic nerve head in amblyopic eyes.

In this study, we used the HRT, optic disc morphology, and retinal nerve fiber layer thickness (RNFLT) in unilateral amblyopia in order to investigate whether or not there are any differences in these measurements between amblyopic and fellow eyes. We also compared patients whose amblyopia had improved and those whose amblyopia had failed to improve.

Materials and methods

We examined 48 patients with a history of amblyopia or who were receiving amblyopia treatment. The patients enrolled in this study had amblyopia caused by strabismus, anisometropia, or both. All of the patients were treated for their amblyopia with refractive correction and patching of the sound eye or atropine instillation in the sound eye when necessary (at least once). Some of the patients were not fully compliant with the prescribed treatment or had been found too late for treatment. The data from four patients were excluded because of poor image quality. Thus, in the end, we analyzed data from 44 patients. These patients were divided into two groups, ie, 54 eyes of 27 patients with persistent amblyopia (13 male and 14 female, aged 6–30 [10.8 ± 5.9] years when the images were taken) and 34 eyes of 17 patients who had recovered from amblyopia (six male and 11 female, whose ages ranged 5 to 15 years (8.2 ± 2.7 years). Resolved amblyopia was defined as improvement in visual acuity of the amblyopic eye to 20/20 or better. All of the patients underwent a complete eye examination including cycloplegic refraction. There were no statistically significant differences in age (P = 0.059 unpaired t-test), gender (P = 0.335, Fisher’s Exact test), or refractive error (P = 0.53) for amblyopic eyes; P = 0.432 for fellow eyes (unpaired t-test) distributions between persistent and resolved amblyopia. The mean final best-corrected visual acuities of the amblyopic eye in persistent and resolved amblyopia were 20/50 (range 20/200–20/30) and 20/20 (range 20/20–20/16), respectively. Both of the mean final best-corrected visual acuities of the fellow eye in persistent and resolved amblyopia were 20/16 (ranged from 20/20 to 20/16). All patients were followed up for more than five years in order to ascertain if the improvement in visual acuity had reached a plateau.

In each patient, optic disc topography was examined using HRT II (version 3.0, Heidelberg Engineering, GmbH, Heidelberg, Germany). Informed consent was obtained from each patient. HRT II uses a diode laser (670 nm wavelength) to scan the retinal surface sequentially in the horizontal and vertical directions on multiple focal planes. By using confocal scanning principles, a three-dimensional topographic image is constructed from a series of optical image sections at consecutive focal planes. The topographic image determined from the acquired three-dimensional image consists of 384 × 384 (147,456) pixels, each of which is a measurement of the surface height of the retina and the optic nerve head at the corresponding location. For every subject in this study, images were obtained through dilated pupils, with a 15-degree field of view. Three topographic images were obtained, combined, and automatically aligned to make a single mean topographic image for analysis. A contour line of the optic disc margin was drawn around the inner margin of the peripapillary scleral ring by an experienced examiner, who had been kept uninformed of all the other clinical information. Seven HRT parameters obtained using routine analysis were analyzed in terms of disc area, cup area, cup volume, cup-to-disc area ratio, rim area, rim volume, and RNFLT. Magnification errors were corrected using the subjects’ refractive status and corneal curvature measurements.

The analysis was restricted to eyes that had valid optic disc measurements with HRT II. Good image quality was defined by appropriate focus, brightness, and clarity, minimal eye movement, optic disc centered in the image, and a standard deviation of the mean topographic image <30 μm. Eyes in which good quality images could not be obtained were excluded from the analysis.

SPSS software (v. 17.0 J; SPSS Japan, Inc., Tokyo, Japan) and MedCalc version 10.1 (MedCalc Software bvba, Mariakerke, Belgium) were used to perform the statistical analysis. A paired t-test was used to assess differences in parameters between amblyopic (previously or persistently) and fellow eyes. An unpaired t-test was used to assess differences in parameters between amblyopic eyes of patients with persistent amblyopia and previously amblyopic eyes of patients with resolved amblyopia. Logistic regression analysis was applied to adjust for effects of refraction, presence of strabismus, and disc area. Because of the large number of tests, the Bonferroni correction for multiple comparisons was also applied.

Results

In patients with persistent amblyopia, the spherical equivalent refractive error in the amblyopic eyes was significantly larger (P < 0.001) than that in the fellow eyes, ie, the amblyopic eyes were significantly more hyperopic (see ). RNFLT in amblyopic eyes was slightly thinner than that in fellow eyes (P = 0.022), but the difference was not significant after Bonferroni correction. There were no significant differences between amblyopic and fellow eyes in other HRT parameters. In patients with resolved amblyopia, there was no significant difference in refraction between previously amblyopic and fellow eyes. Also, there was no significant difference in each HRT parameter between the two eyes.

Table Comparison of refraction and Heidelberg Retina Tomography parameters

Rim volume in amblyopic eyes of patients with persistent amblyopia was greater than that in previously amblyopic eyes of patients with resolved amblyopia, but the P value (P = 0.037) was not significant after correction for multiple comparisons (see ). Logistic regression analysis adjusting for refraction, presence of strabismus, and disc area showed that there was no significant difference in each HRT parameter between the two eyes (P ≥ 0.056).

Discussion

In this study, we have shown that optic discs in amblyopic eyes are morphologically not significantly different from those of fellow eyes. RNFLT was slightly thinner in amblyopic eyes in fellow eyes in patients with persistent amblyopia. However, when compared with previously amblyopic eyes of patients with resolved amblyopia, the RNFLT of amblyopic eyes of patients with persistent amblyopia was not thinner.

The effects of such variables as disc area, axial length, or refraction on HRT parameters must also be considered. There was a statistically significant difference in refractive error between amblyopic and fellow eyes in patients with persistent amblyopia. It is possible that refractive error has influenced measurement of RNFLT because amblyopic eyes were significantly more hyperopic. Since RNFLT has been shown to correlate negatively with refractive error in a large cross-sectional study using HRT,Citation15 the decrease in RNFLT of amblyopic eyes might be explained by greater hyperopia in these eyes. Therefore, it seems premature to conclude that RNFLT in amblyopic eyes is different from that in normal eyes, if such an influence is not taken into account. Thus, we performed a second analysis using logistic regression analysis. However, even after adjusting for refractive errors, presence of strabismus, and disc area, there was no significant difference in each HRT parameter, including RNFLT, between amblyopic eyes of patients with persistent amblyopia and previously amblyopic eyes of patients with resolved amblyopia.

Whether or not the RNFLT values in amblyopic eyes are different from those in normal subjects is very important in elucidating the site responsible for amblyopic visual deficits. Previous studies of RNFLT have shown inconsistent results. The majority of the studies have demonstrated no significant difference between amblyopic and fellow or control eyes,Citation16Citation24 although some studies have shown that RNFLT may be thicker in eyes with refractive (anisometropic) amblyopia.Citation25,Citation26 Yen et al hypothesized that the RNFLT is thicker in amblyopic eyes because amblyopia affects the process of postnatal reduction of ganglion cells.Citation25 On the other hand, Duranoglu found that RNFLT was thinner in amblyopic eyes.Citation27 However, in these studies investigating RNFLT in amblyopia, the difference in RNFLT between the eyes was quite small, at most around 10 μm, even when there was a significant difference. Such a small difference in RNFLT may not be clinically significant.

Do optic discs in amblyopic eyes have any deformity, such as Lempert has shown? Using optic disc images and calculations for magnification, Lempert showed that the rim areas and disc areas of amblyopic eyes were smaller than those of fellow eyes or normal control eyes.Citation10Citation12 He assumed that there is a peripheral cause for visual deficits in amblyopia. Our results did not confirm his data. We did not find any significant differences in optic disc parameters, including disc area, between amblyopic and fellow eyes. On the other hand, Duranoglu found differences in some of the HRT parameters between amblyopic and fellow or control eyes.Citation27 However, there was no clear explanation for those findings. Atilla et al have noted that both hyperopic and amblyopic subjects had a smaller disc area and cup-disc ratio in comparison with a control group, but there was no significant difference between the hyperopic and amblyopic groups.Citation19 They also suggested the influence of refractive errors on optic disc topography.

In conclusion, we did not find any strong evidence for deformity of the optic disc in amblyopic eyes using HRT. Our finding, together with most of those of the previous studies, will not change current practice of amblyopia treatment, but it may have an implication on future treatment targeting the sites affected by amblyopia.

Disclosure

The authors report no conflict of interest in this work.

References

  • von NoordenGKAmblyopia: A multidisciplinary approach. Proctor lectureInvest Ophthalmol Vis Sci19852612170417163934105
  • HessRFAmblyopia: Site unseenClin Exp Optom200184632133612366358
  • von NoordenGKCrawfordMLLevacyRAThe lateral geniculate nucleus in human anisometropic amblyopiaInvest Ophthalmol Vis Sci19832467887906853107
  • von NoordenGKCrawfordMLThe lateral geniculate nucleus in human strabismic amblyopiaInvest Ophthalmol Vis Sci1992339272927321639619
  • MikiALiuGTGoldsmithZGLiuC-SJHaselgroveJCDecreased activation of the lateral geniculate nucleus in a patient with anisometropic amblyopia demonstrated by functional magnetic resonance imagingOphthalmologica2003217536536912913328
  • HessRFThompsonBGoleGMullenKTDeficient responses from the lateral geniculate nucleus in humans with amblyopiaEur J Neurosci20092951064107019291231
  • BarnesGRLiXThompsonBSinghKDDumoulinSOHessRFDecreased gray matter concentration in the lateral geniculate nuclei in human amblyopesInvest Ophthalmol Vis Sci20105131432143819875650
  • KiorpesLMcKeeSPNeural mechanisms underlying amblyopiaCurr Opin Neurobiol19999448048610448162
  • IkedaHTremainKEAmblyopia occurs in retinal ganglion cells in cats reared with convergent squint without alternating fixationExp Brain Res1979353559582456457
  • LempertPOptic nerve hypoplasia and small eyes in presumed amblyopiaJ AAPOS20004525826611040474
  • LempertPThe axial length/disc area ratio in anisometropic hyperopic amblyopia: A hypothesis for decreased unilateral vision associated with hyperopic anisometropiaOphthalmology2004111230430815019380
  • LempertPRetinal area and optic disc rim area in amblyopic, fellow, and normal hyperopic eyes: A hypothesis for decreased acuity in amblyopiaOphthalmology2008115122259226118801577
  • GirkinCAMcGwinGJrMcNealSFDeLeon-OrtegaJRacial differences in the association between optic disc topography and early glaucomaInvest Ophthalmol Vis Sci20034483382338712882785
  • MedeirosFAZangwillLMBowdCWeinrebRNComparison of the GDx VCC scanning laser polarimeter, HRT II confocal scanning laser ophthalmoscope, and stratus OCT optical coherence tomograph for the detection of glaucomaArch Ophthalmol2004122682783715197057
  • AbeHShirakashiMTsutsumiTTajimi Study GroupLaser scanning tomography of optic discs of the normal Japanese population in a population-based settingOphthalmology2009116222323019084274
  • ColenTPde FaberJTLemijHGRetinal nerve fiber layer thickness in human strabismic amblyopiaBinocul Vis Strabismus Q200015214114610893456
  • Baddini-CaramelliCHatanakaMPolatiMUminoATSusannaRJrThickness of the retinal nerve fiber layer in amblyopic and normal eyes: A scanning laser polarimetry studyJ AAPOS200152828411304814
  • BozkurtBIrkecMOrhanMKaraagaogluEThickness of the retinal nerve fiber layer in patients with anisometropic and strabismic amblyopiaStrabismus20031111712789579
  • AtillaHBatioğluFErkamNRetinal nerve fiber analysis in subjects with hyperopia and anisometropic amblyopiaBinocul Vis Strabismus Q2005201333715828869
  • AltintasOYukselNOzkanBCaglarYThickness of the retinal nerve fiber layer, macular thickness, and macular volume in patients with strabismic amblyopiaJ Pediatr Ophthalmol Strabismus200542421622116121551
  • KeeSYLeeSYLeeYCThicknesses of the fovea and retinal nerve fiber layer in amblyopic and normal eyes in childrenKorean J Ophthalmol200620317718117004633
  • DickmannAPetroniSSalerniADell’OmoRBalestrazziEUnilateral amblyopia: An optical coherence tomography studyJ AAPOS200913214815019157939
  • RepkaMXKrakerRTTamkinsSMSuhDWSalaNABeckRWPediatric Eye Disease Investigator GroupRetinal nerve fiber layer thickness in amblyopic eyesAm J Ophthalmol2009148114314719327749
  • HuynhSCSamarawickramaCWangXYMacular and nerve fiber layer thickness in amblyopia: The Sydney Childhood Eye StudyOphthalmology200911691604160919560205
  • YenMYChengCYWangAGRetinal nerve fiber layer thickness in unilateral amblyopiaInvest Ophthalmol Vis Sci20044572224223015223799
  • YoonSWParkWHBaekSHKongSMThicknesses of macular retinal layer and peripapillary retinal nerve fiber layer in patients with hyperopic anisometropic amblyopiaKorean J Ophthalmol2005191626715929489
  • DuranogluYOptic nerve head topographic analysis and retinal nerve fiber layer thickness in strabismic and anisometropic amblyopiaAnn Ophthalmol (Skokie)200739429129518025648