Measurement of the Mechanical Properties of the Tympanic Membrane with a Microtension Tester

, &
Pages 85-91 | Received 13 Dec 1989, Accepted 02 Feb 1990, Published online: 08 Jul 2009
 

Abstract

The mechanical properties of guinea pig and human tympanic membranes were measured by a microtension tester newly developed by us. Guinea pig tympanic membranes were tested under various conditions. When the tensile rate was 4.17×10−4 msec' fresh specimens were 8.8 times stronger for a tension appllied parallel to the radial fibres than for that applied at a right angle to them. No difference in tensile strength was seen when the tension was applied at a right angle to the radial fibres. When the tensile direction was parallel to the radial fibres, the posterior part was stronger than the anterior part. Two human tympanic membranes were measured under formalin fixation. The measured values were calculated into those of the fresh specimens based on the results of the guinea pig material. The adjusted values for fresh human specimens were 2.56 kgfmm2 at a right angle to the radial fibres and 3.28 kgf-mnT2 parallel to them. Applying the thin cylinder and shell theory, the breakage pressure of the human tympanic membrane was calculated to be 1.02-22.4 atmospheric pressure.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.