Reproducibility of thrombelastometry (ROTEM®): Point-of-care versus hospital laboratory performance

, , , , &
Pages 313-317 | Received 30 Sep 2012, Accepted 29 Jan 2012, Published online: 24 Jun 2012
 

Abstract

Thrombelastometry (ROTEM®) has gained wide acceptance in detecting and tailoring acquired hemostatic changes in adults and children. We investigated in this observational trial whether the reproducibility of this point-of-care testing was influenced by performance at the bedside or in the hospital laboratory. In addition, difference in time of performance between both measurements was compared. Perioperative blood samples obtained during major pediatric surgery were run in duplicate on two different ROTEM® devices located in the OR and in the hospital laboratory. The Bland-Altman test was used to compare differences of both measurements. ROTEM® measurements of 90 blood samples obtained from 24 children showed no overall clinically meaningful differences, whether they were performed bedside or in the hospital laboratory. Minor differences were found for the InTEM clot formation time (CFT) showing a mean bias of 10.79 seconds. Time saving was 11 minutes (8–16 minutes) if ROTEM® measurements were performed bedside (p < 0.001). In conclusion, there were minimal effects on ROTEM® measurements irrespective of whether they were performed in the hospital laboratory or at the bedside by a single trained staff member, while the latter saved valuable time.

Acknowledgements

We thank Monica Ceresetti and the entire team of the routine haematology laboratory for excellent collaboration and participation in the study.

Declaration of interest: Dr Haas has received lecturer fee from CSL Behring, Bern, Switzerland. The authors alone are responsible for the content and writing of the paper.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 200.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.