Comparison between ADVIA Chemistry systems Enzymatic Creatinine_2 method and ADVIA Chemistry systems Creatinine method (kinetic Jaffe method) for determining creatinine

, , , , &
Pages 629-636 | Received 20 Dec 2013, Accepted 25 May 2014, Published online: 02 Jul 2014
 

Abstract

Objective. The aim of this study was two-fold: Firstly, to compare the serum creatinine concentration measured by enzymatic method and uncompensated kinetic Jaffe method, and secondly, to compare the effects of certain interfering substances such as glucose, bilirubin, proteins, triglycerides and hemoglobin on creatinine measurement. Methods. The determination of serum creatinine concentrations by enzymatic method and uncompensated kinetic Jaffe method was performed on ADVIA® 2400 analyzer. The interfering substances were tested by adding solutions of interference to serum pool with low (62 μmol/L), medium (221 μmol/L) and elevated (486 μmol/L) creatinine concentration. Results. In the method comparison study, despite the fact that the correlation between both methods for determining serum creatinine is very good (r = 0.998, p < 0.001), the regression analysis revealed that the results are not transferable, as indicated by the slope and intercept, which are significantly different from 1 and 0, respectively. A positive bias of + 14.1% in the determination of serum creatinine by uncompensated kinetic Jaffe method was found, and when the creatinine value is lower than ∼ 180 μmol/L this difference or bis progressively increases. We found a significant positive interference due to proteins and glucose and a significant negative interference due to bilirubin by kinetic Jaffe method, and no interferences by enzymatic method were found. Conclusions. In conclusion, the enzymatic method is the best choice for determining serum creatinine with the ADVIA® 2400 analyzer.

Acknowledgements

The authors are grateful to Fundación Profesor Novoa Santos (Hospital Arquitecto Marcide, Ferrol) for the English review.

Declaration of interest: The authors report no conflict of interest. The authors alone are responsible for the content and writing of the paper.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 200.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.