Assessment of an image-guided neurosurgery system using a head phantom

, , , , &
Pages 606-610 | Received 20 Jan 2015, Accepted 28 Mar 2016, Published online: 21 Apr 2016
 

Abstract

Objective: To acknowledge the challenges and limitations of image-guided neurosurgery systems, we compared the application accuracy of two different image registration methods for one commercial system. (VectorVision, BrainLab, Germany).

Methods: We used an anthropomorphic head phantom for radiosurgery and a custom built add-on to simulate surgical targets inside the brain during an image-guided neurosurgery. We used two image registration methods, fiducial registration using attachable surface markers for computed tomography (CT) and surface registration using infrared laser face scanning. After simulation, we calculated the three-dimensional (3D) distance between the predicted position of a target, and its actual position using a registered pointer and an infrared camera. Deviations were measured for both superficial fiducial markers and internal surgical targets by five different users.

Results: Deviations from the location of fiducial markers after each registration method were 2.15 ± 0.93 mm after CT surface marker registration and 1.25 ± 0.64 mm after infrared face scanner registration. The mean target registration errors were 2.95 ± 1.4 mm using fiducial registration and 2.90 ± 1.3 mm using surface registration. The largest deviations (6.2 mm) were found for the targets in the skull base and posterior cranial fossa. Fiducial deviations and target registration errors were statistically uncorrelated. The total application accuracy was 4.87 ± 0.97 mm after CT surface marker registration and 4.14 ± 0.64 mm after infrared face scanner registration.

Conclusions: Despite others have reported differences, we did not find significant variations between both registration methods for the target registration error, although application accuracy was slightly better after surface face registration. Superficial registration errors, but not the target registration error, can be routinely evaluated in the operating room. Since both errors were uncorrelated, surgeons may neglect the achievable accuracy of the procedure. The described method is recommended to assess application accuracy in the operating room.

Disclosure statement

The authors report no conflict of interest.

Funding information

The authors want to acknowledge the support from project CONACYT-SALUD-2012-181785.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 764.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.