Abstract
Click evoked oto-acoustic emissions (EOAE) at a range of stimulus levels and the auditory brainstem response (ABR) threshold have been measured in 40 infants admitted to a neonatal intensive care unit (NICU). The stimulus levels at which the emissions were first observed had a similar mean value and distribution to those found in previous studies on normal newborns and young adult volunteers. The difference was in the proportion who had no detectable emissions, 16.7% as against 3.9% in the normal newborns and 2.6% in young adult volunteers. The distribution of thresholds for the ABR was again similar to those measured on normal newborns but 8.1 % of the ears of the NICU infants had a threshold worse than 53 dBnHL compared with 0% for the normal newborns. A combined screening method starting with the click evoked oto-acoustic emission is proposed.
A comparison of the mean ABR thresholds for those in whom an EOAE was recorded to those with no EOAE showed a group effect. The stimulus latency relationship of the NV peak showed too much variation between individual results to consider its use to define accurately the degree of conductive hearing loss.