BIODOSIMETRY COMPARISON

Evaluation of the annual Canadian biodosimetry network intercomparisons

, , , , , , , , & show all
Pages 443-451 | Received 23 Jun 2014, Accepted 19 Jan 2015, Published online: 27 Feb 2015

Abstract

Purpose: To evaluate the importance of annual intercomparisons for maintaining the capacity and capabilities of a well-established biodosimetry network in conjunction with assessing efficient and effective analysis methods for emergency response.

Materials and methods: Annual intercomparisons were conducted between laboratories in the Canadian National Biological Dosimetry Response Plan. Intercomparisons were performed over a six-year period and comprised of the shipment of 10–12 irradiated, blinded blood samples for analysis by each of the participating laboratories. Dose estimates were determined by each laboratory using the dicentric chromosome assay (conventional and QuickScan scoring) and where possible the cytokinesis block micronucleus (CBMN) assay. Dose estimates were returned to the lead laboratory for evaluation and comparison.

Results: Individual laboratories performed comparably from year to year with only slight fluctuations in performance. Dose estimates using the dicentric chromosome assay were accurate about 80% of the time and the QuickScan method for scoring the dicentric chromosome assay was proven to reduce the time of analysis without having a significant effect on the dose estimates. Although analysis with the CBMN assay was comparable to QuickScan scoring with respect to speed, the accuracy of the dose estimates was greatly reduced.

Conclusions: Annual intercomparisons are necessary to maintain a network of laboratories for emergency response biodosimetry as they evoke confidence in their capabilities.

Introduction

Biological dosimetry has been employed for many years as a method for estimating the dose of ionizing radiation received by an individual. This information is critical to the medical community as it assists with effective and timely treatment regimens for potentially exposed individuals (CitationWaselenko et al. 2004, International Atomic Energy Agency [IAEA] 2011, CitationSullivan et al. 2013), or for identifying radiation workers who are near or have exceeded their limit for exposure.

To date, several biological markers have been developed to measure radiation-induced damage. Traditionally, the dicentric chromosome assay (DCA), which provides dose estimates based on the frequency of dicentric chromosomes in peripheral blood lymphocytes, has been the method of choice. The DCA is very sensitive due to a low and stable background dicentric frequency (0.5–1 per 1000 metaphase spreads) (IAEA 2011) and is specific to damage from ionizing radiation. Using this assay, dose levels as low as 0.1–0.2 Gy can be detected when 500–1000 metaphase spreads are analyzed, but this requires many hours of analysis (CitationIAEA 2001). In a mass casualty event, however, where medical treatment would be administered only to those receiving more than 2.0 Gy, this level of sensitivity is not required (CitationSullivan et al. 2013). In these situations, the sensitivity of the assay can be reduced by decreasing the number of metaphase cells scored which subsequently greatly reduces the time required for analysis. Standard triage DCA analysis now consists of analyzing only 50 metaphase spreads, providing a threshold of detection of 1–2 Gy; still adequate to guide treatment of acute radiation syndrome (ARS) (CitationLloyd 1997, CitationLloyd et al. 2000, CitationVoisin et al. 2001, CitationInternational Organization for Standardization [ISO] 2008).

The time efficiency of triage-based scoring has been further improved dramatically, without losing accuracy in the dose estimate, by introducing a scoring technique termed ‘DCA QuickScan’ (CitationFlegal et al. 2010, Citation2012). The basis for this method is that individual centromeres are not counted and metaphase spreads are only rapidly examined for obvious damage, thereby eliminating the counting of individual chromosomes to ensure the completeness of the analyzed cell, as done in the conventional DCA (CDCA) method. This method has been demonstrated to be as accurate as conventional triage scoring, while reducing the time for scoring by a factor of about 6 (CitationFlegal et al. 2012).

The cytokinesis block micronucleus (CBMN) assay has also been demonstrated to be a useful tool for biological dosimetry. In this assay, the frequency of micronuclei (MN) in binucleated cells (BNC) is used as a measure of damage from ionizing radiation. Although these MN are not radiation specific, they do increase with dose and have been validated as a technique for the estimation of exposures to radiation (CitationFenech et al. 2003, CitationVral et al. 2011). The advantage of this method over the DCA is that the manual scoring is much faster, as only 200 BNC are required to provide a sensitivity of 1 Gy (CitationMcNamee et al. 2009), and the method requires less technical expertise.

Another strategy for increasing the throughput of biological dosimetry is the development of biodosimetry laboratory networks. Several networks have already been established to improve dose estimation throughput, such as the National Biological Dosimetry Response Plan (NBDRP) in Canada (CitationMiller et al. 2007), the Latin American Biological Dosimetry Network (CitationGarcia et al. 1995) and the Chromosome Network in Japan (CitationYoshida et al. 2007). In addition and the European Network, Realizing the European Network of Biodosimetry (RENEB) is on its way to being established (CitationKulka et al. 2012). When a network is established, it is imperative to perform regular intercomparisons between the laboratories of the network to maintain and assess accuracy and throughput. Many one-time intercomparisons have been conducted over the past few years, both between laboratories within a network (CitationGarcia et al. 1995) and between laboratories from different networks or countries (CitationRoy et al. 2004, CitationWilkins et al. 2008, CitationDi et al. 2011, CitationBeinke et al. 2013) each with different designs. This paper aims to describe the results of intercomparisons held by the National Biodosimetry Response Plan (NBDRP) in Canada over the past 6 years. Exercises of similar design were conducted each year which included the DCA, both conventional and QuickScan and the CBMN assay. These exercises involved each of the four Canadian reference laboratories as well as the occasional participation of two biodosimetry laboratories from the United States. The lessons learned from these intercomparisons will be discussed with an emphasis on demonstrating the importance of repeated intercomparison exercises.

Methods

Blood collection, irradiation and transportation

All blood donors were volunteers who willingly responded to an advertising call for participation in a research protocol approved by Health Canada Research Ethics Board. All donors gave informed consent and none had a recent history of ionizing radiation exposure. For each exercise, blood samples were drawn from each of 10–12 donors (ages 20–60 years), by venipuncture into 4 ml lithium heparinized Vacutainer® tubes (Becton Dickinson, Oakville, ON, Canada). Irradiation of all blood samples was completed ex vivo in the Vacutainer® collection tubes at room temperature. The irradiation system used varied from year to year based on availability as outlined in .

Table I. Irradiation conditions for each intercomparison.

Irradiations at 10–12 different dose points between 0.0 and 5.0 Gy were delivered to each set of samples, respectively, such that each laboratory received matched irradiated samples from the same donors. Immediately after irradiation, samples were incubated for 2 h at 37°C to allow for repair and blinded bar coded samples from each dose-point were sent to each of the participating laboratories: Health Canada (HC), Defence Research and Development Canada-Ottawa Research Centre (DRDC), Canadian Nuclear Laboratories Limited (CNL), McMaster University, Oak Ridge Institute for Science Education (ORISE) and the Armed Forces Radiobiology Research Institute (AFRRI). The laboratories were informed of the radiation quality prior to the intercomparison. Samples were dispatched either by air with FedEx (AFRRI, ORISE, McMaster) or transported by road (DRDC, CNL). Shipment procedures simulated those during an actual event and followed the Canadian Transportation of Dangerous Goods Regulations for Class 6.2, UN3373 and labelling for Biological Substances Category B. Specifically, the Vacutainer® tubes were shipped according to packaging instructions 650 of International Air Transportation Association (IATA). The package was prominently labeled to indicate that it should not be frozen or X-rayed at airport security checkpoints. As a quality control measure, the package included a temperature data logger for monitoring temperature, which showed that the samples remained between 14 and 27°C during transit and an optically stimulated luminescence (OSL) dosimeter to rule out X-ray screening at airports.

Cell culture and harvest

Although sample processing protocols varied slightly from laboratory to laboratory, for both the DCA and CBMN assays, they were based on the general guidance provided by the IAEA (CitationIAEA 2001, Citation2011) and ISO 19238 and 21243 (CitationISO 2004, Citation2008).

Dicentric chromosome assay

Each laboratory set up a whole blood or isolated lymphocyte culture for harvesting metaphase spreads. The cells were added to culture medium (RPMI 1640) containing 15% fetal bovine serum, with L-glutamine, penicillin, streptomycin and 15 μM BrdU and stimulated to cycle by the addition of 2% phytohemagglutinin (PHA). The cells were incubated at 37°C and 5% CO2 for 48 h, mitotic arrest was done using 1% colcemid at 10 mg/ml at 44 h (for HC) and 48 h (for DRDC) and then harvested for metaphase spreads to determine the yield of radiation-induced dicentrics. Only first-division metaphase spreads were used for counting dicentrics. The standard method used to ensure that only first-division metaphase spreads were scored was based on either fluorescence-plus-Giemsa (FPG) staining technique coupled with BrdU and Hoechst 33258 or on the addition of cytochalasin B to the cultures after 24 h to inhibit cytokinesis. The FPG staining allows identification by the differential staining of second-division metaphase spreads. The metaphase spreads were harvested after a brief treatment in a suitable hypotonic solution such as 0.56% (0.075 mM) potassium chloride and fixation in 3:1 methanol: glacial acetic acid Carnoy's fixative. A temperature and humidity-controlled chamber was used to prepare metaphase spreads on glass slides. Multiple slides were prepared for each dose and stained with Giemsa or FPG for chromosome aberration analysis using brightfield microscopy.

In each laboratory, sufficient slides were prepared such that each scorer could score up to 50 metaphase cells or 30 dicentrics on one slide. Conventional DCA scoring followed the recommendations of ISO (CitationISO 2004) and IAEA (IAEA 2011), and the method for QuickScan scoring was as described by CitationFlegal et al. (2012) except without the stipulation that examination stopped if five dicentrics were seen in less than 20 metaphases.

Cytokinesis block micronucleus assay

Cell culture was performed similarly to the DCA except that no colcemid was added. Instead cytochylasin B was added after 48 h of incubation to inhibit cytokinesis, and cells were cultured for a total of 72 h. Cells were first fixed with 5:1 methanol: glacial acetic acid and then in 2.5% of 37% formaldehyde and slides were prepared similar to DCA. Slides were stained immediately before scoring with 50 μg/ml acridine orange. Details of the procedure can be found in CitationMcNamee et al. (2009). Manual scoring was performed by all laboratories under fluorescent microscopy under 630 × magnification. The number of MN in 200 BNC was scored according to the criteria of CitationFenech et al. (2003).

Dose estimates

Dose estimates were made based on each laboratory's own calibration curves for the appropriate assay. The dose-response curve from each laboratory was constructed using the conventional weighted Poisson regression model, Y = c + βD+αD2, where Y is the number of dicentrics/number of metaphase spreads scored or the number of MN/BNC, c is the background value, D is the radiation dose in Gray (Gy); and β and α are dose and dose-squared coefficients used to estimate the rate of dicentrics in metaphase spreads or the rate MN/BNC. Maximum likelihood estimation was used to estimate the parameters of the fitted curves using either CABAS (CitationDeperas et al. 2007) or Dose Estimate (CitationAinsbury and Lloyd 2010).

Intercomparison design

Intercomparisons have been conducted annually within the Canadian biodosimetry network for the past 6 years. Each year the design of the intercomparison varied slightly but in general, it involved 10–12 ex vivo irradiated blood samples being blinded and shipped to each of the participating laboratories. Each laboratory was asked to identify the dose delivered to each sample using CDCA, QuickScan DCA and CBMN depending on which methods had been established in their laboratories. For DCA, data was recorded after scoring 20 and 50 cells (or 30 dicentrics) except for the first year where data was only recorded after 50 cells or 30 dicentrics. For CBMN, 200 BNC were scored. The time to score each sample was also recorded. Each laboratory was requested to have as many scorers as possible score each sample. A summary of the participation is found in . Although a laboratory may have participated in the intercomparison and is included in the table, their results were only included in the analysis if at least two scorers from their laboratory participated in the intercomparison.

Table II. Summary of exercise participation.

Statistical analysis

Analysis was performed to evaluate laboratory performance, dose estimates between methods in general and scoring methods between laboratories. The dose estimation was first assessed for consistency between laboratories within method of scoring and number of cells scored in order to compare the performance of the laboratories. Secondly, performance statistics were evaluated to compare dose estimates from the various methods and number of cells scored to the physical dose of radiation (reference value). Thirdly, the methods of scoring were compared between laboratories. In addition, dose estimates were compared to the actual dose and considered correct when within 0.5 Gy of the actual dose. Slides made from samples exposed to different doses of radiation were analyzed independently, and these are referred to as ‘SampleID’.

Dose estimation

As previously mentioned, after slides were scored, each scorer estimated a dose of exposure based on their laboratory's calibration curve. Scorers only scored one replicate for each method of scoring and SampleID. Due to the limited number of replicates, the analysis to compare dose estimates is restricted at the laboratory level. Estimates of dose based on the calibration curves are assumed to be asymptotically normally distributed (CitationCasella and Berger 2002). For this reason only laboratories with a minimum of two scorers were included in the analysis for comparing dose estimation results between laboratories or methods. Furthermore each year was analyzed separately.

An analysis of variance (ANOVA) model was used to assess differences in dose estimation between and within each method of scoring and number of cells scored. The assumptions for ANOVA (residuals are normally distributed, with constant variance between groups) were verified for each ANOVA model using Anderson-Darling test for normality and Levene's test for homogeneity. When the assumptions were not satisfied for the original scale of the data, then non-parametric Kruskal-Wallis test was applied. If the results were similar for the parametric and non-parametric approaches then the results based on the parametric approaches (assuming normality and constant variance across groups) were followed, indicating that the assumptions were adequately satisfied (CitationMontgomery 2000). Pair-wise comparisons were conducted when the laboratory effect was significant (p < 0.05), using Tukey's multiple comparison tests in order to control the overall Type I error to be less than 0.05.

Performance statistics z and u for dose estimations (at the laboratory level)

The performance statistic z was applied to the dose estimation data to measure the deviation of each laboratory's estimated dose from the robust average. The robust average was determined using Algorithm A from ISO 5725-5:1998 (CitationISO 1998) which is currently suggested for proficiency testing to minimize the influence of outliers.

The performance statistics z is described here for comparing laboratories. For each laboratory, a z value was calculated using Equation (1):

where di is the reported estimated dose from the ith laboratory, and dref, and sref are the robust average and standard deviation (SD), respectively, as obtained from Algorithm A. The reference group from which dref and sref are evaluated is based on the CDCA method after scoring 50 cells. When the physical dose of radiation is known, then this is taken to be the reference value dref. The term uref is the standard uncertainty of dref and is calculated as:

where, p is the number of participating laboratories. The value uref is the uncertainty on the physical dose delivered. For each year of analysis, uref was considered as negligible if the following criterion was satisfied (Equation 3):

To evaluate laboratory performance, z statistics were interpreted as follows: |z| values ≤ 2 were considered to be satisfactory, between 2 and 3 ‘questionable’ and ≥ 3 ‘unsatisfactory’.

Results

Typical results from an intercomparison from a single year are shown in for CDCA, QuickScan DCA and CBMN assays. These figures show the results after scoring 50 cells for DCA and 200 cells for CBMN. Similar results after scoring 20 cells for DCA are not shown. The black solid lines represent the ± 0.5 Gy range from the actual dose delivered. Data from each laboratory is represented by a different symbol. The data from multiple years has been analyzed to provide an overview of the results from several exercises for the purposes of assessing biological trends. Examples of the statistical analysis are shown with a summary of the data in bar chart format below.

Figure 1. Intercomparisons across laboratories of estimated doses obtained using (A) conventional dicentric chromosome assay (CDCA) analysis of 50 metaphase spreads, (B) QuickScan DCA of 50 metaphase spreads, and (C) cytokinesis block micronucleus (CBMN) analysis of 200 binucleated cells. Each data point is a dose estimate from one individual, with scorers from each laboratory shown using the same symbol. The solid lines represent ± 0.5 Gy intervals.

Figure 1. Intercomparisons across laboratories of estimated doses obtained using (A) conventional dicentric chromosome assay (CDCA) analysis of 50 metaphase spreads, (B) QuickScan DCA of 50 metaphase spreads, and (C) cytokinesis block micronucleus (CBMN) analysis of 200 binucleated cells. Each data point is a dose estimate from one individual, with scorers from each laboratory shown using the same symbol. The solid lines represent ± 0.5 Gy intervals.

Comparison of laboratories within each method of scoring and number of cells scored

The first analysis was a comparison of the agreement in the dose estimate from each laboratory within each method and for each number of cells scored. Sample results are shown in for sample 2 from 2008. For each sample from each year, the average dose estimate from each laboratory and method was calculated and compared using ANOVA. In order to generate , a tally was made of the samples in which there was insufficient evidence to reject the null hypothesis in all laboratories (p > 0.05). As demonstrated in , the dose estimate from all laboratories agreed within each method except for the CBMN assay. In this case, the dose estimate from Laboratory 1 was statistically greater than that from Laboratory 2. Where data is missing, either that laboratory did not perform the assay or only one scorer analyzed the sample with that method.

Figure 2. Illustration of the percentage of samples with dose estimates in agreement between laboratories for each year for each endpoint. A tally was prepared of the samples in which all laboratories were in agreement of the dose estimate (p > 0.05). The average dose estimate from each laboratory and method was calculated and compared using ANOVA.

Figure 2. Illustration of the percentage of samples with dose estimates in agreement between laboratories for each year for each endpoint. A tally was prepared of the samples in which all laboratories were in agreement of the dose estimate (p > 0.05). The average dose estimate from each laboratory and method was calculated and compared using ANOVA.

Table III. Sample ANOVA analysis of data from year 2008; dose delivered = 1.8 Gy.

shows the agreement between laboratories for each year for each endpoint. For all assays except the CBMN assay, dose estimates from all laboratories were in agreement for more than 60% of the samples and in 19 out of these 22 cases, agreement occurred in 80% or more of the cases. This can be compared to a similar analysis based on the percent of correct dose estimates as those being within 0.5 Gy of the dose delivered to the sample () which also shows the percentage of samples over and underestimated. Similarly, both variations of the DCA performed better than the CBMN, however, based on this criteria, the CBMN assay was consistently correct at least 55% of the time in all years tested. also demonstrates that a greater number of samples were overestimated rather than underestimated.

Figure 3. Illustration of the percentage of samples with dose estimates (A) within 0.5 Gy of the given dose, (B) more than 0.5 Gy over the given dose, and (C) more than 0.5 Gy under the given dose.

Figure 3. Illustration of the percentage of samples with dose estimates (A) within 0.5 Gy of the given dose, (B) more than 0.5 Gy over the given dose, and (C) more than 0.5 Gy under the given dose.

illustrates the time it took to score one sample averaged over all scorers and all years except 2007 when no times were recorded. It is evident that scoring 50 cells by CDCA scoring was the most time-consuming, requiring almost 1 h to score a single sample. The time to score decreased for all of the other endpoints, with QuickScan scoring of 20 cells being the quickest, requiring about 10 min per sample.

Figure 4. Illustration of the average time required to score one sample for each method.

Figure 4. Illustration of the average time required to score one sample for each method.

Performance statistics z for dose estimations

Performance statistic z for the dose estimates was calculated using the physical dose of radiation as the reference value for the robust average (dref). The reference standard deviation (sref) was based on the CDCA method after scoring 50 cells. The objective was to define a reference interval to allow comparison of the effectiveness of the QuickScan and CBMN method with respect to the CDCA method, as well as compare the effectiveness of scoring 20 versus 50 cells (or 200 cells in the CBMN method). Values of |z| ≤ 2 were considered to be satisfactory. Sample data from this analysis is presented in .

Table IV. Performance statistics z sample analysis of data from year 2008; dose delivered = 1.8 Gy.

Similar to above, the data from this analysis was tallied to get a better overall view of the results. Samples with values of |z| ≤ 2 were tallied as being satisfactorily close to the physical dose. shows the tallied data for all laboratories and all methods as a function of intercomparison year. If the data from 2010 is excluded, it can be seen that scoring 50 cells using CDCA (CON50) was extremely successful (98%) and scoring 20 cells using CDCA (CON20) was in agreement with the physical dose 90% of the time. QuickScan was also in agreement with the physical dose at least 90% of the time after scoring either 20 or 50 cells. Only the CBMN assay fell below 80% agreement in two of the intercomparison years.

Figure 5. A comparison of the percentage of correctly evaluated samples based on a |z| ≤ 2 for each method across all years of the intercomparisons.

Figure 5. A comparison of the percentage of correctly evaluated samples based on a |z| ≤ 2 for each method across all years of the intercomparisons.

In 2010, the data reflects a change in the exposure conditions that was not matched with the appropriate change in calibration curves. This data set was also separated by laboratory to examine how each laboratory performed each method and how their performance changed year to year. shows the analysis of how each laboratory performed for each method averaged over all years. Similarly to , most endpoints resulted in dose estimates similar to the physical dose except in the case of the CBMN assay that had as low as 60% agreement. The data is also presented for each laboratory based on yearly performance over all endpoints (). In general, each laboratory's performance was maintained from year to year (except 2010) with some small fluctuations. Data is missing for Laboratories 5 and 6 due to only one scorer performing the analysis or that laboratory not participating in those years.

Figure 6. A comparison of the percentage of correctly evaluated samples based on a |z| ≤ 2 for each method in each laboratory.

Figure 6. A comparison of the percentage of correctly evaluated samples based on a |z| ≤ 2 for each method in each laboratory.

Figure 7. A comparison of the percentage of correctly evaluated samples based on a |z| ≤ 2 for each year in each laboratory.

Figure 7. A comparison of the percentage of correctly evaluated samples based on a |z| ≤ 2 for each year in each laboratory.

Discussion

The Canadian NBDRP was conceived in 2002 and by 2005 became well established as a network comprised of four reference laboratories in addition to 18 satellite laboratories in existing hospital genetics departments who had been trained to provide surge capacity for the scoring of dicentrics (CitationMiller et al. 2007). By 2007, a program of annual intercomparisons to maintain the expertise and confirm the capabilities and capacity of the network had been established. In 2008, two laboratories in the United States joined the network to create a North American Biodosimetry Network and periodically participated in the Canadian annual intercomparisons. This paper summarizes these outcomes and describes the importance of performing such intercomparisons.

Overall, the results of these intercomparisons demonstrate the success of the NBDRP in terms of capabilities, capacities and accuracy in dose estimates being well maintained from year to year. The number of participating scorers fluctuated, depending mostly on whether or not the US laboratories took part in the intercomparison. There was some minor fluctuation within the Canadian laboratories which was due to expected changes in staffing. In 2010, the exposure conditions were not well matched to the calibration curves at each laboratory resulting in a large number of overestimated doses. This highlights the importance of having the appropriate calibration curves for intercomparisons in order to effectively assess the capabilities of the participating laboratories. However, even with poorly matched calibration curves, samples that were exposed were clearly identified and the results would be sufficient for mass casualty events.

These intercomparisons provided an opportunity to compare the agreement in dose estimates between each laboratory based on scoring methods, numbers of cells scored and assays used. It was clear that, of all the assays performed, the CBMN assay had the lowest agreement between laboratories. Although there was agreement only 30–60% of the time with the CBMN based on the ANOVA models (comparing dose estimates between laboratories), when the percent of satisfactory dose estimates was determined by laboratory based on z-score analysis (comparing dose estimates to the physical dose), the CBMN assay was satisfactory at least 60% and usually over 80% of the time. There are several reasons for the reduced performance of the CBMN assay including greater inter-individual variation in the sample donor background levels and response to ionizing radiation. These can lead to larger variations in the calibration curves between laboratories which will result in an increased range of dose estimates reported (CitationFenech 2010). The results also demonstrated that the QuickScan DCA analysis, when 50 cells were scored, was as accurate, if not better than CDCA scoring based on agreement in dose estimates between laboratories. When compared to conventional scoring of 50 cells, QuickScan scoring of 50 cells produced accurate dose estimates for most samples. Even when only 20 cells were enumerated with either conventional or QuickScan scoring, satisfactory dose estimates were made, on average, in over 90% of the samples. In addition, when scoring time of the samples is taken into account, QuickScan scoring of 20 cells required on average 10 min per sample as compared to nearly 30 min per sample for the conventional scoring of the same number of cells and about 15 min for CBMN scoring of 200 cells. In summary, these results indicate that for a mass casualty situation, scoring 20 cells using QuickScan would provide an excellent triage dose estimation method.

The fluctuation in results from year to year emphasizes the importance of annual intercomparisons. These fluctuations are likely due to changes in staff, technical experience with the assays and amount of biodosimetry a laboratory performed between intercomparisons. The laboratory with the most consistent results from year to year (laboratory 3) was also the laboratory with the smallest turnover in staff and the longest history of performing biological dosimetry. Laboratory 1 also had a small turnover in staff but less experience before the formation of the network. These observations indicate that for laboratories that do not perform biodosimetry on a regular basis, annual testing provides an opportunity to assess their capabilities and practice their skills. Also, when new staff is hired, it can be used to ensure adequate training has been performed.

These intercomparisons are also an excellent opportunity to identify gaps in the processes required to receive and analyze multiple samples on short notice. They have helped the Network draft and revise standard operating procedures as ‘lessons learned’ were identified and addressed each year. For example, many issues with shipping arose with each intercomparison which has resulted in well- defined procedures and protocols for packaging, labelling and shipping biological samples by air, both within Canada and internationally.

Conclusion

Maintenance of a network of laboratories for emergency response biodosimetry needs to be continuous and rigorous in order to have confidence in the dose estimates being generated. One way to maintain the capacity and capability of a network is through annual intercomparisons which evaluate the abilities and expertise within the network and allow for an opportunity to test protocols and practice procedures. This type of intercomparison is necessary to maintain confidence that the network is in a state of readiness for emergency response. This study has demonstrated, through 6 years of intercomparisons, that the Canadian biodosimetry network is capable of producing dose estimations over a variety of different assays quickly and accurately. These findings provide confidence to the medical community, the public and government bodies that in the event of a nuclear accident, biodosimetry can be applied to manage and medically treat casualties to ensure the minimization of health risks.

Acknowledgements

The authors would like to thank those who contributed to the analysis of samples at each participating laboratory: Nancy Ringuette, Pascale Bellier, Dr J. McNamee (Health Canada); Louise Prud'homme-Lalonde, Aimee Jones, Kate Daniel, Erin Redmond (DRDC); Yvonne Devantier, Dale Buchanan, Laura Paterson, Rebecca Mantha (CNL); Lisa Laframboise, Crystal Palumbo, Dr D. Boreham (McMaster University); Mollie Abbot (ORISE); Uma Subramanian, Dr A. Ramakumar (AFFRI).

Declaration of interest

The authors report no conflicts of interest. The authors alone are responsible for the content and writing of the paper.

References

  • Ainsbury EA, Lloyd DC. 2010. Dose estimation software for radiation biodosimetry. Health Phys 98:290–295.
  • Beinke C, Barnard S, Boulay-Greene H, De AA, De SS, Herodin F, Jones A, Kulka U, Lista F, Lloyd D, Martigne P, Moquet J, Oestreicher U, Romm H, Rothkamm K, Valente M, Meineke V, Braselmann H, Abend M. 2013. Laboratory intercomparison of the dicentric chromosome analysis assay. Radiat Res 180:129–137.
  • Casella G, Berger RL. 2002. Statistical inference, 2nd ed. Belmont, CA: Druxbury Press.
  • Deperas J, Szluinska M, Deperas-Kaminska M, Edwards A, Lloyd D, Lindholm C, Romm H, Roy L, Moss R, Morand J, Wojcik A. 2007. CABAS: A freely available PC program for fitting calibration curves in chromosome aberration dosimetry. Radiat Prot Dosimetry 124:115–123.
  • Di GM, Barquinero JF, Vallerga MB, Radl A, Taja MR, Seoane A, De LJ, Oliveira MS, Valdivia P, Lima OG, Lamadrid A, Mesa JG, Aguilera IR, Cardoso TM, Carvajal YC, Maldonado CA, Espinoza ME, Martinez-Lopez W, Mendez-Acuna L, Di Tomaso MV, Roy L, Lindholm C, Romm H, Guclu I, Lloyd DC. 2011. Biological dosimetry intercomparison exercise: An evaluation of triage and routine mode results by robust methods. Radiat Res 175:638–649.
  • Fenech M. 2010. The lymphocyte cytokinesis-block micronucleus cytome assay and its application in radiation biodosimetry. Health Phys 98:234–243.
  • Fenech M, Chang WP, Kirsch-Volders M, Holland N, Bonassi S, Zeiger E. 2003. HUMN project: Detailed description of the scoring criteria for the cytokinesis-block micronucleus assay using isolated human lymphocyte cultures. Mutat Res 534:65–75.
  • Flegal FN, Devantier Y, Marro L, Wilkins RC. 2012. Validation of QuickScan dicentric chromosome analysis for high throughput radiation biological dosimetry. Health Phys 102:143–153.
  • Flegal FN, Devantier Y, McNamee JP, Wilkins RC. 2010. Quickscan dicentric chromosome analysis for radiation biodosimetry. Health Phys 98:276–281.
  • Garcia OF, Ramalho AT, Di GM, Mir SS, Espinoza ME, Manzano J, Nasazzi N, Lopez I. 1995. Intercomparison in cytogenetic dosimetry among five laboratories from Latin America. Mutat Res 327:33–39.
  • International Atomic Energy Agency (IAEA). 2001. Cytogenetic analysis for radiation dose assessment: A manual. Vienna: IAEA. Available from: http://www-pub.iaea.org/MTCD/publications/PDF/TRS405_scr.pdf
  • International Atomic Energy Agency (IAEA). 2011. Cytogenetic dosimetry: Applications in preparedness for and response to radiation emergencies. EPR-Biodose 2011. Vienna: IAEA. Available from: http://www-pub.iaea.org/MTCD/Publications/PDF/EPR- Biodosimetry%202011_web.pdf
  • International Organization for Standardization (ISO). 1998. Accuracy (trueness and precision) of measurement methods and results – Part 5: Alternative methods for the determination of the precision of a standard measurement method. Geneva: ISO 5725-5:1998.
  • International Organization for Standardization (ISO). 2004. Radiation protection – performance criteria for service laboratories performing biological dosimetry by cytogenetics. Geneva: ISO 19238:2014.
  • International Organization for Standardization (ISO). 2008. Radiation protection – performance criteria for laboratories performing cytogenetic triage for assessment of mass casualties in radiological or nuclear emergencies. General principles and application to dicentric assay. Geneva: ISO 21243:2008.
  • Kulka U, Ainsbury L, Atkinson M, Barquinero JF, Barrios L, Beinke C, Bognar G, Cucu A, Darroudi F, Fattibene P, Gil O, Gregoire E, Hadjidekova V, Haghdoost S, Herranz R, Jaworska A, Lindholm C, Mkacher R, Mortl S, Montoro A, Moquet J, Moreno M, Ogbazghi A, Oestreicher U, Palitti F, Pantelias G, Popescu I, Prieto MJ, Romm H, Rothkamm K, Sabatier L, Sommer S, Terzoudi G, Testa A, Thierens H, Trompier F, Turai I, Vandersickel V, Vaz P, Voisin P, Vral A, Ugletveit F, Woda C, Wojcik A. 2012. Realising the European Network of Biodosimetry (RENEB). Radiat Prot Dosimetry 151: 621–625.
  • Lloyd DC. 1997. Chromosomal analysis to assess radiation dose. Stem Cells 15(Suppl. 2):195–201.
  • Lloyd DC, Edwards AA, Moquet JE, Guerrero-Carbajal YC. 2000. The role of cytogenetics in early triage of radiation casualties. Appl Radiat Isot 52:1107–1112.
  • McNamee JP, Flegal FN, Greene HB, Marro L, Wilkins RC. 2009. Validation of the cytokinesis-block micronucleus (CBMN) assay for use as a triage biological dosimetry tool. Radiat Prot Dosimetry 135:232–242.
  • Miller SM, Ferrarotto CL, Vlahovich S, Wilkins RC, Boreham DR, Dolling JA. 2007. Canadian Cytogenetic Emergency Network (CEN) for biological dosimetry following radiological/nuclear accidents. Int J Radiat Biol 83:471–477.
  • Montgomery DC. 2000. Design and analysis of experiments, 5th ed. New York: John Wiley and Sons.
  • Roy L, Buard V, Delbos M, Durand V, Paillole N, Gregoire E, Voisin P. 2004. International intercomparison for criticality dosimetry: The case of biological dosimetry. Radiat Prot Dosimetry 110:471–476.
  • Sullivan JM, Prasanna PG, Grace MB, Wathen LK, Wallace RL, Koerner JF, Coleman CN. 2013. Assessment of biodosimetry methods for a mass-casualty radiological incident: Medical response and management considerations. Health Phys 105:540–554.
  • Voisin P, Benderitter M, Claraz M, Chambrette V, Sorokine-Durm I, Delbos, M, Durand V, Leroy A, Paillole N. 2001. The cytogenetic dosimetry of recent accidental overexposure. Cell Mol Biol (Noisy-le-grand) 47:557–564.
  • Vral A, Fenech M, Thierens H. 2011. The micronucleus assay as a biological dosimeter of in vivo ionising radiation exposure. Mutagenesis 26:11–17.
  • Waselenko JK, Macvittie TJ, Blakely WF, Pesik N, Wiley AL, Dickerson WE, Tsu H, Confer DL, Coleman CN, Seed T, Lowry P, Armitage JO, Dainiak N. 2004. Medical management of the acute radiation syndrome: Recommendations of the Strategic National Stockpile Radiation Working Group. Ann Intern Med 140:1037–1051.
  • Wilkins RC, Romm H, Kao TC, Awa AA, Yoshida MA, Livingston GK, Jenkins MS, Oestreicher U, Pellmar TC, Prasanna PG. 2008. Interlaboratory comparison of the dicentric chromosome assay for radiation biodosimetry in mass casualty events. Radiat Res 169:551–560.
  • Yoshida MA, Hayata I, Tateno H, Tanaka K, Sonta S, Kodama S, Kodama Y, Sasaki MS. 2007. The Chromosome Network for biodosimetry in Japan. Radiat Measure 42:1125–1127.