Point of care testing provides an accurate measurement of creatinine, anion gap, and osmolal gap in ex-vivo whole blood samples with nitromethane

, , , &
Pages 611-617 | Received 11 Feb 2014, Accepted 21 Apr 2014, Published online: 20 May 2014
 

Abstract

Context. Nitromethane interferes with Jaffé measurements of creatinine, potentially mimicking acute kidney injury. Objectives. We determined the proportional contribution of nitromethane in blood samples to creatinine measured by the Jaffé colorimetric and the point-of-care (POC) reactions and determined whether the difference can reliably estimate the concentration of nitromethane. Additionally, we determined whether the presence of nitromethane interferes with anion/osmolal gaps and ascertained the stability of nitromethane in serum after 7 days. Methods. Nitromethane was added to whole blood from four healthy volunteers to achieve concentrations of 0, 0.25, 0.5, 1, and 2 mmol/L. The following tests were performed: creatinine (Jaffé and POC), electrolytes (associated with Jaffé and POC), osmolality and nitromethane concentration (gas chromatography [GC]). Remaining samples were refrigerated and reanalyzed using GC at 7 days. Anion and osmolal gaps were calculated. Proportional recovery and degradation of nitromethane were measured using GC. Data were analyzed for agreement with single-factor ANOVA (p = 0.05). Results. Mean creatinine for POC and Jaff methods were 0.93 vs. 0.76 mg/dL, respectively. Jaff creatinine concentrations increased linearly with increasing nitromethane concentrations (R2 = 1, p = 0.01): measured creatinine (mg/dL) = 7.1*nitromethane (mmol/L) = 0.79. POC creatinine remained unchanged across the range of nitromethane concentrations (p = 0.99). Anion and osmolal gaps also remained unchanged. Nitromethane was reliably identified in all sample concentrations using GC on Day 0. Detection of 0.25 mmol/L nitromethane was not consistently recovered on Day 7. Nitromethane degradation was most pronounced at 2 mmol/L concentrations (81% recovery). Conclusions. Nitromethane alters apparent concentration of creatinine using the Jaffé reaction in a linear fashion but not when using the POC reaction. Measured difference between Jaffé and POC creatinine may identify the presence and estimate concentration of nitromethane. Presence of nitromethane did not alter the anion or osmolal gap; thus it would not potentially interfere with the diagnosis of co-exposure to a toxic alcohol.

Acknowledgments

We would like to thank the laboratory staff and our technician, Jay Weisenberger, for providing insight into the techniques and running our sample data.

Declaration of interest

The authors report no declarations of interest. The authors alone are responsible for the content and writing of the paper.

This study was funded by a research training award from the Emergency Medicine Foundation and Medical Toxicology Foundation to DC, with authors MCB, AMM and WPK serving as mentors for this training award. SM, AMM, WPK and MCB report no declarations of interest.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 1,501.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.