2,139
Views
18
CrossRef citations to date
0
Altmetric
Technical Papers

A Comparative Study of the Performance of Passive Samplers

, , &
Pages 260-268 | Published online: 10 Oct 2011

ABSTRACT

Atmospheric concentrations of benzene, toluene, ethylbenzene, and xylenes (o-xylene and m,p-xylene) were assessed in the Tricity area (Gdańsk-Sopot-Gdynia, Poland) with the use of two types of passive samplers: permeation (homemade passive samplers) and diffusive (Radiello and Orsa 5). Samples were collected during 2008 at selected sites in the Tricity area at monitoring stations belonging to the agency of Regional Air Quality Monitoring Foundation. The field study was conducted to compare the performance of these two different types of passive samplers. A statistical approach was formulated, and the experimental data were evaluated using the paired t test, Wilcoxon signed rank-sum test, and Friedman analysis of variance. All the statistical results confirm the hypothesis that the differences between the performances of the three sampling devices are highly significant. Despite the fact that data obtained with the homemade passive sampler indicated that the results were higher compared with those for the Radiello and Orsa 5 diffusive samplers, the authors note that all differences between the homemade permeation sampler and the Radiello and Orsa 5 diffusive samplers are positive.

IMPLICATION

This study examined the performance of two types of passive samplers (permeation and diffusive type) for the long-term characterization of time-weighted mean concentration of analytes in a given environmental compartment. The results of this evaluation demonstrated that passive sampling was found to be suitable for consideration as part of an emerging strategy for monitoring a wide spectrum of priority pollutants in ambient air. The samples of analytes obtained can serve as a source of information in reference to creating the precise distribution of pollutant concentrations in both time and space and analyzing trends in pollutant concentrations in outdoor air. The data obtained may aid in the identification of analyte sources and modeling air pollution transport.

INTRODUCTION

The elevated levels of volatile organic compounds (VOCs) in the air over many conurbations have spawned numerous attempts on various scales to clean up the air. Air quality management systems have been set up to supply information on concentration levels of selected atmospheric pollutants. Levels of analytes detected in samples collected at particular places and times are increasingly being used in models for predicting the spread of pollutants and in compiling distribution maps of pollutant levels. For the persons responsible for environment quality control, the biggest problem is choosing the proper field sampling technique, which should have the following features: detection of an appropriate concentration range of target analytes, known accuracy and precision of measurements, clearly defined influence of interferents on measurement results, simplicity and rapidity of measurement, and low running costs.

Passive sampling is currently one of main areas of development in analysis, particularly in the monitoring of environmental pollutants. The main advantage of passive sample collection/enrichment technology is that in situ sample collection is very simple, which is particularly important in the case of long-term measurements and is also the reason that passive sampling appears to be an interesting alternative to the routinely used dynamic methods. Passive air sampling can provide atmospheric concentrations of pollutants over a period of months and simultaneously at a number of sites at low cost. With this method, pollutant concentrations are integrated over the whole exposure time, making them immune to accidental, extreme variations. Passive sampling can be used to investigate the atmospheric concentration of organic pollutants not only at the local scaleCitation1,Citation2 but also at the continental and global scales.Citation3 Information obtained in this way is suitable for producing a long-term overview of pollutant concentration levels.Citation4,Citation5 The time-weighted average concentrations of analytes obtained by passive sampling give a better picture than short-term concentrations of the effect of pollutants on human health, because they reflect the long-term influence of pollutants on human health.

The principle of operation of passive samplers is based on mass transport, described by Fick's first law of diffusion. Passive samplers fall into one of two categories, depending on the type of diffusion barrier usedCitation6,Citation7: passive diffusion samplers in which the transport of analytes takes place by way of free diffusion of the analyte through a stagnant gas layer; and passive permeation samplers, in which the transport of analytes takes place by way of permeation of the analyte through a semipermeable membrane.

To a large extent, the type of analytical information obtained using passive sampling technology depends on the accumulation regimen in which passive samplers operate during field exposure. In general, a passive sampling device is designed to operate in two different accumulation regimens:

In the kinetic and time-integrative uptake phase (rate of mass transfer to the receiving phase is linearly proportional to the difference in chemical potential of the contaminant in the receiving phase and sample). These kinds of passive samplers are called linear uptake passive samplers.

In the at-equilibrium regimen, described by the partition coefficient between the receiving phase and the sample matrix. These kinds of passive samplers are called equilibrium passive samplers.

The objective of this study was to obtain information on the level of atmospheric air pollution due to BTEX (benzene, toluene, ethylbenzene, and xylenes) in the Tricity area (Gdańsk-Sopot-Gdynia, Poland) by means of passive sampling (dosimetry). Because of the diversity of passive samplers available on the market, an attempt has been made to assess the applicability of these samplers at different sampling rates of analytes and on the basis of different phenomena (diffusion and permeation) of mass transport for continuous urban air quality control. To accomplish this goal, a statistical approach was formulated, and the experimental data were compared using the paired t test, Wilcoxon signed rank-sum test, and Friedman analysis of variance (ANOVA).

MATERIALS AND METHODS

Study Design

In the assessment of atmospheric air quality, the choice of sampling points determines whether the sample will be representative and whether it will reflect the real state of the monitored air. Therefore, to obtain reliable estimates of background BTEX pollution of the atmospheric air in and around the Tricity conurbation, sampling points that would reflect the actual state of the environment had to be chosen. The locations of the sampling sites were chosen so that no point or linear sources of emission were in the direct vicinity of the sites and surrounding areas. Thus, 10 monitoring stations designated by the Regional Air Quality Monitoring Foundation were used. On the basis of results of long-term studies, the localization of these stations was determined by factors such as weather conditions, human population density, and existing databases on the emissions of pollutants from point and surface sources.

To evaluate the atmospheric average urban background BTEX concentration in the Tricity area, 12 1-month passive sampling campaigns were conducted (for the Orsa 5 passive sampler and homemade permeative passive sampler). To avoid saturation of the sorbent, the exposure time of the Radiello passive sampler was shortened to 2 weeks and, consequently, 24 exposures were conducted.

Passive Sampler

For isolation and enrichment of the analytes from atmospheric air, three sampling devices were used:

A homemade, badge-type permeation passive sampler equipped with a 50-μm-thick semipermeable membrane of polydimethylsiloxane (SSPM100, Speciality Silicone Products Inc.). The sorbent was activated charcoal (40–60 mesh; Gryfskand, Gryfino, Poland), which has a large specific surface area (>1000 m2/g) and hence a large sorption capacity and is not particularly sensitive to variations in temperature or atmospheric pressure. These properties of activated charcoal make it ideally suited for long-term sample collection; analyte enrichment, a significant parameter where passive sampling is concerned, takes place concurrently.Citation8,Citation9

An Orsa 5 (Dräger, Lübeck, Germany), a tube-type diffusive passive sampler with a short diffusion path length (L). The sorbent is activated charcoal (made from coconut husks; granulation 0.4–0.8 mm), bounded on either side by a cellulose acetate diffusion barrier, which defines the diffusion length within the sampler.Citation10

A Radiello radial diffusive passive sampler, designed and developed by the Fondazione Salvatore Maugieri (Padua, Italy). This sampling system is made up of a cylindrical adsorbing cartridge housed coaxially inside a cylindrical diffusive body of polycarbonate and microporous polyethylene. It combines different characteristics common to both types: the diffusive surface is the cylinder itself, the sampling area is large, and the diffusion length is short. Diffusion is radial, and the analyte passes across a microporous cylinder before reaching an inner stainless steel net cylinder containing 300 mg of Carbo-graph 4 (a thermally desorbable adsorbent) as an adsorbent. Because of the high sampling rate, care must be taken not to saturate the sorbent with the component of interest.Citation11,Citation12 Radiello diffusive samplers filled with a thermally desorbable adsorbent have been evaluated for the monitoring of BTEX according to the European Standard EN 13528-2 as a reference passive sampler.Citation13

The use of passive samplers under real conditions for monitoring and assessing atmospheric air quality requires a knowledge of the parameters responsible for the rate of analyte sampling by a particular type of sampler. Then it becomes possible to determine the time-weighted average concentrations of analytes in the air from a knowledge of the sampler exposure time and the mass of compounds retained by the sampler sorbent during exposure. For the commercially available passive diffusion samplers such as the Radiello and Orsa 5, routinely used for monitoring atmospheric air quality, the parameters responsible for the analyte sampling rate are supplied by their manufacturers. The results obtained using the Radiello and Orsa 5 samplers were corrected for temperature to give the numerical value of the rate constant of sampling analytes from atmospheric air.Citation14,Citation15

Passive permeation samplers fitted with a polydimethylsiloxane semipermeable membrane have hitherto been used only to assess the quality of indoor air. For these samplers, the parameter responsible for the rate of analyte sampling from the air, the numerical value of the calibration constants (k) for each compound determined, was measured experimentally on the basis of tests in exposure chambers in an atmosphere of standard gas mixtures exclusively at a temperature of 23 °C.Citation16,Citation17 Although the dosimeter calibration constants for each compound were not corrected for temperature, the possibility of extending the range of applicability of these samplers was examined.

To ensure that the samples of atmospheric air were representative and homogeneous, two exposed samplers of each kind were placed 20 cm from one another at the measuring stations during the monitoring campaignn. They were mounted at a height of ∼3 m above the ground in specially designed acid-resistant steel screens to protect them from the elements (precipitation and strong winds). In addition, in accordance with Good Laboratory Practice, one unexposed (i.e., isolated from the external environment) passive sampler of each type was installed at each station to obtain field blanks. After exposure, the passive samplers were sealed and isolated from the outside air, in accordance with the procedures relevant to each type. Each Orsa 5 passive sampler was placed in a separate glass vessel, which was then sealed off with a polyethylene screw top. The sorption medium in the Radiello samplers was removed from the cylindrical diffusion casing and placed in a glass vessel, which was then tightly stoppered with a polyethylene bung. The permeation samplers were closed with a polyethylene lid. All of the samplers were taken to the laboratory where they were stored at 5 °C until analysis. The analytes were released from the sorbent, and the final determinations were done no later than 1 week after the conclusion of the exposure period.

Sample Analysis

Liberation of Analytes from Sorption Bed and Chromatographic Conditions

For the Orsa 5 passive samplers and the homemade permeation passive samplers, all experiments were performed using an HP 5890 gas chromatograph (Hewlett-Packard, Palo Alto, CA) equipped with a flame ionization detector (280 °C) and a splitless injector (inlet temperature 200 °C; injection volume 2 μL). The analytes were liberated from the sorbent bed by extraction with 1 mL of CS2. To confirm the identity of BTEX, randomly selected CS2 extracts were analyzed by means of a gas chromatograph-mass selective detector.

The analytes in the Radiello diffusive samplers were liberated from the sorbent bed by two-step thermal desorption. All experiments were performed using an Agilent Technologies 6890 gas chromatograph coupled with a 5873 network inert mass spectrometer (mass spectrometry) and a thermal desorber (limited version 2.00, Unity Markes International, South Queensferry, U.K.).

Further details of the measurement methods were described in previous publications.Citation18,Citation19

Chemical Standards

External calibration was performed with a volatile calibration mix (VCM) containing 13 VOCs (including BTEX) at 2000 μg/mL (Supelco). Six standard solutions were prepared by diluting the VCM in methanol for gas chromatography (Merck). For solvent extraction, the concentration of standard solutions ranged between 1 and 25 μg/mL and for thermal desorption ranged between 50 and 2000 μg/mL. Calibration solutions were freshly prepared just before calibration.

Quality Control/Quality Assurance

The quality assurance and quality control procedures included analysis of field blanks and parallel samples (for each type of passive samplers) and triplicate measurements of CS2 extracts (for the homemade and Orsa 5 samplers). The precision estimated from replicate analyses of the standards and samples was within ±10%. For field blanks, unexposed passive samplers were analyzed for BTEX under conditions identical to those used in the analysis of samplers exposed to atmospheric air. The method limit of detection (MDL) was defined using the equations to calculate BTEX concentrations in atmospheric air for each analyte and each type of passive sampler, where the mass of analyte collected on the sorbent bed is the average mass obtained from the unexposed samplers (blanks) being used.Citation18,Citation19 The MDL for benzene for the Orsa 5 and passive permeation samplers for a 1-month exposure was 0.6 μg/m3; for the other compounds monitored it was 0.3 μg/m3. Consequently, the method quantification limit for benzene for the Orsa 5 and the homemade passive permeation sampler was 1.7 μg/m3; for the other compounds monitored it was 1 μg/m3. In general, BTEX concentrations measured using the Orsa 5 and the homemade passive permeation sampler were substantially higher in the samples than in the field blanks.

Statistical Analysis

Statistical analysis was performed using Statistica 7 for Windows (Stat-Soft, Inc.). Pearson and Spearman rank correlation tests were used to examine the strength of associations among the BTEX concentrations. The differences among the samplers were compared with the paired t test, Wilcoxon signed rank-sum test, and Friedman ANOVA nonparametric tests.Citation20–22

RESULTS AND DISCUSSION

The main goal of the research was the field comparison of passive samplers at different sampling rates of analytes () based on different phenomena (diffusion and permeation) to verify the possibility of extending the range of applicability of a homemade permeation passive sampler. Summary statistics (arithmetic mean and median, minimum, maximum, SD, and also skewness and kurtosis) for the final results (12 months, 10 monitoring stations) are shown in and plotted in . The patterns in are in good agreement with the results in and suggest that there is considerable similarity between the Radiello and Orsa 5 diffusive samplers but a large difference between these two and the permeation homemade sampler with respect to the central tendency (mean and median) and the spread of results (SD range, including skewness and kurtosis). Higher concentrations of all compounds were observed for the permeation homemade sampler. In addition, the correlations between the BTEX concentrations and meteorological parameters presented in illustrate a low influence of atmospheric conditions on the results generated.

Table 1. Sampling rate of passive samplers for a given analyte

Table 2. Statistical results

Table 3. Pearson and Spearman rank correlation coefficients between the concentrations of BTEX compounds and meteorological parameters

Figure 1. Box and whisker plot of the results obtained with the three samplers.

Figure 1. Box and whisker plot of the results obtained with the three samplers.

A low influence of atmospheric conditions on the data also results from the data showing fluctuations in monthly average BTEX concentrations. shows monthly average concentrations of analytes at station 10, where the BTEX are measured by BTEX monitor (Chrompack CP 7001).Citation23 It is interesting to note that despite the differences in the BTEX concentration values obtained with different types of passive samplers, the concentration profiles of the compounds monitored for particular stations are similar.

Figure 2. Monthly fluctuations in average concentrations of BTEX in atmospheric air at station number 10 in 2008.

Figure 2. Monthly fluctuations in average concentrations of BTEX in atmospheric air at station number 10 in 2008.

The comparison of monthly BTEX concentrations determined using active sampling and each type of passive samplers was examined by means of the linear regression method.Citation24,Citation25 The BTEX concentrations determined with the Chrompack CP 7001 monitor were plotted versus those determined with each type of passive sampler. The results are presented in and . In addition, the confidence band of the monthly BTEX concentrations determined using active sampling is plotted in each case to help visualize the estimated range of values.Citation26

Figure 3. Comparison of the time-weighted monthly average benzene concentrations in the atmospheric air obtained with passive samplers and the Chrompack CP 7001 monitor at station number 10 (– – –, confidence bands; ——, regression curve).

Figure 3. Comparison of the time-weighted monthly average benzene concentrations in the atmospheric air obtained with passive samplers and the Chrompack CP 7001 monitor at station number 10 (– – –, confidence bands; ——, regression curve).

Figure 4. Comparison of the time-weighted monthly average toluene concentrations in the atmospheric air obtained with passive samplers and the Chrompack CP 7001 monitor at station number 10 (– – –, confidence bands; ——, regression curve).

Figure 4. Comparison of the time-weighted monthly average toluene concentrations in the atmospheric air obtained with passive samplers and the Chrompack CP 7001 monitor at station number 10 (– – –, confidence bands; ——, regression curve).

For the difference between the two values to be insignificant, the dependence should be linear (y =bx +a), the line should pass through the origin of the coordinate system, and the slope should be close to unity. In other words, the parameters used for the validation of the proposed approach to compare monthly concentrations are the slope b, the intercept a, and the regression coefficient r. The previous conditions with respect to the slope b (tb calc [(1 − b)/sec b ] ≤ t cr =2.014) were not fulfilled for all passive samplers () with the exception of the Orsa 5 sampler for ethylbenzene. Thus, the slope b values were significantly different from the expected values of β0 =1, which means that the differences between the BTEX concentrations determined using passive samplers were statistically significant at the probability level p = 95% and for f = n – 2 = 46 df.

Table 4. Statistical parameters of the correlation analysis of results (the concentration levels of BTEX) obtained using a Chrompack CP 7001 monitor and passive dosimeters on the stage of the sampling of analytes from ambient air

The previous conditions with respect to the intercept a (ta calc [(a − 0)/sec a ] ≤ t cr =2.014) were not fulfilled in most cases with the Radiello and Orsa 5 samplers () with the exception of toluene (Radiello) and toluene with ethylbenzene (Orsa 5). Thus, the intercept a values were significantly different from the expected values of β0 =0, which means that the differences between the BTEX concentrations determined using Radiello and Orsa 5 passive samplers were statistically significant. For the homemade passive sampler, it was found that the intercept a was not significantly different from the expected values for benzene, toluene, and ethylbenzene.

The regression coefficients r for the Radiello and homemade passive samplers were higher than r cr at the probability level p = 95% and for f = n – 2 =46 df. For the Orsa 5 sampler, it was found that the values of coefficient r were higher than r cr for toluene, ethylbenzene, and xylenes.

On the basis of the linear regression method if can be concluded that the differences between the concentrations determined using passive samplers and the active sampling technique are statistically significant in most cases at the probability level p = 95% and for f = n – 2 =46df, but the data are statistically correlated.

With regard to the experimental design and data structure, the paired t test and nonparametric tests evidently seem to be the best suited to demonstrate similarities or differences between the results obtained with the three samplers. The statistical results for the paired t test () indicate that the individual concentrations are quite different; that is, the differences between the results from each sampler are highly significant (p < 0.00001), except for the difference between the concentrations of m,p-xylene collected with the Radiello and Orsa 5 diffusive samplers, which is not significant (p > 0.12056). In addition, all differences between the homemade permeation sampler and the Radiello and Orsa 5 diffusive samplers are positive. The Wilcoxon signed rank-sum test was also applied to the differences between measurements. Here, the differences are first ranked without regard for sign. The smallest difference is given rank 1, and ties are assigned average ranks. The sign of the difference is assigned to each rank. The positive and negative ranks aresummed and the absolute value of the smaller rank (T) is then compared with the standard critical values for testing the hypothesis. Because in this case the number of measurements was very large, the statistic Z, which is supposed to follow a standard normal distribution, was used. The results (p < 0.00001, except for the difference between the concentrations of m,p-xylene collected with the Radiello and Orsa 5 diffusive samplers, which is not significant [p > 0.117635]) are very similar to the results obtained with the paired t test and strongly endorse the same conclusion that the difference between the permeation and diffusive samplers is highly significant. Finally, the Friedman ANOVA, a test based on a nonparametric multiple comparison, was applied to compare the differences among the samplers. The results confirm the hypothesis that the differences between the results of the three sampling techniques are highly significant (p <0.00001).

Table 5. Results obtained from applying the paired t test

CONCLUSION

The concentrations of VOC analytes in the atmosphere can vary widely in time as well as spatially. Therefore, proper equipment and analytical methodology for measuring pollution levels have to be designed to obtain reliable and valid data for the given environmental compartments. In recent years, passive sampling has gained broad application as a tool for air quality assessment in urbanized areas. This is a relatively simple and inexpensive technique for the sampling and enrichment of analytes collected from outdoor air. The samples of analytes obtained can serve as a source of information in reference to creating the precise distribution of pollutant concentrations in both time and space and determining the time-weighted average concentration of an analyte. It was confirmed that the passive technique can be used to isolate and enrich analytes to get long-term information on the concentration of an individual BTEX in atmospheric air in the Tricity agglomeration.

The results of investigations of passive sampler performance with regard to BTEX are reported. The analytical data obtained in northern Poland during 2008 (12 months and 10 sampling stations) were evaluated by the paired t test, Wilcoxon signed rank-sum test and Friedman ANOVA. All the statistical results confirm the hypothesis that the differences between the performances of the sampling procedures are highly significant. The homemade permeation sampler seems to be the most efficient concerning the amount retained, whereas the Radiello and Orsa 5 diffusive samplers seem to be similar.

The greater differences observed with the permeation sampler compared with the Orsa 5 and Radiello samplers may result from the fact that in the former the influence of temperature on calibration constant (k) and, hence, on sampling rate was not taken into account. The parameter responsible for the sampling rate in the gaseous phase was determined at a constant temperature (23 °C), and the influence of temperature on the k constant has been not examined in a very wide range of temperatures. For this reason, to extend the range of applicability of these samplers, new dosimeter calibration constants must be assessed. In addition, the greater variation between the passive samplers could be explained by differences in the sampling rates.

On the basis of the linear regression method, it can be concluded that the differences between the concentrations determined using passive samplers and the active sampling technique are statistically significant, in most cases at the probability level p = 95% and for f = n – 2 =46 df, but the data are statistically correlated. This research on the quality of the atmospheric air in the metropolitan Tricity area confirmed the good capability of passive samplers to show fluctuations in BTEX concentrations in atmospheric air, which makes them applicable for air monitoring on the local scale.

ACKNOWLEDGMENTS

The authors are grateful to the agency of Regional Air Quality Monitoring Foundation for access to the monitoring stations.

REFERENCES

  • Harner , T. , Pozo , K. , Gouin , T. , Macdonald , A.M. , Hung , H. , Cainey , J. and Peters , A. 2006 . Global Pilot Study for Persistent Organic Pollutants (POPs) Using PUF Disk Passive Air Samplers . Environ. Pollut. , 144 : 445 – 452 .
  • Gibson , M. , Guernsey , J. , Beauchamp , S. , Waugh , D. , Heal , M. , Brook , J. , Maher , R. , Gagnon , G. , McPherson , J. , Bryden , B. , Gould , R. and Terashima , M. 2009 . Quantifying the Spatial and Temporal Variation of Ground-Level Ozone in the Rural Annapolis Valley, Nova Scotia, Canada Using Nitrite-Impregnated Passive Samplers . Journal of the Air & Waste Management Association , 59 : 310 – 320 . doi: 10.3155/1047-3289.59.3.310
  • Richardson , B.J. , Zheng , G.J. , Tse , E.S.C. and Lam , P.K.S. A . 2001 . Comparison of Mussels (Perna viridis) and Semi-Permeable Membrane Devices (SPMDs) for Monitoring Chlorinated Trace Organic Contaminants in Hong Kong Coastal Waters . Chemosphere , 45 : 1201 – 1208 .
  • Kot-Wasik , A. , Zabiegała , B. , Urbanowicz , M. , Dominiak , E. , Wasik , A. and Namieśnik , J. 2007 . Advances in Passive Sampling in Environmental Studies . Anal. Chim. Acta , 602 : 141 – 163 .
  • Namieśnik , J. , Zabiegała , B. , Kot-Wasik , A. , Partyka , M. and Wasik , A. 2005 . Passive Sampling and/or Extraction Techniques in Environmental Analysis . Anal. Bioanal. Chem. , 381 : 279 – 301 .
  • Górecki , T. and Namieśnik , J. 2002 . Passive Sampling . Trends Anal. Chem. , 21 : 276 – 291 .
  • Zabiegała , B. , Przyjazny , A. and Namieśnik , J. 1999 . Passive Dosimetry as an Alternative Technique to Dynamic Enrichment of Organic Pollutants of Indoor Air . J. Environ. Pathol. Toxicol. Oncol. , 18 : 47 – 59 .
  • Partyka , M. , Zabiegała , B. , Namieśnik , J. and Przyjazny , A. 2007 . Applications of Passive Samplers in Monitoring of Organic Constituents of Air . Crit. Rev. Anal. Chem. , 37 : 51 – 77 .
  • Zabiegała , B. and Namieśnik , J. 2007 . “ Use of Permeation Passive Samplers in Air Monitoring ” . In Comprehensive Analytical Chemistry; Barcelo, D , Vol. 48 , 85 – 106 . Elsevier : Amsterdam . Passive Sampling Techniques in Environmental Monitoring
  • Pfeffer , H.U. and Breuer , L. 2000 . BTX: Measurements with Diffusive Samplers in the Vicinity of a Cokery: Comparison between ORSA-Type Samplers and Pumped Sampling . J. Environ. Monit. , 2 : 483 – 486 .
  • Oury , B. , Lhuillier , F. , Protois , J.C. and Moréle , Y. 2006 . Behavior of the GABIE, 3m 3500, PerkinElmer Tenax TA, and RADIELLO 145 Diffusive Samplers Exposed Over a Long Time to a Low Concentration of VOCs . J. Occup. Environ. Hygiene , 3 : 547 – 557 .
  • Pennequin-Cardinal , A. , Plaisance , H. , Locoge , N. , Ramalho , O. , Kirchner , S. and Galloo , J.C. 2005 . Dependence on Sampling Rates of Radiello® Diffusion Sampler for BTEX Measurements with the Concentration Level and Exposure Time . Talanta , 65 : 1233 – 1240 .
  • European Standard EN 13528-2: Ambient Air Quality—Diffusive Samplers for the Determination of Concentration of Gases and Vapours. Part 2: Specific Requirements and Test Methods . European Committee for Standardization .
  • Ballesta , P.P. , Fild , R.A. , Connolly , R. and Cao , N. 2006 . Population Exposure to Benzene: One Day Cross-Sections in Six European Cities . Atmos. Environ. , 40 : 3355 – 3366 .
  • Shneider , P. , Gebefugi , I. , Richter , K. , Wolke , G. , Schnelle , J. , Wich-mann , H.E. and Heinrich , J. 2001 . Indoor and Outdoor BTX Levels in German Cities . Sci. Total Environ. , 267 : 41 – 51 .
  • Zabiegała , B. , Górecki , T. and Namieśnik , J. 2003 . Calibration of Permeation Passive Samplers with Silicone Membranes Based on Physicochemical Properties of the Analytes . Anal. Chem. , 75 : 3182 – 3192 .
  • Zabiegała , B. , Partyka , M. , Górecki , T. and Namieśnik , J. 2006 . Application of the GC Retention Index System for the Determination of the Calibration Constants of Permeation Passive Samplers with PDMS Membranes . J. Chromatogr. A , 1117 : 19 – 30 .
  • Zabiegała , B. , Urbanowicz , M. , Szymańska , K. and Namieśnik , J. 2010 . Application of Passive Sampling Technique for Monitoring of BTEX Concentration in Urban Air: Field Comparison of Different Types of Passive Samplers . J. Chromatogr. Sci. , 48 : 167 – 175 .
  • Zabiegała , B. , Urbanowicz , M. , Namieśnik , J. and Górecki , T. 2010 . Spatial and Seasonal Patterns of Benzene, Toluene, Ethylbenzene, and Xylenes in the GdaŃsk, Poland and Surrounding Areas Determined Using Radiello Passive Samplers . J. Environ. Qual. , 39 : 896 – 906 .
  • Wachasunder , S.D. and Pandya , G.H. 2002 . Instrumental Determination of Trace Metals in Fly Ash by INAA and AAS Techniques and Its Statistical Analysis . Water Air Soil Pollut. , 135 : 325 – 334 .
  • Wang , Z. and Ng , H.K.T. 2006 . A Comparative Study of Tests for Paired Lifetime Data . Lifetime Data Anal. , 12 : 505 – 522 .
  • Miller , J.C. and Miller , J.N. 1988 . Statistics for Analytical Chemistry , 2nd , 137 – 161 . New York : Ellis Horwood .
  • 2005 . European Standard EN 14662-3 Ambient Air Quality—Standard Method for the Measurement of Benzene Concentrations. Part 3: Automated Pumped Sampling with in Situ Gas Chromatography . European Committee for Standardization . 2005 .
  • Brüggemann , L. and Wennrich , R. 2002 . Evaluation of Measurement Uncertainty for Analytical Procedures Using a Linear Calibration Function . Accred. Qual. Assur. , 7 : 269 – 273 .
  • Heydorn , K. and Anglov , T. 2002 . Calibration Uncertainty . Accred. Qual. Assur. , 7 : 153 – 158 .
  • Miller , J.N. 1991 . Basic Statistical Methods for Analytical Chemistry. Part 2. Calibration and Regression Methods . Analyst , 116 : 3 – 14 .

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.