A gel tumour phantom for assessment of the accuracy of manual and automatic delineation of gross tumour volume from FDG-PET/CT

, , , , &
Pages 636-644 | Received 20 Mar 2012, Accepted 28 Jul 2012, Published online: 17 Oct 2012
 

Abstract

Introduction. Our primary aim was to make a phantom for PET that could mimic a highly irregular tumour and provide true tumour contours. The secondary aim was to use the phantom to assess the accuracy of different methods for delineation of tumour volume from the PET images. Material and methods. An empty mould was produced on the basis of a contrast enhanced computed tomography (CT) study of a patient with a squamous cell carcinoma in the head and neck region. The mould was filled with a homogeneous fast-settling gel that contained both 18F for positron emission tomography (PET) and an iodine contrast agent. This phantom (mould and gel) was scanned on a PET/CT scanner. A series of reference tumour contours were obtained from the CT images in the PET/CT. Tumour delineation based on the PET images was achieved manually, by isoSUV thresholding, and by a recently developed three-dimensional (3D) Difference of Gaussians algorithm (DoG). Average distances between the PET-derived and reference contours were assessed by a 3D distance transform. Results. The manual, thresholding and DoG delineation methods resulted in volumes that were 146%, 86% and 100% of the reference volume, respectively, and average distance deviations from the reference surface were 1.57 mm, 1.48 mm and 1.0, mm, respectively. Discussion. Manual drawing as well as isoSUV determination of tumour contours in geometrically irregular tumours may be unreliable. The DoG method may contribute to more correct delineation of the tumour. Although the present phantom had a homogeneous distribution of activity, it may also provide useful knowledge in the case of inhomogeneous activity distributions. Conclusion. The geometric irregular tumour phantom with its inherent reference contours was an important tool for testing of different delineation methods and for teaching delineation.

Declaration of interest: The authors report no conflicts of interest. The authors alone are responsible for the content and writing of the paper.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.