The bone conduction implant: Clinical results of the first six patients

, , , &
Pages 408-416 | Received 18 Aug 2014, Accepted 24 Nov 2014, Published online: 23 Feb 2015

References

  • Barbara M., Perotti M., Gioia B., Volpini L. & Monini S. 2013. Transcutaneous bone-conduction hearing device: Audiological and surgical aspects in a first series of patients with mixed hearing loss. Acta Otolaryngol, 133, 1058–64.
  • Carlsson P., Håkansson B., Rosenhall U. & Tjellström A. 1986. A speech-to-noise ratio test with the bone-anchored hearing aid: a comparative study. Otolaryngol Head Neck Surg, 94, 421–426.
  • Cochlear Bone Anchored Solutions celebrated 100 000 patients at the Osseo 2013 meeting, Newcastle, UK.
  • Cox R.M. & Alexander G.C. 1995. The abbreviated profile of hearing aid benefit. Ear Hear, 16, 176–186.
  • de Wolf M.J.F., Shival M.L.C., Hol M.K.S., Mylanus E.A.M., Cremers C.W.R.J. et al. 2010. Benefit and quality of life in older bone-anchored hearing aid users. Otol Neurotol, 31, 766–772.
  • de Wolf M.J.F. 2011. Bone anchored hearing aid: Clinical outcomes of the linear incision technique and benefit assissment. Thesis, Radboud University Nijmegen Medical Centre, ISBN 978-90-9026047-1.
  • Desmet J., Wouters K., De Bodt M. & Van de Heyning P. 2014. Long-term subjective benefit with a bone-conduction implant sound processor in 44 patients with single-sided deafness. Otol Neurotol, 35, 1017–1025.
  • Dun C.A.J., Faber H.T.F., de Wolf M.J.F., Mylanus E.A.M., Cremers C.W.R.J. et al. 2012. Assessment of more than 1000 implanted percutaneous bone conduction devices: Skin reactions and implant survival. Otol Neurotol, 33, 192–98.
  • Eeg-Olofsson M., Stenfelt S., Tjellström A. & Granström G. 2008. Transmission of bone-conducted sound in the human skull measured by cochlear vibrations. Int J Audiol, 47, 761–9.
  • Eeg-Olofsson M., Stenfelt S., Taghavi H., Reinfeldt S., Håkansson B. et al. 2013. Transmission of bone conducted sound: Correlation between hearing perception and cochlear vibration. Hear Res, 306, 11–20.
  • Eeg-Olofsson M., Lith A., Håkansson B., Reinfeldt S., Taghavi H., Fredén Jansson KJ., Johansson C. 2014a. Evaluation of Bone Tissue Formation in a Flat Surface Attachment of a Bone Conduction Implant: A Pilot Study in a Sheep Model. Audiol Neurotol Extra, 4, 62–76.
  • Eeg-Olofsson M., Håkansson B., Reinfeldt S., Taghavi H., Lund H. et al. 2014b. The bone conduction implant: First implantation, surgical and audiologic aspects. Otol Neurotol, 35, 679–685.
  • Fredén Jansson K.J., Håkansson B., Taghavi H., Reinfeldt S. & Eeg-Olofsson M. 2014. MRI induced torque and demagnetization in retention magnets for bone conduction implants. IEEE Trans Biomed Eng, 61, 1887–1893.
  • Hagerman B. 1982. Sentences for testing speech intelligibility in noise. Scand Audiol, 11, 79–87.
  • Hagerman B. 1993. Efficiency of speech audiometry and other tests. Br J Audiol, 27, 423–425.
  • Hagerman B. & Kinnefors C. 1995. Efficient adaptive methods for measuring speech reception threshold in quiet and in noise. Scand Audiol, 24, 71–77.
  • Håkansson B. 2003. The balanced electromagnetic separation transducer: A new bone conduction transducer. J Acoust Soc Am, 113, 818–25.
  • Håkansson B., Eeg-Olofsson M., Reinfeldt S., Stenfelt S., Granström G., 2008. Percutaneous vs. transcutaneous bone condution implant system: A feasibility study on a cadaver head. Otol Neurotol, 29, 1132–1139.
  • Håkansson B., Reinfeldt S., Eeg-Olofsson M., Östli P., Taghavi H. et al. 2010. A novel bone conduction implant (BCI): Engineering aspects and pre-clinical studies. Int J Audiol, 49, 203–215.
  • Ihler F., Volbersw L., Blum J., Matthias C. & Canis M. 2014. Preliminary functional results and quality of life after implantation of a new bone conduction hearing device in patients with conductive and mixed hearing loss. Otol Neurotol, 35, 211–215.
  • ISO 8253-1, 2010. Acoustics - Audiometric test methods - Part 1: Pure-tone air and bone conduction audiometry. ISO.
  • ISO 8253-3, 2012. Acoustics - Audiometric test methods - Part 3: Speech audiometry. ISO.
  • Kiringoda R. & Lustig L.R. 2013. A meta-analysis of the complications associated with osseointegrated hearing aids. Otol Neurotol, 34, 790–794.
  • Manrique M., Sanhueza J., Manrique R. & de Abajo J. 2014. A new bone conduction implant: Surgical technique and results. Otol Neurotol, 35, 216–220.
  • Oticon Medical. 2014. Oticon Medical: Taking ideas further. http://www.oticonmedical.com/∼asset/cache.ashx?id = 7229&type = 14&format = web
  • Pfiffner F., Caversaccio M.D. & Kompis M. 2011. Comparisons of sound processors based on osseointegrated implants in patients with conductive or mixed hearing loss. Otol Neurotol, 32, 728–735.
  • Reinfeldt S., Håkansson B., Taghavi H. & Eeg-Olofsson M. 2014. Bone conduction hearing sensitivity in normal-hearing subjects: Transcutaneous stimulation at BAHA vs. BCI position. Int J Audiol, 53, 360–369.
  • Reinfeldt S., Östli P., Håkansson B., Taghavi H., Eeg-Olofsson M. & Stalfors J. 2015a. Study of the feasible size of a bone-conduction implant (BCI) transducer in the temporal bone. Otol Neurotol, Published Ahead-of-Print.
  • Reinfeldt S., Håkansson B., Taghavi H. & Eeg-Olofsson M. 2015b. New developments in bone-conduction hearing implants: A review. Medical Devices: Evidence and Research, 8, 79–93.
  • Robinson K., Gatehouse S. & Browning G.G. 1996. Measuring patient benefit from otorhinolaryngological surgery and therapy. Ann Otol Rhinol Laryngol, 105, 415–422.
  • Snik A., Mylanus E., Proops D., Wolfaardt J., Hodgetts W. et al. 2005. Consensus statements on the BAHA system: Where do we stand at present? Ann Otol Rhinol Laryngol, Suppl 195, 2–12.
  • Sprinzl G., Lenarz T., Ernst A., Hagen R., Wolf-Magele A. et al. 2013. First European multicenter results with a new transcutaneous bone conduction hearing implant system: Short-term safety and efficacy. Otol Neurotol, 34(6), 1076–1083.
  • Stenfelt S. & Goode R. 2005. Transmission properties of bone- conducted sound: measurements in cadaver heads. J Acoust Soc Am, 118, 2373–2391.
  • Stenfelt S. & Reinfeldt S. 2007. A model of the occlusion effect with bone-conducted stimulation. Int J Audiol, 46, 595–608.
  • Taghavi H., Håkansson B. & Reinfeldt S. 2012a. A novel bone conduction implant system: Analog radio frequency data and power link design. In: Proceedings of the 9th IASTED international conference on biomedical engineering, Innsbruck, Austria. 327–335.
  • Taghavi H., Håkansson B. & Reinfeldt S. 2012b. Analysis and design of RF power and data link using amplitude modulation of Class-E for a novel bone conduction implant. IEEE Transactions on Biomedical Engineering, 59, 3050–3059.
  • Taghavi H., Håkansson B., Reinfeldt S., Eeg-Olofsson M. & Akhshijan S. 2012c. Feedback analysis in percutanous bone-conduction device and bone-conduction implant on a dry cranium. Otol Neurotol, 33, 413–420.
  • Taghavi H., Håkansson B., Eeg-Olofsson M., Johansson C., Tjellström A. et al. 2013. A vibration investigation of a flat surface contact to skull bone for direct bone conduction transmission in sheep skulls in vivo. Otol Neurotol, 34, 690–698.
  • Taghavi H., Håkansson B., Reinfeldt S., Eeg-Olofsson M., Fredén Jansson K.J. et al. 2014. Technical design of a new bone conduction implant (BCI) system. Int J Audiol – Accepted after revision.
  • Tjellström A. & Håkansson B. 1995. The bone anchored hearing aid: Design principles, indications, and long-term clinical results. Otolaryngologic clinics of north America, 28(1), 53–72.