Size Measurement of Polystyrene Latex Particles Larger than 1 Micrometer using a Long Differential Mobility Analyzer

, , &
Pages 1178-1184 | Received 21 Apr 2004, Accepted 15 Oct 2004, Published online: 01 Aug 2012
 

Abstract

We describe a newly constructed annular-type differential mobility analyzer (DMA) with an effective electrode length of 60 cm, which is longer than that of our original DMA (40 cm length). This long DMA was developed to extend the classification size of particles measuring up to 1.5 μm. As an application of this DMA, the mean diameters and standard deviations were determined for six samples of monodisperse polystyrene latex (PSL) particles ranging from 0.94 to 1.27 μm in nominal diameter. These PSL particles suspended in double-distilled water were aerosolized by a glass nebulizer and then introduced into the DMA. The mean diameters and standard deviations of these PSL particles were calculated by Ehara's method and compared with the nominal diameters and uncertainties. There was good correlation between the nominal diameters of these samples—particularly for recently certified samples—and the measured diameters. Classification of aerosol particles more than 1 micrometer in diameter using this DMA will be useful for many purposes.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.