Influence of preparation design on fit and ceramic thickness of CEREC 3 partial ceramic crowns after cementation

, , , , , , & show all
Pages 107-113 | Received 20 Feb 2014, Accepted 07 Aug 2014, Published online: 09 Oct 2014
 

Abstract

Objective. This study investigated the influence of preparation design on the marginal and internal gap and ceramic thickness of partial ceramic crowns (PCCs) fabricated with the CEREC 3 system. Materials and methods. Sixteen extracted human mandibular molars were prepared according to two different preparation designs (n = 8): a retentive preparation design with traditional cusp capping (Group I) and a non-retentive preparation design with horizontal reduction of cusps (Group II). PCCs were fabricated from IPS Empress CAD with the CEREC 3 system. The parameters for luting space and minimum occlusal ceramic thickness were set to 30 μm and 1.5 mm, respectively. The fabricated PCCs were cemented to their corresponding teeth with self-adhesive resin cement and were then scanned by micro-computed tomography. The marginal and internal gaps were measured at pre-determined measuring points in five bucco-lingual and three mesio-distal cross-sectional images. The ceramic thicknesses of the PCCs were measured at the measuring points for cusp capping areas. Results. Group II (167.4 ± 76.4 μm) had a smaller overall mean gap, which included the marginal and internal gap measurements, than that of Group I (184.8 ± 89.0 μm). The internal gaps were larger than the marginal gaps, regardless of preparation design. Group I presented a thinner ceramic thickness in the cusp capping areas than the minimum occlusal ceramic thickness parameter of 1.5 mm. Conclusion. Preparation design had an influence on fit, particularly the internal gap of the PCCs. Ceramic thickness could be thinner than the minimum ceramic thickness parameter.

Acknowledgments

This study was supported by the Basic Science Research Program through the National Research Foundation (NRF) of Korea, funded by the Ministry of Education, Science, and Technology (2010-0005090, 2012-009268). The authors thank Dentsply for the kind donation of materials for this investigation.

Declaration of interest: The authors report no conflicts of interest. The authors alone are responsible for the content and writing of the paper.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.