Comparison of Gebauer SLc and Moria CBm Carriazo-Barraquer ALK Microkeratomes for Descemet’s Stripping Automated Endothelial Keratoplasty Preparation

, , , , , & show all
Pages 343-349 | Received 24 Jul 2014, Accepted 30 Jan 2015, Published online: 24 Mar 2015
 

ABSTRACT

Purpose: We compared the hand-guided Moria Carriazo-Barraquer (CBm) microkeratome with the fully automatic SLc microkeratome for Descemet’s stripping automated endothelial keratoplasty (DSAEK)-lamella preparation and storage, vis-à-vis accuracy, endothelial cell loss (ECL), and lamellar surface roughness (LSR).

Methods: A total of 18 human corneas were dissected with both the 300 μm CBm multi-use (n = 9) and the 300 µm SLc (n = 9) single-use heads, after which they were incubated for 6 d in a 5% dextran medium. Before preparation (0 h) and 1, 24, and 144 h after dissection, ECL and corneal thickness (CT) were measured by ultrasound pachymetry (USP) and optical coherence tomography (OCT). LSR was assessed by scanning electron microscopy (SEM) and evaluated by three masked observers.

Results: Prior to cutting, CTs did not differ significantly between OCT or USP measurements, with a high correlation between the two modalities (r2 = 0.94, p < 0.0001). One hour after preparation the anterior lamella showed a significantly higher dissection depth with the CBm (429.4 ± 21.8 µm) than the SLc (311.7 ± 54.8 µm, p = 0.0006), with the variance of the SLc system showing a trend towards higher values (p = 0.07). Anterior and posterior lamellae swelled significantly in the subsequent culture period. Both groups showed a significant ECL 1 h after preparation (p < 0.0001) with no significant difference between the systems (1 h: p = 0.44; CBm: − 9.4%, SLc: −11.7%), which stabilized over 144 h (144 h CBm: −13.9%, 144 h SLc: −10.3%).

LSR did not differ significantly between both systems (p = 0.60).

Conclusions: The SLc system agrees more with the designated cutting depth than the CBm. The dissection produced a comparable LSR and a ∼10% ECL independently of the system. Further incubation of the prepared lamellae led to a swelling, but no further ECL.

DECLARATION OF INTEREST

The authors report no conflicts of interest and have no proprietary interest in any of the materials mentioned in this article. Ethics approval: Institutional review board, University Hospital RWTH Aachen. This research received no specific grant from any funding agency in the public, commercial or not-for-profit sectors.

Reprints and Corporate Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

To request a reprint or corporate permissions for this article, please click on the relevant link below:

Academic Permissions

Please note: Selecting permissions does not provide access to the full text of the article, please see our help page How do I view content?

Obtain permissions instantly via Rightslink by clicking on the button below:

If you are unable to obtain permissions via Rightslink, please complete and submit this Permissions form. For more information, please visit our Permissions help page.