189
Views
1
CrossRef citations to date
0
Altmetric
Systematic Review

Review of safety reports of cardiac MR-imaging in patients with recently implanted coronary artery stents at various field strengths

, , , , &
Pages 83-90 | Received 25 May 2020, Accepted 02 Dec 2020, Published online: 18 Dec 2020
 

ABSTRACT

Background: Aim of this study was to review current literature and data regarding the effects of MRI-examination post stent implantation on re-occlusion rates.

Methods: We focused on representative studies in the database MEDLINE. Inclusion criteria were: clinical studies with the main focus on the safety of coronary artery stents after MRI-examination in the time interval of 8 weeks post stent implantation. During a follow up period the incidence of cardiac events was recorded. In addition, the time interval between stent implantation and MRI-examination should be defined.

Results: Our search resulted in a total of relevant 17 studies. There were in-vivo as well as in-vitro studies and in addition three further publications f.e. guidelines. Concerning the patients, we differentiated between MRI performed shortly after acute cardiac event and in stable CAD. MRI-examinations were performed at different field strengths and reported different stent types. Considered were the incidences of cardiac events.

Conclusion: Independent of MRI field strength (1.5 Tesla or 3.0 Tesla) or used stent type (BMS or DES), there was no increased rate for cardiac events in patients, who underwent MRI < 8 weeks after stent placement. MRI < 8 weeks after stent placement seems to be safe.

Declaration of interest

The authors have no relevant affiliations or financial involvement with any organization or entity with a financial interest in or financial conflict with the subject matter or materials discussed in the manuscript. This includes employment, consultancies, honoraria, stock ownership or options, expert testimony, grants or patents received or pending, or royalties.

Reviewer disclosures

Peer reviewers on this manuscript have no relevant financial or other relationships to disclose.

Additional information

Funding

This paper was not funded.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 99.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 570.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.