The effect of pearl spacing on single-cycle load-to-failure and cyclic loading parameters of 2.0 mm pearl locking plates

ORCID Icon, ORCID Icon &
Pages 337-342 | Received 06 Feb 2021, Accepted 01 Jun 2021, Published online: 05 Jul 2021
 

ABSTRACT

Aims

To compare the mechanical performance and mode of failure in four-point bending of two different 2.0 mm “string of pearls” locking plates that differ in dimensions.

Methods

Ten *2.0 mm, 82 mm long, 10-hole (Plate A) and ten 2.0 mm, 69 mm long, 12-hole (Plate B) Cortical Pearl Systems were secured to plate extenders and centred beneath an Instron tensile tester in four-point bending. In all constructs, a simulated fracture gap was maintained at 33 mm. Due to differences in plate dimensions, 33 mm corresponded to four pearls (Plate A) and six pearls (Plate B). Following an initial preload of 10 N, ramped single-cycle load-to-failure at 0.1 mm/second was performed in five Plate A and five Plate B constructs. Load and displacement were recorded. Constant frequency sinusoidal cyclic loading (33 N) at 20 mm/minute was performed on five Plate A and five Plate B constructs following 10 N of preload. Maximum moment and cycle count were recorded. Testing and data analysis were completed in accordance with the American Society for Testing and Materials F382-14 guidelines. Differences in performance and mode of failure were compared.

Results

: Plate A constructs produced higher mean values for bending stiffness (19.8 (SD 2.0) N/mm vs. 10.1 (SD 0.6) N/mm; p < 0.001), bending structural stiffness (0.77 (SD 0.08) Nm2 vs. 0.39 (SD 0.02) Nm2; p < 0.001), yield point (64.1 (SD 4.2) N vs. 54.6 (SD 3.9) N; p = 0.01), proof load (65.4 (SD 3.2) N vs. 55.6 (SD 4.0) N; p = 0.005), and bending strength (1.3 (SD 0.1) Nm vs. 1.1 (SD 0.08) Nm; p = 0.005) when compared to Plate B constructs in single cycle load-to-failure. Plate A constructs had a greater (p = 0.001) mean cycle count to failure (26,178 (SD 4,061) cycles) when compared with Plate B constructs (15,550 (SD 1,291) cycles). All plates failed by non-catastrophic plastic deformation.

Conclusions

Plate A, which is wider, thicker and has a greater spacing between pearls, was mechanically superior to Plate B in four-point bending under single-cycle load-to-failure and sinusoidal cyclic loading.

Clinical relevance

Although mechanical differences were identified in four-point bending, in vivo biomechanical performance remains undetermined. By selecting Plate B, the clinician may gain bone purchase through a greater number of pearls and thus screws per unit length, however, the inferior mechanical characteristics, as evaluated in four-point bending, should also be considered. Further research into the mechanical and biomechanical performance of these plating systems is warranted.

Acknowledgements

Anthony Wade, from the School of Engineering and Advanced Technology, Massey University, (Palmerston North, NZ), generously provided time and expertise in the implementation of mechanical testing. Furthermore, Mr Wade designed the custom Labview programme that was used for test configuration and data acquisition for all mechanical tests. All testing was performed in facilities and using equipment provided by the School of Engineering and Advanced Technology, Massey University. Veterinary Orthopedic Implants donated the Cortical Pearl Systems that were destructively tested in this study

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 61.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 213.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.