396
Views
3
CrossRef citations to date
0
Altmetric
Innovation

Pull-out strength comparison of a novel expanding fastener against an orthopaedic screw in an ovine vertebral body: an ex-vivo study

, , , , , & show all
Pages 43-51 | Received 05 Jul 2015, Accepted 29 Nov 2015, Published online: 13 Jan 2016
 

Abstract

The purpose of this study was to mechanically test a novel Unthreaded Expandable Fastener (UEF), manufactured using Selective Laser Melting, which was designed for fixation in the cervical lateral mass. The pull-out strength and stiffness of the prototype UEFs was evaluated in a non-osteoporotic ovine bone model against equivalent screws. The prototype UEF demonstrated a 41% increase in failure force and a 60% reduction in failure force standard deviation compared to the screws. All bone samples were micro CT-scanned and no significant differences in bone microstructural properties was found between the screw and UEF sample sets, indicating that the UEFs may be less sensitive to bone quality variation. This increased performance can potentially translate into improved surgical outcome and reduced surgical risk for lateral mass fixation. With further design optimisation, additional improvement in performance over screws may be possible in future studies.

Acknowledgements

The authors acknowledge the help of Astrid Armitrage, a Senior Technician in Charge at the Large Animal Facility, UWA, for facilitating the acquisition of the sheep spines. The authors acknowledge the facilities and the scientific and technical assistance of the National Imaging Facility at the Centre for Microscopy, Characterisation & Analysis, UWA, a facility funded by the University, State and Commonwealth Governments. The authors also acknowledge the use of the Realizer SLM100 machine at UWA and the Instron 8874 servo-hydraulic materials testing machine at RPH.

Disclosure statement

Matthew Oldakowki, Intan Oldakowska, Thomas B. Kirk and Philip Hardcastle are named inventors in the patent application (PCT/AU2015/000099) related to the design of the fastener tested in this experiment. Chris T. Ford, Tim B. Sercombe and Robert E. Day report no conflicts of interest.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 706.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.