Portable Nanoporous Electrical Biosensor for Ultrasensitive Detection of Troponin-T

, , , , &
Article: FSO24 | Published online: 17 Jun 2015

Figures & data

Figure 1. Components of the developed sensor platform.

(A) Assembled sensor platform. (B) SEM micrograph of nanoporous nylon membrane with average pore diameter of 200 nm. (C) Characteristic representation of immunoassay inside nanowells.

cTnT:Cardiac Troponin-T; DSP:Dithiobis succinimidyl propionate; EDL:Electrical double layer; IHP:Inner Helmholtz plane; OHP:Outer Helmholtz plane.

Figure 1.  Components of the developed sensor platform.(A) Assembled sensor platform. (B) SEM micrograph of nanoporous nylon membrane with average pore diameter of 200 nm. (C) Characteristic representation of immunoassay inside nanowells.cTnT:Cardiac Troponin-T; DSP:Dithiobis succinimidyl propionate; EDL:Electrical double layer; IHP:Inner Helmholtz plane; OHP:Outer Helmholtz plane.

Figure 2. Calibration dose response analysis for cardiac Troponin-T antigen diluted in 7% bovine serum albumin.

Cardiac Troponin-T concentrations in the range of 0.001 pg/ml to 10,000 pg/ml were analyzed for n = 3 replicates. Error bars indicate two standard deviation for n = 3 replicates.

Figure 2.  Calibration dose response analysis for cardiac Troponin-T antigen diluted in 7% bovine serum albumin.Cardiac Troponin-T concentrations in the range of 0.001 pg/ml to 10,000 pg/ml were analyzed for n = 3 replicates. Error bars indicate two standard deviation for n = 3 replicates.
Figure 3.  Regression plot for linearity experiments.
Figure 3.  Regression plot for linearity experiments.

Figure 4. Interference experiments with triglyceride concentrations in the range of 0 mg/dl–2000 mg/dl.

Triglyceride aliquots with cTnT concentration of (A) 1000 pg/ml; (B) 4000 pg/ml; (C) 7000 pg/ml and (D) 10,000 pg/ml.

cTnT:Cardiac Troponin-T.

Figure 4.  Interference experiments with triglyceride concentrations in the range of 0 mg/dl–2000 mg/dl.Triglyceride aliquots with cTnT concentration of (A) 1000 pg/ml; (B) 4000 pg/ml; (C) 7000 pg/ml and (D) 10,000 pg/ml.cTnT:Cardiac Troponin-T.

Table 1.  Table representing the comparison of cardiac Troponin-T concentrations between electrochemical and chemiluminescent immunoassay detection method.