Evaluation of the Difference between Predicted and Measured Central Corneal Thickness Reduction after SMILE and Femtosecond Laser-assisted LASIK for Myopia

, &
Pages 1089-1095 | Received 02 Mar 2020, Accepted 27 Dec 2020, Published online: 26 Jan 2021
 

ABSTRACT

Purpose

To evaluate the agreement between the predicted and measured central corneal thickness (CCT) reduction after the small incision lenticule extraction (SMILE) surgery and femtosecond laser-assisted in situ keratomileusis (FS-LASIK) surgery.

Methods

A total 165 patients were enrolled in this prospective study. Eighty patients with a mean spherical equivalent (SE) of −4.72 ± 1.80 Diopters (D) were treated with the FS-LASIK procedure and Eighty-five patients with a mean SE of −4.78 ± 1.63 D were treated with SMILE procedure. One eye for each patient was randomly selected and included for statistical analysis. Ultrasound pachymetry measurement and Scheimpflug camera corneal topography were performed preoperatively and 3 months postoperatively. The measured CCT reduction was calculated by comparing the preoperative examinations with postoperative examinations. Comparative statistics and linear regression analyses were performed.

Results

The mean predicted CCT reduction was 95.02 ± 21.39 μm in FS-LASIK group and 103.49 ± 22.87 μm in SMILE group (P = .015). The prediction of laser platform was found to overestimate the measured CCT reduction for both FS-LASIK group (ultrasound 13.20 ± 9.34 μm) and SMILE group (ultrasound 13.12 ± 8.68 μm). The prediction of laser platform was found to systematically overestimate the measured CCT reduction in FS-LASIK group. In SMILE group, the difference between predicted and measured CCT reduction were found significantly related with the predicted CCT reduction (P < .001 for ultrasound; and P = .004 for Pentacam).

Conclusion

A systematic overestimation of measured CCT reduction in FS-LASIK group did not influence the refractive precision of FS-LASIK. Due to the different biomechanical distributions in post-SMILE cornea, the measured CCT reduction was influenced as the changes in refractive correction. Nomogram adjustment for high myopic correction needs further research.

Declaration of interest

The authors have no conflicts of interest to declare in the conducting of this study.

Additional information

Funding

Supported by the National Natural Science Foundation of China (Grant No. 81300757)

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 65.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 555.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.