Abstract
Objective
To assess the slim modiolar array as a hearing preservation electrode.
Methods
Retrospective chart review of adult, post-lingual CI recipients implanted with slim modiolar array Sept 2016 to July 2017 in a tertiary referral center. Baseline audiograms were obtained within six months of initial CI evaluation. Patients with low frequency pure tone average (LFPTA) (125, 250, 500 Hz) <80 dB were considered HP candidates. Postoperative audiograms were obtained within 48 h before activation. Successful HP was considered as (1) retention of LFPTA threshold <80 dB and (2) change in threshold from pre- to post-operative.
Results
Sixty-three patients received the slim perimodiolar array and 42 were HP candidates. Post-operative audiograms were obtained for 39 of 42 patients an average of 28.92 days after surgery. 56.4% of HP candidates retainedLFPTA <80 dB. Mean ΔLFPTA was 24.15 dB (±16.14; p < 0.001). 56.4% of HP candidates experienced Δ LFPTA <20 dB; 69.2% <30 dB. Functional hearing preservation was more successful in lower frequencies where starting thresholds were better – 78% with LFPTA <50 dB retained serviceable hearing at activation. The postoperative change was similar in each low frequency (Δ125 Hz: mean 21.25 +/− 14.76 (N = 28); Δ250 Hz: 26.28 +/− 19.29 (N = 39); Δ500 Hz: 25.00 +/− 17.73 (N = 39)).
Conclusions
The slim perimodiolar array is moderately effective at immediate hearing preservation. In subjects with preoperative audiometric profiles similar to those in prior EAS trials, immediate HP is comparable.
Keywords: