Towards the quantification of residual milk allergens in caseinate-fined white wines using HPLC coupled with single-stage Orbitrap mass spectrometry

, , , &
Pages 1304-1314 | Received 26 Mar 2011, Accepted 30 May 2011, Published online: 21 Jul 2011
 

Abstract

A method based on LC-ESI-high-resolution (HR)-MS analysis, using a single-stage Orbitrap mass spectrometer, was developed for the quantification of casein allergens potentially present in white wines as a result of fining by caseinate. The method consists of (1) extraction from the matrix by ultrafiltration, (2) digestion with trypsin and (3) detection/quantification of residual caseins, obtained by monitoring the LC-MS response of representative tryptic peptides (peak areas in extracted-ion chromatograms). Method linearity was assessed first on caseinate solutions prepared either in water or in wine matrix (the ultrafiltration residue of a protein-free white wine). Limits of detection (LOD) ranged from 0.1 to 0.3 µg ml−1 (S/N = 3) in water, and between 0.15 and 0.7 µg ml−1 in wine matrix, depending on the selected peptide. Method repeatability and reproducibility, measured as response variability (standard deviation) due to LC-MS analysis alone and to both enzymatic digestion and LC-MS analysis, were assessed on caseinate standard solutions in water and ranged from 5 to 12% and from 8 to 20%, respectively. A higher variability was usually observed for the peptide marker response in the case of matrix-matched samples, the only exception being peptide GPFPIIV from β-casein, the marker also providing the highest sensitivity. The method was finally applied to a casein-free white wine (‘Greco di Tufo’) fined with caseinate at different concentrations, after discarding the precipitate due to casein–wine components aggregation. Minimum detectable added caseinate concentrations (i.e. those corresponding to responses with S/N = 3) were estimated between 39 and 51 µg ml−1, according to the peptide marker chosen. These limits are compatible with caseinate concentrations typically adopted for wine-fining purposes. Moreover, a cross-check with the calibration performed in wine matrix led to an estimation of the concentration of dissolved caseinate to be in the low ng ml−1 range.

Acknowledgements

The authors are grateful to Elisabetta De Angelis and Barbara Introna for their valuable help in performing LC-MS analyses.

Log in via your institution

Log in to Taylor & Francis Online

PDF download + Online access

  • 48 hours access to article PDF & online version
  • Article PDF can be downloaded
  • Article PDF can be printed
USD 61.00 Add to cart

Issue Purchase

  • 30 days online access to complete issue
  • Article PDFs can be downloaded
  • Article PDFs can be printed
USD 799.00 Add to cart

* Local tax will be added as applicable

Related Research

People also read lists articles that other readers of this article have read.

Recommended articles lists articles that we recommend and is powered by our AI driven recommendation engine.

Cited by lists all citing articles based on Crossref citations.
Articles with the Crossref icon will open in a new tab.