Mammalian cell culture density determination using a laser through-beam sensor

, , , , , & show all
Pages 224-226 | Received 25 Apr 2018, Accepted 07 Aug 2018, Published online: 04 Oct 2018

Abstract

High-throughput protein expression platforms are increasingly used to produce proteins for many applications: to support studies in structure/function, regulation and proteomics, as well as for direct use as potential biotherapeutic agents for medical applications. Here we describe a device that we refer to as the flask density reader (FDR) consisting of a through-beam laser and sensor, and a customized culture flask-receiving nest. The FDR has been integrated onto GNF System™’s automated protein expression platform to enable rapid, noninvasive, fully automated spectrophotometric determination of cell densities in suspension mammalian cell cultures. The FDR reduces the risk of culture contamination from frequent flask sampling and greatly reduces the time and effort needed to count cells using off-line methods.

METHOD SUMMARY

A device called the flask density reader consisting of a through-beam laser and sensor, and a custom cell culture flask-receiving nest was built and integrated onto an automated protein production platform for rapid, noninvasive, fully automated cell density determination in suspension mammalian cell cultures.

Automated workflows for both mammalian transient protein production and protein expression from stable cell lines are used in many laboratories [Citation1,Citation2]. We have previously described fully automated cell culture workflows performed on GNF System’s (CA, USA) protein expression and purification platform (PEPP) for the production and purification of endogenous and recombinant proteins from mammalian cells such as hybridomas, FreeStyle 293-F, (ThermoFisher Scientific #R79007, MA, USA ) and CHO [Citation3]. On PEPP, cells are maintained in CELLSTAR® AutoFlasks (Greiner Bio-One #779190-J, Wein, Austria), which conform to the ANSI SLAS standard footprint compatible with common laboratory automation.

Suspension-adapted cell lines are frequently used for recombinant protein expression from mammalian cells because they grow to high densities and have high cell-specific productivities. Cell lines used on PEPP are suspension-adapted and maintained in shaking incubators. While suspension-adapted lines are easy to handle, cell density and viability must be monitored frequently to follow growth kinetics and to ensure that the culture does not undergo significant cell death prior to harvest. Trypan blue exclusion is a standard assay for determining density and viability of mammalian cells; several companies sell off-line instruments that automate this procedure [Citation4]. While commercially available cell counters confer advantages such as count reproducibility, accuracy and cell viability determination, the user must frequently withdraw samples from the culture vessel. This increases risk of culture contamination and reduces final culture volumes, affecting product yield. Additionally, off-line automated counters typically operate at lower throughput, so as sample numbers increase, more time and effort are spent monitoring cell densities. To address these issues, we worked to develop a high-throughput, noninvasive way to determine cell density. Spectrophotometry is often employed for density determination in bacterial, yeast, algal and human cell cultures [Citation5–7], and we applied that approach here. GNF Systems developed a device called the flask density reader (FDR), which consists of a Keyence Intelligent-B Laser Thrubeam Sensor (Keyence IB-05) housed in a custom nest designed to hold the AutoFlask (). The FDR laser operates at 660 nm wavelength, and the sensor measures transmitted intensity values for each flask. As the laser beam passes through the AutoFlask (1.85 cm path length), light is absorbed and scattered due to the presence of cells in the culture media. The resulting attenuated light, as detected by the sensor, is inversely related to cell density. The laser beam diameter is 5 mm (only 0.05% of the area of the AutoFlask). Like flow cytometry lasers, the FDR laser is rated Class 1, and we have not observed any impact on cells. The FDR is fully integrated onto PEPP and determines cell density in a 500-ms read per AutoFlask.

Figure 1.  The flask density reader.

Front (A) and side (B) views of the flask density reader with an AutoFlask seated in the custom nest.

Figure 1.  The flask density reader.Front (A) and side (B) views of the flask density reader with an AutoFlask™ seated in the custom nest.

To interpret FDR output values for CHO cells, FreeStyle 293-F cells and hybridomas grown in AutoFlasks on PEPP, we generated a standard curve for each cell type by plotting a range of cell culture densities measured on a stand-alone automated cell counter (Beckman Coulter Vi-CELL XR) against FDR output values (). Curves were fitted using the nonlinear regression analysis with a two-phase decay equation in GraphPad Prism 7.03. All three resulting curves visually yielded satisfactory fits with R2 values > 0.99, and they can be used to interpolate cell densities (X values) from FDR output (Y) values generated for cultures using the Interpolate function in GraphPad Prism (not shown). We recommend that the user establish a standard curve for each combination of cell type, media and labware used to account for factors that could impact FDR output values (e.g., flask material, color changes in phenol red-containing media). GFP expression does not impact FDR reads (not shown).

Figure 2.  Standard curves for three mammalian cell types grown in AutoFlasks.

Standard curves created in GraphPad Prism from data generated by reading mammalian cell suspension cultures grown in AutoFlasks side-by-side on the Vi-CELL and the flask density reader. (A) FreeStyle 293-F cells, (B) hybridoma cells and (C) CHO cells.

Figure 2.  Standard curves for three mammalian cell types grown in AutoFlasks™.Standard curves created in GraphPad Prism from data generated by reading mammalian cell suspension cultures grown in AutoFlasks side-by-side on the Vi-CELL™ and the flask density reader. (A) FreeStyle™ 293-F cells, (B) hybridoma cells and (C) CHO cells.

The FDR can also enable measurement of cell densities in vessels that do not conform to the ANSI SLAS format. For example, we used the FDR for automated determination of cell density in CHO cultures grown in CELLSTAR T-75 flasks (Greiner Bio-One #658 175), despite the longer path length (3.5 cm) (Supplementary Figure 1).

Other commercially available noninvasive devices for cell density determination exist, such as those from BugLab (http://buglab.com [Citation8]) and IPRASENSE (www.iprasense.com [Citation9]). The BugLab devices use an infrared or near-infrared light source and multiple detectors or laser/detector pairs, respectively, to provide accurate measurements over a wide range of cell concentrations. The IPRASENSE devices use lens-free microscopy with holographic imaging to determine both cell density and viability. The BugLab devices detect scattered light, so cells can be cultured in virtually any clear vessel, while the FDR detects transmitted light, necessitating the use of vessels with parallel flat surfaces and a defined short path length. The IPRASENSE devices are also built to accommodate vessels with parallel flat surfaces. The noninvasive BugLab devices are designed for either manual operation or surface mounting to a flask or bioreactor with corded data transmission to a base unit, while the FDR and the IPRASENSE devices can accommodate a plurality of flasks, making them more suitable for fully automated, higher-throughput cell density determination.

The FDR cannot be used for cell viability determination, and low cell viability negatively affects FDR data quality (not shown). Due to these limitations, we recommend that the user first monitor both cell density and viability over time using a Vi-CELL or similar method for his/her specific cell line, culture conditions and labware to determine how long high-density cultures can be maintained before a drop in cell viability is observed. The FDR can then be used to monitor cell growth up to that defined density and time-point. Additionally, at FDR output values below 10 (corresponding to ∼3e6 cells/ml for CHO and hybridoma cultures and ∼2.5e6 cells/ml for 293-F cultures), estimated cell densities become less accurate. For all cell lines tested, FDR measurements are not useful at cell densities <0.1e6 cells/ml. Accordingly, the optimal working range for monitoring CHO stable pool recovery from selection to determine when pools reach target density for cryo-archiving and inoculating protein production cultures (1e6 cells/ml) is from 0.1e6 to ∼3e6 cells/ml. Hybridoma and 293-F cultures reach lower cell densities than CHO cultures, so the FDR is also used to monitor growth of these cells.

The FDR can be purchased through GNF Systems (http://gnfsystems.com).

Author contributions

SR prepared the manuscript with editing by SG. SL and SF conceived of the FDR concept and SF performed proof-of-concept experiments and designed the device. JC built the FDR and integrated it onto the automated platform. JM, PA and SR generated standard curves using ViCell data and FDR output values.

Competing & financial interests disclosure

The authors were all GNF employees at the time of preparation of this manuscript, and this manuscript may help GNF Systems, a company associated with GNF, to sell devices or platforms. The authors have no other relevant affiliations or financial involvement with any organization or entity with a financial interest in or financial conflict with the subject matter or materials discussed in the manuscript apart from those disclosed.

No writing assistance was utilized in the production of this manuscript.

Supplemental material

Supplementary Figure 1

Download PDF (309.8 KB)

Supplementary data

To view the supplementary data that accompany this article please visit the journal website at: www.tandfonline.com/doi/suppl/10.2144/btn-2018-0059

References

  • Koehn J , HuntI. High-throughput protein production (HTPP): a review of enabling technologies to expedite protein production. Methods Mol. Biol.498, 1–18 (2009).
  • Battle T , AntonssonB, FegerG, BessonD. A high-throughput mammalian protein expression, purification, aliquoting and storage pipeline to assemble a library of the human secretome. Comb. Chem. High Throughput Screen.9(9), 639–649 (2006).
  • Rue SM , AndersonPW, GaylordMR, MillerJM, GlaserSM, LesleySA. A high-throughput system for transient and stable protein production in mammalian cells. In: Methods in Molecular Biology. VincentelliR (Ed.). (2018).
  • Louis KS , SiegelAC. Cell viability analysis using trypan blue: manual and automated methods. Methods Mol. Biol.740, 7–12 (2011).
  • Myers JA , CurtisBS, CurtisWR. Improving accuracy of cell and chromophore concentration measurements using optical density. BMC Biophysics6(1), 4 (2013).
  • Brescia P , RaymondK, BanksP. Monitoring cell growth in 2 μl volumes using a microplate reader (2011). www.biotek.com/resources/application-notes/monitoring-cell-growth-in-2-l-volumes-using-a-microplate-reader/.
  • Ude C , Schmidt-HagerJ, FindeisM, JohnGT, ScheperT, BeutelS. Application of an online-biomass sensor in an optical multisensory platform prototype for growth monitoring of biotechnical relevant microorganism and cell lines in single-use shake flasks. Sensors14(9), 17390–17405 (2014).
  • Debreczeny MP , RomeroJ, PetersenE. US 2009/0075248 A1 (2009).
  • Allier C , BordyT, HervéLet al. Label-free cell viability assay using lens-free microscopy. International Society for Optics and Photonics10497, doi:10.1117/12/2289811 (2018).