Feasibility and challenges of using multiple breath washout in COPD

, , &
Pages 2113-2119 | Published online: 10 Jul 2018

Abstract

Background

Lung clearance index (LCI), derived from multiple-breath washout (MBW), is a well-established assessment of ventilation inhomogeneity in cystic fibrosis but has not been widely applied in other conditions characterized by heterogeneous airways disease, such as COPD. The aim of this study was to evaluate the sensitivity, repeatability, and practicality of LCI in patients with COPD.

Methods

Fifty-four COPD patients completed MBW using nitrogen as the washout tracer gas (MBWN2, measured using an Exhalyzer™ device), spirometry, and plethysmography. Twenty patients repeated MBWN2, MBWSF6 (using a separate Innocor™ gas analyzer to measure washout of the exogenous trace sulphur hexafluoride), and spirometry at a second visit ≥24 hours later.

Results

Mean (SD) COPD LCI measured by nitrogen washout (LCIN2) was 12.1 (2.2); mean (SD) LCI Z-score 5.8 (2.0). LCIN2 increased across Global Initiative for Obstructive Lung Disease stages 1 to 3 and was abnormal (Z-score >1.65) in all COPD patients, including those with forced expiratory volume in 1 second (FEV1) ≥80% predicted. LCI was repeatable (median intra-test coefficient of variation 4.1%) and reproducible (limits of agreement −1.8 to 1.6) after mean of 16 days. Functional residual capacity (FRC) measurements were significantly greater using nitrogen than SF6 or plethysmography: mean FRC measured by nitrogen washout (FRCN2) 139% predicted versus FRC measured by plethysmography 125% predicted, p<0.0001.

Conclusion

LCI is most suitable as a measure of early airways disease in COPD in those with well-preserved FEV1, with similar repeatability and limitations to that observed in cystic fibrosis. Using the Exhalyzer system to perform MBWN2, however, appeared to substantially over-read FRC. This discrepancy needs addressing before FRCN2 measurements made using this device can be reliably deployed.

Introduction

Multiple-breath washout (MBW) testing is now well established in cystic fibrosis (CF) research as a practical and reproducible measure of airways physiology that is both more sensitive at detecting the presence of disease and at monitoring treatment responses compared with conventional spirometry.Citation1 MBW involves tidal breathing to wash an inert tracer gas from the lungs and generates measures of lung volume (functional residual capacity; FRC) as well as of gas mixing efficiency.Citation2 The most important and commonly reported outcome is the lung clearance index (LCI), a summary measure of ventilation heterogeneity. It has been known for some time that COPD results in abnormal gas mixing,Citation3 and also that ventilation heterogeneity is an early feature of smoking-related lung disease.Citation4 Ventilation heterogeneity has been demonstrated in COPD using ventilation-MRI, and seems to relate to other measures of disease severity.Citation5 However, the applicability of current MBW technologies for measuring ventilation heterogeneity in COPD has not been studied.

The European CF Society have recently published recommendations for use of MBW in clinical trials in CFCitation6 and there is a drive for clinical application of these measurements.Citation7 MBW is now available in several CF centers in the US and Europe, the majority using nitrogen washout systems, which derive nitrogen concentration from measurements of expired O2 and CO2. Nitrogen washout is performed by breathing 100% oxygen. This differs from the majority of recent formative MBW studies where SF6 was used as the tracer gas, with washout by breathing room air. Different technologies seem to deliver different measures of LCI and FRC,Citation8 and these are not interchangeable even between nitrogen washout systems.Citation9

The role of MBW in COPD may differ from that in CF, being potentially more relevant in disease phenotyping and assessing treatment response than in detecting and preventing early disease progression. For MBW to be adopted in this way, it must be sensitive to disease identification, practical to perform,Citation10 and reproducible.Citation11 The applicability of MBW in adults with COPD has not previously been assessed.

We hypothesized that MBW would be a practical, repeatable, and potentially informative test in adults with COPD. The aims of this study were, therefore, as follows:

  1. To assess practical limitations of MBW in patients diagnosed with COPD, including test time and success rates.

  2. To assess intra-visit repeatability and inter-visit reproducibility of LCI and FRC.

  3. To compare MBW outcomes with other lung function measures in patients diagnosed with COPD.

  4. To compare indices from nitrogen washout with those from SF6 washout.

Materials and methods

This study was performed at the Medicines Evaluation Unit, University Hospital of South Manchester, UK. COPD patients were aged >40 years and had a diagnosis of COPD in accordance with the 2013 Global Initiative for Obstructive Lung Disease (GOLD) recommendations.Citation12 This included a smoking history ≥10 pack-years with typical symptoms (1 or more of productive cough, breathlessness, and wheeze) and evidence of airflow obstruction (forced expiratory volume in 1 second [FEV1]/forced vital capacity <0.7). Additional exclusion criteria included any respiratory tract infection or COPD exacerbations within 6 weeks of assessment.

Patients diagnosed with COPD continued all regular pharmacological therapies but restricted short-acting inhaled bronchodilators for a minimum of 8 hours, and long-acting inhaled bronchodilators for a minimum of 24 hours prior to assessments. The study was approved by the Greater Manchester South research ethics committee (reference: 10/H1003/108) and written informed consent was obtained from each subject.

Study design

Fifty-four patients diagnosed with COPD performed impulse oscillometry, MBW, gas transfer, whole-body plethysmography, spirometry, and a 6-minute walk test (6MWT) of exercise capacity (in that order). Subjects were invited to return for repeat MBW testing >24 hours later.

Assessments

Detailed descriptions of each of the measurements can be found in the online supplement. Multiple-breath nitrogen washout (MBWN2) was performed as previously describedCitation8 using an open circuit, bias flow system (Exhalyzer D®, Eco-Medics AG, Duernten, Switzerland). Analysis was performed using the associated software (Spiroware 3.1 EcoMedics AG). MBWN2 was performed 3 times and reported values for FRC measured by nitrogen washout (FRCN2) and LCI measured by nitrogen washout (LCIN2) are the average of at least 2 reproducible repeats. Age-matched Z-scores for LCIN2 were taken from those provided by the manufacturers.

A subset of patients performed additional MBW of SF6 (MBWSF6). This was carried out using an open circuit washin of 0.2% SF6 followed by washout breathing room air, using a modified Innocor analyzer as previously described.Citation13 Both MBWN2 and MBWSF6 were performed according to consensus recommendations.Citation14

Spirometry was performed according to the American Thoracic Society/European Respiratory Society recommendationsCitation15 using a Sensormedics Vmax spirometer (Sensormedics Corporation, Yorba Linda, CA, USA). Reference values used were those of the 2012 Global Lung Initiative.Citation16

Impulse oscillometry was performed using a Masterscreen impulse oscillometer (IOS; Erich Jaeger, Hoechenberg, Germany) in accordance with consensus recommendations.Citation17 Plethysmographic lung volumes were measured in accordance with the American Thoracic Society/European Respiratory Society recommendationsCitation16 using a constant volume whole-body plethysmograph (Autobox 6200 DL, Sensor-medics Corporation). Carbon monoxide diffusing capacity (DLCO) and carbon monoxide transfer coefficient (KCO) were measured using the Vmax 22 instrument (Sensormedics Corporation), according to ATS/ERS recommendations.Citation18 The reported value is the mean of at least 2 successful readings within 10% of the highest value. The 6MWT test was conducted to a standardized protocol in accordance with current guidelines using a 20-meter course.Citation19

Statistical analysis

A sample size of 10 patients would give 99% power to detect a 20% difference in LCI over historical control data and a sample size of 22 would be required to detect a 10% difference in LCI. We were interested, however, in how LCI compared with lung physiology across a range of disease severity and estimated that 50 patients would be sufficient to make reasonable inferences about how airway markers compared, and would give us enough subjects to get repeat LCI measurements on at least 20.

Data were analyzed using Prism (GraphPad Software Inc., San Diego, CA, USA). Parametric data were expressed as mean (SD) and non-parametric data expressed as median with interquartile range (IQR). For comparison of physiology parameters with GOLD stage, a 1-way analysis of variance was conducted, followed by Tukey’s multiple comparison test to compare pairs of data. Correlations were assessed using a Pearson’s correlation coefficient or a Spearman’s rank correlation coefficient. Intra-visit repeatability was determined by calculating the coefficient of variation (CV) of the 3 recordings for each subject. Inter-visit reproducibility was assessed using Bland–Altman analysisCitation20 and intra-class correlation coefficient (ICC). ICC was calculated using SPSS (version 22, IBM SPSS inc, Armonk, NY, USA) using a 2-way mixed analysis of absolute agreement. For the comparison of MBW methods, outcomes were compared using a paired t-test and the agreement between the 2 systems assessed using Bland–Altman analysis.Citation20 Significance level was set at p=0.05.

Results

Fifty-four patients diagnosed with COPD (37 male) performed assessment of lung function and MBW. Mean (SD) age of subjects was 66 (7) years, with a median smoking history of 42 pack-years (full demographics presented in ). Seventeen (31%) of the COPD patients were frequent exacerbators, defined as having ≥2 exacerbations in the last 12 months.

Table 1 Clinical characteristics and lung function measurements

Practicality of MBWN2

Four out of the 54 patients diagnosed with COPD (7%) were unable to complete MBWN2 due to leak at the mouthpiece, dry throat, or a sensation of dyspnea when breathing high-flow O2. Nine repeat measurements were discarded for failure to meet quality control criteria. Patients with high minute ventilation would sometimes re-inspire expired air or room air from the end of the flow-past circuit. Overall, success rate was similar to that achieved for the other physiological assessments (Table S1). Median (IQR) total test time in COPD patients was 36.3 (23.8–44.9) minutes. Test time increased in those with higher LCI, rCitation2=0.24, p=0.0006 (Figure S1).

Repeatability and reproducibility of MBW in COPD

The median (IQR) intra-test CV of triplicate repeat FRCN2 measurements on the same visit was 4.3 (1.7–6.3) for FRC and 4.1 (2.2–7.1) for LCI.

Twenty clinically stable patients diagnosed with COPD performed repeat measurements of MBWN2 in triplicate after a mean (SD) interval of 16 (13) days. There was a mean difference in FRCN2 between visits of −0.4% predicted (absolute), 95% limits of agreement −28.9% to 28.1% (Figure S2). ICC for repeat FRC was 0.90 (95% CI: 0.76–0.96). For LCIN2, the mean difference between visits was −0.11 with limits of agreement −1.79 to 1.59, approximately equivalent to ±15% of initial LCI (Figure S3). ICC for LCI was 0.94 (95% CI: 0.86–0.98).

Comparison of LCI with other lung physiology assessments

LCIN2 was raised in all patients with COPD (Z-score >1.65), including 10 (20% of the COPD patients) with GOLD stage 1 (defined as FEV1 >80% predicted).Citation21 LCIN2 showed significant inverse correlation with FEV1 Z-score (rCitation2=0.36, p<0.0001; ) and forced expiratory flow (FEF)25%–75% (rCitation2=0.36, p<0.0001). LCIN2 also showed significant positive correlations with R5–R20 (rCitation2 =0.17, p=0.009), residual volume/total lung capacity (RV/TLC [%]) (rCitation2=0.16, p=0.02), Raw (rCitation2=0.30, p=0.0006), and significant negative correlations with X5 (rCitation2=0.35, p=0.0006), sGaw (rCitation2 =0.18, p=0.012) and DLCO (rCitation2=0.34, p=0.0001). There was no correlation between LCIN2 and 6MWT distance. Full details of these correlations are presented in Table S2 and Figures S4–S11.

Figure 1 COPD LCIN2 Z-score versus FEV1 Z-score.

Notes: Z-scores were calculated using the GLI 2012 reference equations for FEV1Citation17 and those provided by Ecomedics for LCI. The horizontal dotted line represents the upper limit of normal for LCIN2. The vertical dotted line represents the FEV1 Z-score of −1.65 (lower limit of normal range).
Abbreviations: FEV1, forced expiratory volume in 1 second; GLI, global lung initiative; LCIN2, lung clearance index measured by nitrogen washout.
Figure 1 COPD LCIN2 Z-score versus FEV1 Z-score.

LCIN2 also increased significantly with greater GOLD stage. Mean (SD) LCIN2 in GOLD stage 1 was 10.2 (1.2), versus GOLD stage 2 (FEV1 50%–80%), 11.9 (1.8) p=0.048, or GOLD stage 3 (FEV1 30%–50%), 14.19 (2.3) p<0.0001 (). There was only a single subject with GOLD stage 4 COPD, omitted from this analysis. Difference between lung physiology outcomes for the different GOLD stages (1–3) is presented in the OLS (Table S3). In addition to LCI, there was statistically significant separation between all 3 GOLD stages for sGaw and FEF25%–75%.

Figure 2 LCIN2 versus COPD GOLD stage.

Abbreviations: GOLD, Global Initiative for Obstructive Lung Disease; LCIN2, lung clearance index measured by nitrogen washout
Figure 2 LCIN2 versus COPD GOLD stage.

Comparison of FRC assessments

Seventeen patients diagnosed with COPD completed separate sequential assessments of FRC and LCI using both N2 and SF6 as the washout gases (). There was no significant difference in LCI measured by sulphur hexafluoride washout (LCISF6) and LCIN2 (11.9 [3.1] versus 12.3 [2.2], p=0.4). FRCN2, however, was significantly elevated compared with FRC measured by sulphur hexafluoride washout (FRCSF6). Mean (SD) of FRCN2 was 136.8 (42.8) % predicted compared with 112.4 (28.0) % for FRCSF6, p=0.002. The difference between FRCN2% and FRCSF6% increased disproportionally as FRC% increased. The mean difference was 24.4 (26.4) % predicted with limits of agreement −27.3 to 76.1 % predicted (Figure S12).

Table 2 Clinical characteristics and lung function measurements of COPD patients with paired MBW assessments

FRC values measured by plethysmography and nitrogen washout were also compared for the whole cohort at visit 1. Overall, FRCN2 was significantly greater than FRC measured by plethysmography (FRCpleth) (139.3% versus 124.5%, p<0.0001). Mean difference between FRCN2 and FRCpleth was 14.8% predicted (limits of agreement −17.3 to 46.8; ), with a disproportionate increase in FRCN2 in those with higher FRC % predicted.

Figure 3 Bland–Altman comparison between FRCN2 and FRCpleth in subjects with COPD.

Notes: The central dotted line represents the mean difference, and the upper and lower dotted lines represent the limits of agreement. The vertical line represents hyper-inflation, defined as FRCN2 >120% predicted. Data points plotted above the x-axis represent subjects in which FRCN2 was greater than FRCpleth.
Abbreviations: FRC, functional residual capacity; FRCN2, FRC measured by nitrogen washout; FRCpleth, FRC measured by plethysmography.
Figure 3 Bland–Altman comparison between FRCN2 and FRCpleth in subjects with COPD.

Discussion

This is the first study to compare different MBW techniques, alongside detailed physiological phenotyping, in patients diagnosed with COPD. The specific objectives were to understand more about the repeatability and practical limitations of MBW in COPD. This study found LCI to be sensitive to detect mild COPD, being elevated even in those in GOLD stage 1. Intra-visit repeatability was similar to that reported in CF studiesCitation8,Citation13 and inter-visit reproducibility was similar to that of CF adults.Citation22 Finally, LCI has physiologically valid correlations with other small airways measurements (including X5, R5–R20, RV/TLC, Raw) as well as other markers of COPD severity, such as DLCO. This accords with recent work showing that smokers with preserved FEV1 had evidence of small airways damage, gas trapping, and impaired gas transfer.Citation23,Citation24 Abnormality in single-breath nitrogen washout in symptomatic smokers and in MBW parameters in asymptomatic smokers (>10 pack years) has also been previously demonstrated.Citation24,Citation25 The particular advantages of LCI as an outcome for assessing these early changes is that the methodology is standardized,Citation14 with good reproducibility, and with apparently good sensitivity. This last feature is harder to quantify, however, since there is no gold-standard definition of COPD that does not involve spirometry. It is nevertheless reassuring that LCI was abnormal (Z-score >1.65) in all the subjects with previously defined COPD and that it appeared to be more sensitive in this regard at detecting early airway change than the other lung physiology assessments. As with severe CF, however, the measurement appears less useful in those with more advanced COPD. In these patients, the test takes much longer to complete and the overall ventilation heterogeneity signal likely reflects a mixed pattern of disease. Thus, any signal from potentially reversible small airways disease will be overwhelmed by the signal from other lung regions where damage is more advanced and irreversible. These data and observations, therefore, support a role for MBW in measuring disease severity in patients with mild-to-moderate (GOLD stage 1–2) COPD, and possibly in detecting disease before it fulfills current spirometry-driven criteria.

An important caveat, however, exists for the specific nitrogen washout system used in this study. An unanticipated finding was that the measures of FRCN2 were far greater not only than those of FRCSF6 but also FRCpleth. Plethysmography measures all gas within the thoracic cavity and should, therefore, always measure a greater value for “FRC” than that which is ventilated during the quiet tidal breathing of a MBW test. FRCpleth and FRCN2 are similar in healthy subjects,Citation26 but increasing FRCpleth should be apparent in disease as gas trapping worsens, leading to an even greater discrepancy between tidally ventilated and total end-expiratory lung volumes.Citation27 In this study, however, we have observed the opposite. Although both FRCpleth and FRCN2 were elevated in COPD, FRCN2 was higher than FRCpleth with an offset of almost 20%. This difference also worsened in those with more severe disease, implying that tidally ventilated lung volume becomes increasingly greater than total lung volume as disease progresses. Clearly, this conclusion cannot be correct.

A recent study compared 2 similar indirect N2 devices (Exhalyzer D versus EasyOne Pro, NDD Medical Technologies, Zurich, Switzerland). A difference of 30% in FRC values was found between these systems, with the Exhalyzer D generating higher mean values,Citation9 consistent with the observations in this study. This suggests an over-reading of FRC by the Exhalyzer system, which may be device-specific. Recent attempts to correct for excretion of body nitrogen did not fully account for this error,Citation28,Citation29 and are not included in this analysis. The limitations of this same MBWN2 system for measuring lung volumes have also recently been recognized in patients with CF,Citation30 and seem to have occurred despite promising in vitro data for this method.Citation7 Importantly, accurate FRC measurement is essential for all MBW outcome measurements, including LCI. Reassuringly, however, the SF6 washout gave similar values for LCI despite much lower values for FRC. This may be because both FRC and cumulative expired volume are overestimated to the same extent with the nitrogen washout, but importantly suggests no loss of sensitivity by using SF6.

In this study, we have assessed patients diagnosed with COPD over a wide range of severity, including at repeat time points. The lack of a simple positive control for COPD, however, means that we cannot easily assess what the response to therapy should be. We also did not include a healthy control population, but have instead used the reference data from previous age-matched cohorts provided by the manufacturer to calculate Z-scores. LCI was assessed on site, using the on-board software of the Exhalyzer-D device. Measurements were not subject to external over-read as recently recommended in clinical trials,Citation6 although the research team has considerable experience with MBW as a technique.

Conclusion

We have shown that LCI offers substantial promise as a measurement in COPD. The key practical features of MBW assessment in COPD are similar to those in CF, and LCI appears to work best as a measure of early airways disease in those with well-preserved FEV1. However, the specific nitrogen washout system used here exhibited major limitations in terms of the accuracy of the FRC measurements, something that has not been adequately accounted for and which has significant implications for clinical trials and trial networks already in progress. Considerable caution should, therefore, be employed in interpreting nitrogen washout outcomes until the reasons for this error are better understood.

Acknowledgments

The authors gratefully thank the patients who kindly took part in this study and acknowledge Paul Hitchen in aiding data collection. This study received no specific grant from funding agencies in the public, commercial, or not-for-profit sectors. AH received funding to support the study from the National Institute of Health Research (NIHRS012-13). This report, therefore, presents independent research funded by the National Institute for Health Research (NIHR). The views expressed are those of the author(s) and not necessarily those of the NHS, the NIHR, or the Department of Health.

Disclosure

ASB and PJL were supported by the Medicines Evaluation Unit, University Hospital of South Manchester, UK. DS reports personal fees from Apellis, Peptinovate, Skyepharma, Cipla, and Genentech, and grants and personal fees from AstraZeneca, Boehringer Ingleheim, Chiesi, Glenmark, Merck, Mundipharma, Novartis, Pfizer, Pulmatrix, Teva, Therevance and Verona, all outside the submitted work. AH reports personal fees from Celtaxys, Vertex, Chiesi, Boehringer Ingelheim, and collaboration agreement with Innovision ApS, outside the submitted work. The authors report no other conflicts of interest in this work.

References

  • HorsleyASiddiquiSPutting lung function and physiology into perspective: cystic fibrosis in adultsRespirology2015201334525219816
  • RobinsonPDGoldmanMDGustafssonPMInert gas washout: theoretical background and clinical utility in respiratory diseaseRespiration200978333935519521061
  • FlemingGMChesterEHSaniieJSaidelGMVentilation inhomogeneity ising multibreath nitrogen washout: comparison of moment ratios and other indexesAm Rev Respir Dis198012157897947406312
  • VerbanckSSchuermansDMeysmanMPaivaMVinckenWNoninvasive assessment of airway alterations in smokers: the small airways revisitedAm J Respir Crit Care Med2004170441441915130906
  • SwiftAJWildJMFicheleSEmphysematous changes and normal variation in smokers and COPD patients using diffusion 3He MRIEur J Radiol200554335235815899335
  • KentLReixPInnesJAEuropean Cystic Fibrosis Society Clinical Trial Network (ECFS-CTN) Standardisation CommitteeLung clearance index: evidence for use in clinical trials in cystic fibrosisJ Cyst Fibros201413212313824315208
  • SingerFHoultzBLatzinPRobinsonPGustafssonPA realistic validation study of a new nitrogen multiple-breath washout systemPLoS One201274e3608322558338
  • JensenRStanojevicSGibneyKMultiple breath nitrogen washout: a feasible alternative to mass spectrometryPLoS One201382e5686823457632
  • PoncinWSingerFAubriotASLebecquePAgreement between multiple-breath nitrogen washout systems in children and adultsJ Cyst Fibros201716225826627919570
  • HuqueMFAloshMBhoreRAddressing multiplicity issues of a composite endpoint and its components in clinical trialsJ Biopharm Stat201121461063421516560
  • FuchsSIEllemunterHEderJFeasibility and variability of measuring the Lung Clearance Index in a multi-center settingPediatr Pulmonol201247764965722170866
  • VestboJHurdSSAgustiAGGlobal strategy for the diagnosis, management, and prevention of chronic obstructive pulmonary disease: GOLD executive summaryAm J Respir Crit Care Med2013187434736522878278
  • HorsleyARGustafssonPMMacleodKALung clearance index is a sensitive, repeatable and practical measure of airways disease in adults with cystic fibrosisThorax200863213514017675315
  • RobinsonPDLatzinPVerbanckSConsensus statement for inert gas washout measurement using multiple- and single- breath testsEur Respir J2012413507522
  • MillerMRHankinsonJBrusascoVATS/ERS Task ForceStandardisation of spirometryEur Respir J200526231933816055882
  • WangerJClausenJLCoatesAStandardisation of the measurement of lung volumesEur Respir J200526351152216135736
  • OostveenEMacLeodDLorinoHERS Task Force on Respiratory Impedance Measurements. The forced oscillation technique in clinical practice: methodology, recommendations and future developmentsEur Respir J20032261026104114680096
  • MacintyreNCrapoROViegiGStandardisation of the single-breath determination of carbon monoxide uptake in the lungEur Respir J200526472073516204605
  • CrapoROEnrightPLZeballosRJATS Committee on Proficiency Standards for Clinical Pulmonary Function Laboratories. American Thoracic Society Statement: guidelines for the six-minute walk testAm J Respir Crit Care Med2002166111111712091180
  • BlandJMAltmanDGStatistical methods for assessing agreement between two methods of clinical measurementLancet1986184763073102868172
  • VogelmeierCFCrinerGJMartinezFJGlobal strategy for the diagnosis, management, and prevention of chronic obstructive lung disease 2017 report. GOLD executive summaryAm J Respir Crit Care Med2017195555758228128970
  • HorsleyAShawcrossAOladapoMP255 Reproducibility of lung clearance index (LCI) in clinically stable adults with mild cystic fibrosis (CF)Thorax201671A226
  • ElbehairyAFFaisalAGuenetteJACanadian Respiratory Research Network (CRRN)Resting physiological correlates of reduced exercise capacity in smokers with mild airway obstructionCOPD201714326727528368706
  • Di MarcoFTerraneoSJobSCardiopulmonary exercise testing and second-line pulmonary function tests to detect obstructive pattern in symptomatic smokers with borderline spirometryRespir Med201712771328502420
  • VerbanckSSchuermansDMeysmanMPaivaMVinckenWNoninvasive assessment of airway alterations in smokers: the small airways revisitedAm J Respir Crit Care Med2004170441441915130906
  • DuboisABBotelhoSYBedellGNMarshallRComroeJHJrA rapid plethysmographic method for measuring thoracic gas volume: a comparison with a nitrogen washout method for measuring functional residual capacity in normal subjectsJ Clin Invest195635332232613295396
  • CoatesALPeslinRRodensteinDStocksJMeasurement of lung volumes by plethysmographyEur Respir J1997106141514279192953
  • NielsenNNielsenJGHorsleyAREvaluation of the impact of alveolar nitrogen excretion on indices derived from multiple breath nitrogen washoutPLoS One201389e7333524039916
  • KaneMRaymentJHJensenRMcDonaldRStanojevicSRatjenFCorrecting for tissue nitrogen excretion in multiple breath washout measurementsPLoS One20171210e018555329020072
  • BayfieldKJSaundersCAltonEWFWDaviesJCComparison of CF and Non CF LCI results using the Exhalyzer D and InnocorTM devicesThorax201570Suppl 3A115