3,682
Views
80
CrossRef citations to date
0
Altmetric
Review

New developments in prosthetic arm systems

, &
Pages 31-39 | Published online: 07 Jul 2016

Abstract

Absence of an upper limb leads to severe impairments in everyday life, which can further influence the social and mental state. For these reasons, early developments in cosmetic and body-driven prostheses date some centuries ago, and they have been evolving ever since. Following the end of the Second World War, rapid developments in technology resulted in powered myoelectric hand prosthetics. In the years to come, these devices were common on the market, though they still suffered high user abandonment rates. The reasons for rejection were trifold – insufficient functionality of the hardware, fragile design, and cumbersome control. In the last decade, both academia and industry have reached major improvements concerning technical features of upper limb prosthetics and methods for their interfacing and control. Advanced robotic hands are offered by several vendors and research groups, with a variety of active and passive wrist options that can be articulated across several degrees of freedom. Nowadays, elbow joint designs include active solutions with different weight and power options. Control features are getting progressively more sophisticated, offering options for multiple sensor integration and multi-joint articulation. Latest developments in socket designs are capable of facilitating implantable and multiple surface electromyography sensors in both traditional and osseointegration-based systems. Novel surgical techniques in combination with modern, sophisticated hardware are enabling restoration of dexterous upper limb functionality. This article is aimed at reviewing the latest state of the upper limb prosthetic market, offering insights on the accompanying technologies and techniques. We also examine the capabilities and features of some of academia’s flagship solutions and methods.

Introduction

A significant portion of the injuries treated in the emergency rooms around the globe involve upper extremities.Citation1Citation3 The majority of them occur at home,Citation2 during work,Citation4,Citation5 or when performing sports.Citation6 Considering that almost all our everyday activities depend on manipulation by the hands, severe hand injuries can truly be devastating. Consequences of such incidents can lead to long-term disabilities, also affecting the mental and social state, with difficult reintegration in the society.Citation7,Citation8

The severe consequences of upper limb loss have been recognized centuries ago,Citation9 and the ideas of artificial substitution have been since then very appealing. Transition from simple cosmetic prostheses to a more functional solution was inevitable and in high demand, resulting in the development of early body-powered and cable-driven systems. These simple devices proved themselves to be very useful and, with modern materials, are still currently in use. Approximately half of the current market of upper limb prosthetics is indeed based on cable-driven systems.Citation10

The first pneumatic hand was developed at the beginning of the 20th century, soon followed by the first electric-powered hand. At the end of the Second World War, early concepts of myoelectric prostheses were introduced.Citation11 These devices, which linearly translated the electrical activity of the residual muscles of the stump into the velocity of closing and opening of gripers, started to be used in research laboratories and to be sold on the market in the late 1950s. The concept of direct proportional control is still present in current commercially available systems, due to its simplicity and robustness.

By the late 1960s, pneumatic prostheses were able to drive and control several joints and grip types. However, the control was inefficient and not robust enough, requiring specific anatomical features, dexterity, and cognitive effort of the patient.Citation12 Myoelectric systems have tried to face these issues with state-based control. Accordingly, the patient would control the prosthesis using two control sites, as in the single degree of freedom (DoF) case. When there was the need to control a different joint or grip type, a co-contraction of the muscles under the two recording sites changed the control state of the prosthesis. This quite cognitively demanding system is still dominant on the market of dexterous prosthetics, mostly due to its robustness.

Numerous studies have been conducted in the past two decades, still indicating high rejection rates of all types of upper limb prosthetic devices across a variety of users.Citation13Citation18 Depending on the study population, rejection rates vary from 25% to >50% for myolelectric and up to 35% for body-powered devices. However, these figures have leveled off with respect to previous periods,Citation19,Citation20 mainly due to modern technology. However, it is possible that the trend will eventually reverse.

Hand transplantation is an alternative to the prosthetic devices, offering functionality, superior visual appeal, and integrated sensory function.Citation21 However, it is associated with the lifelong immunosuppressant therapy, lengthy rehabilitation, loss of grip force, and high risk of complications, leading to the possible rejection.Citation22 These issues are then combined with very high costs.

Advances in micromachining and material design have enabled construction of versatile lightweight prosthetic hands and wrists. These market products, in combination with precise, small-sized, low-consumption electromotors, corresponded to highly actuated systems. Development of high-speed processing units with the top-end battery management and large memories in small housing propelled the research into more advanced and intuitive control systems. Greater understanding of the human neuromuscular system yielded new surgical and reconstructive techniques, which now provide access to high-quality and intuitive electromyography (EMG) sources even in high-level amputations. Socket design has also benefited from this rapid development of technology and now can offer solutions that are able to host multiple surface sensors, facilitate the use of implanted electrodes, and, in combination with surgical advancements, provide direct link with the skeletal system in the form of osseointegration. Finally, three-dimensional printing is quickly becoming a viable alternative for production of highly customizable products that are lightweight and inexpensive. Several open-source hand prosthetic projects are available for personal printing, and an ever growing number of companies are already using this method for building certain components of their own products.

Considering the latest rate of innovation in the field of prosthetics, significant increase in funding, and the number of new competitors on the market, this review aims to present the latest state of functional, myoelectrically controlled upper limb prosthetic solutions. The goal is to make a literature overview of the new developments of representative hardware, control algorithms, and interfaces from both the medical and technical perspective. Even though highly significant for prosthetic applications, solutions and research in the area of sensory feedback are beyond the scope of this review.

Hardware

The current market of actuated myoelectrically controlled upper limb prosthetic devices can roughly be divided into systems that address transradial/transcarpal, transhumeral, and shoulder disabilities. Each group features specific requirements, and, in general, the options for replacement are more advanced the more distal the level of impairment is.

Transradial/transcarpal solutions

Due to the complex anatomical nature of hands and yet crucial role in object handling and manipulation, their prosthetic counterparts have undergone an important evolution and functional advances in recent years. Although simple grippers are still dominant in the market, multi-actuated hands that provide several grip types or even fully controllable individual digits and finger joints are now common ().

Figure 1 Examples of Ottobock prosthetic hands and cosmetics (from left to right): small System Inner Hand, small MyoHand VariPlus Speed, and medium Michelangelo hand.

Figure 1 Examples of Ottobock prosthetic hands and cosmetics (from left to right): small System Inner Hand, small MyoHand VariPlus Speed, and medium Michelangelo hand.

Some of the current common characteristics of the most promising commercial hand products – i-limb Quantum by Touch Bionics, RSL Steeper’s BeBionic v3, and Ottobock’s Michelangelo – are presented in . For the sake of putting these products into perspective, the same table lists three different, already established, products aiming at children with hand disabilities, adult long stump cases, and usual transradial users – Centri Child Myoelectric Hand, Transcarpal-Hand by Ottobock Healthcare, and Select Electric Hand by Liberating Technologies, respectively.

Table 1 Features of three already established devices aiming at children population, transcarpal and transradial cases, and three commercially available flagship prosthetic hands

From , it is evident that the increase in functionality did not significantly influence the size, the weight, and the power grasp force of prosthetic hands. Moreover, the listed products all offer certain wrist solutions, indicating an overall tendency of the market for further development of this joint. Besides the features listed, each of the hands in is also equipped with a variety of product-specific features, mostly focusing on the different grasp types and safety measures (this is further discussed in the “Control strategies” section).

Prosthetic hand devices developed by smaller companies or research laboratories include the world’s first touch sensing hand prosthesis from Vincent Systems, Evolution 2, and DARPA founded DEKA Arm RC. Evolution 2 combines the sensory feedback information with the individually motorized digits and fully actuated thumb in a compact and light package (~400 g).Citation23 The DEKA Arm RC is the heaviest of the products so far listed (1,270 g)Citation24 and, in its third generation, offers highly actuated digits and thumb with an included compound wrist that can be actuated in three DoFs.Citation25

In order to provide a more natural object manipulation and therefore increase functional benefits to the users,Citation26 modern prosthetic hand devices are frequently accompanied by an actively or passively controllable wrist joint. Passive versions of prosthetic wrists can be manually adjusted by the user in order to be in either compliant mode or to lock in one of the predefined positions in flexion/extension direction (Ottobock AxonWrist, Touch Bionics Flexion Wrist, BeBionic Flexion-Wrist) or along the rotation axis (Ottobock AxonRotation, Touch Bionics QWD, BeBionic Short Wrist). Additionally, certain vendors offer wrists that allow prosthetic hands to be positioned in any direction (MyolinoWrist 2000, BeBionic Multi-Flex). Commercial active units are almost exclusively focused on the wrist rotation (MC Wrist Rotator, Ottobock Electric Wrist Rotator).

Transhumeral and shoulder disarticulation solutions

The more proximal the upper limb impairment is, the greater the disability. In transhumeral amputations, the absence of an elbow requires an additional prosthetic joint to substitute the missing two DoFs. Even though the rotation of the forearm can be compensated using the wrist rotation unit, the major setback at this level of disability is the reduced number of sources and methods for controlling all the necessary prosthetic components. As for the wrist, prosthetic elbows can be passive or active with several locking positions. Passive or body-powered elbows are dominant on the market, though several major vendors also provide electrically powered, myoelectrically controllable devices. Some representative products and their characteristics are listed in . The DEKA Arm HC offers a motorized elbow solution with a limited range of motion to prevent reaching the face for safety reasons.Citation25 This characteristics and the absence of the free swing mode have not been well received by the users.Citation25

Table 2 Features of commercially available myoelectric elbows

Even though improvements in academia have indicated possible solutions in designing elbow joints with anthromorphic characteristics,Citation27 one of the current main challenges is the design of a device fully compliant with all the standards and yet light enough to be suited for children and frail adults.

Powered shoulder joints are more complex and currently not present on the market. Nonetheless, promising solutions have been presented in research laboratories.Citation25,Citation28 The DEKA arm HC has recently received Food and Drug Administration approval,Citation29 and it offers four movement directions in the shoulder joint.Citation25 Moreover, for exploring the full possibilities of the targeted muscle reinnervation (TMR) approach, a modification of the LTI-Collier shoulder joint has been developed with an added electronic lock/unlock feature.Citation28

Socket technology

The amount of time a user wears a prosthesis mainly depends on the socket fit and its design.Citation30 Inadequate prosthetic fit might lead to limited range of motion, discomfort, and general poor performance,Citation31 usually resulting in the abandonment of the device.Citation32

After the development of the Otto Bock Muenster style socket in the 1960sCitation33 and the Northwestern University socket in 1972,Citation34 the transradial powered prosthesis socket design has not significantly changed. The introduction of flexible thermoplastics has indeed been the only major improvement.Citation35 At the end of 1990s, silicon liners have been introduced, and the transparent, moldable plastics allowed better analysis of the inside socket dynamics, resulting in a tighter contouring around the fitted stumpCitation30 (). Nowadays, new textile materials allow better, more hygienic, and less obstructing harnesses to be developed and custom fit to higher level upper limb amputees. The general tendency over the past few years, which was enabled through these novel materials, is the design of anatomically contoured sockets for all levels of amputation.Citation36

Figure 2 Example socket design with custom pattern, silicon liner, and Ottobock Energy Pack housing.

Figure 2 Example socket design with custom pattern, silicon liner, and Ottobock Energy Pack housing.

Based on the osseointegration technique introduced ~60 years ago,Citation37 an advanced mounting concept of the upper limb prosthesis has been developed and applied.Citation38 The main idea behind this technique is to exploit the direct structural and functional connection between the skeletal bone and the surface of the titanium implant,Citation39 which would further be connected to an additional implant penetrating the skin.Citation40 In this way, a point for a direct, rigid connection of the prosthesis and the skeletal system is created.Citation41 Osseointegration offers numerous advantages over the traditional sockets by providing a more intimate fit, increased range of motion, and osseoperception.Citation42,Citation43 However, it also requires additional surgery and poses potential risks of infection, implant fracture, or incomplete integration requiring revision surgeries.Citation44,Citation45

The introduction of implantable sensing technologies for EMG detection and control of prosthetic devices such as implantable myoelectric sensorCitation46 requires certain revisions of the standard sockets. Namely, implantable myoelectric sensor compatible shafts are equipped with transmitter/receiver coil capable of enclosing the stump and receiving the signal transmitted by the implanted EMG electrodes.Citation47

Other, more research-oriented, invasive solutions, such as implantable electrode arrays,Citation48 fine wire-based systems,Citation49 and epimysial electrodes,Citation50 are at a prototype stage. Nerve interfacing solutionsCitation51 for efferent decoding remain currently highly complex without sufficient benefits.

Control strategies

Myoelectric prosthetic upper limbs have been known for over half a century.Citation11 As previously elaborated, advances have been made in many aspects of these devices. Yet, the commercially available control systems up until recently have practically remained unchanged. The initial two-channel control allowing direct proportional steering of a single DoF has been extended to sequentially drivable multiple DoFs using a state machine approach.Citation52 With this approach, the user can decide which DoF of the device to control by cycling through states using co-contractions or by quickly repeating a specific muscle activation pattern.Citation53,Citation54 For example, Michelangelo hand users may switch between DoFs by modulating the contraction speed.Citation54 Regardless, all these approaches, even though quite robust, are unintuitive and cumbersome especially in the high-level amputation cases, which require articulation of numerous DoFs.Citation55

Some prosthetic hand vendors proposed a decrease in the direct control by the user, with the introduction of supportive technologies. For example, BeBionic v3 has a plain switch for alternating between groups of preset grip types.Citation56 Touch Bionics offers i-mo™ technology (Touch Bionics Inc., Livingston, UK) that utilizes gyroscopes in order to detect sudden, user-elicited direction changes of the device that is used to select preset gestures.Citation57 Moreover, users of modern prosthetics are now given access to applications that can run on external devices capable of fine tuning and setting up gestures or gesture patterns.Citation56,Citation58,Citation59 This allows high-level customization. Touch Bionics even proposed the exchange of control settings through QR codes.Citation60 Finally, the same company introduced the so-called “grip chips” that can automatically preshape the hand once the prosthetic finds itself in the close proximity of one of these devices. Chips can then be placed in the characteristic spots within the user’s everyday environment.Citation61 Even though promising, all these novel features are yet to be clinically tested and their impact on the improvement of the quality of life remains to be demonstrated.

Academia has been extensively working on numerous prosthetic control solutions in the form of machine-learning approaches that are mostly based on the assumption that distinguishable and repeatable signal patterns exist among different motor tasks.Citation62,Citation63 There are several classification schemes proposed in the past several decades. However, the transition from the laboratory tests to the clinical and everyday practice has been very challenging. Only recently, a first pattern recognition-based system, COAPT,Citation64 has become commercially available, but it is yet to be seen how well this system will be accepted in the market.

The main issue of EMG pattern recognition-based systems is that they rely on repeatable matching of the produced EMG patterns during prosthetic manipulation to those used for system training. On the other hand, these patterns tend to significantly change due to environmental factors such as sweat or electrode shift, as well as fatigue, load, limb position, or simply due to the user’s change of focus.Citation65Citation69 Another drawback of these solutions is their limited ability to successfully cope with simultaneous motions, which makes them still not fully intuitive and somewhat cognitively demanding.Citation70

In order to overcome some of these issues, a new direction in myoelectric control research has been taken and biologically inspired algorithms have been designed. They combine techniques of mathematical regression with physiologically based models that allow intuitive control.Citation71,Citation72 These systems may enable the users to intuitively control multiple DoFs in a simultaneous and proportional way.Citation73,Citation74 These techniques have shown promising potential, though they are limited in the number of DoFs that can be controlled, at least without relying on nonintuitive motions and high-density EMG recordings.Citation75

Recently, a system that combines signal classification and regression for the control of the hand and wrist, respectively, has been proposed,Citation76 although full clinical testing is still lacking.Citation77

Surgical techniques for improved prosthetic experience

The importance of a proper stump management has been recognized long time ago, and its influence on prosthetic fitting, control, and acceptance can be significant.Citation78 Various surgical techniques have been used for salvation or restoration of the affected hand and arm function, thus enabling simpler and more efficient prosthetic solutions. Among others, these surgical methods include surgical tendon transfers,Citation79,Citation80 a variety of flaps,Citation81Citation84 digit and toe transfers,Citation85Citation87 and skin grafting.Citation88,Citation89

Mastering of the aforementioned techniques and broadening of the knowledge about nerve transfers in the past decade led to the development of TMR that transfers residual nerves that have lost their original targeted muscles to alternative muscle sites.Citation90 By treating muscles as biological amplifiers of the signals transmitted through nerves,Citation91 TMR allows reformation of several well-separated sources of intuitive EMG signals, which can be of high value for prosthetic control.Citation92 The Ottobock DynamicArm Plus in combination with the MyoHand Vari PlusSpeed terminal device accompanied by Wrist Rotator and a custom TMR socket enables the control of up to six DoFs following successful TMR procedure. To our knowledge, this is currently the only strictly TMR dedicated commercially available prosthetic solution on the market.

Recently, the concept of bionic reconstruction has been proposed for various scenarios after hand/arm trauma, extending the reconstructive options beyond the biological scope. In patients suffering from severe brachial plexus injuries presenting permanent hand dysfunction, which is beyond restoration using usual biological means alone, bionic reconstruction can offer a valid solution for restoring hand use where even the simplest of the aforementioned terminal devices are now allowing them to have better quality of life.Citation93 The success of this technique depends on numerous factors. One of the crucial aspects is the neurorehabilitation program that is delivered to the patients throughout the procedure.

Conclusion

By observing the current state of the upper limb prosthetic market, the most rapid development occurred during the last decade as the result of advances in technology, surgical techniques, and increased knowledge of human anatomy and physiology.

We have presented some of the latest prosthetic solutions on the market and have given insights in what academia is able to offer in the upcoming period. These systems offer numerous features when equipped with the state-of-the-art technology. Nowadays, prosthetic hands closely resemble the anatomical dimensions of their biological counterparts, while the solutions for higher level impairments are yet to match these specifications. Prosthetic solutions replacing certain DoFs, such as shoulder movements, are still missing.

The introduction of modern materials has allowed better and more intimate socket design that enables improved prosthetic experience for the users. This facilitated development of new socket fitting techniques allow orthopedic technicians to custom-match the residual limb anatomy of each user. Advances in implant design have come to a point where prosthetic fitting through osseointegration is a viable option. This kind of interface, directly coupled to the user’s skeletal system, provides increased range of motion and osseoperception. Following the advances of implantable sensor technologies, the new socket designs include features harvesting the significant potential of such solutions through embedded transmitter/receiver coils.

In order to improve the functional potential of upper limb-impaired patients, new surgical techniques have been recently proposed. TMR and bionic reconstruction, for example, provide important functional benefits to the user, in particular for high-level amputees and patients with brachial plexus injuries.

The current bottleneck of the upper limb prosthetic development seems to be the control of the robotic limbs. The majority of commercially available devices are still relying on the cumbersome mode-switching approaches dating some decades ago, while novel techniques seem to continuously fail in making a stable transition into the market. There are new solutions that have recently emerged but that have not yet been tested clinically on a large scale.

High prosthetic abandonment rates are still present, though they have been stagnating compared to the trends from about two decades ago. This can be attributed to the significant improvement in robustness and hardware design. Nevertheless, in order for these figures to start dropping, advances are still needed in prosthetic control, general system simplification, and custom, user-oriented solutions.

Acknowledgments

This work was supported by the Christian Doppler Research Foundation of the Austrian Federal Ministry of Science, Research and Economy and by the European Research Council Advanced Grant DEMOVE (contract #267888).

Disclosure

The authors report no conflicts of interest in this work.

References

  • NiskaRBhuiyaFXuJNational Hospital Ambulatory Medical Care Survey: 2007 emergency department summaryNatl Health Stat Report. 201026131 Available from: http://www.ncbi.nlm.nih.gov/pubmed/20726217Accessed June 1, 2016
  • OotesDLambersKTRingDCThe epidemiology of upper extremity injuries presenting to the emergency department in the United StatesHand (N Y)201271182223449400
  • LarsenCFMulderSJohansenAMStamCThe epidemiology of hand injuries in the Netherlands and DenmarkEur J Epidemiol200419432332715180102
  • U.S. Department of Labor, Bureau of Labor StatisticsWork Related Hand Injuries and Upper Extremity Amputation0073Washington, DCU.S. Department of Labor1982 Bulletin 2
  • KahnAPThe Encyclopedia of Work-Related Hand Injuries and Upper Extremity AmputationsNew York, NYFacts on File2004
  • RettigACAthletic injuries of the wrist and handAm J Sports Med200432126227314754754
  • GrunertBKSmithrJSDevineCAEarly psychological aspects of severe hand injuryJ Hand Surg Am198813-B2177180
  • GrobMPapadopulosNAZimmermannABiemerEKovacsLThe psychological impact of severe hand injuryJ Hand Surg Eur Vol200833335836218450795
  • PuttiVHistorical ProsthesesJ Hand Surg200530310325
  • MerrillDRLockhartJTroykPRWeirRFHankinDLDevelopment of an implantable myoelectric sensor for advanced prosthesis controlArtif Organs201135324925221371058
  • ChildressDHistorical aspects of powered limb prosthesesClin Prosthet Orthot198591213
  • MarquardtEThe Heidelberg pneumatic arm prosthesisJ Bone Jt1965473425
  • BiddissEChauTUpper-limb prosthetics: critical factors in device abandonmentAm J Phys Med Rehabil2007861297798718090439
  • PylatiukCSchulzSDöderleinLResults of an Internet survey of myoelectric prosthetic hand usersProsthet Orthot Int200731436237018050007
  • EgermannMKastenPThomsenMMyoelectric hand prostheses in very young childrenInt Orthop20093341101110518636257
  • McFarlandLVHubbard WinklerSLHeinemannAWJonesMEsquenaziAUnilateral upper-limb loss: satisfaction and prosthetic-device use in veterans and servicemembers from Vietnam and OIF/OEF conflictsJ Rehabil Res Dev201047429931620803400
  • ØstlieKLesjøIMFranklinRJGarfeltBSkjeldalOHMagnusPProsthesis rejection in acquired major upper-limb amputees: a population-based surveyDisabil Rehabil Assist Technol20127429430322112174
  • SilcoxHRooksMVogelRRFlemingLLMyoelectric prostheses. A long-term follow-up and a study of the use of alternate prosthesesJ Bone Joint Surg Am19937512178117898258548
  • BurroughSFBrookJAPatterns of acceptance and rejection of upper limb prosthesesOrthot Prosthet19853924047
  • GlynnMGalwayHHunterGSauterWManagement of the upper-limb-deficient child with a powered prosthetic deviceCurr Orthop Pract1986209202205
  • JonesJWGruberSABarkerJHBreidenbachWCSuccessful hand transplantation one-year follow-upN Engl J Med2000343746847310950668
  • SalmingerSSturmaARocheA[Hand function after hand transplant and prosthetic reconstruction] Handfunktion nach Handtransplantation und prothetischer Rekonstruktion38 Seminar – Surgery meets Biomedical Engineering. Austrian Society for Surgical ResearchWagrain, Austria2014
  • Vincent Systems [webpage on the Internet]Vincent Evolution 22016 Available from: www.vincentsystems.de/en/prosthetics/vincent-evolution-2/Accessed June 21, 2016
  • ResnikLKlingerSLEtterKThe DEKA arm: its features, functionality, and evolution during the veterans affairs study to optimize the DEKA armProsthet Orthot Int201438649250424150930
  • ResnikLKlingerSLEtterKUser and clinician perspectives on DEKA arm: results of VA study to optimize DEKA armJ Rehabil Res Dev2014511273824805891
  • SearsHHIversenEArcherSJacobsTWrist innovations to improve function of electric terminal devicesProc 2008 MyoElectric Control Prosthetics Symp20081316
  • BennettDAMitchellJGoldfarbMDesign and characterization of a powered elbow prosthesis2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC). IEEE201524582461
  • LipschutzRDKuikenTAMillerLADumanianGAStubblefieldKAOtrLShoulder disarticulation externally powered prosthetic fitting following targeted muscle reinnervation for improved myoelectric controlProsthet Orthot Int2006182245253
  • FDA [webpage on the Internet]FDA allows marketing of first prosthetic arm that translates signals from person’s muscles to perform complex tasks Available from: http://www.fda.gov/NewsEvents/Newsroom/PressAnnouncements/ucm396688.htmAccessed February 24, 2016
  • LakeCThe evolution of upper limb prosthetic socket designJPO J Prosthet Orthot200820385
  • DalyWVooLRosenbaum-ChouTArabianABooneDSocket pressure and discomfort in upper-limb prostheses: a preliminary studyJPO J Prosthet Orthot201426299106
  • BiddissEAChauTTUpper limb prosthesis use and abandonment: a survey of the last 25 yearsProsthet Orthot Int200731323625717979010
  • HeppOKuhnGGUpper extremity prosthesesProsthetics International, Proceedings of the Second International Prosthetics Course. Committee on Prosthetics, Braces, and Technical Aids, International Society for the Welfare of CripplesCopenhagen, Denmark1960133181
  • BillockJNThe Northwestern University supracondylar suspension technique for below-elbow amputationsOrthot Prosthet19722641623
  • UellendahlJERamdialSCustom silicone sockets for myoelectric prosthesesMEC ’05 Intergrating Prosthetics and Medicine Proceedings of the 2005 MyoElectric Controls/Powered Prosthetics SymposiumFredericton, NB20052124
  • AlleyRDAdvancement of upper extremity prosthetic interface and frameMEC ’02 The Next Generation Proceedings of the 2002 MyoElectric Controls/Powered Prosthetics SymposiumFredericton, NB2002
  • BranemarkPIOsseointegration and its experimental backgroundJ Prosthet Dent19835033994106352924
  • BrånemarkRBrånemarkP-IRydevikBMyersRROsseointegration in skeletal reconstruction and rehabilitation: a reviewJ Rehabil Res Dev200138217518111392650
  • TjellstromALindstromJHallenOAlbrektssonTBrånemarkP-IOsseointegrated titanium implants in the temporal bone. A clinical study on bone-anchored hearing aidsAm J Otol1981243043106894824
  • BrånemarkP-IChienSThe Osseointegration Book: From Calvarium to CalcaneusHanover Park, ILQuintessence Publishing Company2005
  • JönssonSCaine-WinterbergerKBrånemarkROsseointegration amputation prostheses on the upper limbs: methods, prosthetics and rehabilitationProsthet Orthot Int201135219020021697201
  • GallagherPDesmondDMacLachlanMPsychoprostheticsBerlinSpringer-Verlag2007
  • KlinebergIMurrayGOsseoperception: sensory function and proprioceptionAdv Dent Res19991312012911276734
  • ZuoKJOlsonJLThe evolution of functional hand replacement: from iron prostheses to hand transplantationCan J Plast Surg20142214451
  • SheltonTJBeckPJBloebaumRDBachusKNPercutaneous Osseo-integrated prostheses for amputees: limb compensation in a 12-month ovine modelJ Biomech201144152601260621920525
  • WeirRFFTroykPRRDeMicheleGKuikenTKuTImplantable myoelectric sensors (IMES) for upper-extremity prosthesis control-preliminary workProc 25th Annu Int Conf IEEE Eng Med Biol Soc. IEEECancun, Mexico2003215621565
  • WeirRTroykPDemicheleGKernsDTechnical details of the implantable myoelectric sensor (IMES) system for multifunction prosthesis controlConf Proc IEEE Eng Med Biol Soc200577337734017281975
  • MuceliSPoppendieckWNegroFAccurate and representative decoding of the neural drive to muscles in humans with multi-channel intramuscular thin-film electrodesJ Physiol2015593173789380426174910
  • CiprianiCSegilJLBirdwellJAWeirRFDexterous control of a prosthetic hand using fine-wire intramuscular electrodes in targeted extrinsic musclesIEEE Trans Neural Syst Rehabil Eng201422482883624760929
  • FarnsworthBDTalyorDMTrioloRJYoungDJWireless in vivo EMG sensor for intelligent prosthetic controlTRANSDUCERS 2009 – 15th Int Conf Solid-State Sensors, Actuators Microsystems. IEEEDenver, CO, USA2009358361
  • MiceraSNavarroXCarpanetoJOn the use of longitudinal intrafascicular peripheral interfaces for the control of cybernetic hand prostheses in amputeesIEEE Trans Neural Syst Rehabil Eng200816545347218990649
  • ScottRNParkerPAMyoelectric prostheses: state of the artJ Med Eng Technol19881241431513057209
  • SteeperRSLBeBionic v3 Product BrochureLeedsRSL Steeper2013
  • Otto Bock Healthcare GmbHFDA 510(k) Premarket Notification: Summary Axon-Bus Prosthetic SystemSilver Spring, MDOtto Bock Healthcare GmbH2014 20993-0002
  • WilliamsTW3rdPractical methods for controlling powered upper-extremity prosthesesAssist Technol19902131810149040
  • SteeperRSLUpper Limb Prosthetic Components – Technical Product BrochureLeedsRSL Steeper2014
  • Touch Bionics [webpage on the Internet]How the i-limb works Available from: http://www.touchbionics.com/products/how-i-limb-worksAccessed April 29, 2016
  • Touch BionicsBiosim-i My i-limb DatasheetMansfield, MATouch Bionics2014
  • Otto Bock Healthcare GmbHMichelangelo Technican Product BrochureDuderstadtOtto Bock Healthcare GmbH2013
  • Touch Bionics [webpage on the Internet]i-limb QR codes Available from: http://www.touchbionics.com/resources/qr-codesAccessed April 29, 2016
  • Touch BionicsGrip Chips – Product OverviewMansfield, MATouch Bionics2015
  • HudginsBParkerPScottRNA new strategy for multifunction myoelectric controlIEEE Trans Biomed Eng199340182948468080
  • GraupeDSalahiJKohnKHMultifunctional prosthesis and orthosis control via microcomputer identification of temporal pattern differences in single-site myoelectric signalsJ Biomed Eng19824117227078136
  • Coapt LLC [homepage on the Internet]Coapt Engineering2016 Available from: http://www.coaptengineering.com/Accessed April 29, 2016
  • YoungAJHargroveLJKuikenTAThe effects of electrode size and orientation on the sensitivity of myoelectric pattern recognition systems to electrode shiftIEEE Trans Biomed Eng20115892537254421659017
  • HargroveLEnglehartKHudginsBA training strategy to reduce classification degradation due to electrode displacements in pattern recognition based myoelectric controlBiomed Signal Process Control200832175180
  • StangoANegroFFarinaDSpatial correlation of high density EMG signals provides features robust to electrode number and shift in pattern recognition for myocontrolIEEE Trans Neural Syst Rehabil Eng201523218919825389242
  • Ortiz-CatalanMBrånemarkRHåkanssonBDelbekeJOn the viability of implantable electrodes for the natural control of artificial limbs: review and discussionBiomed Eng Online20121113322715940
  • RocheADRehbaumHFarinaDAszmannOCProsthetic myoelectric control strategies: a clinical perspectiveCurr Surg Rep20142344
  • JiangNDosenSFarinaD webpage on the InternetMyoelectric control of artificial limbs: is there the need for a change of focus?IEEE Signal Process Mag20121215 Available from: http://www.lifesciences.ieee.org/articles/174-myoelectric-control-of-artificial-limbs-is-there-a-need-to-change-focusAccessed June 27, 2013
  • JiangNEnglehartKBParkerPAExtracting simultaneous and proportional neural control information for multiple-DOF prostheses from the surface electromyographic signalIEEE Trans Biomed Eng20095641070108019272889
  • JiangNRehbaumHVujaklijaIGraimannBFarinaDIntuitive, online, simultaneous, and proportional myoelectric control over two degrees-of-freedom in upper limb amputeesIEEE Trans Neural Syst Rehabil Eng201422350151023996582
  • JiangNVujaklijaIRehbaumHGraimannBFarinaDIs accurate mapping of EMG signals on kinematics needed for precise online myoelectric control?IEEE Trans Neural Syst Rehabil Eng201422354955824235278
  • HofmannDJiangNVujaklijaIFarinaDBayesian filtering of surface EMG for accurate simultaneous and proportional prosthetic controlIEEE Trans Neural Syst Rehabil Eng Epub201511201
  • IsonMVujaklijaIWhitsellBFarinaDArtemiadisPHigh-density electromyography and motor skill learning for robust long-term control of a 7-DoF robot armIEEE Trans Neural Syst Rehabil Eng201624442443325838524
  • AmsuessSVujaklijaIGobelPContext-dependent upper limb prosthesis control for natural and robust useIEEE Trans Neural Syst Rehabil Eng Epub2015791
  • RocheADVujaklijaIAmsuessSStructured rehabilitation training for improved multifunctional prosthetic control : a case studyJ Vis Exp2015105e52968
  • DederichRPlastic treatment of the muscles and bone in amputation surgery a method designed to produce physiological conditions in the stumpJ Bone Joint Surg Am196345B6066
  • LieberRLSkeletal muscel architecture implications for muscle function and surgical tendon transferJ Hand Ther1993621051138343877
  • CooneyWPLinscheidRLAnKNOpposition of the thumb: an anatomic and biomechanical study of tendon transfersJ Hand Surg Am1984967777866512185
  • MayJWGordonLPalm of hand free flap for forearm length preservation in nonreplantable forearm amputation: a case reportJ Hand Surg Am1980543773807419882
  • UlusalBGLinY-TUlusalAELinC-HFree lateral arm flap for 1-stage reconstruction of soft tissue and composite defects of the hand: a retrospective analysis of 118 casesAnn Plast Surg200758217317817245144
  • BrandtKKhouriRKUptonJFree flaps as flow-through vascular conduits for simultaneous coverage and revascularization of the hand or digitPlast Reconstr Surg19969823213278764721
  • NeumeisterMHeggeTAmalfiASauerbierMThe reconstruction of the mutilated handSemin Plast Surg20102417710221286307
  • BlairVPByarsLTToe to finger transplantAnn Surg194011228729017857641
  • MorrisonWAO’BrienBMacLeodAMRing finger transfer in reconstruction of transmetacarpal amputationsJ Hand Surg Am19849A14116693742
  • FoucherGMossALHMicrovascular second toe to finger transfer: a statistical analysis of 55 transfersBr J Plast Surg199144287902018911
  • CantyTJBleckEESplit-thickness skin grafting of amputation stumpsPlast Reconstr Surg195210299115
  • WoodMRHunterGAMillsteinSGThe value of stump split skin grafting following amputation for trauma in adult upper and lower limb amputeesProsthet Orthot Int198711271743309883
  • KuikenTADumanianGALipschutzRDMillerLAStubblefieldKAThe use of targeted muscle reinnervation for improved myoelectric prosthesis control in a bilateral shoulder disarticulation amputeeProsthet Orthot Int200428324525315658637
  • HofferJLoebGImplantable electrical and mechanical interfaces with nerve and muscleAnn Biomed Eng1980843513607027837
  • KuikenTALiGLockBATargeted muscle reinnervation for real-time myoelectric control of multifunction artificial armsJAMA2009301661962819211469
  • AszmannOCRocheADSalmingerSBionic reconstruction to restore hand function after brachial plexus injury : a case series of three patientsLancet201538599832183218925724529
  • Hosmer Dorrance CorporationHosmer Upper Extremity Product Catalog2012 Available from: http://hosmer.com/products/PDFs/Fil-lauer Product Catalog.pdfAccessed June 1, 2016
  • Hosmer Dorrance CorporationHooks, Hands, and Wrists – Product Brochure2003158 Available from: http://hosmer.com/products/book3/pdfs/2-HooksHandsWrists.pdfAccessed June 1, 2016
  • Otto Bock Healthcare GmbHUpper Limb Prosthetics Catalog4DuderstadtOtto Bock Healthcare GmbH2014177184
  • Liberating Technologies IncSelect Electric Hands for Adults Product CatalogHolliston, MALiberating Technologies Inc2013
  • Liberating Technologies Inc [homepage on the Internet]Select Electric Hands & Gloves Data Sheet2010 Available from: http://www.liberatingtech.com/products/documents/LTI_Data_Sheet_-_Electric_Hands_&_Wrists.pdfAccessed June 21, 2016
  • Touch Bionicsi-Limb Quantum Clinician ManualMansfield, MATouch Bionics2015MA 01333 Issue No 1
  • Touch Bionicsi-Limb Quantum User Manual201511121 Available from: http://www.touchbionics.com/sites/default/files/files/i-limb quantum user manual July 2015.pdfAccessed June 1, 2016
  • Touch Bionicsi-Limb Quantum Product SheetMansfield, MATouch Bionics201512
  • Touch BionicsFlexion Wrist Product SheetMansfield, MATouch Bionics201512
  • FillauerUtah Arm 3 Plus Sales SheetChattanooga, TNFillauer2012
  • HussainIMasoodAIqbalJKhanUSDesign of high torque elbow joint for above elbow prosthesisInt J Mech Aerosp Ind Mechatron Eng2011511318321
  • Boston DigitalBoston Digital Arm System CatalogAucklandBoston Digital2013
  • Otto Bock Healthcare GmbHElbow Components – At a GlanceDuderstadtOtto Bock Healthcare GmbH2014
  • Hosmer Dorrance Corporation Externally Powered Systems Product CatalogCampbell, CA201518 Available at: http://hosmer.com/products/book3/pdfs/4-ExternallyPoweredSystems.pdfAccessed June 21, 2016